dentall unit
DESCRIPTION
DentistryTRANSCRIPT
-
0197
IMPORTANTThis manual provides operating instructions for the CLESTA. The instructions contained in this booklet should be thoroughly read and
understood before operating the unit. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance.
If you have any questions about this Manual or this product, please contact us. If manual becomes unreadable or is lost, please request a new manual by contacting your dealer.
Installation should be conducted by authorized personnel only. Follow instructions on installation manual.
DENTAL UNIT
INSTALLATION andOPERATING INSTRUCTIONS
-
Page
SAFETYPRECAUTIONS------------------------------------------------- 1
1.GENERALINTRODUCTIONS---------------------------------------- 5
2.INSTALLATIONINSTRUCTIONS----------------------------------- 8 3.OPERATINGINSTRUCTIONS--------------------------------------- 19
4.DAILYMAINTANANCE----------------------------------------------- 23
5.MAINTANANCEANDINSPECTION5-1.GUIDEFORDAILYMAINTENANCEANDINSPECTION---- 27(Maintenanceandinspectionbyuser)5-2.DUIDEFORPERIODICALCHECK-UP------------------------ 29
6.ELECTROMAGNETICCOMPATIBILITY(EMC)----------------- 30
7.LISTOFCOMPATIBLEHANDPIECES----------------------------- 33
TABLE OF CONTENTS
-
Intended Use of the ProductThisproductisanactivetherapeuticdeviceintendedfortheexclusiveusefordiagnoses,treatmentsandrelativeproceduresofdentistry.The product must be operated or handled by the qualified dentists or by dental staffs under
thesupervisionofthedentist.Suchdentistsordentalstaffsshouldinstructand/orassistthepatientstoapproachtoandleavefromtheproduct.Patientsshouldnotbeallowedtooperateorhandletheproductunlesshe/sheissoinstructed.Theproductissuppliedtogetherwiththehandpieceslikeelectricmicromotor,airturbineand/ormotor,scalerandsoon.
Compatibility of HandpiecesUsethecompatiblehandpiecesasshownontheattachedlistforthisunit.(Listofcompatiblehandpieces).
Important Notes In case of the troubles, please contact Takara Belmont offices or your dealers.
Donotdisassembleorattempttorepair. Disassembly, repair or modifications should only be done by a qualified repair technician.
Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.
Disposal of residue material Pleaserequestaspecialcontractorwhenyoudisposeamalgam.
In case of disposal of equipment Whendisposingtheunit,appropriatelydisposecomplyingwithallcurrentapplicableregulationsandlocalcodes.InEUarea,EUdirective2002/96/EConwasteelectricalandelectronicequipment(WEEE)isappliedonthisproduct.Inthisdirective,environmentconsciousrecycling/abandonmentisobligated.
-
EC REP
SN
Protective earth (ground) ON (power) OFF (power)
Chair last position Chair auto return Chair preset 1 Chair preset 2
To raise the chair To recline the backrest To lower the chairTo raise the bacrest
Chairmanual control
Fiber optic handpiece light on/off
Micro motor rotation speed
Scaler power control
Micro Motor Forward/Reverseselect
Syringe Saliva ejector Bowl flush
Cupfiller Service outlet (water) Water Air
Water heater Serial number
Authorizedrepresentative in the European community
Manufacturer Date of manufacture
Non-ionizingradiaion
Separatecollection for electrical and electronicequipment
Type Bapplied parts
alternating current
CautionIt means caution, warnings, or possibility to danger.
Refer to instructionmanual/booklet
Inthismanual,onthelabelsoronthecontrolpanelofCLESTA,followingsymbolsareused.Confirm the meaning of each symbol.
SYMBOLS
-
Beforeuse,readtheSafetyprecautionscarefullytoensureproperuse.Thefollowinginformationisdesignedtoensuresafeuseofthisproductandtopreventinjuryanddamage to you and others. The precautions contained here are classified depending on the severity and degree of
imminenceofpossibleinjuryordamageresultingfromimproperuse.Besuretofollowalltheinformation,whichisimportantforsafety.
SAFETY PRECAUTIONS
Thissymbolindicatesthatitisrecommendedtofollowtheseprecautionsforsafety.
Classification of precautions Severityanddegreeofimminenceofpossibleinjuryordamage
WARNING
CAUTION
NOTICE
Thissymbolindicatesthatignoranceoftheseprecautionsmayleadtosevereinjuryorevendeathasaresultofimproperuse.Thissymbolindicatesthatignoranceoftheseprecautionsmayleadtomildormoderatephysicalinjuryordamagetopropertyasaresultofimproperuse.
WARNING1.BesuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftimeBesuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime(followingthecompletionofwork,duringthesuspensionofbusiness,etc.).Insulationdegradationmaycause electrical fire.
2.BesuretoturnoffthemainswitchuponcompletionofworkorduringworkbreaksBesure to turnoff themainswitchuponcompletionofworkorduringworkbreaks.Thisprevents incorrectoperationduetoaccidentalcontactandassociatedhazards.
4.DonotplaceanundueloadonthearmDonotgetonorplaceanundueloadonthearmofthisunitordentalchairarmrest.Thiscouldcausetheunittotoppleorotheraccidents.5.BesuretoestablishagroundingconnectionBesure toestablishapropergroundingconnection.(Refer toavendorforgroundingconnection.)Failureorelectricleakagemayleadtoelectricshock.6.Neverdisassemble,repairormodifythisproductIndividuals other than certified repair technicians should not disassemble or attempt to repair and modify this product. This could lead to an accident, failure, electric shock or fire.
3.DonotsitonotherthanseatWhenthebackrestisattheforwardposition.donotsitonorplaceanundueloadontheheadrestorlegreatofdentalchair.Thiscouldcausetheunittotoppleorcoulddamagetheunit.
7.UsewithcautioninthepresenceofelectromagneticinterferencewavesDonotplace thisproductaroundequipmentgeneratingelectromagneticwaves(includingcommunicationsequipment,elevators,etc.)asincorrectoperationofthisproductmayoccurinthepresenceofelectromagneticinterferencewaves.Donotuseequipmentgeneratingelectromagneticwaves,suchasmobilephones,aroundthisproduct.
-1-
-
SAFETY PRECAUTIONS
WARNING8.BesuretoturnoffthemainswitchwhenelectrocauteryisinuseBesuretoturnoffthemainswitchwhenelectrocauteryisinuse,becausenoisemaycauseincorrectoperationofthisproduct.
10.Donotplaceobjectsweighing3kgormoreontheDoctor'stable
Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.Refertothesectionofmaintenance.
11.BesuretousethemirrorcoverBesuretousethemirrorcoverofthedentallightwhenthelightisturnedon.Directcontactwithlampsmaycauseburns.SeetheInstructionManualofthedentallightforfurtherinformation.12.Besuretoturnoffthepowerwhenreplacinglamps
9.Ensurethemaintenanceofthisproduct
DonotplaceobjectsweighingthreekilogramsormoreontheDoctor'stable.ThiscouldcausedamagetotheDoctor'stable,defectivefunctionoraccidents.
