CERTIFICATENo. P16221549
Order No. 316649 Page 1 of 2
Date of issue 10-11-2016
Okhyun JeonCertification Department
Nemko ASGaustadalléen 30, P.O. Box 73 Blindern, 0314 Oslo, NorwayTEL +47 22 96 03 30 FAX +47 22 96 05 50 EMAIL [email protected] NUMBER NO974404532 nemko.com
Product Medical Power Supply (for building-in)
Applicant Bel Fuse Inc.206 Van Vorst St.Jersey City, NJ 07302USA
Manufacturer Bel Fuse Inc.206 Van Vorst St.Jersey City, NJ 07302USA
Factory EOS Power India PVT Ltd.Unit #57, SDF II, Seepz-Sez, Andheri (East), Mumbai,400 096India
Ratings Input: 4.0A Max, 100- 240V~, 47-63Hz, Output: 275W Max
Trade mark
Model / Type Ref. MBC275-ZVXXL-YYY
Principal characteristics Cl. l, The model designation variables are defined as follows: Z= 1 is formodels without PG/PF and 0 is with PG/PF, V= T or 0 to 9 denoting the outputconnector. X = any alpha character or numeric from 0-9 (Representing theoutput voltage which can be any voltage in the range 12V to 58V, max currentas stated in ref table) Y = any alpha or numeric character or blank (denotesminor output variation and /or minor SELV circuit variations)
A sample of the product was testedand found to be in conformity with
MED EN 60601-1:2006;A1
Validity This certificate documents conformity with the standards shown, and also applies as license foruse of Nemkos name and certification mark. The certificate and license is valid as long as theapplicable conditions are complied with, and provided that any changes to the product are notifiedto Nemko for acceptance prior to implementation.New standards or amendments to the standards may imply that the product design must beupdated and/or that re-testing and re-certification is necessary.
Additional information
Compliance with requirements regarding building-in, protection against electric shock andElectromagnetic Compatibility (EMC) must be checked when the equipment is built-in a completedproduct or forms a part of a complete system.
Additional model(s) (1) See page 2
CERTIFICATENo. P16221549
Order No. 316649 Page 2 of 2
Date of issue 10-11-2016
Okhyun JeonCertification Department
Nemko ASGaustadalléen 30, P.O. Box 73 Blindern, 0314 Oslo, NorwayTEL +47 22 96 03 30 FAX +47 22 96 05 50 EMAIL [email protected] NUMBER NO974404532 nemko.com
Additional model(s)
