compensation

16
COMPENSATION FITC FITC - PE PE - FITC - - + + gain=1 gain=1 detector detector PE FITC FL 1 FL 2 FL2-%FL1

Upload: ansel

Post on 05-Jan-2016

26 views

Category:

Documents


0 download

DESCRIPTION

COMPENSATION. FL1. FITC. +. detector. FITC -. PE. gain=1. -. FL2-%FL1. FITC. FL2. PE. +. PE -. FITC. gain=1. detector. -. COMPENSATION. FL1. FITC. +. detector. FITC -. PE. gain=1. -. PE. FL2. PE. +. PE -. FITC. gain=1. detector. FL1-%FL2. -. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: COMPENSATION

COMPENSATION

FITC

FITC - PE

PE - FITC

-

-

+

+gain=1

gain=1detector

detectorPE

FITC

FL1

FL2

FL2-%FL1

Page 2: COMPENSATION

COMPENSATION

FITC

FITC - PE

PE - FITC

-

-

+

+gain=1

gain=1detector

detectorPE

PE

FL1

FL2

FL1-%FL2

Page 3: COMPENSATION

Uncompensated vs Compensated

FL1

FL2

Page 4: COMPENSATION

FL1

FL2

COMPENSATION IS INTENSITY DEPENDENT

uncompensated partially compensated

fully compensated

Page 5: COMPENSATION

COMPENSATE INSTRUMENT USING STAINED CELLS

1. Adjust PMT voltages using unstained cells

2. Adjust compensation for each fluorochrome

Page 6: COMPENSATION

0 256 512 768 1024

030695001FSC ->

0 256 512 768 1024

030695001FSC ->

Lymphogated Cells for Four Color Single Laser Compensation

A B

FSC FSC

SS

C

SS

C

Page 7: COMPENSATION

10 10 10 10 100 1 2 3 4

042095001FL1 ->

10 10 10 10 100 1 2 3 4

042095001FL4 ->

10 10 10 10 100 1 2 3 4

042095001FL3 ->

10 10 10 10 100 1 2 3 4

042095001FL1 ->

10 10 10 10 100 1 2 3 4

042095001FL3 ->

10 10 10 10 100 1 2 3 4

042095001FL3 ->

FITC

FITC

PE-TxRd PE-Cy5 or PerCP

PE-Cy5 or PerCP PE-Cy5 or PerCP

PE

PE

PE

PE-TxRd

PE-TxRd

FITC

Setting the High Voltage

Page 8: COMPENSATION

10 10 10 10 100 1 2 3 4

042095002FL1 ->

10 10 10 10 100 1 2 3 4

042095002FL4 ->

10 10 10 10 100 1 2 3 4

042095002FL3 ->

10 10 10 10 100 1 2 3 4

042095002FL1 ->

10 10 10 10 100 1 2 3 4

042095002FL3 ->

10 10 10 10 100 1 2 3 4

042095002FL3 ->

Compensation adjustment using FITC stained lymphocytes

PE

PE

PE

PE-TxRd

PE-TxRd

FITC-CD45

FITC-CD45

PE-TxRd PE-Cy5 or PerCP

PE-Cy5 or PerCP PE-Cy5 or PerCP

FITC-CD45

Page 9: COMPENSATION

10 10 10 10 100 1 2 3 4

042095003FL1 ->

10 10 10 10 100 1 2 3 4

042095003FL4 ->

10 10 10 10 100 1 2 3 4

042095003FL3 ->

10 10 10 10 100 1 2 3 4

042095003FL1 ->

10 10 10 10 100 1 2 3 4

042095003FL3 ->

10 10 10 10 100 1 2 3 4

042095003FL3 ->

Compensation adjustment using PE stained lymphocytes

FITC PE-TxRd PE-Cy5 or PerCP

FITC PE-Cy5 or PerCP PE-Cy5 or PerCP

PE-CD4

PE-CD4

PE-CD4

PE-TxRd

PE-TxRd

FITC

Page 10: COMPENSATION

10 10 10 10 100 1 2 3 4

042095004FL1 ->

10 10 10 10 100 1 2 3 4

042095004FL4 ->

10 10 10 10 100 1 2 3 4

042095004FL3 ->

10 10 10 10 100 1 2 3 4

042095004FL1 ->

10 10 10 10 100 1 2 3 4

