IST Co., Ltd. EMC LABORATORY TEST REPORT NO. : 08-IST-0005 1 of 10 EMC TEST REPORT Electromagnetic Compatibility Emission Immunity Test Report File No. : 08-IST-0005 Date of Issue : September 10, 2008 Basic Model(s) : ULTra U-801 Kind of Product : Power Saving Appliance Applicant : KESECO CO.,LTD. Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu, Seoul, Korea (133-120) Manufacturer : KESECO CO.,LTD. Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu, Seoul, Korea (133-120) Test Result Positive Negative Reviewed By Approved By S.J.CHO / EMC Group Manager B.S.Kim / Chief Comment(s) -The test regulations are in page 3. -This test report consists of 10 pages with appendix. -The test result only responds to the tested sample. This report shall not be reproduced except in full, without the allowance of the test laboratory.

Upload: alejandro-arce

Post on 12-Mar-2016

218 views

Category:

Documents


0 download

DESCRIPTION

S.J.CHO / EMC Group Manager B.S.Kim / Chief Manufacturer : KESECO CO.,LTD. Comment(s) IST Co., Ltd. EMC LABORATORY TEST REPORT NO. : 08-IST-0005 Basic Model(s) : ULTra U-801 Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu, Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu, Test Report File No. : 08-IST-0005 Date of Issue : September 10, 2008 Kind of Product : Power Saving Appliance Electromagnetic Compatibility Seoul, Korea (133-120) Seoul, Korea (133-120) 1 of 10

TRANSCRIPT

Page 1: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

1 of 10

EMC TEST REPORT Electromagnetic Compatibility

□Emission ■ Immunity

Test Report File No. : 08-IST-0005 Date of Issue : September 10, 2008

Basic Model(s) : ULTra U-801

Kind of Product : Power Saving Appliance

Applicant : KESECO CO.,LTD.

Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu,

Seoul, Korea (133-120)

Manufacturer : KESECO CO.,LTD.

Address : 8F, NanYoung Bldg., 284-49 SungSuDong2GA, SungDong-Gu,

Seoul, Korea (133-120)

Test Result □ Positive ■ Negative

Reviewed By Approved By

S.J.CHO / EMC Group Manager B.S.Kim / Chief

Comment(s)

-The test regulations are in page 3. -This test report consists of 10 pages with appendix. -The test result only responds to the tested sample. This report shall not be reproduced except in full, without the allowance of the test laboratory.

Page 2: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

2 of 10

TABLE OF CONTENTS

Table of contents 2

Test regulations 3

Information of test laboratory, EUT Installed System,

Product Information

4

Equipment Under Test and test setup 5

Summary 6

□ Test Conditions and Data - Emissions

◇ Conducted Emissions 0.15 MHz - 30 MHz Not applicable

Test Conditions / Data and Plots

◇ Radiated Emissions 30 MHz - 1 GHz Not applicable

Test Conditions / Data and Plots

□ Test Conditions and Data – Generic

◇ Voltage Fluctuations Not applicable

Test Conditions / Data and Plots 14-16

■ Test Conditions and Data - Immunity

◇ Electro-Static Discharge Not applicable

Test Conditions / Data and Plots

◇ Radiated Susceptibility Not applicable

Test Conditions / Data and Plots

◇ Electrical Fast Transient (Burst) Not applicable

Test Conditions / Data and Plots

◆ Surge Transient Applicable

Test Conditions / Data and Plots 7-8

◇ Conducted Susceptibility Not applicable

Test Conditions / Data and Plots

◇ Voltage Dips, interruptions Not applicable

Test Conditions / Data and Plots

Appendix A. The EUT Photos 9

B. The Test Setup Photos 10

Page 3: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

3 of 10

TEST REGULATIONS

The tests were performed according to 2004/108/EC directive and following

regulations :

□ Emissions

◇ EN55022:1998 +A1:2000 ,+A2:2003

■ Immunity

◆ EN55024:1998 +A1:2001, +A2:2003

◇ EN 61000-4-2:1995 +A1:1998, +A2:2001

◇ EN 61000-4-3:2006

◇ EN 61000-4-4:2004

◆ EN 61000-4-5:1995 +A1:2001

◇ EN 61000-4-6:1996 +A1:2001

◇ EN 61000-4-8:2003 +A1:2001

◇ EN 61000-4-11:2004

□ Generic

◇ EN61000-3-2:2006

◇ EN61000-3-3:1995 +A1:2001, +A2:2005

Note :

- ◆ means the test is applicable,

- ◇ is not applicable.