Product Medical Power Supply (for building-in)Pos. No 1Model / Type Ref. MBC180-ZVXXL-YYYTrade mark (if different from page 1)Rating Input: 2.5A Max, 100- 240V~, 47-63Hz, Output: 180W MaxPrincipal characteristics Cl. l, The model designation variables are defined as follows: Z= 1 is for
models without PG/PF and 0 is with PG/PF, V= T or 0 to 9 denoting the outputconnector. X = any alpha character or numeric from 0-9 (Representing theoutput voltage which can be any voltage in the range 12V to 58V, max currentas stated in ref table) Y = any alpha or numeric character or blank (denotesminor output variation and /or minor SELV circuit variations)
λºò Ý»®¬·ºò Ò±ò
ÒÑçìéçë
Ù¿«¬¿¼¿´´7»² íðÒÑóðíéí Ñ´±ô Ò±®©¿§
Ü¿¬»æ ðéóïïóîðïê Í·¹²¿¬«®»æ Í©»»¬ Ç«¿²Ý»®¬·º·½¿¬·±² Ü»°¿®¬³»²¬
׫»¼ îððéóðì ïñï
×ÛÝ ÍÇÍÌÛÓ ÚÑÎ ÓËÌËßÔ ÎÛÝÑÙÒ×Ì×ÑÒ ÑÚ ÌÛÍÌ ÝÛÎÌ×Ú×ÝßÌÛÍ ÚÑÎ ÛÔÛÝÌÎ×ÝßÔÛÏË×ÐÓÛÒÌ ø×ÛÝÛÛ÷ ÝÞ ÍÝØÛÓÛ
ÍÇÍÌÛÓÛ ÝÛ× ÜßÝÝÛÐÌßÌ×ÑÒ ÓËÌËÛÔÔÛ ÜÛ ÝÛÎÌ×Ú×ÝßÌÍ ÜÛÍÍß×Í ÜÛÍÛÏË×ÐÛÓÛÒÌÍ ÛÔÛÝÌÎ×ÏËÛÍ ø×ÛÝÛÛ÷ ÓÛÌØÑÜÛ ÑÝ
ÝÞ ÌÛÍÌ ÝÛÎÌ×Ú×ÝßÌÛ ÝÛÎÌ×Ú×ÝßÌ ÜùÛÍÍß× ÑÝ
Ю±¼«½¬Ð®±¼«·¬
Ó»¼·½¿´ б©»® Í«°°´§ øº±® ¾«·´¼·²¹ó·²÷
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» ¿°°´·½¿²¬Ò±³ »¬ ¿¼®»» ¼« ¼»³¿²¼»«®
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» ³¿²«º¿½¬«®»®Ò±³ »¬ ¿¼®»» ¼« º¿¾®·½¿²¬
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» º¿½¬±®§Ò±³ »¬ ¿¼®»» ¼» ´ù«·²»
Ò±¬»æ ɸ»² ³±®» ¬¸¿² ±²» º¿½¬±®§ô °´»¿» ®»°±®¬ ±² °¿¹» îÒ±¬»æ Ô±®¯«» ·´ § °´« ¼ù«²» «·²»ô ª»«·´´»¦ «¬·´·»® ´¿ ¼»«¨·8³» °¿¹»
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ±² °¿¹» î
ο¬·²¹ ¿²¼ °®·²½·°¿´ ½¸¿®¿½¬»®·¬·½Ê¿´»«® ²±³·²¿´» »¬ ½¿®¿½¬7®·¬·¯«» °®·²½·°¿´»
ײ°«¬æ ìòðß Ó¿¨ô ïððóîìðÊ¢ô ìéóêíئôÑ«¬°«¬æ îéëÉ Ó¿¨Ý´ò ´
Ì®¿¼»³¿®µ ø·º ¿²§÷Ó¿®¯«» ¼» º¿¾®·¯«» ø· »´´» »¨·¬»÷
̧°» ±º Ó¿²«º¿½¬«®»®� Ì»¬·²¹ Ô¿¾±®¿¬±®·» «»¼Ì§°» ¼» °®±¹®¿³³» ¼« ´¿¾±®¿¬±·®» ¼�»¿· ½±²¬®«½¬»«®
Ó±¼»´ ñ ̧°» λºòλºò Ü» ¬§°»
ÓÞÝîéëóÆÊÈÈÔóÇÇÇ
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ø·º ²»½»¿®§ ³¿§ ¿´± ¾»®»°±®¬»¼ ±² °¿¹» î÷Ô» ·²º±®³¿¬·±² ½±³°´7³»²¬¿·®» ø· ²7½»¿·®»ô°»«ª»²¬ 6¬®» ·²¼·¯«7 «® ´¿ ¼»«¨·8³» °¿¹»
Æã ï ±® ðô Êã Ì ±® 𠬱 çô È ã ¿²§ ¿´°¸¿ ½¸¿®¿½¬»® ±® ²«³»®·½ º®±³ ðóçô Ç ã¿²§ ¿´°¸¿ ±® ²«³»®·½ ½¸¿®¿½¬»® ±® ¾´¿²µ
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ±² °¿¹» î
ß ¿³°´» ±º ¬¸» °®±¼«½¬ ©¿ ¬»¬»¼ ¿²¼ º±«²¼¬± ¾» ·² ½±²º±®³·¬§ ©·¬¸Ë² 7½¸¿²¬·´´±² ¼» ½» °®±¼«·¬ ¿ 7¬7 »¿§7 »¬ ¿ 7¬7½±²·¼7®7 ½±²º±®³» @ ´¿
×ÛÝ êðêðïóïæîððëô ×ÛÝ êðêðïóïæîððëñßÓÜïæîðïî
ß ¸±©² ·² ¬¸» Ì»¬ λ°±®¬ λºò Ò±ò ©¸·½¸ º±®³ °¿®¬±º ¬¸· Ý»®¬·º·½¿¬»Ý±³³» ·²¼·¯«7 ¼¿² ´» ο°°±®¬ ¼»¿· ²«³7®± ¼»®7º7®»²½» ¯«· ½±²¬·¬«» °¿®¬·» ¼» ½» Ý»®¬·º·½¿¬
