carestream 9600 cbct integration - kb.tdo4endo.com

4
Carestream 9600 CBCT Integration 731 TDO KB March 24, 2021 Carestream , CBCT 0 2588 Overview The CS9600 is the newest CBCT unit from Carestream Dental. Users of TDO 12.338 or newer can integrate their CS 9600 directly into TDO. Detailed Steps 1. Open [Setup > Site Configuration > Device Configuration] 2. Fill in the information for the Device Configuration. Leave the ’CSDAT’ field blank. 3. Go to Acquire, and then click ’CBCT’, you should see the CS9600 window appear.

Upload: others

Post on 09-Dec-2021

11 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Carestream 9600 CBCT Integration - kb.tdo4endo.com

Carestream 9600 CBCT Integration731 TDO KB March 24, 2021 Carestream, CBCT 0 2588 Overview

The CS9600 is the newest CBCT unit from Carestream Dental. Users of TDO 12.338 or newer canintegrate their CS 9600 directly into TDO.

Detailed Steps

1. Open [Setup > Site Configuration > Device Configuration]

2. Fill in the information for the Device Configuration. Leave the ’CSDAT’ field blank.

3. Go to Acquire, and then click ’CBCT’, you should see the CS9600 window appear.

Page 2: Carestream 9600 CBCT Integration - kb.tdo4endo.com

4. Once the main screen open, click on the horizontal lines, then select General Settings.

5. Select ’No’ for ’Display Quality Control for panoramic modality’, ’Display Quality Control for 3Dmodality’, and ’Display Quality Control for Facial Scan modality’.

Page 3: Carestream 9600 CBCT Integration - kb.tdo4endo.com

6. Save these settings.

7. Reboot the CS9600.

8. Open Acquire and click on the ’CBCT’ button again. If you select “MAR” (Metallic Artefact Removal)mode in the CS 9600 unit you will have 2 volumes generated: one for the standard volume, and the otherfor the MAR volume. In Acquire you will see the 3D icon (see the example below). The viewer will openafter reconstruction for the standard volume is complete. It will take a few more minutes to complete theMAR reconstruction, but you can review the standard volume in the viewer while the MAR volume isreconstructing. Once the MAR volume is complete you can now use the MAR functionalities in yourviewer.