Discovering quality in radiotherapy through transmission detection
Mark Geurts, Researcher and Adjunct ProfessorUniversity of Wisconsin, Madison, USA
January 20, 2018
![Page 2: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/2.jpg)
Presentation outline
1. Introduction to University of Wisconsin
2. Review goals of patient-specific Quality Assurance
3. Explore role of transmission detection in QA
4. Evaluation of the ScandiDos Delta4 Discover
![Page 3: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/3.jpg)
Author Disclosure
• ScandiDos AB and HGPT funded travel to this meeting
and provided an honorarium
• The University of Wisconsin is a Center of Excellence
for ScandiDos products
• Images in this presentation are from delta4family.com
![Page 4: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/4.jpg)
University of Wisconsin, Madison
![Page 5: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/5.jpg)
UW medical physics history
John Cameron
Thermo-Luminescent Dosimetry (TLD)
Bone densitometry
Frank Attix
Spencer-Attix cavity theory
![Page 6: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/6.jpg)
UW medical physics history
Charles Mistretta
Digital Subtraction Angiography (DSA)
Vascular imaging
Thomas “Rock" Mackie
Pinnacle treatment planning system
TomoTherapy treatment system
![Page 7: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/7.jpg)
UW Radiation Oncology by the Numbers
Delivery
• 4 Varian
• 3 TomoTherapy
• 2 Elekta HDR
• 1 ViewRay
Planning
• RayStation
• Pinnacle
• Eclipse
• TomoTherapy
• Oncentra
Quality Assurance
• 5 Delta4
phantoms
• Mobius3D
![Page 8: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/8.jpg)
Delta4 history at the UW
• First unit commissioned in 2008
• Established as primary tool in 2012
• Upgraded to Delta4+ in 2016
• All UW centers now use Delta4
![Page 9: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/9.jpg)
UW IMRT QA policy
• Performed prior to treatment start, or
before 3 fractions/10% on transfers
• >95% data points must pass 3%/3 mm
global Gamma, using 20% threshold
• Median absolute dose < 3%
![Page 10: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/10.jpg)
Delta4 performance
• 2135 clinical plans since 2012
• Pass rate (3%/3mm): 96.3%
• Mean pass rate: 98.9%
• Absolute dose: 0.2% ± 1.3%
![Page 11: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/11.jpg)
Delta4+ upgrade
• Pass rate increased from
95.8% to 98.8%
• Standard deviation
decreased 1.4% to 1.0%
• No service in 2 years
![Page 12: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/12.jpg)
Goals of Patient Specific QA
![Page 13: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/13.jpg)
How safe is radiotherapy?
• WHO survey from 1976-2007
identified 3125 incidents1
• Patients died in 38 of them
• Other studies2 report error rates
between 0.1% to 4.7%
1WHO Radiotherapy Risk Profile (2008). www.who.int2Ford (2009). IJROBP 74:852-8
![Page 14: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/14.jpg)
Radiation therapy complexity
Huq (2016). Med Phys 43: 4209-62
![Page 15: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/15.jpg)
What is patient specific QA?
Quality Management
Quality Assurance
Machine QA Patient Specific QA
Pre-treatment Review
Delivery QAOn-
treatment Review
Quality Control
• Quality controls force the
desired level of quality
• Quality assurance
demonstrates a level of
quality for a process
American Society for Quality. asq.org
![Page 16: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/16.jpg)
Is delivery QA beneficial?In house ILS found delivery QA only detected 1.4% of all incidents:
Ford (2012). IJROBP 84: e263-9
![Page 17: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/17.jpg)
Is delivery QA beneficial?
1WHO Radiotherapy Risk Profile (2008). www.who.int2Gopan (2016). Med Phys 43: 5181-7
• WHO1 found 10% of all errors
occurred during treatment
delivery, 9% in data transfer
• In a study of SAFRON
events2, 23 of 81 (28%) had
incorrect field parameters
(angle, MU, wedge, etc.)
PlanningCommissioning
Delivery
Data Transfer
Source of Incident
![Page 18: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/18.jpg)
Plan quality sensitivity to delivery error
Fredh (2013). Med Phys 40: 031716
![Page 19: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/19.jpg)
Plan quality sensitivity to delivery error
Arumugam (2016). Physica Medica 32: 1238-44
![Page 20: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/20.jpg)
Recent surveys of medical physicists
Yes
No
Have you ever changed a plan based on IMRT QA?
On every
patient
Routinely
Not at all
Is absolute dose measurement necessary?
www.surveymonkey.com
![Page 21: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/21.jpg)
Transmission Detection QA
![Page 22: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/22.jpg)
What Is Transmission QA?
• Transmission QA systems place
an array of detectors between
the collimated beam and patient
• They allow intra-fraction
measurement of machine
parameters during treatment
![Page 23: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/23.jpg)
Strengths and Weaknesses
Strength Weakness
Real time QA measurement Difficult to compare to TPS
Independent of treatment system Low energy discrimination
Output, gantry angle, and collimation Surface dose, beam quality affected
Accurate MLC position verification*
*some systems
![Page 24: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/24.jpg)
Transmission vs. exit dosimetry
• 21st Century Oncology
evaluated 6.4 million EPID
exit doses across 84 centers
• 13% to 26% of all images
failed using 3% or 5% criteria50%
60%
70%
80%
90%
100%
H&NPelv
isBrea
stLu
ng
Prostat
e
Average Weekly Portal Pass Rate
Olivera et al. (2016) www.standardimaging.com
![Page 25: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/25.jpg)
What about log file analysis?