íïêêìè
̸· ÝÞ Ì»¬ Ý»®¬·º·½¿¬» · ·«»¼ ¾§ ¬¸» Ò¿¬·±²¿´ Ý»®¬·º·½¿¬·±² Þ±¼§Ý» Ý»®¬·º·½¿¬ ¼»¿· ÑÝ »¬ 7¬¿¾´· °¿® ´ùÑ®¹¿²·³» Ò¿¬·±²¿´ ¼» Ý»®¬·º·½¿¬·±²
λºò Ý»®¬·ºò Ò±ò
ÒÑçìéçé
Ù¿«¬¿¼¿´´7»² íðÒÑóðíéí Ñ´±ô Ò±®©¿§
Ü¿¬»æ ðéóïïóîðïê Í·¹²¿¬«®»æ Í©»»¬ Ç«¿²Ý»®¬·º·½¿¬·±² Ü»°¿®¬³»²¬
׫»¼ îððéóðì ïñï
×ÛÝ ÍÇÍÌÛÓ ÚÑÎ ÓËÌËßÔ ÎÛÝÑÙÒ×Ì×ÑÒ ÑÚ ÌÛÍÌ ÝÛÎÌ×Ú×ÝßÌÛÍ ÚÑÎ ÛÔÛÝÌÎ×ÝßÔÛÏË×ÐÓÛÒÌ ø×ÛÝÛÛ÷ ÝÞ ÍÝØÛÓÛ
ÍÇÍÌÛÓÛ ÝÛ× ÜßÝÝÛÐÌßÌ×ÑÒ ÓËÌËÛÔÔÛ ÜÛ ÝÛÎÌ×Ú×ÝßÌÍ ÜÛÍÍß×Í ÜÛÍÛÏË×ÐÛÓÛÒÌÍ ÛÔÛÝÌÎ×ÏËÛÍ ø×ÛÝÛÛ÷ ÓÛÌØÑÜÛ ÑÝ
ÝÞ ÌÛÍÌ ÝÛÎÌ×Ú×ÝßÌÛ ÝÛÎÌ×Ú×ÝßÌ ÜùÛÍÍß× ÑÝ
Ю±¼«½¬Ð®±¼«·¬
Ó»¼·½¿´ б©»® Í«°°´§ øº±® ¾«·´¼·²¹ó·²÷
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» ¿°°´·½¿²¬Ò±³ »¬ ¿¼®»» ¼« ¼»³¿²¼»«®
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» ³¿²«º¿½¬«®»®Ò±³ »¬ ¿¼®»» ¼« º¿¾®·½¿²¬
Ò¿³» ¿²¼ ¿¼¼®» ±º ¬¸» º¿½¬±®§Ò±³ »¬ ¿¼®»» ¼» ´ù«·²»
Ò±¬»æ ɸ»² ³±®» ¬¸¿² ±²» º¿½¬±®§ô °´»¿» ®»°±®¬ ±² °¿¹» îÒ±¬»æ Ô±®¯«» ·´ § °´« ¼ù«²» «·²»ô ª»«·´´»¦ «¬·´·»® ´¿ ¼»«¨·8³» °¿¹»
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ±² °¿¹» î
ο¬·²¹ ¿²¼ °®·²½·°¿´ ½¸¿®¿½¬»®·¬·½Ê¿´»«® ²±³·²¿´» »¬ ½¿®¿½¬7®·¬·¯«» °®·²½·°¿´»
ײ°«¬æ îòëß Ó¿¨ô ïððó îìðÊ¢ô ìéóêíئôÑ«¬°«¬æ ïèðÉ Ó¿¨Ý´ò×
Ì®¿¼»³¿®µ ø·º ¿²§÷Ó¿®¯«» ¼» º¿¾®·¯«» ø· »´´» »¨·¬»÷
̧°» ±º Ó¿²«º¿½¬«®»®� Ì»¬·²¹ Ô¿¾±®¿¬±®·» «»¼Ì§°» ¼» °®±¹®¿³³» ¼« ´¿¾±®¿¬±·®» ¼�»¿· ½±²¬®«½¬»«®
Ó±¼»´ ñ ̧°» λºòλºò Ü» ¬§°»
ÓÞÝïèðóÆÊÈÈÔóÇÇÇ
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ø·º ²»½»¿®§ ³¿§ ¿´± ¾»®»°±®¬»¼ ±² °¿¹» î÷Ô» ·²º±®³¿¬·±² ½±³°´7³»²¬¿·®» ø· ²7½»¿·®»ô°»«ª»²¬ 6¬®» ·²¼·¯«7 «® ´¿ ¼»«¨·8³» °¿¹»
Æã ï ±® ðô Êã Ì ±® 𠬱 çô È ã ¿²§ ¿´°¸¿ ½¸¿®¿½¬»® ±® ²«³»®·½ º®±³ ðóçô Ç ã¿²§ ¿´°¸¿ ±® ²«³»®·½ ½¸¿®¿½¬»® ±® ¾´¿²µ
ß¼¼·¬·±²¿´ ·²º±®³¿¬·±² ±² °¿¹» î
ß ¿³°´» ±º ¬¸» °®±¼«½¬ ©¿ ¬»¬»¼ ¿²¼ º±«²¼¬± ¾» ·² ½±²º±®³·¬§ ©·¬¸Ë² 7½¸¿²¬·´´±² ¼» ½» °®±¼«·¬ ¿ 7¬7 »¿§7 »¬ ¿ 7¬7½±²·¼7®7 ½±²º±®³» @ ´¿
×ÛÝ êðêðïóïæîððëô ×ÛÝ êðêðïóïæîððëñßÓÜïæîðïî
ß ¸±©² ·² ¬¸» Ì»¬ λ°±®¬ λºò Ò±ò ©¸·½¸ º±®³ °¿®¬±º ¬¸· Ý»®¬·º·½¿¬»Ý±³³» ·²¼·¯«7 ¼¿² ´» ο°°±®¬ ¼»¿· ²«³7®± ¼»®7º7®»²½» ¯«· ½±²¬·¬«» °¿®¬·» ¼» ½» Ý»®¬·º·½¿¬
íïêêìè
̸· ÝÞ Ì»¬ Ý»®¬·º·½¿¬» · ·«»¼ ¾§ ¬¸» Ò¿¬·±²¿´ Ý»®¬·º·½¿¬·±² Þ±¼§Ý» Ý»®¬·º·½¿¬ ¼»¿· ÑÝ »¬ 7¬¿¾´· °¿® ´ùÑ®¹¿²·³» Ò¿¬·±²¿´ ¼» Ý»®¬·º·½¿¬·±²
Test Report issued under the responsibility of
www.nemko.com
This Test Report, when bearing the Nemko name and logo is only valid when issued by a Nemko laboratory, or by a laboratory having special agreement with Nemko.
