delta test report - audiologi · 2017. 4. 11. · the report must not be reproduced, except in...

34
!""#$%&’( **+ , -./&$0 12 *34 50 256768$0$&$0 DELTA Danish Electronics, Light & Acoustics Edisonsvej 24 5000 Odense C Denmark Tel. (+45) 72 19 41 00 Fax (+45) 72 19 41 01 www.delta.dk This report is issued under the rules of DANAK (Danish Accreditation) and ILAC (International Laboratory Accreditation Cooperation) including its MRA (Mutual Recognition Arrangement). Further information can be found at www.danak.dk and www.ilac.org. The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report TEST Reg. no. 100 Acoustic Test Report according to IEC 60118-0, 1, 2, 6 on Epoq XW manufactured by Oticon

Upload: others

Post on 27-Feb-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

!""#$%&'()**+),)-./&$0)12)*34)50)256768$0$&$0)

DELTA

Danish Electronics,

Light & Acoustics

Edisonsvej 24

5000 Odense C

Denmark

Tel. (+45) 72 19 41 00

Fax (+45) 72 19 41 01

www.delta.dk

This report is issued under the rules of DANAK (Danish Accreditat ion) and ILAC (International Laboratory Accreditat ion Cooperation) including i ts MRA (Mutual Recognit ion Arrangement). Further information can be found at www.danak.dk and www.i lac.org. The report must not be reproduced, except in ful l , without the writ ten approval of DELTA.

DELTA Test Report 1.

TEST Reg. no. 100

Acoustic Test Report according to IEC 60118-0, 1, 2, 6 on

Epoq XW manufactured by Oticon

Page 2: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 93 PR079_10

Oticon Epoq XW

Name and type Oticon Epoq XW

Serial number # 15768593

General description BTE

Report number PR079_10

Number of pages 17

Client DELTA TAL

Edisonsvej 24

5000 Odense

Danmark

Contact Person LFN

Reception date 26-03-2010

Test date 05-05-2010

Climatic conditions Temperature : 22°C

Barometric pressure : 101 kPa

Humidity : 40 %

Test methods IEC 60118 - 0, 1, 2, 6 & DQP-86001

Software version TAL HM 5.0.5

Test program version TAL008

Co-Reader CVJ

Test personnel

Responsible

LFN

Lars F. Nielsen

M.Sc.E.E.

DELTA Danish Electronic, Light & Acous-

tics

Technical-Audiological Lab.

The present report describes results of measurements on hearing instruments, identified by name, type and serial number.

The laboratory can not guarantee, that these data applies on the entire production.

Page 3: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 94 PR079_10

Oticon Epoq XW

Electro Acoustical Properties

Climatic requirements

The climate in the testing laboratory is registered. The IEC 60118-0 standard states the following requirements

which were fulfilled during test (see page 2):

Temperature: 23 ± 5°C

Air pressure: 101.3 +5/-20 kPa

Humidity: 40 - 80 %

Test setup

The measurements are carried out in a free field, in accordance with IEC 60118-0 substitution method.

The test object is connected to the Ear Simulator (IEC 60318-4) using tubing with dimension Ø2 x 25 mm.

The loudspeaker and the test object are placed in an anechoic chamber 1.2 metres above the grating floor 0.8

metres apart.

Hearing aid adjustments and list of accessories

The hearing aid is programmed to deliver unity gain.

Measurements

Reference test frequency is 1600 Hz.

All sound pressures are expressed in dB relative 20 µPa.

Uncertainties

Frequency: ± 0.5 %

Current: ± 0.5 %

Sound Pressure Level, 200 Hz - 4000 Hz: ± 1.5 dB

Sound Pressure Level, 4000 Hz - 8000 Hz: ± 1.9 dB

The uncertainty primarily due to tolerances on the control microphone and the IEC 60318-4 coupler.

Attack time: ± 1.1 ms The uncertainty primarily due to “slew rate” on the loudspeaker.

Recovery time : ± 10 ms The uncertainty primarily due to Test system.

Page 4: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 95 PR079_10

Oticon Epoq XW

DATA (Primary results)

Maximum output : 101 dB SPL at 3981 Hz

Maximum acoustic gain : 15 dB at 3019 Hz

Reference test gain : 7 dB at 1600 Hz

Reference measurements

# 15768593

OSPL90 (dB SPL)

Maximum 101

1600 Hz 95

Unity gain (dB)

Maximum 15

1600 Hz 7

Reference measurement are made on 1 hearing instruments, identified by the serial number stated in the table

above (marked by *).