Besuretoturnoffthepowerwhenreplacingthedentallight.Thiscouldresultinelectricshock.Useonlydedicatedhalogenlamps.Immediatelyafterahalogenlamphasburntout,thelampandthelampholderarestillhot.Replacethelampaftertheycooldown.Donottouchhalogenlampswithbarehands.SeetheInstructionManualofthedentallightforfurtherinformation.13. Immediately wipe off any water spills or leakage on the floor
Immediately wipe off any water spills or leakage on the floor. Decreased strength of the floor may lead tophysicalinjuryincludingfall,orpropertydamage.
14.UsewithcautiononpatientswithacardiacpacemakerUsethisproductwithextremecautiononpatientswithacardiacpacemaker.Inthecaseofanyabnormalitiesinpatientsduringuse,immediatelyturnoffthisproductanddiscontinueuse.
CAUTION1.OnlyexperiencedpersonnelshouldusethisproductOnlydentistsorotherdentalprofessionalsshouldusethisproduct.
2. Confirm safety before use.
Before use, confirm that the parts are correctly and safely operating and that there are no obstacles around this product.
4.Discontinueuseifyoufeelthatsomethingiswrong
Keepyoureyesonpatients(especially,children)sothatmischieforinadvertentoperationofequipmentwillnotleadtounexpectedaccidents.
3.Payattentiontopatientsandchildren
Alwaysbecarefultoinspectthisproductforlooseness,rattling,tilting,wobbling,sounds,temperature,odors,etc. Immediately discontinue use at the first feeling that something is wrong.
-2-
-
SAFETY PRECAUTIONS
CAUTION5.keepyoureyesonthepatientduringoperation.
6.PayattentionduringmovementoftheDoctor'stablePayattentiontosurroundingswhenyoumovetheDoctor'stable.Injurybythetipsofhandpieces,etc.,mayoccur.BesuretomovetheDoctor'stablebyholdingthehandleoftheunit.Besuretomoveup&downtheDoctor'stablebyreleasingthebalancearmbrakeoftheunit.7.DonotplaceanythinghotontheDoctor'stable
12.PrecautionsforcleaningtheresincoverForcleaning,donotusecleaningagentscontainingsolventorabrasives,thinnersoroil-basedalcohol(butanolandisopropylalcohol),whichmaycausecracks.
DonotplaceanythinghotontheDoctor'stable.Thiscouldcausedeformationordiscoloration.
10.DonotsmackorrubthisproductDonotsmackorrubthisproductforcefully.Thiscouldcausedamagetocoversordefectivefunction.
8.Cautionswhenadjustingtheheightofthecart-typeinstrumenttable(Carttypeonly)Afteradjustingtheheight,besuretofastentheloosenedlockscrew.Failuretoheedthiswarningmaycausethetabletodrop,resultinginanaccident.9.Observethecarthose(Carttypeonly)Donottreadonthecarthose.Failuretoheedthiswarningmayresultindamagetothehose,aswellaspeopletrippingover.
11.Precautionsforcleaningaspittoonbowl
Confirm that the patient is seated in the proper position. Keep your eyes on the patient during the operation.Payspecialattentiontosurroundingsatautomaticoperationofthedentaltreatmenttable.Damagetothebackrest,stoolorDoctor'stablemayoccur.
Thespittoonbowlismadeofglass.Handlewithcare.Donotwashitwithhotwater.Otherwise,itmaybreak.
13.ImmediatelywipeoffdrugsolutionwhenitcomesintocontactwiththisunitShoulddrugsolutionorwatercomesintocontactwiththisunit,immediatelywipeitoffwithadrysofttowel,etc.Thiscouldresultindefectivefunctionorelectricleakageaswellasspottingorrusting.
14.ClosethewatermainvalveandturnoffthemainswitchuponcompletionofworkBesuretoclosethewatermainvalveandthemainswitchattheendofeachworkdaytopreventwaterleakagefromoccurring.
16.Payattentionduringtheheadrestoperation
Besuretooperateswitcheswithyourhands,except thefootcontroller,whichisoperatedwithyourfoot.Operationwithbodypartsotherthanhandsmaycausedamageorincorrectoperation.
17.BecarefulnottoheatanemptywaterheaterExercise caution as heating of an empty water heater may result in burning of the heater, leading to fire.
Do not allow hands, fingers, or hair to become entangled in the moving parts of the headrest during operationg
15.Besuretooperateswitcheswithyourhands
-3-
-
SAFETY PRECAUTIONS
CAUTION18.Precautionsforcleaningtheoperationpanel(membraneswitches)
The water tank is intended only for use with purified water, distilled water and pure water. Do not use mouthwash orelectrolyzedwater,suchasConCoolorpovidoneiodine,astheymaycausecloggedtubingoraffectinternalvalvesandequipment.22. Precaautions for insert the bowlflush nozzle and the cupfiller nozzle to the spittoon bowl
Penetrationofdropletsofsanitizingsprayintothebackoftheoperationpanelmaybeassociatedwithswitchfailure.Useapapertowelsoakedwithsanitizingsolutiontocleanthesurfaceoftheoperationpanel.
20.Setthepressureofthewatertankat200kPaorlessAdjust theairsupplypressurefor thewater tankto200kPaorless.Anexcessivelyhighpressuremaycausedamagetothewatertank.
19.Precautionsforcleaning
21. Do not use water other than purified water, distilled water or pure water for the water tank
NOTICE1.TroubleshootingandcontactinformationInthecaseofanyproblems,discontinueuse,turnoffthemainswitchandcontactthedealerorourcompany.
2.CheckoperationofthecompressorWithnoairsupplied,thisproductdoesnotoperateevenafterturningonthemainswitch.Turnonthepowerofthecompressorbeforeoperatingthisproduct.
4.Usetheturbinewithawatercheckvalve
Thewaterunitisintendedforusewithtapwater.Cautionshouldbeexercisedastheuseofwaterotherthantapwater(waterthroughasterilizerofwatersystems,etc.)mayresultinfailureofequipment.
5.HandlingofequipmentinthecaseofapowerfailurePutthehandpieceintheholderandturnoffthemainswitchifequipmentstopsworkingduringuseduetoapowerfailureorotherreasons.
3.Precautionswhenusingwaterotherthantapwater
Usetheturbinewithawatercheckvalve.Contactthedealerorourcompanywhenaturbinewithoutawatercheckvalvewillbeused.
Neverusesandpaper,metalscrubbrushesandabrasivecleaningagentstocleantheunit.Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagentstoavoidcorrosionofmetals,etc.
23.ReadthedocumentsaccompanyingthevariouspiecesofequipmentBeforeuse,besure tocarefullyreadthepackageinsertsandInstructionManualsaccompanyingthevariouspiecesofequipment(includingoptionalarticles)toensureproperuse.