Several groups have discovered errors not caught by log file:
Agnew (2014). Phys Med Bio 59: N49-63
![Page 26: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/26.jpg)
What about log file analysis?
Neal (2016). Med Phys 43: 2933-5
Several groups have discovered errors not caught by log file:
![Page 27: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/27.jpg)
What about log file analysis?Several groups have discovered errors not caught by log file:
Rowshanfarzad (2015). Br J Radiol 88: 20140581
![Page 28: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/28.jpg)
Transmission detector comparison
*6 MV, 10x10
Li (2016). JACMP 17: 235-44
![Page 29: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/29.jpg)
ScandiDos Delta4 Discover
![Page 30: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/30.jpg)
Delta4 design
• 4040 p-type diodes with 1 mm active diameter
• Active area 25 x 20 cm at isocenter
• Diode spacing 1.6 (2.5) mm along MLC, 3.2 (5) mm against
• Overall system thickness 23 mm
Li (2016). JACMP 17: 235-44
![Page 31: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/31.jpg)
Delta4 Discover evaluation
Parameters
• Transmission
• Surface dose
• Beam quality
• Beam profile
• Output sensitivity
• IMRT reproducibility
• MLC error sensitivity
References
• Riley (2011). Med Phys 38: 3534
• Li (2016). JACMP 17: 235-44
• Saenz (2016). Med Phys 43: 3539
• Hoffman (2016). Med Phys 43: 3538
• Gonod (2017). Phys Med 44: 30
• Cheung (2017). PRO 7: e559-67
![Page 32: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/32.jpg)
Transmission
6 MV 10 MV 15 MV 18 MV Reference
0.989 0.993 Li (2016). JACMP 17: 235-44
0.988 0.993 Saenz (2016). Med Phys 43: 3539
0.985 Gonod (2017). Phys Med 44: 30
0.983 0.986 0.988 Hoffman (2016). Med Phys 43: 3538
![Page 33: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/33.jpg)
Surface dose
Author Cheung Li
Field (cm2) 10x10, 40x40 4x4, 10x10, 30x30
SSD (cm) 80, 100 90
Energy (MV) 6, 10 6, 15
Detector OSLD Diode
Result 10 x 10 40 x 40
80 +35% +63%
100 +4% +32%
![Page 34: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/34.jpg)
Beam quality
Author Hoffman Cheung Li Saenz
Field 1x1, 30x30 3x3 to 40x40 4x4 to 30x30 10x10
SSD 80, 100 90
Energy 6, 10, 15 6, 10 6, 15 6, 18
PDD 1% (> 2.5 cm)1.2% (100)
3% (80)
PDD(10):
-0.5%
PDD(10):
-0.7% (6)
+0.6% (18)
Dmax
4 mm (100)
8 mm (80)1.2 mm
![Page 35: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/35.jpg)
Beam profile
Author Hoffman Cheung Li
Field (cm2) 30x30 3x3 to 40x40 30x30
SSD (cm) 80, 90, 100 90
Energy (MV) 6, 10, 15 6, 10 6, 15
Result 1.25%2.6% (100)
6.3% (80)
Flatness -0.3%
Symmetry +0.5%
![Page 36: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/36.jpg)
Output sensitivityAuthor Li Saenz
MU error +1 to +5% -5 to +5%
Modality Static MLC field Static MLC field
Result
![Page 37: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/37.jpg)
IMRT reproducibility
Author Hoffman Gonod
Modality VMAT VMAT
Plan type Prostate Cervix, Head and Neck
Frequency Measurements over 9 days
Metric Gamma pass rate (3%/3mm)
StandardDeviation
0.5% 1.1-1.5%
![Page 38: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/38.jpg)
MLC error sensitivity
Author Li Saenz
Modality Static MLC Prostate VMAT Conformal arc
Leaf error 1-5 mm 0.5-5 mm 0-4 mm
Sensitivity 0.7 mm 1.2 mm
Result
![Page 39: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/39.jpg)
Combination with Delta4 Phantom
0
1
2
3
Discover Phantom+
Discover
Phantom+
Phantom+
Phantom+
Discover
Discover
![Page 40: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/40.jpg)
Error sensitivity relative to Delta4 Phantom
Riley (2011). Med Phys 38: 3534
![Page 41: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/41.jpg)
3DCRT evaluation
Without a Delta4 Phantom measurement, Discover can still measure
and compare to the DICOM RTPLAN:
• Monitor Units
• MLC position
• Gantry angle
• Collimator angle
![Page 42: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/42.jpg)
Anatomic evaluation
• Delta4 results can be used to
reconstruct dose to the planning CT
• Several groups have investigated:
Hu (2013). Med Phys 40: 213
Stambaugh (2014). JACMP 15: 169-180
Hauri (2014). JACMP 15: 181-90
Crockett (2015): Med Phys 42: 3411-2
Bängtsson (2016). Med Phys 43: 3370-1
![Page 43: Discovering quality in radiotherapy through transmission ... · Is delivery QA beneficial? 1WHO Radiotherapy Risk Profile (2008). 2Gopan (2016).Med Phys 43: 5181-7 • WHO1 found](https://reader033.vdocuments.mx/reader033/viewer/2022053000/5f04f6c87e708231d41094a5/html5/thumbnails/43.jpg)
Thank you for your time