Page 5: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 96 PR079_10

Oticon Epoq XW

IEC 60118-0.7.2 : Output sound pressure level response, OSPL90

HA adjustment: Unity gain.

Input = 90 dB SPL.

Frequency response of the sound pressure level obtained in the ear simulator using an input of 90 dB SPL.

Page 6: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 97 PR079_10

Oticon Epoq XW

IEC 60118-0.7.3 : Full-on acoustic gain

HA adjustment: Unity gain.

Input = 50 dB SPL.

Measurement of full-on acoustic gain obtainable with the hearing aid.

Page 7: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 98 PR079_10

Oticon Epoq XW

IEC 60118-0.7.4 : Comprehensive frequency response

HA adjustment: Unity gain.

Input = 40, 50, 60, 70, 80 and 90 dB SPL.

The curve for 60 dB SPL input is the ‘Basic frequency response’.

Reference test gain = 7 dB.

Page 8: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 99 PR079_10

Oticon Epoq XW

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 60 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument 2nd order (Black), 2nd & 3rd order (Red).

Harmonic distortion is measured by means of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 9: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 100 PR079_10

Oticon Epoq XW

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 70 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument 2nd order (Black), 2nd & 3rd order (Red).

Harmonic distortion is measured by means of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 10: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 101 PR079_10

Oticon Epoq XW

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 65, 90 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument THD input=90 dB SPL (Black), THD in-

put=65 dB SPL (Red). Harmonic distortion is measured by means of Test system with a frequency resolution of

10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 11: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 102 PR079_10

Oticon Epoq XW

IEC 60118-0.7.12.2 : Measurement of nonlinearities (intermodulation distortion)

HA adjustment: Unity gain.

Input = 2 x 54 dB SPL.

Measurement of amplitude nonlinearities in the hearing aid. Intermodulation products of 2nd order (Black) and

3rd order (Red), relative to the level of the fundamental tone. F2 = F1+125 Hz. Distortion is measured by means

of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 12: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 103 PR079_10

Oticon Epoq XW

IEC 60118-0.7.12.2 : Measurement of nonlinearities (intermodulation distortion)

HA adjustment: Unity gain.

Input = 2 x 64 dB SPL.

Measurement of amplitude nonlinearities in the hearing aid. Intermodulation products of 2nd order (Black) and

3rd order (Red), relative to the level of the fundamental tone. F2 = F1+125 Hz. Distortion is measured by means

of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 13: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 104 PR079_10

Oticon Epoq XW

IEC 60118-0.7.14.1 : Internal noise generated in the hearing instrument (1/3 octave)

HA adjustment: Unity gain.

Input = 50 dB SPL.

The purpose of the measurement is to determine the internally generated noise in the hearing aid (input related).

Page 14: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 105 PR079_10

Oticon Epoq XW

IEC 60118-2.5.1 : Steady-state input/output graph (AGC systems)

HA adjustment: Unity gain.

Test frequency = 1600 Hz.

Measurement of the relation between input sound pressure level and output sound pressure level.

Page 15: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 106 PR079_10

Oticon Epoq XW

IEC 60118-2.5.1 : Steady-state input/output graph (AGC systems)

HA adjustment: Unity gain.

Test frequency = 1600 Hz.

Measurement of the relation between input sound pressure level and output sound pressure level.

Page 16: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 107 PR079_10

Oticon Epoq XW

IEC 60118-2.6.2.1 : Dynamic output characteristics for speech levels

HA adjustment: Unity gain.

Input = 55 dB - 80 dB (speech level) at 1600 Hz.

Input signal stable for 0.5 seconds prior to level shift.

AGC attack time with speech level signal: < 1 ms.

Input = 80 dB - 55 dB (speech level) at 1600 Hz.

Input signal stable for 0.2 seconds prior to level shift.

AGC recovery time with speech level signal: 0.007 s.

Measurement of the hearing aid attack and recovery time.

Page 17: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 108 PR079_10

Oticon Epoq XW

IEC 60118-2.6.2.2 : Dynamic output characteristics for high level input

HA adjustment: Unity gain.

