calibration of clinometers for the alignment of muon chambers in the cms experiment. moschoutis...
TRANSCRIPT
Calibration of clinometers Calibration of clinometers for the alignment of muon for the alignment of muon
chambers in the CMS chambers in the CMS experiment. experiment.
MOSCHOUTIS ThomasMOSCHOUTIS Thomas
FAS meeting – Orlando, 12 March 2004
Global view of the Compact Muon Global view of the Compact Muon SolenoidSolenoid
View of one DiskView of one Disk
Dual-axis clinometers for monitoring orientation of transfer pates:
Calibration benchCalibration bench
CIinometer
sensing Liquid
calibrated Linear mover
Θ
First setupFirst setup
Clinometer’s placeLinear mover
Error calculationError calculation
Error in Error in φφ::• φφ =Tan =Tan-1-1(L/H) (L/H) • ΔφΔφ=(∂=(∂φφ/∂L)*/∂L)*ΔΔL+(∂L+(∂φφ/∂H)*/∂H)*ΔΔH H
=(1/(L+H))*=(1/(L+H))*ΔΔL-2L/(HL-2L/(H22+H*L)*+H*L)*ΔΔHH Using the same formula:Using the same formula:
• ΔΩΔΩ=(1/(L+M))*=(1/(L+M))*ΔΔL-2L/(ML-2L/(M22+M*L)*+M*L)*ΔΔMM Taking Taking ρρ=cos-1(M/K) and M=0=cos-1(M/K) and M=0
• ΔρΔρ==ΔΔM/KM/K So the total error is:So the total error is:
• ΔθΔθ =( =(ΔφΔφ22++ΔΩΔΩ22++ΔρΔρ22))1/21/2
MK
ρρ
φφ
LL
HH
First test curveFirst test curve
Angle vs clinometer voltage output
-5
0
5
10
15
20
25
30
-3.0 -2.5 -2.0 -1.5 -1.0 -0.5 0.0 0.5 1.0 1.5 2.0
Response Voltage [V]
An
gle
[D
eg
ree
s]
Zero degree measurementZero degree measurement
Pendulum
clinometer
Long rod to get more precision
Problem induce by the plate Problem induce by the plate thicknessthickness
Second test curve Second test curve
Angle vs clinometer voltage output
0
2
4
6
8
10
12
14
16
-2.0 -1.5 -1.0 -0.5 0.0 0.5 1.0 1.5
Response voltage [Volts]
An
gle
[D
egre
es]
The 6 different mounting platesThe 6 different mounting plates
Clinometer’s place
Next setupNext setupLinear mover
Adapter
Tilting base
Mounting plate
Click here to see movie