Page 4: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

4 of 10

INFORMATION OF TEST LABORATORY

IST EMC LABORATORY

400-19, Singal-dong, Giheung-gu, Yongin-si, Gyeonggi-do, 446-599, Korea

TEL : +82 31 326 6700 FAX : +82 31 326 6799

Measurement Uncertainty

The measurement uncertainty was evaluated for all test items listed in

this report. Also it, “ The evaluation and treatment of uncertainty “,

is described in IST Quality Manual according to ISO17025 Guide. The data

and results described in this report are true and include evaluated

uncertainty. It may cause some deviation of uncertainty by change component

or process of the test for similar products.

PRODUCT INFORMATION Input Power 220 Vac, 60 Hz

-EMC suppression device is not used during the test.

-Find information in user manual.

Page 5: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

5 of 10

Equipment Under Test

EUT Type :

Table-Top. Floor-Standing.

Table-Top and Floor-Standing(Combination).

Operation – mode of the E.U.T. :

The equipment under test was operated during the measurement under following conditions :

Standby Mode

Operational Condition : Normal operation mode.

Configuration of the equipment under test :

Following peripheral devices and interface cables were connected during the measurement :

Equipment Type Brand Serial No.

Connecting Interface Cables :

Page 6: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

6 of 10

SUMMARY

Emissions □ Conducted Emissions (From 0.15 MHz To 30 MHz)

The requirements are ○ MET ○ Not MET

Minimum limit margin

Maximum limit exceeding

Remarks : Limits are kept with more than 3dB margin

□ Radiated Emissions (From 30 MHz To 1GHz)

The requirements are ○ MET ○ Not MET

Minimum limit margin

Maximum limit exceeding

Remarks : Limits are kept with more than 3dB margin

Generic Test □ Harmonic current

The requirements are ○ MET ○ Not MET

□ Voltage Fluctuations

The requirements are ○ MET ○ Not MET

Immunity

□ Electrostatic Discharge ○ MET ○ Not MET

□ Radiated Electromagnetic Fields ○ MET ○ Not MET

□ Electrical Fast Transient(Burst) ○ MET ○ Not MET

■ Surge Transient ○ MET ● Not MET

□ Conducted Susceptibility ○ MET ○ Not MET

□ Power Frequency Magnetic ○ MET ○ Not MET

□ Voltage Dips, interruption ○ MET ○ Not MET

Test date

Begin of test date: September 10, 2008

End of test date: September 10, 2008

Note : -■ means the test is applicable, □ is not applicable.

Prepared By

S.J.OH / EMC Engineer

Page 7: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

7 of 10

TEST CONDITIONS AND DATA

Surge Transient

[Applicable]

◆ Test Equipment Used

The test equipment used is calibrated in regular for every year. Model Name Description Manufacture Calibration Date Serial No.

PSURGE 4.1 Surge Generator Haefely May 21, 2007 082876-05

◆ Environmental Conditions

Temperature 25 ℃

Humidity 47 %

Atmosphere pressure 1016 mbar

◆ Test Program Normal operation mode.

◆ Test Area EMS Test Room

◆ Test Date September 10, 2008

Note :

Page 8: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

8 of 10

Surge Transient

[Applicable]

◆ Test Specification

Pulse amplitude

– Power line :

■ 1.0 kV : Line to line

■ 3.0 kV : Line to line■ 2.0 kV : Line to line

Pulse amplitude

– Data line :

□ 0.5 kV

□ 2.0 kV □ 1.0 kV

Source impedance : ■ 18 µF : Line to line

□ 40 Ω + 0.1 µF ■ 10 Ω + 9 µF: Line to ground

□ 40 Ω + 0.5 µF

Number of surges : □ 3 surges/angle ■ 5 surges/polarity

Angle ■ 0° ■ 90°

■ 180° ■ 270°

Repetition rate : □ 10 sec. ■ 60 sec.

Polarity : ■ Positive ■ Negative

Location of Coupling:

name of lines: Power cable length : 2.5 m

type of lines □ – shielded ■ - unshielded

status of lines: □ – passive ■ - active

kind of transmission: ■ - analog □ - digital

◆ Result :

□ No degradation of function Met Criterion A

■ Distortion of function Met Criterion B – incandescent lamp

■ Error of function Met Criterion C – fluorescent lamp

□ Loss of function Unrecoverable Failure -

Note : 1 kV – A fluorescent and incandescent lamp wear flickering. : Pass

2 kV - A incandescent lamp was flickering and ballast of fluorescent

was broken. : incandescent lamp – Pass, fluorescent lamp - Fail

3 kV - A incandescent lamp was flickering. : incandescent lamp – Pass

Page 9: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

9 of 10

Appendix A. The EUT Photos

EUT View

EUT Set-up

Page 10: Certificado_EMC_Test_U-801

IST Co., Ltd. EMC LABORATORY

TEST REPORT NO. : 08-IST-0005

10 of 10

Appendix B. Test Setup Photos

Surge Transient(Fluorescent)

Surge Transient(Incandescent)