Input = 60 dB - 100 dB (high level) at 1600 Hz.

Input signal stable for 0.5 seconds prior to level shift.

AGC attack time with high level signal: < 1 ms.

Input = 100 dB - 60 dB (high level) at 1600 Hz.

Input signal stable for 0.2 seconds prior to level shift.

AGC recovery time with high level signal: 0.047 s.

Measurement of the hearing aid attack and recovery time.

Page 18: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

!""#$%&'()**+),)-./&$0)12)*34)50)256768$0$&$0)

DELTA

Danish Electronics,

Light & Acoustics

Edisonsvej 24

5000 Odense C

Denmark

Tel. (+45) 72 19 41 00

Fax (+45) 72 19 41 01

www.delta.dk

This report is issued under the rules of DANAK (Danish Accreditat ion) and ILAC (International Laboratory Accreditat ion Cooperation) including i ts MRA (Mutual Recognit ion Arrangement). Further information can be found at www.danak.dk and www.i lac.org. The report must not be reproduced, except in ful l , without the writ ten approval of DELTA.

DELTA Test Report 1.

TEST Reg. no. 100

TEST Reg. no. 100

Acoustic Test Report according to IEC 60118-0, 1, 2, 6 on

Mind 440 M4-m CB manufactured by Widex

Page 19: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 110 PR078_10

Widex Mind 440 M4-m CB

Name and type Widex Mind 440 M4-m CB

Serial number # 032884

General description BTE

Report number PR078_10

Number of pages 17

Client DELTA TAL

Edisonsvej 24

5000 Odense

Danmark

Contact Person LFN

Reception date 26-03-2010

Test date 05-05-2010

Climatic conditions Temperature : 22°C

Barometric pressure : 101 kPa

Humidity : 40 %

Test methods IEC 60118 - 0, 1, 2, 6 & DQP-86001

Software version TAL HM 5.0.5

Test program version TAL008

Co-Reader CVJ

Test personnel

Responsible

LFN

Lars F. Nielsen

M.Sc.E.E.

DELTA Danish Electronic, Light & Acous-

tics

Technical-Audiological Lab.

The present report describes results of measurements on hearing instruments, identified by name, type and serial number.

The laboratory can not guarantee, that these data applies on the entire production.

Page 20: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 111 PR078_10

Widex Mind 440 M4-m CB

Electro Acoustical Properties

Climatic requirements

The climate in the testing laboratory is registered. The IEC 60118-0 standard states the following requirements

which were fulfilled during test (see page 2):

Temperature: 23 ± 5°C

Air pressure: 101.3 +5/-20 kPa

Humidity: 40 - 80 %

Test setup

The measurements are carried out in a free field, in accordance with IEC 60118-0 substitution method.

The test object is connected to the Ear Simulator (IEC 60318-4) using tubing with dimension Ø2 x 25 mm.

The loudspeaker and the test object are placed in an anechoic chamber 1.2 metres above the grating floor 0.8

metres apart.

Hearing aid adjustments and list of accessories

The hearing aid is programmed to deliver unity gain.

Measurements

Reference test frequency is 1600 Hz.

All sound pressures are expressed in dB relative 20 µPa.

Uncertainties

Frequency: ± 0.5 %

Current: ± 0.5 %

Sound Pressure Level, 200 Hz - 4000 Hz: ± 1.5 dB

Sound Pressure Level, 4000 Hz - 8000 Hz: ± 1.9 dB

The uncertainty primarily due to tolerances on the control microphone and the IEC 60318-4 coupler.

Attack time: ± 1.1 ms The uncertainty primarily due to “slew rate” on the loudspeaker.

Recovery time : ± 10 ms The uncertainty primarily due to Test system.

Page 21: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 112 PR078_10

Widex Mind 440 M4-m CB

DATA (Primary results)

Maximum output : 105 dB SPL at 2897 Hz

Maximum acoustic gain : 16 dB at 2897 Hz

Reference test gain : 7 dB at 1600 Hz

Reference measurements

# 032884

OSPL90 (dB SPL)

Maximum 105

1600 Hz 96

Unity gain (dB)

Maximum 16

1600 Hz 7

Page 22: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 113 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.2 : Output sound pressure level response, OSPL90

HA adjustment: Unity gain.