Aftercleaningofthespottoonsection,besuretoinsertthenozzlesstraightlytospittoonbowl.Ifnot,itmaybedamagetheo-ringanditmayleakthewater.
-4-
-
-5-
Bowl Flush Nozzle
Cuspidor Top
Assistant Holder Bar
Saliva Ejector
Assistant Holder
Light Pole
Cupfiller Nozzle
Syringe
High Volume Evacuator
Foot Control
Dental Light
Horizontal Arm
Light Post
Cuspidor Unit
Balance Arm
Dental Size Film Viewer
Doctor Table TrayControl Panel
Handle
Umbilical Hose
Junction Box
Syringe
Waste Receptacle
Handpiece Holder
Handpiece Tubing
Handpieces(Air Turbine, Air Motor, Scaler and etc.)Manufacturers recommend to use the handpieces with CE markings.
[1] GENERAL INTRODUCTIONS(1)UNITTYPEVARIATIONS&MAJORPARTSIDENTIFICATIONS1.UNITMOUNTTYPE
Fig.1-1Overview
-
-6-
Dental Size Film Viewer
Doctor Table Tray
Caster
Cart Base
Cart Hose
Junction Box
Foot Control
Handle
Umbilical Hose
2.SPLITCARTHOLDERTYPEWITHCUSPIDOR
Fig.1-2Overview
-
188
171 287 364
294493
76
529
98
98250
610
52
860
783 65
0
423
521
562
905
330
14
38
180
330
9090
(2)DIMENSIONS
Fig.1-3Dimensions
Ratedpowersupply-------------------------------------------------AC230V50/60HZ0.7AAirmainpressure---------------------------------------------------0.45to0.5MPaWatermainpressure------------------------------------------------0.1to0.2MPaNetweight-----------------------------------------------------------80.5kg(withoutdentallight)Dentallight----------------------------------------------------------AL-501Workingenvironment-----------------------------------------------Temperature:(Operating)+5~+40(Storage)-10~+50Humidity:10~80%Atmosphericpressure:600hPa~1060hPaServiceLife-----------------------------------------------------------10Years
Classification of foot controller ------------------------------------IPX1(applicablestandardIEC60529)Protectionclassagainstelectricshock----------------------------ClassIequipmentAppliedparts---------------------------------------------------------TypeBappliedparts:Handpieceforunit(Listofcompatiblehandpieces)Equipment that is not suitable for use in air, flammable anesthetic gas, oxygen or nitrous oxide
(3)SPECIFICATIONS
-7-
-mm-
-
-8-
220
130
527
85 121
135
11055 70
70
311
243
175
CM/OP MIN.1000 SC MIN.1200
MIN.800
MIN
.600
CENTER OF CLESTA CHAIR
VACUUMDRAIN
VACUUMOPERATINGPOWER
SOURCE
AIR
WATER
[2] INSTALLATION INSTRUCTIONS(1)PREPARINGFORINSTALLATIONTherecommendedlocation,plumbingandlayoutofjunctionboxareshowninFug.2-1,2-2and2-3respectively.1.LOCATIONANDPLUMBNGLAYOUTOFJUNCTIONBOX
Fig.2-1LocationandPlumbingLayout
StandardType
-
-9-
22013
0
527
80 121
140
110
7070
311
243
175
CM/OP MIN.1000 SC MIN.1200
MIN.800M
IN.6
00
VACUUM
DRAIN
VACUUMOPERATING
POWER SOURCE
AIR
WATER
CENTER OF CLESTA CHAIR
BelmontType
NOTE:CM/OP.....CHAIRMOUNTOVERTHEPATIENTTYPESC............SPLITCARTTYPE
Fig.2-2LocationandPlumbingLayout
-
-10-
40
20 30
80
F
A
D EG
B
C
VACUUM DRAIN POWER SOURCE &VACUUM OPERATING
AIR & WATER
2.PLUMBINGHEIGHTOFJUNCTUONBOX
Fig.2-3PlumbingHeight
A:NIPPLEB:SOCKETJOINTC:AIR
WATER
D:DRAIN
E:POWERSOURCE
SUPPLYCABLEEARTHWIREF:VACUUM
G:VACUUMOPERATING
;PT1/2-1/2(Brass);UnplasticiziedPolyvinylChloridePipeFitting;UnplasticiziedPolyvinylChloridePipe(ShockResistingType)'HI13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipeforWaterWorksService'VP13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipe'VP40'48mmO.D.-40mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.;PolyvinylChlorideInsulatedandSheathedCable'VVF2.0-2C';PolyvinylChlorideInsulatedWire'VE1.6';UnplasticiziedPolyvinylChloridePipe'VP20'26mmO.D.-20mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.
-
-11-
1.CHECKTHEPACKINGCheckeachquantityofthepartsforinstallationaslistedbelow. *Standard Specification
1)SHUTOFFVALVESFORAIR&WATER 1SET2)WOODSCREW 4PCS.3)FLEXIBLEPIPE 2PCS.4)DRAINELBOW 1PCE.5)VACUUMELBOW 1PCE.6)DRAINCAP 1PCE.7)BASKETSTRAINER 1PCE.8)CUPFILLERNOZZLE 1PCE.9)BOWLFLUSHNOZZLE 1PCE.10)SALIVAEJECTORTIP 1PCE.11)VACUUMTIPS(BENT&STRAIGHT) 1SET12)SILICONTIP 10PCS.13)LEVELADJUSTMENTSCREW 4PCS.14)WASTERECEPTACLE 1PCE.15)LIGHTPOLE 1SET16)SILICONMATFORDR.TABLE 1PCE.17)WIRECONNECTOR 3PCS.18)SEAL 4PCS.19)PATIENTCUP 1PCE.20)VIALSWITHSTAND 1SET21)COTTONCONTAINERWITHSTAND 1SET21)MOUNTINGBRACKETCOVER 1PCE.
2.ATTACHUNITTOCHAIR2-1.MountingbracketAType(Fig.2-4)1)Raisetheupperstructureofdentalchairtothehighestposition.2)InstallthecuspidorunitonMountingBracketATypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.3)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.4)SticktheSealsovertheFlatHeadScrews.5)Removethestopperscrewfromthemountingbracket.6)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopper
screw.
[2] INSTALLATION PROCEDURES
Fig.2-4AttachUnittoChair
Flat Head Screw
MountingBracket A
Level Adjustment Screw
Cuspidor Unit
NutSpring Washer
Mounting Bracket
Seal
StopperScrew
Mounting BracketCover
StopperScrew
-
-12-
Dental LightPlastic Wahser
Light Pole
Set Screw
Cover
Cable from Dental Light
Cable from Cuspidor Unit
2-2.MountingBracketBType(Option)(Fig.2-5)1)Raisetheupperstructureofthedentalchairtothehighestposition.2)AttachtheAdaptortoMountingBracketBTypewith4pcs.oftheM10-40SocketHeadCapScrewsandSpringWashers.3)InstallthecuspidorunitonMountingBracketBTypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.4)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.5)SticktheSealsovertheFlatHeadScrews.6)Removethestopperscrewfromthemountingbracket.7)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopper
screw.