IEC 60601-1 Medical electrical equipment
Part 1: General requirements for basic safety and essential performanceReport Reference No. .................... : 316648
Date of issue .................................. : 04 November 2016
Total number of pages .................. : 303
CB Testing Laboratory .................. : Nemko USA Inc. Phone: +1 760 444 3500
Address .......................................... : 2210 Faraday Avenue, Suite 150, Carlsbad, CA 92008, USA
Applicant’s name ........................... : EOS Power India PVT Ltd.
Address .......................................... : Unit #57, SDF II, Seepz-Sez, Andheri (E), Mumbai, 400096, India
Test specification:
Standard ......................................... : IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint)
Test procedure ............................... : CB Scheme
Non-standard test method………..: N/A
Test Report Form No. .................... : IEC60601_1K
Test Report Form Originator ......... : UL(US)
Master TRF ..................................... : 2015-11
Copyright © 2015 Worldwide System for Conformity Testing and Certification of Electrotechnical Equipment and Components (IECEE), Geneva, Switzerland. All rights reserved.
This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context.
If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo and the reference to the CB Scheme procedure shall be removed.
This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory and appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02.
General disclaimer:
The test results presented in this report relate only to the object tested. This report shall not be reproduced, except in full, without the written approval of the Issuing CB testing laboratory. The authenticity of this Test Report and its contents can be verified by contacting the NCB, responsible for this Test Report.
Page 2 of 303 Report No. 316648
TRF No. IEC60601_1K
Test item description ....................... : Medical Power Supply (for building-in)
Trade Mark ........................................ :
Manufacturer .................................... : EOS Power India PVT Ltd.
Model/Type reference ...................... : MULP275-ZVXX-YYY, MULP180-ZVXX-YYY
MBC275-ZVXXL-YYY, MBC180-ZVXXL-YYY
Ratings ............................................. : MULP275-ZVXX-YYY, MBC275-ZVXXL-YYY Input: 4.0A Max, 100- 240V~, 47-63Hz, Output: 275W Max MULP180-ZVXXL-YYY, /MBC180-ZVXXL-YYY Input: 2.5A Max, 100- 240V~, 47-63Hz, Output: 180W Max
(Refer to General product information for output rating conditions and variable definition)
Testing procedure and testing location:
Nemko USA Inc. Nemko USA Inc.
Testing location/ address ............................ : 2210 Faraday Ave. Suite 150, Carlsbad, CA 92008, USA
Testing location/ address ............................ :
Tested by (name + signature) ..................... : Luis Gonzalez
Approved by (name + signature) ................ : David Atkinson
Testing procedure: CTF Stage 1:
Testing location/ address ............................ :
Tested by (name, function, signature) ....... :
Approved by (name, function, signature) .. :
Testing procedure: CTF Stage 2:
Testing location/ address ............................ :
Tested by (name, function, signature) ....... :
Witnessed by (name, function, signature) . :
Approved by (name, function, signature) .. :
Testing procedure: CTF Stage 3:
Testing procedure: CTF Stage 4:
Testing location/ address ............................ :
Tested by (name, function, signature) ....... :
Witnessed by (name, function, signature) . :
Approved by (name, function, signature) .. :
Supervised by (name, function, signature) :
Page 3 of 303 Report No. 316648
TRF No. IEC60601_1K
List of Attachments (including a total number of pages in each attachment):
Attachment 1: National differences (23 pages)
Summary of testing:
Clause Requirement + Test Comment
All Information about the standards, documents considered
European Countries, EN 60601-1 (2006) +A11 (2011) + AM1:2013 +AC:2014 +A12:2014.
Riting for updated from SDP, efer also to National Differences attachment
All General The equipment is an open-frame, Class I switch mode power supply with universal AC input and multiple DC voltage outputs for building-in. This report covers multiple models and all comments / tests apply to all models unless otherwise indicated. Testing was conducted on various models as indicated.
All Risk Management The manufacturer performed the Risk Management process according to ISO 14971, 2nd edition.
All Rated ambient temperature range The equipment is rated for an operational environment of: Temperature: 0°C - 70°C, refer to General product info. Humidity: 5% - 95% RH, non-condensing. The Tma for the maximum output rating is 50°C for most models. Power supplies can operate in up to a 70°C ambient with a derated output load, refer to General product information and appended table 11.1.1.
4.3 Essential performance functions identified according to manufacturer’s policy for risk acceptability in risk management file
Essential performance shall be determined when installed in the end-use equipment as part of the risk management process for the end product. The equipment was determined to have no essential performance for this evaluation.
4.6 Risk management process identifies parts that can come into contact with patient but not defined as applied parts subjected to the requirements for applied parts, except for 7.2.10
Output of the PSU was not considered for direct contact to the patient (as part of this evaluation); additional evaluation will be required in the end installation, however the power supply was designed to meet the requirements of a BF Applied Part with insulation evaluated according to Means of Patient Protection (MOPP).
