reaF.T4r^-^r^^^jr^^-J^Z^gjr£^ -• t s s I I Cesky metrologicky institut Okruzni'31,638 00 Brno tel. +420 545 555 111 www.cmi.cz 2202 Calibration laboratory No. 2202 accredited by the Czech Accreditation Institute according to ISCMEC 17025:2005 Laboratory: Regional Inspectorate Brno, Okruzni 31, 638 00 Brno Department of primary metrology of liquid flow, flow velocity and heat tel. +420 545 555 111, fax. +420 545 555 183 CERTIFICATE OF CALIBRATION 6015-KL-P0094-17 This certificate is consistent with the capabilities that are included in Appendix of the MRA drawn up by the International Committee for Weights and Measures (CIPM). Under the MRA, all participating institutes recognize the validity of each other's calibration and measurement certificates for the quantities, ranges and measurement uncertainties specified in Appendix (for details see www.bipm.org ). Date of issue: Customer: Meas. instrument: Manufacturer: Type: Serial No.: Specification: 13.2.2017 Page 1 of 2 Remix Bulgaria Ltd. 2 Eng. Georgi Belov St. 1712 Sofia Mass flowmeter Micro Motion CNG050S / 3500 13001689 K-factor: 140.404.50; Kl: 4085,13403 us; K2: 4250.217777 us; mass flow rate factor: 0,9990; zero: 0,02621 us; range: (2-12) kg/min. The results of the calibration have been obtained following the procedures reported in this Certificate and are related only to the date, place and conditions of the calibration. Measurement standards used: Gravimetric test rig SENSUS PREMATEST 32Z/E-95-SP, scale Mettler Toledo, type WM3002-L22, Ser. No. 4230440891, calibrated by CMI OI Brno, certificate of calibration No. 6051-KL-H0188-16; scale Mettler Toledo, type WMHCC300-22, Ser. No. 3113910, calibrated by CMI OI Brno, certificate of calibration No. 6012-KL-V0001-16; scale Mettler Toledo, type WMHA32-S-22, Ser. No. 3113909, calibrated by CMI OI Brno, certificate of calibration No. 6051-KL-H0187-16. Used standards are traceable to national standard of Czech Republic. Date of calibration: 9.-10.2.2017 Calibrated Vedouci oddele I b S ;l sssssds Jafoslav Foltynek Jindffch Bilek This document mayonly be reproduced infull, except with theprior written permission by the issuing laboratory. iKriy-^glg.g-r^glg^g^^g'g-^^y^^

Upload: others

Post on 22-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CERTIFICATE OF CALIBRATION · This certificate is consistent with the capabilities that are included in Appendix Сofthe MRA drawn up by the International Committee for Weights and

reaF.T4r^-^r^^^jr^^-J^Z^gjr£^

-•

tssI

I

Cesky metrologicky institutOkruzni'31,638 00 Brno

tel. +420 545 555 111

www.cmi.cz '"/.,|,Л'ЛК 2202

Calibration laboratory No. 2202 accredited by the Czech Accreditation Institute accordingto ISCMEC 17025:2005

Laboratory: Regional Inspectorate Brno, Okruzni 31, 638 00 BrnoDepartment of primary metrology of liquid flow, flow velocity and heattel. +420 545 555 111, fax. +420 545 555 183

CERTIFICATE OF CALIBRATION

6015-KL-P0094-17

This certificate is consistent with the capabilities that are included in Appendix С of the MRA drawn up by theInternational Committee for Weights and Measures (CIPM). Under the MRA, all participating institutesrecognize the validity of each other's calibration and measurement certificates for the quantities, ranges andmeasurement uncertainties specified in Appendix С (for details see www.bipm.org ).

Date of issue:

Customer:

Meas. instrument:

Manufacturer:

Type:

Serial No.:

Specification:

13.2.2017 Page 1 of 2

Remix Bulgaria Ltd.

2 Eng. Georgi Belov St.

1712 Sofia

Mass flowmeter

Micro Motion

CNG050S / 3500

13001689

K-factor: 140.404.50; Kl: 4085,13403 us; K2: 4250.217777 us; mass flow rate factor:0,9990; zero: 0,02621 us; range: (2-12) kg/min.

The results of the calibration have been obtained following the procedures reported in this Certificate and arerelated only to the date, place and conditions of the calibration.

Measurement

standards used:

Gravimetric test rig SENSUS PREMATEST 32Z/E-95-SP, scale Mettler Toledo, typeWM3002-L22, Ser. No. 4230440891, calibrated by CMI OI Brno, certificate of calibrationNo. 6051-KL-H0188-16; scale Mettler Toledo, type WMHCC300-22, Ser. No. 3113910,calibrated by CMI OI Brno, certificate of calibration No. 6012-KL-V0001-16; scaleMettler Toledo, type WMHA32-S-22, Ser. No. 3113909, calibrated by CMI OI Brno,certificate of calibration No. 6051-KL-H0187-16.

Used standards are traceable to national standard of Czech Republic.

Date of calibration: 9.-10.2.2017

Calibrated Vedouci oddele

I

b

S— ;lsssssds

Jafoslav Foltynek Jindffch Bilek

This document mayonly be reproduced infull, except with theprior written permissionby the issuing laboratory.iKriy-^glg.g-r^glg^g^^g'g-^^y^^

Page 2: CERTIFICATE OF CALIBRATION · This certificate is consistent with the capabilities that are included in Appendix Сofthe MRA drawn up by the International Committee for Weights and

CERTIFICATE OF CALIBRATION 6015-KL-P0094-17 Page 2 of 2

Calibration procedure:

Meter output:

Place of calibration:

Ambient conditions:

Calibration conditions:

Results of calibration:

Flying start gravimetric method acording to procedure no. 615-MP-C142.

Pulse output (1000 imp/kg).

CMI OI Brno

Temperature (23 ± 2) °C; RH (20 ± 10) %

Water temperature (20 ± 2)

Before adjustment: mass flow rate factor 1,0030

Flowrate Error Expandeduncertainty

[kg/min] [%] [%]0,13 0.41 0.10

0,46 0,41 0,10

0,71 0,42 0,10

After adjustment: mass flow rate factor 0,9990

Flowrate Error Expandeduncertainty

[kg/min] [%] [%]0,13 0.06 0,10

0,44 0,04 0,10

0,70 0,02 0,10

The standard uncertainty of measurement has been determined in accordance with EA^t/02 document. Thereported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied bythe coverage factor к corresponding to a coverage probability of approximately 95 %, which for normaldistribution corresponds to a coverage factor к = 2.

-

End of calibration certificate.

Cesky metrologicky institutOblastni inspektorat Brno

Okruzni 31

638 00 Brno

-18-

This document may only be reproduced infull, except with the prior written permission by the issuing laboratory.