Fig.2-5AttachUnittoChair
Fig.2-6InstallDentalLight
3.INSTALLDENTALLIGHT(Fig.2-6)1)RemoveCoverfromthebottomoftheunit2)PutPlasticWasherontheLightPole.3)RuntheCablethroughLightPoleandattachtheDentalLighttotheLightPole.4)RuntheCablethroughLightPostandinstallthePole,thentighten2pcs.oftheSetScrews.5)ConnectthewiresfromtheDentalLightandCuspidorUnitinaccordancewiththe identification color.
6)Re-attachtheCover.
Flat Head Screw
Socket Head ScrewSpringWasher
Mounting Bracket BLevel Adjustment Screw
Nut
Adaptor
Cuspidor Unit
Spring Washer
Seal
Mounting Bracket
Stopper
Mounting BracketCover
Stopper
-
1 2 3 4 4 5 5 6 7 8
1 2 3 4 4 5 5 6 7 8
4 5
4 5
L
C C
L
Umbilical Hose
Umbilical Hose Clamp
Brown 1/8Blue 5/16
Yellow 1/4
Light Green 1/4Wirings for Water Heater & Dental Light
White Black
White Black
Brown 1/8Blue 1/16
Yellow 1/4
Light Green 1/4
Wirings for Water Heater
Wirings for Dental Light
Blue Brown
BlueBrown
24V Line Voltage
Small thin Screw Driver
Electric Terminal
Wiring
-13-
4.CONNECTUMBILICALHOSEWITHJUNCTIONBOX(Fig.2-7,2-8&2-9)ThissectionisonlyforSPLITCARTTYPE.Forothertypeofunit,proceedto5.1)OpenthecoverofJunctionBoxbylooseningtwoscrews.2)LoosenthescrewofUmbilicalHoseClamp.3)Insertallhoses,tubingsandwiringsofUmbilicalHosefromCuspidorunitintoJunctionBoxthroughtheUmbilicalHoseClamp.4)CorrectthetwistoftheUmbilicalHoseastheJunctionBoxisplacedattheplannedlocation.5)SettheendofUmbilicalHoseattheUmbilicalHoseClampthenre-tightenthescrewfortheUmbilicalHoseClamp.6)Connecttubingswithnylonsleevestothe fittings as shown in Fig.2-8 or 2-9.
7)ConnectwiringstoElectricTerminalasshowninFig.2-8or2-9.
*HowtoconnectwiringtotheElectricTerminal(Fig.2-10)DuringpressingtheslotoftheElectricTerminalbytheSmallThinScrewDriver,insertthewiringtotheElectricTerminal.
Fig.2-7ConnectUmbilicalHose
Fig.2-8 Standard Specification
Fig.2-9 Low Voltage SpecificationFig.2-10HowtoConnectWiring
-
-14-
Standard Type Belmont Type
Center of Clesta Chair Center of Clesta Chair
Water Shut Off Valve
Air Shut Off Valve
Water Shut Off Valve
Air Shut Off Valve
ElectricalReceptacle
Shut Off Valve
Nipple
ElectricalReceptacle
Wood ScrewVacuum Drain Hose
Drain Hose
Drain Elbow
Vacuum Elbow
Vacuum Drain HoseVacuum Elbow Drain HoseDrain Elbow
Vacuum Pipe Drain Pipe
Drain Bush (Option)
Drain Joint (Option)
Drain Bush (Option)
5.CONNECTUTILITYSECTION(Fig.2-11&2-12)1)AttachShutOffValvestotheairandwatersupplylines,pointingtoproperdirectionsforfurtherconnections.2)AttachElectricalReceptacletoMainPowerSupplyCable.(TheElectricalReceptaclemustbepreparedlocally.)3)OpeneachShutOffValvetoremoveanychipsordirtintheairandwatersupplyline,thenclosetheValves.4)OpenthecoverofJunctionBoxbylooseningtwoscrews,andplaceJunction Box at the planned location, then fix it on
the floor with 4pcs. of Wood Screws.
5)CuttheDrainHoseandtheVacuumDrainHoseattheproperlength,andconnectandgluethemtoeachelbow.6)ConnecttheVacuumElbowtotheSuctionPipe.7)InserttheDrainElbowintotheDrainPipe.
*AttachingDrainBushing(Option)totheDrainPipeisdesirable,andDrainElbow should be bonded firmly into the Drain Pipe.
* For Air Vacuum specification, prepare Drain
Joint(Option)asshowninFig.2-13,sothatthedrainlineandthevacuumdrainlinemaybeinsertedtogetherintotheDrainPipe.
Fig.2-11ConnectSupplyLines
Fig.2-12ConnectDrainLines(1)
Fig.2-13ConnectDrainLines(2)
-
-15-
Shut Off ValvePacking
Filter
Filter
Standard Type Belmont Type
Air Filter Air FilterWater Filter Water Filter
1 2 3 4 4 5 5 6 7 8
1 2 3 4 4 5 5 6 7 8
Cable for Chair Control (9P)
Cable for Chair Safety Device (4P)
Cable from Chair
Main Power Supply Cable
VacuumOperatingCable (9P)
Plug
VacuumOperating Cable
Cable for Chair Control (9P)
Cable for Chair Safety Device (4P)
Cable from Chair
Main Power Supply Cable
VacuumOperatingCable (9P)
Plug
VacuumOperating Cable
SeeFig.2-14&2-15.8)ConnectWaterShutOffValveandWaterFilterwithStainlessFlexiblePipeandPacking.ConnectAirShutOffValveandAirFilterinthesameway.
SeeFig.2-16or2-17.9)PreparetheElectricalPlug,andconnectittoMainPowerSupplyCable.(ElectricalPlugmustbepreparedlocally.)10)ConnectVacuumOperatingCabletotheElectricTerminalasshowninFig.2-16or2-17. (For Central Vacuum Specification only)
11)Runthethreecablesfromchair(PowerSupplyCablewithplug,Cablewith9PconnectorandCablewith4Pconnector),throughtheholeoftheJunctionBoxwallabovetheUmbilicalHoseintotheJunctionBox.12)ConnectthePowerSupplyCableofthechairtothereceptacle,andconnectothertwocablestothecorrespondingconnectors.
Fig.2-14ConnectFilters(1)
Fig.2-15ConnectFilters(2)
Fig.2-16ConnectElectricalcable(1)
Fig.2-17ConnectElectricalcable(2)
-
-16-
6.ATTACHFILMVIEWER(Fig.2-18&2-19)1)RemovetheDoctorTableTraybylooseningeightFlatHeadScrews.2)RunthecablefromFilmViewerintoDoctorTable.3)AttachFilmViewertothebracketforDoctorTablewithtwoPhilipsScrews.4)ConnectthecablefromFilmViewertotheElectricalTerminal.