Input = 90 dB SPL.

Frequency response of the sound pressure level obtained in the ear simulator using an input of 90 dB SPL.

Page 23: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 114 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.3 : Full-on acoustic gain

HA adjustment: Unity gain.

Input = 50 dB SPL.

Measurement of full-on acoustic gain obtainable with the hearing aid.

Page 24: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 115 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.4 : Comprehensive frequency response

HA adjustment: Unity gain.

Input = 40, 50, 60, 70, 80 and 90 dB SPL.

The curve for 60 dB SPL input is the ‘Basic frequency response’.

Reference test gain = 7 dB.

Page 25: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 116 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 60 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument 2nd order (Black), 2nd & 3rd order (Red).

Harmonic distortion is measured by means of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 26: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 117 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 70 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument 2nd order (Black), 2nd & 3rd order (Red).

Harmonic distortion is measured by means of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 27: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 118 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.12.1 : Measurement of nonlinearities (harmonic distortion)

HA adjustment: Unity gain.

Input = 65, 90 dB SPL.

Measurement of amplitude nonlinearities in the hearing instrument THD input=90 dB SPL (Black), THD in-

put=65 dB SPL (Red). Harmonic distortion is measured by means of Test system with a frequency resolution of

10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 28: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 119 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.12.2 : Measurement of nonlinearities (intermodulation distortion)

HA adjustment: Unity gain.

Input = 2 x 54 dB SPL.

Measurement of amplitude nonlinearities in the hearing aid. Intermodulation products of 2nd order (Black) and

3rd order (Red), relative to the level of the fundamental tone. F2 = F1+125 Hz. Distortion is measured by means

of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 29: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 120 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.12.2 : Measurement of nonlinearities (intermodulation distortion)

HA adjustment: Unity gain.

Input = 2 x 64 dB SPL.

Measurement of amplitude nonlinearities in the hearing aid. Intermodulation products of 2nd order (Black) and

3rd order (Red), relative to the level of the fundamental tone. F2 = F1+125 Hz. Distortion is measured by means

of Test system with a frequency resolution of 10 points per octave.

Conversion between dB and % THD

dB %

-10 32

-20 10

-26 5

-30 3

-40 1

Page 30: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 121 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-0.7.14.1 : Internal noise generated in the hearing instrument (1/3 octave)

HA adjustment: Unity gain.

Input = 50 dB SPL.

The purpose of the measurement is to determine the internally generated noise in the hearing aid (input related).

Page 31: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 122 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-2.5.1 : Steady-state input/output graph (AGC systems)

HA adjustment: Unity gain.

Test frequency = 1600 Hz.

Measurement of the relation between input sound pressure level and output sound pressure level.

Page 32: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 123 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-2.5.1 : Steady-state input/output graph (AGC systems)

HA adjustment: Unity gain.

Test frequency = 1600 Hz.

Measurement of the relation between input sound pressure level and output sound pressure level.

Page 33: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 124 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-2.6.2.1 : Dynamic output characteristics for speech levels

HA adjustment: Unity gain.

Input = 55 dB - 80 dB (speech level) at 1600 Hz.

Input signal stable for 0.5 seconds prior to level shift.

AGC attack time with speech level signal: < 1 ms.

Input = 80 dB - 55 dB (speech level) at 1600 Hz.

Input signal stable for 0.2 seconds prior to level shift.

AGC recovery time with speech level signal: 0.008 s.

Measurement of the hearing aid attack and recovery time.

Page 34: DELTA Test Report - Audiologi · 2017. 4. 11. · The report must not be reproduced, except in full, without the written approval of DELTA. DELTA Test Report 1. TEST Reg. no. 100

Page 125 PR078_10

Widex Mind 440 M4-m CB

IEC 60118-2.6.2.2 : Dynamic output characteristics for high level input

HA adjustment: Unity gain.

Input = 60 dB - 100 dB (high level) at 1600 Hz.

Input signal stable for 0.5 seconds prior to level shift.

AGC attack time with high level signal: 22 ms.

Input = 100 dB - 60 dB (high level) at 1600 Hz.

Input signal stable for 0.2 seconds prior to level shift.

AGC recovery time with high level signal: 0.247 s.

Measurement of the hearing aid attack and recovery time.