4.8 Components of ME equipment The optocouplers have been separately tested and approved to the relevant component standard (EN 60747-5-5) by VDE. The component certification is considered to provide equivalent safety for two means of patient protection (2 MOPPs), refer to Insulation Diagrams for defined operator and patient protection. The MOVs are certified, however the needed level of surge protection is dependent on the end-product installation. The PCB, transformer and bridging Y-capacitors comply with IEC 60601-1 (ed.3) requirements for patient protection. Acceptance is at the discretion of the accepting NCB or agency. Acceptance of the optocoupler is at the discretion of the accepting NCB or agency. The PCB and transformer comply with IEC 60601-1 (ed.3) requirements for patient protection.
Page 4 of 303 Report No. 316648
TRF No. IEC60601_1K
Summary of testing:
Clause Requirement + Test Comment
4.11 Power input This input testing was conducted on the worst-case models (highest output power and highest input current draw) and test results are representative of the entire series and derated output testing for elevated ambient temperatures, refer to appended table 4.11.
5.9 Determination of applied parts and accessible parts
Output of the PSU was not considered for direct contact to the patient (as part of this evaluation); additional evaluation will be required in the end installation, however the power supply was designed to meet the requirements of a BF Applied Part with insulation evaluated according to Means of Patient Protection (MOPP).
7 Identification, markings and documents
Instructions and marking related to safety shall be in a language which is acceptable in the country in which the equipment is to be sold. English version evaluated.
7.9 Accompanying documents The equipment is a PSU for building-in. The end product evaluation shall ensure that the requirements related to accompanying documents complying with this standard are met in which is provided in the language required for the market where the equipment is sold.
8.3 Classification of applied parts Output of the PSU was not considered for direct contact to the patient (as part of this evaluation); additional evaluation will be required in the end installation, however the power supply was designed to meet the requirements of a BF Applied Part with insulation evaluated according to Means of Patient Protection (MOPP). Additional evaluation will be required in the end-use equipment.
8.4 Limitation of voltage, current or energy
Output power of modeles MULP275 series exceeded 240VA, therefore the output is considered to be an energy hazard. The equipment is a PSU for building-in. To be further evaluated in the end use product. For MULP180 series the output does not exceeded 240VA, therefore the output is considered not to be an energy hazard.
8.4.1 Accessible part including applied parts.
Output of the PSU was not considered for direct contact to the patient (as part of this evaluation); additional evaluation will be required in the end installation, however the power supply was designed to meet the requirements of a BF Applied Part with insulation evaluated according to Means of Patient Protection (MOPP).
8.5 Separation of Parts - Means of Protection (MOP)
The insulation was evaluated according to Means of Patient Protection (MOPP).
8.5.2.1 F-type Applied Part isolation (MOPP)
Output of the PSU was not considered for direct contact to the patient (as part of this evaluation); additional evaluation will be required in the end installation, however the power supply was designed to meet the requirements of a BF Applied Part with insulation evaluated according to Means of Patient Protection (MOPP).
8.7 Leakage currents and patient auxiliary currents
The equipment is a PSU for building-in. Earth and touch leakage current, (patient and voltage on applied part conducted by the request of manufacturer) were performed. To be re-evaluated when installed in the end product.
8.9 Creepage distances and air clearances
The equipment is a PSU for building-in. The creepage and clearance distance need to be evaluated when installed in the end equipment.
Page 5 of 303 Report No. 316648
TRF No. IEC60601_1K
Summary of testing:
Clause Requirement + Test Comment
11.1 Protection against excessive temperatures and other hazards
This report covers a series of power supply models. Testing was performed on various models as indicated in this report. This testing was considered representative for all models. Unless otherwise stated, all tests were performed on worst case Model. Note: The models can be mounted in various positions within an end system as indicated in the data/installation sheet with the worst case being horizontal standard position.
11.1.1 Maximum temperature during normal use
The Tma for the maximum output rating is 50°C for most models. Power supplies can operate in up to a 70°C ambient with a derated output load, refer to General product information and appended table 11.1.1.
11.3b Constructional requirements for fire enclosures of ME equipment
The equipment is a PSU for building-in which provide >15W output. The PSU must be provided with a suitable fire enclosure. The fire enclosure shall be evaluated in the end product.
11.7 ME equipment, ME system, and accessories coming into direct or indirect contact with biological tissues, cells, or body fluids assessed and documented according to ISO 10993
The equipment is a PSU for building-in. The end-use equipment must be evaluated to determine compliance with the requirements of ISO 10993-1.
17 Electromagnetic compatibility of ME equipment and ME systems
The equipment is a PSU for building-in. The end-use equipment must be evaluated to determine compliance with the latest EMC requirements according to IEC 60601-1-2.
Tests performed (name of test and test clause):
4.11: Power input 5.7: Humidity and preconditioning treatment 7.1.3: Durability of marking 8.4.2: Working voltage / Power measurement 8.4.3: Plug-measurement of voltage 8.6.4 Protective earthing 8.7: Leakage current 8.8.3: Dielectric strength 8.8.4.1: Mechanical strength and resistance to heat 11.1: Excessive temperatures 11.8: Interruption of power supply 13.2: Single fault conditions 15.5: Transformers
Testing location: See Page 2
Summary of compliance with National Differences List of countries addressed:
EN 60601-1 (2006) +A11 (2011) + AM1:2013 +AC:2014 +A12:2014. The list of countries recognizing the CB Scheme is actively updated on the IECEE website. (http://www.iecee.org/)
All CB members as listed in the IECEE Online Bulletin Canada, Switzerland, Japan, Korea, United States.