7)SETUPCUSIPDORTOP(Fig.2-20)1) Insert Cupfiller Nozzle and Bowl Flush
Nozzle,andtighteneachnut.2)PutBasketStrainerandDrainCapintothecuspidorbowl.
Flat Head Screw
Philips Screw
DoctorTableTary
ElectricalTerminal
FilmViewer
Electrical Terminal
Cupfiller NozzleBowl Flush Nozzle
Drain Cap
Basket Strainer
Fig.2-18AttachFilmViewer
Fig.2-19ConnectWiringsforFilmViewer
Fig.2-20CuspidorTop
-
-17-
(3)ADJUSTMENTBEFOREUSE1.AIR&WATERSUPPLYLINE(Fig.2-21&2-22)1)ChecktheAirPressureGaugeonthesideofJunctionBox.ThePressureshouldbeapproximately0.45Mpato0.5Mpaforair.Foradjustmentthepressure,pullupandturntheknobofAirRegulatorwhiletheair is flowing (use 3-way Syringe for example).
Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise. After adjustment, push down the knob to fix it.
2)CheckiftheWaterPressureGaugeindicatesthepressureataround0.1Mpato0.2Mpa.AdjustmentcanbedonebyWaterRegulator.Toadjustthepressure,liftandturnthewater regulator knob until the water starts flowing.
(use3-waySyringeforexample).Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise.Afteradjustment,lowertheknobbackinplace.
3)Astheunitisreadytorun,turnontheMasterSwitchlocatedunderneaththedoctortable(seeFig.2-23.),thenrunthesalivaejectorfor a while to flow the water through the unit and
makesurenoleakagetakesplaceattheutilitysection.SeeFig.2-23&26.4)TurnallCoolantWaterControlKnobscounter-clockwisefullyandpickupallhandpiecetubingfromtheholder,thendepressDrivePedalofthefootcontrol(SpraySwitch-ON)foramoment to flush out the water from the tubing into the
cuspidorbowl.
Pull Up
Push DownKnob
Increase decrease
Air Regulator
Water RegulatorIncrease
decrease
Knob
Pull UpPush Down
Water Air
Air Pressure Gauge
Water Pressure Gauge
Master Switch
ONOFF
Coolant WaterControl Knob
Increase decrease
Fig.2-21InsideofJunctionBox
Fig.2-22SideViewofJunctionBox
Fig.2-23BottomViewofDoctorTable
-
-18-
2.DriveAir&SprayforHandpieces(Fig.2-24,2-25&2-26)1)Connecthandpiecestoeachtubing,andmakeadjustmentonebyone.2)Toadjustthedriveairpressureforeachhandpiece,turntheappropriateAdjustScrewforDriveAironetheAuto-SelectValveinthedoctortable.Turnthescrewclockwisetoclosethelinefully.ThenduringdepressingDrivePedalofFootControlfully,graduallyincrease the sir flow by turning the screw counter-
clockwiseuptoHandpiecePressureGaugereadsthefullrecommendedpressure.3)Toadjustsprayofhandpiece,useAdjustScrewforCoolantAirontheAuto-SelectValveandCoolantWaterControlKnob.
Flat Head Screw
DoctorTableTray
Auto-SelectValve
PressureGauge
Adjust Screw for Drive Air
Adjust Screw for Coolant Air
Drive Pedal
Fig.2-24InsideofDoctorTable
Fig.2-25Auto-SelectValve
Fig.2-26FootControl
-
-19-
(1)DOCTORUNITSECTION1.MASTERSWITCH(Fig.3-1&3-2)1)TurnontheMasterSwitchandthePowerIndicatorislitingreen.2)Aftertheoperation,turnofftheMasterSwitch,thenthesuppliesofair,waterandelectricityareallshutdownattheutilitysection.
2.TABLEPOSITION(Fig.3-1)Forstandardmodelofthechairmountoverpatienttype,theheightofthedoctortableisalways fixed by Air Brake. The height can
beeasilychangedbypressingBrakeReleaseButton.
3.CHAIRCONTROL(Fig.3-3)ThechairmovementcanbecontrolledbypressingChairControlonMembraneSwitchA.Pleaserefertooperatinginstructionsforchair.
4.CUPLILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-3)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds at
factory.For bowl flush flushing time is set for 6 seconds
atfactory.
5.3-WAYSYRINGE(Fig.3-2&3-4)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedunderneaththedoctortable.
[3] OPERATION INSTRUCTIONSFilm ViewerPower Indicator
Membrane Switch A
Brake Release Button
Membrane Switch B
Handpiece Holder
Master Switch ONOFFAir
Water
Flow Control Knobfor 3-Way Syringe
Coolant Water Control Knob
2
LP0
1
Cupfiller Switch
Chair Manual Control
Chair Automatic Control
Bowl Flush Switch
w A
Air Button
Water Button
Syringe Nozzle
Fig.3-1DoctorUnit
Fig.3-2BottomView
Fig.3-3MembraneSwitchA
Fig.3-43-WaySyringe
-
-20-
6.OPERATINGHANDPIECE(Fig.3-5)1)Pickupahandpiecefromthepandpieceholder,anddepressDrivePedalofFootControl.2)Forthedry-cutting,turnoffSpraySwitchofFootControl.3)WhiledepressingChipPedal,onlythechipblowcomesoutfromthehandpiece.
7.FILMVIEWER(Fig.3-6)FilmViewerSwitchislocatedontherightside of film viewer. Press the switch once to
illuminatetheviewer,pressitagaintoturnitoff.
8.FIBEROPTICHANDPIECE(OPTION)(Fig.3-7)FiberOpticHandpieceSwitchislocatedontheMembraneSwitchB.Pickupaoptichandpiecefromthehandpieceholderandthehandpiceislit.Returnthehandpiecetotheholderanditisputout.
9.ELECTRICSCALER(OPTION)(Fig.3-7)PowerControlforElectricScalerislocatedundertheMembraneSwitchB.
ON
OFFCoolant Water Switch
Drive Air Pedal
Chip Blower Button
Film Viewer Switch
Fig.3-5FootControl
Fig.3-6FilmViewer
Fig.3-7MembraneSwitchB
1
0
2 3
4
5
Fiber Optic Handpiece Switch
Power Control for Electric Scaler
HandpiecesRefertohandpiecemanufacturersoperatinginstructions.
CAUTION
-
-21-
(2)ASSISTANTUNITSECTION1.CUPFILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-8)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds a
factory.For bowl flush. Flushing time is set for 5 seconds
atfactory.
2.3-WAYSYRINGE(Fig.3-4&3-9)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedinthecuspidorbody.
3.MASTERVALVEBLOCK(Fig.3-9)Openthesidepanelofthecuspidorbody.Flow Control knobs for Cupfiller and
BowlFlushareaccessible.
