1/21

©Eastman Kodak Company, 2005

Vorschlag einer einheitlichenDefinition “Dosisindikator” in der

Digitalen Projektions Radiographie

Walter StrengKodak Health Group, TIC Genoa

Richard VanMetter and John YorkstonHealth Imaging Research Laboratory

Eastman Kodak Company

2/21 ©Eastman Kodak Company, 2005

S-Wert SpeedE.I. Wert

?

Exposure Index: Heute

Log M

3/21 ©Eastman Kodak Company, 2005

Das Problem Heute

•Verschiedene Kalibrationstechniken

•Ungleiche ”pixel weighting methods”

•Unterschiedliche Methoden zum Ausschlussvon Direkbestrahlung in derBildberechnung•Kein einheitlicher Maßstab für

den Index•Keinerlei Möglichkeit von

Vergleichen diverser Hersteller

4/21 ©Eastman Kodak Company, 2005

Das Ziel

Exposure Index = 1510

SPEED CLASS = 320

Dosisindex = 320

SPEED Index = 320

5/21 ©Eastman Kodak Company, 2005

AAPM Task Group 116 onStandardization of an Image ReceptorDose Index for Digital Radiography

CHARGE:To identify a method of providing

feedback, in the form of astandard index, to operators ofdigital radiography systems,which reflects the adequacy ofthe exposure that has reachedthe image receptor (IR) afterevery exposure event.

6/21 ©Eastman Kodak Company, 2005

Charakteristica eines idealen Index

•Herstellerunabhängig•Vorteile verfügbar für alle Hersteller

•Einheitlich Systemkalibrierung•Standard Kalibrierung der Pixelwerte

•Klar Definiert•Meßbar und Reproduzierbar

•Technologieunabhäng•CR / DR

•Robust•Unabhängig vom Bildinhalt

•Konstanz•Klinische- sowie Phantomaufnahmen

•Einfach•Berechnung und Verständnis

7/21 ©Eastman Kodak Company, 2005

Histogram Originalbild

0

1x105

2x105

3x105

4x105

1000 1500 2000 2500 30000

4000

8000

12000

16000

Pixel Value

Anatomy Only

Entire Receptor

1 Gy 10 Gy3 Gy

100:1

30 Gy

8/21 ©Eastman Kodak Company, 2005

Der mid-Value

0

1

2

250 500

mid-value

DICOM GSDFLightbox Luminance 1000 cd/m2

DICOM p-value

Net

Den

sity

9/21 ©Eastman Kodak Company, 2005

1. Systemkalibration in Kodak

180 cm0.5 mm Cu1.0 mm Al

2.6 mm AlInherent Filtration

100.0 2 µGy@ 80 kVp

Blei Unterlage

10/21 ©Eastman Kodak Company, 2005

0

1024

2048

3072

4096

1000 1500 2000 2500 3000

Original Pixel Value

Dis

play

-rea

dyP

ixel

Val

ue

Pfiducial

Poriginal(KS)

} 1% range

Pdisplay(Poriginal)

2. Original & DisplayReady-Image(Analyse)

PixelValuesdisplayed as a function of PixelValues original

original display ready

11/21 ©Eastman Kodak Company, 2005

3. Berechnung Dosisindikator

•Notwendig sind zur Berechnung Originalund Display-Ready Images•Bildanalyse•System property: Porginal(KS) (calibration)•Image property: Pdisplay(Poriginal) (image proc)•Berechnung Pdisplay(KS) = Pdisplay(Poriginal(KS))

•Berechnung Dosis oder Speed Index•KS Pdisplay(KS) = P2048 or [Dose Class]•S = K0 / KSor [Speed Class]

(1000 : KS)

12/21 ©Eastman Kodak Company, 2005

4. Empfehlung

Hersteller kalibrieren ihre digitalen System nachden Standards aufgezeigt in ISO 9236-1

Die Projektionen müssen den ISO Gruppen in9236 zugeordnet werden

Eventuelle Berechnung der Porig und Pdispl alssubsampled Bild (event >105)

Berechnung der Pixel im MidValue +/-1% für dasDisplay-Ready-Image wobei der MidValue einerNetto Dichte von 1.0 im Film/Screen entspricht.

Documentation des Dosisindikators im DICOMHeader.

13/21 ©Eastman Kodak Company, 2005

Original Image Poriginal(KS)

500

1000

1500

2000

2500

1 2 5 10 20

Extremity (ISO-I)Skull, Lumbar Spine and Chest(ISO- II, III and IV)

ISO 9236-1

Kerma in Air (Gy)

Exp

osur

eC

ode

Val

ue

Kodak CR

14/21 ©Eastman Kodak Company, 2005

5. Nächste Schritte

Abstimmung der Industrie über einheitlicheDosisindikatoren

DIN-NAR Arbeitskreis für Dosisindikatoren

Kontakt zur AAPMTask Group 116 zur Abstimmung

15/21 ©Eastman Kodak Company, 2005

DIAGNOSTICPLEASE

Danke