chemical hydrogen generator type gip ™ 3 training module
DESCRIPTION
CHEMICAL HYDROGEN GENERATOR TYPE GIP ™ 3 Training Module. PROGRAMME DE FORMATION. PRACTICE. THEORY. First time start up of unit Charging and production Regular operations Periodical operations Handling and start up of unit by trainees Maintenance of the plant - PowerPoint PPT PresentationTRANSCRIPT
CHEMICAL HYDROGEN GENERATOR
TYPE GIP™ 3
Training Module
THEORY PRACTICE
First time start up of unit Charging and production Regular operations Periodical operations Handling and start up of unit by
trainees Maintenance of the plant Breakdowns, causes and remedies Various questions
Design of the hydrogen generator Operating principle Hydrogen generator layout
Safety, gas, installation instructions
Use of the equipment Earthing Charging Hydrogen production Drawing off the hydrogen Emptying Rinsing
GIP™ reactive cartridge Periodical operations and maintenance Instructions for cylinders reproof test
PROGRAMME DE FORMATION
DESCRIPTION
Chemical hydrogen generator GIP™ 3
Closing assemblyAnd draw off valve
Set of accessories:
1 bucket13 litres 1 Straight funnel 1 bent funnel 1 silicon basket3 m3 1 shovel 1 poker 1 brass brush 1 spanner12” 1 neoprene protective gloves 1 protective goggle 1 earthing device
OPERATING PRINCIPLE
Chemical hydrogen generatortype GIP™ 3
The process of hydrogen production with the GIP™ 3 generators is based on the attack of silicon by the caustic soda in aqueous solution under pressure and high temperature.
Its production per cartridge is 3 m3 of hydrogen (measured in conditions at 15 °C and 760 mm of mercury).
The caustic soda undergoes the action of pure silicon
in autoclave.
Pressure is produced by the reaction itself, which is not reversible.
A local rise in temperature enables the liquid to reach a temperature high enough to compensate fully the retarding action of the pressure and to insure that the pure silicon is completely attacked.
The hydrogen is stored into the cylinder at high pressure.
HYDROGEN GENERATOR LAYOUT
The hydrogen generator has to be used outside.
In case of use inside, the building shall be equipped of:
Sufficient dimensions to permit an easy handling,
Ventilation openings at the top and bottom to ensure
effective replacement of ambient air,
Hydrogen approved artificial lighting,
Appropriate fire safety measures,
Drinking or raw water supply,
Pit for the elimination of the residues,
Earth point for earthling connection.
Chemical hydrogen generatortype GIP™ 3
USE OF THE EQUIPMENT
EARTHING Earth rod shall be embedded of 1 meter minimum into the
ground.
Earthing wire (4 mm² – length = 20 m) between the earth rod and the cradle of the generator
Earthing wire (4 mm² – length = 10 m) between the earth rod and the inflating sleeve
Earthing wire (4 mm² – length = 1.1 m) between the cradle and the draw off valve after having introduced the reactive products into the generator and firmly screwed the closing assembly.
CHARGING Make sure that the generator is empty and dry.
Fill completely the filter basket with granulated silicon.
Mix the GIP™ alkaline cartridge with the powder silicon into the bucket.
Pour the powdered mixture into the cylinder of the generator.
Place the basket into the cylinder of the generator.
Pour 13 litres of water into the cylinder of the generator and screw immediately after the closing assembly.
Chemical hydrogen generatortype GIP™ 3
EARTHING DEVICECable 4 mm2
Lenght 1,1 m
To inflation nozzle balloon
Cable 4 mm2
Lenght 10 mCable 4 mm2
Lenght 20 m
Ground
Earthing rod embedded 1 munderground
Hydrogen generator GIP™ 3
USE OF THE EQUIPMENT
PRODUCTION D’HYDROGÈNE
Set the generator in vertical position and shake during 10 seconds.
Check if the closing assembly is firmly screwed.