4.CUSPIDORBODYROTATION&TELESCOPICASSISTANTHOLDERBAR(Fig.3-10)1)CuspidorBodycanrotateby90degreesforassistantsconvenience.2)Fortwo-handstreatment,AssistantHolderBarcanbeextendedby50cm.
5.ASSISTANTINSTRUMENTHOLDER(Fig.3-10)Pickuptheinstrument(SalivaEjectororHighVolumeEvacuator)fromtheholder,anditstartsworking.
Bowl Flush Switch
Cupfiller Switch
Cupfiller Flow Control Knob
Bowl Flush Flow Control Knob
Air Flow Control Knob for 3-Way Syringe
Water Flow Control Knob for 3-Way Syringe
Air Vacuum Flow Control Knob
Increase decrease
TelescopicAssistantHolder bar
AssistantHolder
Fig.3-8MembraneSwitchforAssistant
Fig.3-9MasterValveBlock
Fig.3-10CuspidorUnit
-
-22-
Water Heater Switch
Bowl Flush Switch
Service Water OutletFlow Control Knob for Service Water Outlet
- +Cup Base
MANUAL SENSOR
H
M LH
Touchless Switch
6.WATERHEATERSWITCH(Fig.3-11)TurnontheWaterHeaterSwitchwhenthewarmwaterforthepatientcupisrequired.
7.SERVICEWATEROUTLETWITHFLOWCONTROL(Fig.3-11)WaterOutletwithFlowControlislocatedonthepanelwhichadaptstothequickconnector.
8.BOWLFLUSHSWITCH(OPTION)(Fig.3-11)TurnontheBowlFlushSwitchandthewaterisflushed to the bowl continuously.
9.AUTOMATICCUPFILLER&BOWLFLUSH(OPTION)(Fig.3-12)1)Placethepatientcupweighting45to80gonCupBaseandwaterissuppliedfora certain quantity to the patient cup and flushed
foracertaintimetothebowl.2)QuantityofwaterinpatientcupcanbeadjustedbyturningCupBase.
(3)LIGHTSECTION(MODELIO5000)1.MODESELETIONSWITCH(Fig.3-13)TurntheswitchtoSENSORside,andON/OFFofthelightcanbemadebyapproachinghandofthesensor.(TouchlessSwitch).
2.INTENSITYSWITCH(Fig.3-14)ThenintensityoflightcanbeselectedL(Low),M(Medium)orH(High).RefertothemanualofIO5000.
Fig.3-11CuspidorControlPanel
Fig.3-12 Automatic Cupfiller
Fig.3-13ModeSelectionSwitch
Fig.3-14IntensitySwitch
-
-23-
Drain Cap
BasketStrainer
Solid Collector
Solid Collector Filter
(1)DRAIN&SUCTIONLINEOFUNIT(Fig.4-1)1.BasketStrainer.RemovetheBasketStrainerCap,andcleantheBasketStrainer.2.SolidCollectorFilterOpenthesidepanelofthecuspidorbodyandtakeoutSolidCollectorFilterandcleanit.PullingontheSolidCollectorCapwillbringtheSolidCollectorFilterwithit.
(2)AIRSUPPLYLINE(Fig.4-2)1.AirFilterDrainValve 1) Release the water in the air filter by opening
DrainValveofJunctionBox.2)Turnoffthemainswitchofthecompressor.Releasecompressedairanddrainaccordingtotheinstructionsofthecompressor. Confirm the air Pressure Gauge indicates
0Mpa.
[4] DAILY MAINTENANCE AND CARE
Drain Valve
Air Pressure Gauge
Fig.4-1BasketStrainer&SolidCollector
Fig.4-2JunctionBox
Fig.4-3VacuumHndpiece&SalivaEjectorHandpiece
Vacuum Tip
Saliva EjectorHandpiece
VacuumHandpiece
Vacuum
Vacuum Cap
Saliva EjectorHose
Body
Body
Hose
Slide Knob Slide Knob
SalivaEjectorCap
Hose Connector
Filter
(3)HANDPIECE1-1.VacuumHandpieceandSalivaEjectorHandpieceForeffectivesterilization,washingforremovingcontaminationandimmersionbyacleaningagentarerequired.Then,rinsebywaterinordertoremoveresidualcleaningagentonmedicaldevice.Takefollowingproceduresfromcleaningtosterilization.*Usethedisposablesalivaejectortip.
Disassembly (Fig.4-3)Disassemblethehandpieceforthepreparationof cleaning as the following figures show.
Pullthehoseconnectortodisconnectthevacuumhose.
-
CAUTION
Sterilization pouch
Fig.4-7VacuumHandpieceandSalivaEjectorHandpiece
-24-
Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat405degrees.(Fig.4-4) Scrub the intubation or hole, slide groove and filter by a
cleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat405degrees.(Fig.4-5)Wipeoffbyaclothfortheareawhichbrushisunabletoreach.(Fig.4-6)
B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.
Sterilization (Fig.4-7)VacuumTip/VacuumCap/VacuumHandpieceBody/SalivaEjectorHandpieceBodycanbeautoclave.Vacuumhandpiecebodyandsalivaejectorbodyhavetoassemblebeforeautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121 or 4 min. at 135.
Sterilizationwithautoclaveispermittedupto250times.
Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece forfollowing cases. Any waste material can not be removed by clogged hole. Contamination and solid material attached to handpiece can not be removed.
StorageAftercleaningthevacuumtip,keepitinthecleanplace.
Sterilization by class B cycles. Sterilization temperature is 135 degrees or less. The cap, filter and body are made of resin. They may become deteriorated if they are sterilized in an autoclave many times. After autoclave sterilization, the cap, filter, body and valve are subject to discoloration, which does not have a negative effect on performance. The slide knob can be autoclave 100 times and is expendable supplies. Skip the drying process if the temperature is to exceed 135. If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch.
CAUTION
Cloth
Fitting cleaning brushor Tooth brush
Slide groove
Cloth
Fig.4-4Surfacecleaning
Fig.4-5Cleaningofholeandslidingpart
Fig.4-6Cleaningthearea(brushisunabletoreach)
-
-25-
2.VacuumHose/SalivaEjectorHose(Fig.4-9)Itispracticaltoremovethehosesinordertocleanthem.Note:Attheendoftheday,cleanthevacuumandsalivaejectorhandpiecesbysuckingintwocupsworthofwater.
3.AIRMOTOR/TURBINE/3-WAYSYRINGESterilizethehandpieceaccordingtomanufacturersoperatingmanual.
VacuumHose
SalivaEjectorHose
Fig.4-9VacuumHose&SalivaEjectorHose
CAUTION
4.Belmont77Syringe(Fig.4-10)DisassemblyRemovethenozzlefromsyringebyturningitindirectionA.
Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat405degrees. Scrubthetipandjointpartofnozzlebyacleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat405degrees.(Fig.4-11)B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.
Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece for following cases. Contamination and solid material attached to handpiece can not be removed.
