dissolved inorganic carbon in seawater (vindta 3d)marianda.com/downloads/vindta 3d product...
TRANSCRIPT
PRODUCT INFORMATION
Dissolved inorganic carbon in seawater (VINDTA 3D):
The VINDTA 3D is a system for the
extraction of CO2 from a sea water
sample and quantitative transfer of
released CO2 to a coulometric titration
unit (e.g. coulometer 5011, UIC Inc.,
USA) for the measurement of total
dissolved inorganic carbon (DIC) in
sea water. Sample handling and
coulometric titration is controlled by
the software.
In principle, the VINDTA 3D is the DIC - part of the VINDTA 3C; see there for
further details. The VINDTA 3D is prepared for the optional available separate
add-on units for gas calibration with pure CO2 and for salinity
determination from the same sample. Another option is a Peltier cooler option
for the condenser (this saves one circulation bath).
The VINDTA 3D is comparable to the former SOMMA system. Optional we offer
another version of the VINDTA 3D, that uses the SOMMA glassware and partially
also SOMMA hardware (e.g. the 110V AC valves; this 'SOMMA version' makes
use of 'leftovers' and spare parts from an old SOMMA system).
features:
• automated sample handling and coulometric titration
• easily transportable and sea-going
• only one gas (carrier gas) required
• precision typically ± 1 µmol/kg (open ocean water)
• user friendly software interface using LabVIEW™
• optional control PC (Shuttle-PC) tested with the hardware and with all software
and drivers installed
• gas calibration and salinity measurement as option
LabVIEW™ is a trade mark of National Instruments Inc.
Additional hardware required or recommended:
UIC coulometer model 5011, 5012, or 5014 (with 5011 emulator option)
circulation bath for temperature control (20 or 25°C) of the system equilibration of sample bottles
second circulation bath for the condenser (~ 2°C; not required with Peltier option)
carrier gas supply (nitrogen or dry and CO2-free air)
marine analytics and data
Total alkalinity and dissolved inorganic carbon in
seawater (VINDTA 3C):
VINDTA 3C (VINDTA # 007)
The VINDTA 3C (Versatile INstrument for the Determination of Total
inorganic carbon and titration Alkalinity) combines the proven VINDTA
alkalinity titration concept with a simplified extraction unit for coulometric
dissolved inorganic carbon (DIC) measurement.
The alkalinity part of the VINDTA 3C is nearly identical to the VINDTA 3S
system; an additional part for the extraction of CO2 from seawater for
coulometric titration has been added to allow for determination of
dissolved inorganic carbon from the same sample simultaneously (the
DIC-only system is available as VINDTA 3D).
The sample transport is performed with peristaltic pumps and the acid
addition is done using a membrane pump; therefore no pressurizing
system is required (in contrast to the SOMMA system) and only one gas
supply (nitrogen or dry and CO2-free air) is necessary. As option, a Peltier
cooler system for the condenser (this saves one circulation bath) is
available.
features:
• determine alkalinity and dissolved inorganic from the same sample
simultaneously
• easily transportable and sea-going
• only one gas (carrier gas) required
• precision typically ± 1 µmol/kg ( AT and/or DIC; open ocean water)
• simplified operation for DIC compared to the SOMMA
• user friendly software interface using LabVIEW™
• optional control PC (Shuttle-PC) tested with the hardware and with
all software and drivers installed
LabVIEW™ is a trade mark of National Instruments Inc.
Additional hardware required or recommended:
UIC coulometer model 5011, 5012, or 5014 (with 5011 emulator option)
METROHM Titrino (719SET or similar) with stirrer 728 and 5 mL burette unit
Metrohm reference electrode # 60729.100
Thermo Orion ROSS electrode #8101BN
circulation bath for temperature control (20 or 25°C) of the system equilibration of
sample bottles
second circulation bath for the condenser (~ 2°C; not required with Peltier option)
carrier gas supply (nitrogen or dry and CO2-free air)