Open the draw-off valve during 2 minutes in order to de-aerate the generator.
Wait until the hydrogen production rises 80-100 barg.
Shake again the generator during 10 seconds.
Wait until the end of the reaction.
MAIN CRITERIA
Approximate reaction time 45 minutes Min. / Ave. / Max. temp. elevation 200°C / 250°C /
280°C
Min. / Ave. / Max. pressure elevation 90 b./ 130 b./ 160 b.
Bursting pressure 185 b.
Hydrogen production 3 m3 theoretical Chemical hydrogen generator
type GIP™ 3
USE OF THE EQUIPMENT
HYDROGEN DRAWING OFF
Hydrogen may be used immediately after the reaction.
Hydrogen pressure Apparatus cool Approx. 100 barg Apparatus hot Approx. 130 barg
EMPTYING AND RINCING
The residues have to be drained completely from the cylinder into the bucket.
The residues have to be mixed with 50% of raw water for total elimination.
The cylinder has to be washed and cleaned with raw water.
Chemical hydrogen generatortype GIP™ 3
POSITION TO USE
Position of the generator during the
reaction
Position of the generator in draining
and rincing
Generator in charging position
Preparation of the generator before
draining
PERIODICAL OPERATIONS BEFORE EACH REACTION.
Grease the threaded neck of the cylinder. Grease the threaded plug of the closing assembly.
AFTER EACH REACTION Drain the residues contained inside the cylinder. Wash the internal and external parts of the cylinder. Clean the filter silicon basket.
ONCE A WEEK Grease the wheels and trolley axes. Check the state of the plastic washers of the closing
assembly.
ONCE A YEAR Check the tightness of the draw-off valve and replace it if
necessary.
EVERY SIX MONTHS Replace the plastic washer of the closing assembly. Replace the silicon basket filter. Check the state of the safety copper strip; replace if
necessary.
EVERY FIVE YEARS Retest the GIP™ hydrogen cylinder
Chemical hydrogen generatortype GIP™ 3
INSTRUCTION FOR GIP™
CYLINDERS REPROOF TEST Periodical pressure proof tests every 5 years at 1.5
time the service pressure.
The proof test has to be witnessed by an expert from the competent National Department.
Proof test certificate shall be issued after the stamping of the cylinder.
Proof test has to be performed with a hydraulic pump.
Proof test has to be made in case of visible defects on the cylinder such as :
Notch or groove 5% deeper than the thickness of the cylinder,
Lamination (fold, overlap, entrapment, flaking of the metal),
Punch marks, Collar not fixed, Grinding recess or suspect stain.
Systematic reject has to be decided in case of visible defects on the cylinder such as :
Projection on the surface of the cylinder visible to the naked eye,
Dent greater than 0.5% of the outside diameter of the cylinder,
Crack or tearing of metal, Damage due to fire, Arcing or blow torch damage.
Chemical hydrogen generatortype GIP™ 3
GIP™ REACTIVE CARTRIDGES
COMPOSITION 1 GIP™ alkaline cartridge
1 GIP™ priming cartridge
1 GIP™ silicon cartridge
IDENTIFICATION GIP™ alkaline cartridge Eutectic mixture of caustic soda and caustic
potash
GIP™ priming cartridge Pure silicon powder form
GIP™ silicon cartridge Pure silicon grain form
Customs tariffs 28 15 11 00 00 00 J + 28 04 69 00 00 00 W
NATO Code 36 55 14 507 39 70
STORAGE The GIP™ reactive cartridges have to be stored in a dry area and in their original
packaging.
The maximum storage period allowed is 12 months.
SAFETY Corrosive product which causes immediate burns on mucous membranes such
as eyes and skin.
Contact with strong acids may provoke violent reactions or explosion.
Contact with some metals may produce hydrogen gas.
The handling has to be done by informed people equipped with adequate protections.
The handling shall be accompanied by collective safety measures located near and signaled (shower, eyes fountain).