1-2.Cleaningvacuumandsalivaejectorlines(Fig.4-8)Thesuckingunitcomesintocontactwithsecretions,spitandbloodthatcontainbacteriaeveryday.Besuretocleanandsterilizeitattheendofeachworkday.Recommendedcleaner:OrotolPlusmadebyDURR.
Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagents,whichmaycausecorrosionofmetals,etc/
CAUTION Fig.4-8CleaningVacuum/SalivaElectorLines
Fig.4-11CleaningBelmont77SyringeNozzle
A
B
w A
Nozzle
Nut
Syringe body
Fig.4-10Belmont77Syringe
Cleaning brushor Tooth brush
-
(8)FILTERREPLACEMENT(Fig.4-13)The water filter in the junction box needs to bereplacedatleastonceayear.The air filter in the junction box needs to bereplacedatleastonceeverythreeyears.Contactyourlocalservicerepresentativeforreplacement.
-26-
Water Filter
Air Filter
Fig.4-13Filter
(6)LUBRICATINGHANDPIECESRefertothemanualofthehandpieces.
(4)HOSESANDTUBINGCleanhosesandtubingusingDURRFD333.
(5)CLEANINGUNITAllsurfacescanbecleanedwithDURRFD333cleaner.Spraythecleaner(FD333)onclothandwipethesurfaceswiththecloth.
CAUTIONDonotdrenchtheunit.Wipeallsurfacesdryaftercleaning.
StorageAftercleaningthenozzle,keepitinthecleanplace.
Sterilization by class B cycles. Sterilization temperature is 135 degrees or less. Skip the drying process if the temperature is to exceed 135. If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch. Before use, make sure that the nut is firmly tightened.
Sterilization (Fig.4-12)Thenozzlecanbesterilizedwithautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121 or 4 min. at 135.Sterilizationwithautoclaveispermittedupto250times.ApplyVaselinethinlyandevenlytothetwoO-ringsaftersterilization. Tighten the nut firmly in direction B to undo the nozzle.
CAUTIONFig.4-12SterilizationBelmont77SyringeNozzle
Sterilization pouch
(7)MASTERSWITCHAfterthedailyoperationandmaintenance,turnofftheMasterSwitch.
-
[5] MAINTANANCE AND INSPECTION5-1. Guide for daily maintenance and inspection (Maintenance and inspection by user)Managementofmaintenanceandinspectionofmedicalequipmentshouldbeimplementedbytheuser(medicalinstitution).Incasetheuserdoesnotimplementsuchmanagement,itispermitted that such management is outsourced to a qualified entity such as a medical equipment repaircompany. For safe use of this product, it is necessary that inspection should be conducted in the specified frequencyonthetemsdescribedbelow.
Air turbine revolution, water flow, air flow and so forth shall be free of abnormality.Micromotor revolution, water flow and so forth shall be free of abnormality. Scaler vibration, water flow and so forth shall be free of abnormality.
No. Item Frequency Inspection method and diagnosis
Influence if inspection not conducted
Maintenance required in case of nonconformity
Check of safety functions
Before start
Before start
Before start
Before start
For each patient
For each patient
For each patient
After closing
Unexpected personal injury and troubles may arise due to motion of the chair during medical treatment and due to pinching between doctor section and chair.
Contact your dealer or our office if any abnormality arises.
Check for leakage of water and air
Leakage of water and air shall not be observed around the product.
The product will not normally work, and troubles may arise.
Contact your dealer or our office if any abnormality arises.
Cupfiller When a paper cup is placed on the cupfiller, the cup shall be detected and cupfilling shall be executed.* Malfunction may arise if the cup is of another material grade (such as stainless steel and plastics) or if the paper cup is of dark color or pattern.
Cupfilling may not be executed.
Execute re-inspection in accor-dance with "Method for operation" described in the instruction manual. Contact your dealer or our office if recovery is not achieved as a result of re-inspection.
Check of motions of equipment
Troubles such as injury in patient's oral cavity and equipment failure may arise.
Control the water flow in accor-dance with "Control of compo-nents" described in the instruction manual. If any other abnormality arises, refer to the instruction manual attached to individual equipment. Contact your dealer or our office if recovery is not achieved.
Check of air turbine bar
Appropriate bar shall be positively mounted.Make sure to refer to the instruction manual attached to individual equipment.
The bar will not normally work and troubles may arise.
If abnormality such as flaw and deformation is found on the bar, replace the bar in accordance with the instruction manual attached to individual equipment.
Check of scaler tip
Appropriate tip shall be positively mounted and be correctly used.Make sure to refer to the instruction manual attached to the scaler.
The tip will not normally work and troubles may arise.
If the tip was worn or deformed, replace the tip in accordance with the instruction manual attached to the scaler.Contact your dealer or our office if any other trouble arises.
Check of tightness of syringe nut
The nut for fixing the nut of Type 77, 3-way syringe shall be positively tightened.
Troubles may arise if the nut comes off.
Turn and positively retighten the nut that fixes the nozzle.
Matters attached to micromotor
Excessive handpiece oil or the like shall not be attached to the motor section.
The motor section will not work normally and troubles may arise.
Execute care in accordance with the instruction manual attached to individual micromotor.
Make sure the chair movement stops by any of the following actions.when foot controller pedal is depressed. During chair auto movement, depress of any operation switch. While setup is in progress with function switch on the doctor membrane switch panel.When the spittoon bowl is turned to patient side (Pedestal type)
-27-
-
No. Item Frequency Inspection method and diagnosis
Influence if inspection not conducted
Maintenance required in case of nonconformity
After closing
CareVacuum and saliva ejector handpiece
Flush the suction line, and then clean the filter of the vacuum or saliva ejector handpiece.
Faulty suction may arise.
Clean the suction line and filter in accordance with "Method for care" described in the instruction manual.
9
CareCuspidor section
Clean the cuspidor and dust filter.
After closing
Faulty water drainage may arise.
Execute dust removal and cleaning in accordance with "Method for care" described in the instruction manual.
CareSolid collector
Clean the filter of the solid collector.
Vacuum suction will become weak.
Clean the filter in accordance with "Method for care" described in the instruction manual.
After closing
CareExterior
Chemical, filthy water and so forth shall not be found (attached or remaining) on the product exterior.
Discoloration and deterioration to the exterior, and corrosion and rusting to metallic components may arise.
Execute wiping in accordance with "Method for care" described in the instruction manual.
Check of main switch and main valves
The product main switch shall be off, and water/air main valves shall be closed.
Product failure and troubles may arise.
Contact your dealer or our office if the main switch cannot be turned off or if the main valve cannot be closed.
Product's moving parts
No abnormal noise or the like shall be produced from product's moving parts when the product is operated.
The product will not normally work and troubles may arise.
Contact your dealer or our office if any abnormality arises.
CareDrain valve
Drain the water from the air filter drain valve.
Water may enter the air line, and equipment failure may arise.
