Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca Paolo Pozzi Neuroscience Department University of Pavia

Upload: felicity-armstrong

Post on 13-Jan-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Digital two photon microscopy for multiple fast signals acquisition

Laboratory of Advanced Biological Spectroscopy (L.A.B.S.)University of Milan - Bicocca

Paolo PozziNeuroscience Department

University of Pavia

Page 2: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Flow Cross Correlation in Zebrafish Vascular System

Neuronal Network Analysis in Cerebellar Slices

Page 3: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Galvanometric Mirrors

Scan head

Phototube

MicroscopeObjective

Pulsed IR Laser

Two Photon Microscope

Sample

Page 4: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Raster scanning

Page 5: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Raster scanning

I

t

Page 6: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Raster scanning

I

t

Page 7: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

𝐾 +¿ ¿

𝐶𝑙−

𝑁𝑎+¿ ¿

𝐾 +¿ ¿𝐶𝑙−𝑁𝑎+¿ ¿

𝐴−

Ionic ChannelsDendrites

Axon

𝐶𝑎2+¿ ¿

𝐶𝑎2+¿ ¿

Neurons

Page 8: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Ca2+

Ca2+

Ca+

Ca +

Ca +Ca2+

Ca2+

Ca2+

Ca2+

Ca2+

Ca2+

Ca2+

Ca2+

Neurons

Page 9: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

From olympusconfocal.com

Neuronal Networks

Page 10: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Experiment requirements:

• Confocal imaging

• Largest possible field of view

• Diffraction limited resolution

• High sensitivity to fluorescence signals fluctuations

• Millisecond time scale

Neuronal Networks

Page 11: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Experiment requirements:

• Confocal imaging

• Largest possible field of view

• Diffraction limited resolution

• High sensitivity to fluorescence signals fluctuations

• Millisecond time scale

Over large fields of view, two photon imaging can achieve 2-3 frames per second.

We have to improve by three orders of magnitude!

Neuronal Networks

Page 12: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Luckily, we don’t really need an image

Calcium induced calcium release

Signal in the cellular body varies uniformly on the microsecond scale

Page 13: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Luckily, we don’t really need an image

Experiment outline:

• Generation of a single confocal image

• Selection of 1 pixel (POI) in every cell

• Simultaneous illumination of all POIs

• Simultaneous acquisition of all signals

Page 14: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Spatial light modulation

Back focal plane

Focal Plane

Fourier Transform

User generate phase shift pattern

Spatial light modulator

Page 15: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Gerchberg-Saxton Algorithm

SLM Plane Sample Plane

IFFT

FFT

Desired pattern

START

END

Laser intensity

AmplitudeAmplitude Phase Phase

Page 16: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Galvanometric Mirrors

Scan head

Phototube

State of the art

Camera

Laser

Beam expander

SLM

MicroscopeObjective

Nikolenko et Al. 2008

Page 17: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Some problems:

• Coordinates matching!!!

• Zero order of diffraction

• Excitation power wasted

• Very complicated (Remember, biologists must use it!!!)

Page 18: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

How we do it:

Camera

Laser

SLMMicroscopeObjective

Page 19: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Experimental setup

Phase:

Intensity:

Phase:

Intensity:

Page 20: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Experimental setup

Phase:

Intensity:

Phase:

Intensity:

+

Page 21: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

How to make an image without a galvo scanner?

12

N…

Page 22: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

12

N…

Digital two photon imaging

It is a very long calculation, but you do it once, and forget about it!

Page 23: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Digital two photon imaging

12 …

Each bright spot is a pixel of the final image.

N

Page 24: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Digital two photon imaging

And here comes the image (in convenient SLM input coordinates)

Page 25: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Some numbers

• SLM refresh rate: 60 Hz

• Camera acquisition rate: 100 Hz @full frame, 1600 Hz in a 128x128 ROI

• Laser power: 4 W @ 800 nm

• Scan grid:-28 x 28 focuses in a 12x12 scan pattern (2.4 s)-30 x 30 focuses in a 20x20 scan pattern (6.7 s)-25 x 25 focuses in a 40x40 scan pattern (26.7 s)

• Time for single hologram calculation: ~30 s

Page 26: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

So, does it work?

Page 27: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca

Averange

So, is it useful?

Page 28: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca
Page 29: Digital two photon microscopy for multiple fast signals acquisition Laboratory of Advanced Biological Spectroscopy (L.A.B.S.) University of Milan - Bicocca