The product fulfils the requirements of IEC/EN 60601-1 Edition 3.0 + AM1
Page 6 of 303 Report No. 316648
TRF No. IEC60601_1K
Copy of marking plate
The artwork below may be only a draft. The use of certification marks on a product must be authorized by the respective NCBs that own these marks.
MULP180 series
Page 7 of 303 Report No. 316648
TRF No. IEC60601_1K
Copy of marking plate
The artwork below may be only a draft. The use of certification marks on a product must be authorized by the respective NCBs that own these marks.
MULP275 series
MBC180 series (representative of other models)
Page 8 of 303 Report No. 316648
TRF No. IEC60601_1K
Copy of marking plate
The artwork below may be only a draft. The use of certification marks on a product must be authorized by the respective NCBs that own these marks.
MBC275 series (representative of other models)
Calibration All instruments used in the tests given in this test report are calibrated and
traceable to national or international standards. Further information about traceability will be given on request.
Measurement uncertainty
Measurement uncertainties are calculated for all instruments and instrument set-ups given in this report. Calculations are based on the principles given in the standard EA-4/02 (Dec. 1999), IEC Guide 115:2007 and other relevant internal Nemko-procedures. Further information about measurement uncertainties will be given on request.
Evaluation of results If not explicitly stated otherwise in the standard, the test is passed if the measured value is equal to or below (above) the limit line, regardless of the measurement uncertainty. If the measured value is above (below) the limit line, the test is not passed - ref IEC Guide 115:2007. The instrumentation accuracy is within limits agreed by IECEE-CTL.
Page 9 of 303 Report No. 316648
TRF No. IEC60601_1K
GENERAL INFORMATION
Test item particulars (see also Clause 6):
Classification of installation and use ................................ : Class I, PSU for building-in
Device type (component/sub-assembly/ equipment/ system) ................................................................................. :
Component, PSU for building-in
Intended use (Including type of patient, application location) ............................................................................... :
PSU for building-in
Mode of operation .............................................................. : Continuous
Supply connection ............................................................. : PSU for building-in
Accessories and detachable parts included ..................... : None
Other options include .......................................................... : None
Testing
Date of receipt of test item(s)…………………………………: September 2016
Dates tests performed………………………………………….: October 2016
Possible test case verdicts:
- test case does not apply to the test object ........................ : N/A
- test object does meet the requirement .............................. : Pass (P)
- test object was not evaluated for the requirement ................ : N/E (collateral standards only)
- test object does not meet the requirement ......................... : Fail (F)
Abbreviations used in the report:
- normal condition .................................................... : N.C. - single fault condition .................. : S.F.C. - means of Operator protection ............................... : MOOP - means of Patient protection ...... : MOPP
General remarks: "(See Attachment #)" refers to additional information appended to the report. "(See appended table)" refers to a table appended to the report. The tests results presented in this report relate only to the object tested. This report shall not be reproduced except in full without the written approval of the testing laboratory. List of test equipment must be kept on file and available for review. Additional test data and/or information provided in the attachments to this report. Throughout this report a comma / point is used as the decimal separator.
Manufacturer’s Declaration per sub-clause 4.2.5 of IECEE 02:
The application for obtaining a CB Test Certificate includes more than one factory location and a declaration from the Manufacturer stating that the sample(s) submitted for evaluation is (are) representative of the products from each factory has been provided .....................................................:
Yes
Not applicable
When differences exist; they shall be identified in the General product information section.