042095004FL3 ->

10 10 10 10 100 1 2 3 4

042095004FL3 ->

Compensation adjustment using PE-TxRd stained lymphocytes

PE

PE

PE

ECD-CD4

FITC

FITC

ECD-CD4 PE-Cy5 or PerCP

PE-Cy5 or PerCP PE-Cy5 or PerCP

FITC

ECD-CD4

Page 11: COMPENSATION

10 10 10 10 100 1 2 3 4

042095005FL1 ->

10 10 10 10 100 1 2 3 4

042095005FL4 ->

10 10 10 10 100 1 2 3 4

042095005FL3 ->

10 10 10 10 100 1 2 3 4

042095005FL1 ->

10 10 10 10 100 1 2 3 4

042095005FL3 ->

10 10 10 10 100 1 2 3 4

042095005FL3 ->

Compensation adjustment using PE-CY5 stained lymphocytes

PE

PE

PE

FITC

FITC TC-CD45TC-CD45

TC-CD45

FITC

PE-TxRd

PE-TxRd

PE-TxRd

Page 12: COMPENSATION

10 20 30 40 50 60 70 80 90 100 120

FSC -->

1020

3040

5060

7080

90100

120

SS

C -->

10 1 10 2 10 3 10 4

FL1 -->

101

102

103

104

FL

2 --> R1

1 2

3 4

FSC

SS

C

R1

R2 R3

R4

FITC-CD45

PE-C

D4

10 1 10 2 10 3 10 4

FL4 -->

10

110

210

310

4

FL

2 --> R9

9 10

11 12

PE-CY5-CD8

PE-C

D4

R4 R5

R6

10 1 10 2 10 3 10 4

FL1 -->

101

102

103

104

FL

3 --> R13

13 14

15 16

PE-CY5-CD8

AP

C-C

D8

R8

R9

R10

COMPENSATION UP TO FOUR COLORS WITH TWO LASERS

Page 13: COMPENSATION

10 10 10 10

FITC-CD3

PE

-CD

8

10

TC

-CD

2

FITC-CD3

PE

-CD

8

TC-CD22 10 10

10 2 10 4

EC

D-C

D4

FITC-CD310 10 10 10

EC

D-C

D4

TC-CD2

10 2 10 10

PE

-CD

8

ECD-CD4

UNCOMPENSATED

Page 14: COMPENSATION

SOFTWARE COMPENSATION

101 102 103 104

10

1

10

2

10

3

10

4

PE

CD

8 -

->

PE

-CD

8

TC-CD2101 102 103 104

10

1

10

2

10

3

10

4

PE

-CD

8

ECD-CD4

101 102 103 104

10

1

10

2

10

3

10

4

TC

-CD

2

FITC-CD3101 102 103 104

10

1

10

2

10

3

10

4

EC

D C

D4 -

->

EC

D-C

D4

FITC-CD3101 102 103 104

10

1

10

2

10

3

10

4

EC

D-C

D4

TC-CD2

101 102 103 104

10

1

10

2

10

3

10

4

PE

-CD

8

FITC-CD3

R1

R2

R3

R4

R5

R6

Page 15: COMPENSATION

101 102 103 104

FITC CD3 -->

10

1

10

2

10

3

10

4

PE

CD

8 -

->

PE

-CD

8

FITC-CD3101 102 103 104

10

1

10

2

10

3

10

4

PE

CD

8 -

->

PE

-CD

8

TC-CD2

INSTRUMENT COMPENSATION

101 102 103 104

ECD CD4 -->

10

1

10

2

10

3

10

4

PE

CD

8 -

->

PE

-CD

8

ECD-CD4

101 102 103 104

10

1

10

2

10

3

10

4

TC C

D2

-->

TC

-CD

2

FITC-CD3101 102 103 104

10

1

10

2

10

3

10

4

EC

D C

D4

-->

EC

D-C

D4

FITC-CD3101 102 103 104

10

1

10

2

10

3

10

4

EC

D-C

D4

TC-CD2

Page 16: COMPENSATION

PE, TX-RED & PE-CY5COMPENSATION

TC CD2 -->101 102 103 10

10

1

10

2

10

3

10

PE

-CD

8

TC-CD2101 102 103 104

ECD CD4 -->1

0

1

10

2

10

3

10

4

PE

-CD

8

ECD-CD4

POOR

GOOD

101 102 103 104

10

1

10

2

10

3

10

4

PE

-CD

8

TC-CD2101 102 103 104

ECD CD4 -->

10

1

10

2

10

3

10

4

PE

-CD

8

ECD-CD4

R3

R3

A C

B D