Drain the water from the air filter in accordance with "Method for care" described in the instruction manual.
Check of water pressure and air pressure
Check the water pressure and air pressure by reading pressure gauges in the U-Box section.Basic set pressure : Water: 0.1 to 0.2 MPaAir: 0.45 to 0.5 MPa
The product will not normally work, and troubles may arise.
Contact your dealer or our office if the set pressure is abnormally high or low.
Check of conditions of table section
The table shall be free of inclination, and water shall not flow on the table.
Injury caused by falling of goods located on the table and other troubles may arise.
Contact your dealer or our office if any abnormality arises.
Care Oil mist separator
The oil level in the oil mist separator shall be lower than the red line.
Normal output will not be produced due to inferior handpiece exhaust.
Discharge the oil in accordance with "Method for care" described in the instruction manual.
After closing
After closing
10
11
12
Once every week
13
14
15
16
17
18
Once every week
Once every month
Once every month
Once every month
-28-
-
5-2. Guide for Periodical Check-upSomepartsandcomponentsoftheproductsaredegradedordeteriorateddependingonthe
frequencyofuse.Annualcheck-upandmaintenance,aswellasreplacementofconsumableparts,arerequired.Therequiredparts(includingconsumableparts)arelistedbelow.Itmaybedifferentfromthe
followinglistdependingontheoptionoftheunit.Forcheck-upandrepair,callatechnicianofourauthorizeddealer.
Parts and components that require periodical check-up
Parts Description
Vacuum handpiece body
Saliva ejector handpiece body
Foot controller
Water supply hose
Drain hose
Air supply hose
Electric wiring of moving part
Standard LifetimeNo.
3 years 1
2 3 years
3 5 years
3 years
3
4
4
5 3 years
Consumable parts
No. Parts Description
1
2
Parts Description Standard LifetimeNo.
3 years 8
9 3 years
10 5 years
5 years11
12 3 years
Regulator
Valves
Switches
Film viewer body part
Pressure gauge
Arm section of moving part
Control PCBs.
6 3 years 13 7 years
7 5 years 14 5 years
Valve for vacum handpiece body
Vacuum tip
Handpiece tubings
Vacuum hose
Saliva ejector hose
8
No. Parts Description
6
7
Filter for oil mist separator
Filter (Air & Water)
O-ring, Packing, Diaphragm
5
Executethemaintenanceinaccordancewiththisinstractionmanualandoperatingmanualattachedtoeachindividualequipment(Dentallight,Handpiece,etc..).Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.
WARNING
-29-
-
[6] ELECTROMAGNETIC COMPATIBILITY(EMC)MedicalelectricalequipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledandputintoserviceaccordingtotheEMCinformationprovidedinthismanual.PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.Theequipmentorsystemshouldnotbeusedadjacenttoorstackedwithotherequipment.Ifadjacentorstackeduseisnecessary, theequipmentorsystemshouldbeobservedtoverifynormaloperationintheconfiguration in which it will be used.
Guidance and manufactures declaration electromagnetic emissionsThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.
Emissions test Compliance Electromagnetic environment - guidanceRFemissionsCISPR11 Group1
TheCLESTAUnitusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
RFemissionsCISPR11 ClassB
TheCLESTAUnitissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
HarmonicemissionsIEC61000-3-2 ClassAVoltage fluctuations/FlickeremissionsIEC61000-3-3
Complies
Guidance and manufactures declaration electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.
Immunity test IEC 60601test level Compliance levelElectromagnetic environment-
guidanceElectrostaticdischarge(ESD)IEC61000-4-2
6kVcontact8kVair
6kVcontact8kVair
Floorsshouldbewood,concreteorceramic file. If floors are covered withsyntheticmaterial,therelativehumidityshouldbeatleast30%.
Electricalfasttransient/burstIEC61000-4-4
2kVforpowersupplylines1kVforinput/outputlines
2kVforpowersupplylines1kVforinput/outputlines
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
SurgeIEC61000-4-5
1kVdifferentialmode2kVcommonmode
1kVdifferentialmode2kVcommonmode
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.
Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlinesIEC61000-4-11
95%dipinUT)for0.5cycle40%UT(60%dipinUT)for5cycle70%UT(30%dipinUT)for25cycle95%dipinUT)for5s
95%dipinUT)for0.5cycle40%UT(60%dipinUT)for5cycle70%UT(30%dipinUT)for25cycle95%dipinUT)for5s
Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.IftheuseroftheCLESTAUnitrequirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthattheCLESTAUnitbepoweredfromanuninterruptiblepowersupplyorabattery.
Powerfrequency(50/60Hz)magnetic fieldIEC61000-4-8
3A/m 3A/m Power frequency magnetic fields shouldbeatlevelscharacteristicofatypicallocationinatypicalcommercialorhospitalenvironment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationsofthetestlevel.
-30-
-
Guidance and manufactures declaration electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.Immunity test IEC 60601 test level Compliance level Electromagnetic environment -guidance
PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartoftheCLESTAUnit,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicationstotheFrequencyofthetransmitter.
Recommended separation distance
ConductedRFIEC61000-4-6
3Vrms150kHzto80MHzoutsideISMbandsa
3Vrms d = 1.2P
RadiatedRFIEC61000-4-3
3V/m80MHzto2.5GHz
3V/m d = 1.2P 80MHzto800MHzd = 2.3P 800MHzto2.5GHz
WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmetres(m).
Field strengths from fixed RF transmitters, as determinedbyanelectromagneticsitesurvey,ashouldbelessthanthecompliancelevelineachfrequencyrange.b
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:
NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedby adsorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CLESTA Unit isusedexceedstheapplicableRFcompliancelevelabove,theCLESTAUnitshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingtheCLESTAUnit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Essential performance (purpose of IMMUNITY testing)Unless operated by the switches for chair control, the chair connected to the CLESTA does not make any movements, except for sounding a buzzer and switching on/off the indicator.
-31-
-
Recommended separation distances betweenPortable and mobile RF communications equipment and the CLESTA Unit
TheCLESTAUnitisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuseroftheCLESTAUnitcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheCLESTAUnitasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitterm
150 kHz to 80 MHzd = 1.2P
80 MHz to 800 MHzd = 1.2P
800 MHz to 2.5 GHzd = 2.3P
0.01 0.12 0.12 0.230.1 0.38 0.38 0.731 1.2 1.2 2.310 3.8 3.8 7.3100 12 12 23
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmetres(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.NOTE1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyadsorption and reflection from structures, objects and people.
-32-
-
[7] LIST OF COMPATIBLE HANDPIECES
MODEL
SYRINGELUZZANI(3-way) Minilight w/Light
DCI(3-way)
TURBINE BIEN AIR BORA S36L / UNIFIX with LIGHT
NSK Ti-Max X
AIR MOTOR BIEN AIR Aquilon 830 / UNIFIX with LIGHT/PM1132NSK EX-203/EX-6
SCALER SATELEC SP4055 w/Light
-33-
-