Name and address of factory (ies) ........................ :
EOS Power India PVT Ltd., Unit #57, SDF II, Seepz-Sez, Andheri (East), Mumbai 400096, India
Page 10 of 303 Report No. 316648
TRF No. IEC60601_1K
General product information:
MULP275-ZVXX-YYY and MULP180-ZVXX-YYY are a series of open frame Switch Mode Power Supply intended for use in medical equipment end-products These are component power supply intended for building-in Class I end-products. The model series has an input range of 100-240V ac, with manufacturer tolerances of -20% to +10% (ranging from 80-264V ac).The power supply provides single output, ranging from 12V to 58Vdc that is based on configuration (different voltage and current ratings) of model and 12Vdc auxiliary (for fan). The equipment is provided with a PCB-mount input connector, for mains voltage (or equivalent). Model Differences: Models MULP275-ZVXX-YYY and MULP180-ZVXX-YYY are all similar in construction to each other, except for the power ratings (275W, 180W) and differences in the output windings in the transformer, PCB layouts, secondary components, and minor primary component value changes. Unless otherwise noted, all comments are for all models. 1. Model MULP275-1XXX-YYY series is identical to Model MULP275-0XXX-YYY series, except width of PCB reduces by 12.7 (0.5”) due to no PG/PF signal (power good /power fail circuitry), therefore two different layouts are provided. 2. Model MULP180-1XXX-YYY series is identical to Model MULP180-0XXX-YYY series, except width of PCB reduces by 12.7 (0.5”) due to no PG/PF signal (power good /power fail circuitry), therefore two different layouts are provided. Optional configurations: 1) Fuse - Equipment can be configured with a single fuse (F1) in line phase and optionally provided with a second fuse (F2) in neutral phase. When fuse is not provided a jumper is provided in its place. 2) Ferrite beads - Equipment can be optionally configured with a ferrite beads on the leads of the input fuse (F1 or F2) in addition to the ferrite beads, insulation sheet is provided to cover the ferrite beads to maintain 1 MOOP between the input (non-fused circuit) and output leads (fused circuit).EOS model (MULP275-ZVXX-YYY / MULP180-ZVXX-YYY) designations format is as follows: Z= 1 is for models without PG/PF and 0 is with PG/PF. V= 0 to 9 denoting the output connector. X = any numeric from 0-9 (Representing the output voltage which can be any voltage in the range 12V to 58V, max current as stated in ref table) Y = any alpha or numeric character or blank (denotes minor output variation and /or minor SELV circuit variations) Alternative trademark Bel Power Solution is added to this report with the same construction. The only difference is in the makings for marketing purposes. Testing on the EOS models is representative of the Bel Power Solution models. The model designation variables for Bel Power Solution MBC275-ZVXXL-YYY / MBC180-ZVXXL-YYY are defined as follows: Z= 1 is for models without PG/PF and 0 is with PG/PF, V= T or 0 to 9 denoting the output connector. X = any alpha character or numeric from 0-9 (Representing the output voltage which can be any voltage in the range 12V to 58V, max current as stated in ref table) Y = any alpha or numeric character or blank (denotes minor output variation and /or minor SELV circuit variations) EOS to Bel Power Solution model designations.
Models EOS Terminal Output connector SIGNAL PF/PG Output V Output A Models BEL MUPL180-1012 Screw NO 12 15/10.00 MBC180-1T12L MUPL180-1015 Screw NO 15 12/8.0 MBC180-1T15L MUPL180-1024 Screw NO 24 7.5/5.0 MBC180-1T24L MUPL180-1030 Screw NO 30 6/4.0 MBC180-1T30L MUPL180-1048 Screw NO 48 3.75/2.5 MBC180-1T48L MUPL180-1058 Screw NO 58 3.10/2.07 MBC180-1T58L MUPL180-1312 Header NO 12 15/10.00 MBC180-1012L MUPL180-1315 Header NO 15 12/8.0 MBC180-1015L MUPL180-1324 Header NO 24 7.5/5.0 MBC180-1024L MUPL180-1330 Header NO 30 6/4.0 MBC180-1030L MUPL180-1348 Header NO 48 3.75/2.5 MBC180-1048L MUPL180-1358 Header NO 58 3.13/2.07 MBC180-1058L MUPL180-0012 Screw YES 12 15/10.00 MBC180-0T12L MUPL180-0015 Screw YES 15 12/8.0 MBC180-0T15L
Page 11 of 303 Report No. 316648
TRF No. IEC60601_1K
General product information:
MUPL180-0024 Screw YES 24 7.5/5.0 MBC180-0T24L MUPL180-0030 Screw YES 30 6/4.0 MBC180-0T30L MUPL180-0048 Screw YES 48 3.75/2.5 MBC180-0T48L MUPL180-0058 Screw YES 58 3.10/2.07 MBC180-0T58L MUPL180-0312 Header YES 12 15/10.00 MBC180-0012L MUPL180-0315 Header YES 15 12/8.0 MBC180-0015L MUPL180-0324 Header YES 24 7.5/5.0 MBC180-0024L MUPL180-0330 Header YES 30 6/4.0 MBC180-0030L MUPL180-0348 Header YES 48 3.75/2.5 MBC180-0048L MUPL180-0358 Header YES 58 3.10/2.07 MBC180-0058L MUPL275-1012 Screw NO 12 22.92/13.33 MBC275-1T12L MUPL275-1015 Screw NO 15 18.33/10.67 MBC275-1T15L MUPL275-1024 Screw NO 24 11.46/6.67 MBC275-1T24L MUPL275-1030 Screw NO 30 9.17/5.33 MBC275-1T30L MUPL275-1048 Screw NO 48 5.73/3.33 MBC275-1T48L MUPL275-1058 Screw NO 58 4.74/2.76 MBC275-1T58L MUPL275-1312 Header NO 12 22.92/13.33 MBC275-1012L MUPL275-1315 Header NO 15 18.33/13.67 MBC275-1015L MUPL275-1324 Header NO 24 11.46/6.67 MBC275-1024L MUPL275-1330 Header NO 30 9.17/5.33 MBC275-1030L MUPL275-1348 Header NO 48 5.73/3.33 MBC275-1048L MUPL275-1358 Header NO 58 4.74/2.76 MBC275-1058L MUPL275-0012 Screw YES 12 22.92/13.33 MBC275-0T12L MUPL275-0015 Screw YES 15 18.33/10.67 MBC275-0T15L MUPL275-0024 Screw YES 24 11.46/6.67 MBC275-0T24L MUPL275-0030 Screw YES 30 9.17/5.33 MBC275-0T30L MUPL275-0048 Screw YES 48 5.73/3.33 MBC275-0T48L MUPL275-0058 Screw YES 58 4.74/2.76 MBC275-0T58L MUPL275-0312 Header YES 12 22.92/13.33 MBC275-0012L MUPL275-0315 Header YES 15 18.33/10.67 MBC275-0015L MUPL275-0324 Header YES 24 11.46/6.67 MBC275-0024L MUPL275-0330 Header YES 30 9.17/5.33 MBC275-0030L MUPL275-0348 Header YES 48 5.73/3.33 MBC275-0048L MUPL275-0358 Header YES 58 4.74/2.76 MBC275-0058L
See below for representative Model Ratings Table; for 50°C ambient with 13 LFM forced air cooling from Inductor L3 side blowing inward. Derated power ratings at 70°C ambient are also specified below. The power supply total output power ratings vary based on cooling, output voltage, input voltage and ambient temperature, refer to the tables below and on the following page regarding operating and testing conditions for different configurations. Model:MULP275 SERIES
Vac= 100 to 264Vac
Models Voltage Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP275-1012 12 275 (A1) 137.5 (A2) 160 (A3) 80 (A4) MULP275-1015 15 275 137.5 160 80 MULP275-1024 24 275 137.5 160 80 MULP275-1030 30 275 137.5 160 80 MULP275-1048 48 275 137.5 160 80MULP275-1058 58 275 (A5) 137.5 (A6) 160 (A7) 80 (A8)
Page 12 of 303 Report No. 316648
TRF No. IEC60601_1K
General product information:
Model:MULP275 SERIES
Vac= 90 - < 100Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP275-1012 12 240 (B1) 120 (B2) 120 (B3) 60 (B4) MULP275-1015 15 240 120 120 60 MULP275-1024 24 240 120 120 60 MULP275-1030 30 240 120 120 60 MULP275-1048 48 240 120 120 60 MULP275-1058 58 240 (B5) 120 (B6) 120 (B7) 60 (B8)
Model:MULP275 SERIES Vac= 85 - < 90Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP275-1012 12 220 (C1) 110 (C2) 115 (C3) 57.5 (C4) MULP275-1015 15 220 110 115 57.5 MULP275-1024 24 220 110 115 57.5 MULP275-1030 30 220 110 115 57.5 MULP275-1048 48 220 110 115 57.5 MULP275-1058 58 220 (C5) 110 (C6) 115 (C7) 57.5 (C8)
Model:MULP275 SERIES Vac= 80 - < 85Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP275-1012 12 200 (D1) 100 (D2) 110 (D3) 55 (D4) MULP275-1015 15 200 100 110 55 MULP275-1024 24 200 100 110 55 MULP275-1030 30 200 100 110 55 MULP275-1048 48 200 100 110 55 MULP275-1058 58 200 (D5) 100 (D6) 110 (D7) 55 (D8)
Model: MULP180 SERIES
Vac= 100 to 264Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP180-1012 12 180 (E1) 90 (E2) 120 (E3) 60 (E4)MULP180-1015 15 180 90 120 60 MULP180-1024 24 180 90 120 60 MULP180-1030 30 180 90 120 60 MULP180-1048 48 180 90 120 60 MULP180-1058 58 180 (E5) 90 (E6) 120 (E7) 60 (E8)
Page 13 of 303 Report No. 316648
TRF No. IEC60601_1K
General product information:
Model: MULP180 SERIES Vac= 90 - < 100Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP180-1012 12 160 (F1) 80 (F2) 90 (F3) 45 (F4) MULP180-1015 15 160 80 90 45 MULP180-1024 24 160 80 90 45 MULP180-1030 30 160 80 90 45 MULP180-1048 48 160 80 90 45 MULP180-1058 58 160 (F5) 80 (F6) 90 (F7) 45 (F8)
Model: MULP180 SERIES Vac= 85 - < 90Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP180-1012 12 150 (G1) 75 (G2) 85 (G3) 42.5 (G4) MULP180-1015 15 150 75 85 42.5 MULP180-1024 24 150 75 85 42.5 MULP180-1030 30 150 75 85 42.5 MULP180-1048 48 150 75 85 42.5 MULP180-1058 58 150 (G5) 75 (G6) 85 (G7) 42.5 (G8)
Model: MULP180 SERIES Vac= 80 - < 85Vac
Models Voltage
Maximum Individual Continuous Power with 13 CFM air flow & Power Derating at Elevated Temperatures
Maximum Individual Continuous Power with Convection cooling & Power Derating at Elevated Temperatures
50˚C 70˚C 40˚C 70˚CMULP180-1012 12 140 (H1) 70 (H2) 80 (H3) 40 (H4) MULP180-1015 15 140 70 80 40 MULP180-1024 24 140 70 80 40 MULP180-1030 30 140 70 80 40 MULP180-1048 48 140 70 80 40 MULP180-1058 58 140 (H5) 70 (H6) 80 (H7) 40 (H8)
The alphabets in parenthesis represent the overall series loading.
The power supply is not provided with a fire enclosure.
Maximum manufacturer recommended operating attitude: 5000m