digital techniques for myocardial perfusion spect
DESCRIPTION
SPECT Techniques for data acquisition, processing and display in Nuclear CardiologyTRANSCRIPT
![Page 1: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/1.jpg)
Digital Techniques for Myocardial Perfusion SPECT
Dr. Muhammad AyubDiplomate Certification Board of Nuclear CardiologyDiplomate Certification Board of Cardiovascular CT
Punjab Institute of Cardiology, Lahore
![Page 2: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/2.jpg)
SINGLEPHOTONEMISSIONCOMPUTEDTOMOGRAPHY
A Modern SPECT Gamma Camera
![Page 3: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/3.jpg)
SPECT Equipment
• Camera– Single Detector– Multi detector
• Analog to digital converter• Computer interface
– Data storage– Data processing– Display
![Page 4: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/4.jpg)
Gamma Camera• Gantry• Camera Heads
– Collimator– Scintillation
Crystal– Photo multiplier
tubes– PHA– ADC Crystal
PMTs
Lead Shield
Pulse Processing Electronics
Collimator
PHAx y
Display
![Page 5: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/5.jpg)
Cardiac SPECT Acquisition
• Gamma camera with 1 or 2 heads (90 ° angle)
• 180° rotation• 32 steps (16 per head
if 2)• Planar projection
acquisition at each step
Step and shoot acquisition
![Page 6: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/6.jpg)
Workstation
• Types– Unix Workstations– PCs, Macs
• Storage• Memory• Graphics• Software
![Page 7: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/7.jpg)
PHA
![Page 8: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/8.jpg)
Tomographic Acquisition
• Matrix• Angular Range• Angular Steps• Time per Step• Gated Vs non gated acquisition• Continuous Vs Step and Shoot
acquisition
![Page 9: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/9.jpg)
Tomographic Acquisition
• Matrix– 64 x 64 Standard– 128 x 128
![Page 10: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/10.jpg)
![Page 11: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/11.jpg)
![Page 12: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/12.jpg)
Tomographic Acquisition
• Angular Range– 180 degrees– 360 degrees
215 LPO
RAO
135
![Page 13: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/13.jpg)
Tomographic Acquisition
Angular Steps – 3 degrees Recommended for Tc Studies– 6 degrees Recommended for Tl Studies– 12 degrees Not Recommended
![Page 14: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/14.jpg)
Image Acquisition
• Orbit– Circular Standard– Non-circular Optional
![Page 15: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/15.jpg)
Tomographic AcquisitionGated SPECT
![Page 16: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/16.jpg)
Technical Considerations
• QC– Energy– Uniformity– Linearity– COR
![Page 17: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/17.jpg)
Problems
• Motion Artifacts• COR defects• Soft tissue attenuation• Foreign Objects
![Page 18: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/18.jpg)
Image Processing
• Filtered Back projection• Filtering• Iterative Reconstruction• Corrections
– Attenuation Correction– Scatter Correction
![Page 19: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/19.jpg)
Theory of Back Projection 1
Imaging a point source• Rotating camera head• Step and shoot at 8
angles over 180°• Only photons to
camera face are detected
• Planar ‘projection’ acquisition at each step
0°22.5 °
45°
67.5°
90°
180°
Point source
![Page 20: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/20.jpg)
45°
67.5°
22.5°0°
Back Projection of 4 Ray Sums
Camera anglex
Depth in transaxial slice
Transaxial Plane
![Page 21: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/21.jpg)
Filtering
• Pre filtering• 3D Post Filtering• Types
– Low pass– Metz– Wiener– Band
![Page 22: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/22.jpg)
Types of Filters Used
• Smoothing filter–E.g. Hanning or Butterworth–Smooths image (removes noise
generated by ramp filter)
1.0
0.0
Frequency
Magnitude• Ramp filter –Always applied–Removes star-burst effect (good!)–Amplifies high frequency noise (bad!)
![Page 23: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/23.jpg)
Pre filtration
![Page 24: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/24.jpg)
3 D Post FilterLow pass Filter
![Page 25: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/25.jpg)
Metz Filter
![Page 26: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/26.jpg)
Wiener Filter
![Page 27: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/27.jpg)
Band Pass Filter
![Page 28: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/28.jpg)
Filter Example: Butterworth
Filtered Raw image
Butterworth [Cutoff = 0.3 Order =5]
Filter Reconstructed slice
Too smooth!
![Page 29: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/29.jpg)
Filter Example: Butterworth
Butterworth [Cutoff = 0.5 Order =5]
Filter Filtered Raw image
Reconstructed slice
Good quality
![Page 30: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/30.jpg)
Image Reconstruction
• Short Axis• Vertical Long Slices• Horizontal Long Slices• Slice Thickness
![Page 31: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/31.jpg)
Transverse Reconstruction
![Page 32: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/32.jpg)
• Initial estimate – e.g. uniform object• Compare forward projection to real projections• Perform several iterations:
– Each time, alter original transverse estimate until pseudo-projections converge with real ones.
The Iterative Loop
Flow chart of a loop in iterative process
![Page 33: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/33.jpg)
Copyright ©2008 American College of Cardiology Foundation. Restrictions may apply.
Sharir, T. et al. J Am Coll Cardiol Img 2008;1:156-163
New MethodsWIDE BEAM RECONSTUCTION
![Page 34: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/34.jpg)
Oblique Reformatting
![Page 35: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/35.jpg)
Image Display
• Cardiac Stress Rest Display Format• Colored and Black & White Display• 3D display• Cine Display• Animation
![Page 36: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/36.jpg)
The use of different colors (as opposed to intensities) in a translation table may produce pseudocontouring: the false perception of an edge when none, in fact, is present. An image thatshows a linear transition from maximal to minimal counts is shown using 3 different color tables. The steady transition of the image is shown most faithfully by the gray scale.
![Page 37: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/37.jpg)
Visual scoring1
0
4
2
4
4
1
2
2
3
4
4
100%
70
50
30
10
0
4 normal 100 - 70%
3 mild 70 - 50%
2 moderate 50 - 30%
1 severe 30 - 10%
0 absent 10 - 0%
Score
![Page 38: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/38.jpg)
Normal Scan
![Page 39: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/39.jpg)
3D Display
![Page 40: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/40.jpg)
Quantitative Display
• Bulls Eye Reconstruction– Cequal– PTQ map
• Quantitative Gated SPECT Display
![Page 41: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/41.jpg)
PTQ Map
![Page 42: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/42.jpg)
![Page 43: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/43.jpg)
3 : no uptake ; defect3 : no uptake ; defect2 : severely decreased uptake2 : severely decreased uptake1 : slightly decreased uptake1 : slightly decreased uptake0 : normal uptake0 : normal uptake
Segmental analysisSegmental analysis
A four-point grading system
A four-point grading system
apicalapical basalbasal
V-long axisV-long axisshort-axisshort-axis
1717
1818
1919
2020
22
3344
11
5566
7788
1111
99
1010
12121313
1414
1515
1616
![Page 44: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/44.jpg)
Quantitative Gated SPECT
![Page 45: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/45.jpg)
QGS
![Page 46: Digital Techniques For Myocardial Perfusion Spect](https://reader035.vdocuments.mx/reader035/viewer/2022062220/5560e31fd8b42a3d768b4bb9/html5/thumbnails/46.jpg)
Thank You