  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    1/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 1 of 8.

    Goal:To Upgrade the old and obsolete Seres ODME MK-III Hengstler Printer with the latest Seres and Seil-SeresMegatron one, with out affecting the performance of the ODME system.

    In order to reach this Goal the following two boards has to be removed from the Calculator send to DarlasGreece, workshop for Firmware and Hardware modification and then returned to the ship together with theComplete P/N: S-300-1, Darlas-Seres Printer upgrade kit, which includes all the necessary parts, screwscables, assembly instructions and fastening materials in order to realize the final Printer Upgrade Modification.Of course the kit includes a Brand New Megatron Printer and its embedded Darlas Printer Interface PCB.

    1 x S-C538R PCB ANNEX 804/11, HENGSTLER, SERES MK3 ODME ( Located inside the MK-III Calculator Unit).1 x S-C337R PCB M/P-CESSOR HENGSTLER, SERES MK3 ODME ( Located inside the MK-III Calculator Unit).

    Special Note: We guarantee to sh ip back to the sh ip the com ple te Dar las Seres Pr in ter Upgrade k i t wi th in 72 hours a f te r Receip t of you r boards accompanied wi th yo ur pre-payment and w r i t ten order.

    CAUTION

    Before any dismantling job make sure that both 3-Phase as well as 1-Phase Power supply to ODME is cut off.

    Step 1

    Go to Cargo Control Room and locate the Calculator Unit as it is shown on Photo 1 below.Unscrew the four (4) screws located in the four (4) (North-South & East-West) corners of the facial panel. (Rightside Part of Facial Panel containing the LCD display and the Keyboard).

    PHOTO 1

    Remove and savescrew for later use.

    Remove and savescrew for later use.

    Remove and savescrew for later use. Remove and save

    screw for later use.

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    2/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 2 of 8.

    Step 2

    Unscrew all four (4) screws and remove carefully the facial panel from its chassis as shown in photo 2 below.Please Secure it safely aside, as shown in the below picture:

    PHOTO 2

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    3/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 3 of 8.

    Step 3

    Disconnect the two (2) ribbon cable connectors from the P-cessor board (1 st from left) as shown in photo 3and photo 4 below:

    PHOTO 3

    Ribbon CableConnector No.: 1

    Ribbon CableConnector No.: 2

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    4/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 4 of 8.

    PHOTO 4

    Ribbon CableConnector No.: 1

    Ribbon CableConnector No.: 2

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    5/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 5 of 8.

    Step 4

    Remove the Processor PCB (P/N: ES-65120A) from its Guide Rails by pressing down the red U stopper gently. Please see below photo Number 5.

    PHOTO 5

    Press down the red PCBstopper and gently pull outthe P PCB.

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    6/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 6 of 8.

    Step 5

    Remove the Printer Interface PCB (P/N: S80411 ANNEX) from its Guide Rails by pressing down the red Ustopper gently. Please see below photo Number 6.

    PHOTO 6

    Press down the red PCBstopper and gently pull outthe Printer/Interface PCB.

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    7/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 7 of 8.

    Step 6

    After you have removed both left side fitted PCBs as shown below in Photo 7, please go ahead andfit back the Operators/Calculator Cabinet Facial Panel with the help of the four (4) screws saved fromStep 1. So the Cabinet will be closed again as it is shown in Photo 1.

    Both microprocessor and Printer Interface PCBs removed from ODME Operators Cabinet.

    PHOTO 7

    CAUTION

    Please do not apply power to Operators / Calculators Cabinet until you will receive back from us and for thefinal installation both the removed boards as well as your complete DARLAS-SERES printer upgrade kit.Normal ly we sh ip back to the sh ip th e com ple te Dar las Seres Pr in ter Upgrade k i t wi th in 72 hours a f te r receip t of you r boards accompanied wi th y our pre-payment and wr i t ten ord er.

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]
  • 8/12/2019 DARLAS SERES ODME S 300 2 PCBs Removal Instructions for MK3 Hengstler Printer Upgrade

    8/8

    DRAFT INSTRUCTIONS FOR HOW TO REMOVE THE TWO PCBS :ES-65120A Microprocessor PCB and S80411P Annex 804/11 PCB

    FROM THE SERES MK-III ODME CALCULATOR UNIT Ver. 0Fitted with an obsolete Hengstler Printer

    IN ORDER TO UPGRADE IT WITH THE LATEST MEGATRON PRINTER

    5, Kimonos Str., 122 44 Egaleo Athens Greece.Telephones: +30-210-5986179, +30-210-5986213, +30-210-5310551, +30-210-5310674, +30-210-5310775Fax: +30-210-5911161, E-mail: [email protected] , [email protected] Internet: http://www.darlas.gr

    Established in Athens-Greece as a Societal Anonym (SA), Reg. No: 16494/01 /B/88/343, Fully Paid Capital: 58 694.06Bankers: National Bank of Greece SA, Egaleo Branch (105), IBAN: GR16 0110 1050 0000 1054 7023 795.

    Page 8 of 8.

    Step 7 You are now done with your two PCBs removal :

    1. ES-65120A Microprocessor PCB and 2. S80411P Annex 804/11 PCBPlease go ahead and carefully wrap separately each one of the boards with any suitable bag as shown inphotos 8, 9 & 10 below. If none of the below bags are available then please use first normal A4 size businessmail paper envelopes to wrap each individual board and then pack both of the boards in a suitable for air freight carton box. The parcel should be send to our workshop for Firmware and Hardware modification sothey can be returned back to the ship together with the Complete P/N: S-300-1, Darlas-Seres Printer upgradekit, which includes all the necessary parts, screws cables, assembly instructions and fastening materials inorder to realize the final Printer Upgrade Modification. Of course the kit includes a Brand New Megatron Printer and its embedded Darlas Printer Interface PCB.

    ES-65120A Microprocessor PCB S80411P Annex Printer/Interface PCB

    PHOTO 8

    Antistatic Air Bubble Bag Paper Jiffy Bubble Bag

    PHOTO 9 PHOTO 10

    mailto:[email protected]:[email protected]://www.darlas.gr/http://www.darlas.gr/mailto:[email protected]:[email protected]