CHEM 213Instrumental Analysis
Lab Lecture – Fe and Extraction
2
Chem 102 – used Crystal field theory to explain colour of coordination complexes
Chem 213 – use intensity of colour to determine concentration of metal ion
3 + Fe2+ Fe(bipy)32+
Bipyridyl(bipy)
red
Determination of Trace Iron (LM, p14-22)
Spectrophotometry – any method using light to measure chemical composition
3
P/P0 %T A 1 100 0 0.1 10 10.01 1 2
Fe: Concentration to Colour
Intensity of colour = Absorbance (A)
Measure absorbance of soln to determine the iron concentration.
Transmitted
Transmittance (T) = P/Po
Absorbance A = -log(T) = abc
a = absorptivity
b = pathlength
c = concentration
4
Spectrophotometer (Text p410, LM p19-21)
Spec 20D+pg 21
On/Off Zero control
Readout Mode
Sample comp.
Readout On/Off
Genesys 20pg 21
5
Spectrophotometer (Text p410, LM p19-21)
Sample cell matching LM p 19 (1 hour)- Need 8 test tubes with the same pathlength (b)- prepare dilute amaranth dye soln
- Set wavelength to 510 nm
- Insert test tube in sample compartment
- Adjust trans/absorb control to 90%
- Insert 2nd tube in spec
- Put on alignment line
- Rotate tube to get T=90%
- Repeat until you have 8 matched tubes
Record spectrometer number
- Mark tube with line
6
• Reduce any Fe(III) to Fe(II) so all iron is in the Fe(II) state, because only Fe(II) forms coloured complex
Determination of Trace Iron (LM, p14-22)
3 + Fe2+ Fe(bipy)32+
Bipyridyl(bipy)
red
1. Fe3+acidic
2. xs + Fe2+ Fe(bipy)32+
3. Adjust pH to ~ 5 by adding sodium acetate
+ xs NH2OH Fe2+
hydroxylamine reducing agent
7
Determination of Trace Iron(LM, p14-22)
Measure absorption spectrum of the complex to determine the wavelength of max. absorbance.
Absorbance spectrum – graph of absorbance of light vs. wavelength
A = abc
Measure A for most conc standard from 640 to 400 nm at 20 nm intervals.
The 0 %T (20D Only) and the 100 %T (A = 0) settings must be reset for each new wavelength.
8
Why is 100 %T (A = 0) reset at each wavelength?
Determination of Trace Iron (LM, p14-22)
Light intensity (Po) varies with wavelength)
A is maxBest sensitivity
9
Standards = pipet 5, 10, 15 & 20 mL Fe3+ into flasks
Blank = soln with all of the reagents but no analyte
[Fe3+]=0
Zero (set A =0) instrument using blank
Measure absorbance of all standards and samples
Re-zero
Take second absorbance reading for all stds and samples
Prepare Calibration Curve
10
Prepare Calibration Curve
Blank
standards
Asample
Csample
11
Calibration Curves LM, Appendix A
12
Calibration Curves
Calibration Curve – graph of instrument response vs. conc of analyte
- when response for sample is measured, its conc is determined by interpolating from the graph
- blank value should be part of the calibration curve, i.e. signal for [C] = 0.0
-standards must cover the complete range of expected unknowns
- plot signal on the Y axis and conc on the X axis.
- Normally response is linear
or with math transformation can be made linear
Signal = m(conc) + b
m = slope b = intercept
13
Drawing a Calibration Curve
Manually on graph paper
14
0 1 2 3 4 50
0.1
0.2
0.3
0.4
0.5
0.6
Concentration, ppm
Inst
Res
po
nse
Useful for spotting:
Manual Graphing of Calibration Curve
15
0 1 2 3 4 50
0.1
0.2
0.3
0.4
0.5
0.6
Concentration, ppm
Inst
Res
po
nse
Useful for spotting: outliers
Manual Graphing of Calibration Curve
16
0 1 2 3 4 50
0.1
0.2
0.3
0.4
0.5
0.6
Concentration, ppm
Inst
Res
po
nse
Manual Graphing of Calibration Curve
Useful for spotting: outliers curvature
17
Mathematically calculate line using least squares
fits data to y = mx + b ?
determines m and b
determines error in m and b?
allows accurate calculation of sample conc
Manually on graph paper
- best for spotting outliers, curvature, etc.
- disgard outliers
Manual Graphing of Calibration Curve
Calculate least squares line
18
Solvent extraction
Dextromethorphancough suppressant
Expectorant Glyceryl guaiacolate
Antihistamines Chloropheniramine maleate Diphenyhydramine HCl
Decongestants Phenylephrine HCl
Analgesics Acetaminophen
Preservatives
Flavouring
Colouring
Determination of Dextromethorphan in Cough Syrup
toluenewater
19Dextromethorphan
DMH+
Bromothymol blueBTB-
+ (DMH+BTB-)
ion pair
(DMH+ BTB-)org
yellow
B. Solvent Extraction System
Ion-association Extraction
water
toluene
Colour = Absorbance @ 400 nm [BTB-]org
xs
[DMH+]aq
Calibration curve: (absorbance-blank) vs. [DMH+]aq
20
First week solution preparation (pg 50)
Do NOT add BTBDo NOT prep sample
21
b) Mohr pipet – use like a buret
d) Fluted filter paper – allows liquid to pass through quickly