calibration of balance
DESCRIPTION
Bangladesh BEST Programme. Calibration of Balance. Sri Lanka. Nihal Gunasekara. Types of Balances. Bangladesh BEST Programme. Electromagnetic force balance (Top loading) Single-pan, two-knife-edge balance Two-pan, three-knife-edge balance - PowerPoint PPT PresentationTRANSCRIPT
Calibration of Balance
Sri Lanka
1
Nihal Gunasekara
Bangladesh BEST Programme
Types of Balances• Electromagnetic force balance (Top loading)• Single-pan, two-knife-edge balance• Two-pan, three-knife-edge balance• Triple beam balance• Weighing scale
2
Bangladesh BEST Programme
Analytical Balance
Balance with an enclosed weighing chamber and a resolution 0.01 or 0.1 mg
3
Micro Balance
Balance with a resolution of 1 microgram
Bangladesh BEST Programme
Micro Balance
4
Bangladesh BEST Programme
Semi Micro Balance
Balance with a resolution of 10 microgram
5
Ultra (Super) Micro Balance
Balance with a resolution of 0.1 microgram
Bangladesh BEST Programme
Resolution or Discrimination
The smallest change in mass which can be detect by the balance
6
Sensitivity
This is a measure of the ability of the balance to detect changes in the load applied to it
Bangladesh BEST Programme
Response time
Time taken for the weighing instrument to indicate the value of a change in load
7
Stability
A measure of the time for which the reading on the balance remains unchanged
Bangladesh BEST Programme
Correction
Correction = True Value - Reading
8
Error
Error = Reading - True Value
Bangladesh BEST Programme
Departure From Nominal Value
The amount by which the reading on an instrument depart from its nominal value
9
Hysteresis
The indications of a measuring instrument when the same value of the quantity measured is reached by increasing or decreasing that quantity
Bangladesh BEST Programme
Standby
Operating state of an electronic instrument where one part is switched off for energy saving reasons
10
Accuracy class. (mass)
Legal metrology term. Classification of weights for calibration purposes by OIML R111-1 weights of Accuracy class E1, E2, F1, F2, M1, M2, M3 from 1 mg to 50 kg
Bangladesh BEST Programme
11
Different accuracy classes of mass Bangladesh BEST Programme
Repeatability
Closeness of the agreement between the results of successive measurements of the same measurement carried out under the same conditions of measurement.
Repeatability condition include :• the same observer• the same measuring instrument, used under the same
conditions• the same location• repetition over a short period of time
12
Bangladesh BEST Programme
Reproducibility Closeness of the agreement between the results of measurements of
the same measurement carried out under changed conditions of measurement.
A valid statement of reproducibility requires specification of the conditions change
The changed conditions may include – principle of measurement– method of measurement– observer – measuring instrument– reference standard– location– conditions of use– Time
13
Bangladesh BEST Programme
Influence Factors
14
Bangladesh BEST Programme
Air movements and heat sources
Support surfaces and vibration
Electrical and electromagnetic interference
Magnetic effects
Buoyancy effects
Influence factor
15
Use of a high-mass support, set on a solid surface, for the weighing machine, independent of the adjacent flooring, Eg fixed to bedrockNote : low frequency, vibration is likely to be more serious in its effects than high (acoustic) frequenciesRubber pads used to damp high frequencies may pass low frequency vibration, and may cause tilting of the weighing machine
PrimaryStability of reading under constant load SecondaryAccelerometer measurements
Changes forces on the weighing machine due to acceleration
Various, Eg. Machinery, roads, railways, aircraft operator Vibration causes damages
Vibration Environmental
PrecautionsTest for effectEffectSourceParameterInfluencefactor
Bangladesh BEST Programme
Influence factor
16
Avoid direct sunlight and other connections to the external environment; avoid changes in heat and lighting input. Leave balance cases ajar when not in use. Avoid excessive handling of equipment during weighing. Allow weighted object to acclimatize
Temperature monitoring instruments
Temperature changes and differentials can change the weighing machine response, cause convection effects and change buoyancy
Sunlight, heating systems, operator’s body, greenhouse heating in the weighing chamber
Temperature: air, weight, weighing machine
Environmental
PrecautionsTest for effectEffectSourceParameterInfluencefactor
Bangladesh BEST Programme
Influence factor
17
Avoid liquid water in the working area unless rh is too low .Air conditioning, humidity controller
Rh monitoring instruments
Buoyancy, absorbed moisture, condensation
Air supply ,air conditioning, changes in activity levels
HumidityEnvironmental
PrecautionsTest for effectEffectSourceParameterInfluencefactor
Bangladesh BEST Programme
Influence factor
18
PrecautionsTest for effectEffectSourceParameterInfluencefactor
Avoid temperature gradients. draught free working space, enclose weighing
machines
Primary
Stability of reading under constant load Secondary
Enclose weighing machine and observe change of reading stability
Varies force on weighing machine
Air conditioning, wind, convection, operator movement
Airflow Environmental
Bangladesh BEST Programme
Influence factor
19
PrecautionsTest for effectEffectSourceParameterInfluencefactor
Avoid weights with residual magnetism (de-gauss) Test with compass or gaussmeter Limit magnetic permeability of weights weighed check for magnetic fields with gaussmeter Screen with mumetal
PrimaryUse a magnet to check for induced reading changesSecondaryMonitoring readings when weights are rotated and at varying heights above the weight receptor (place on/under non-magnetic mount)
Change of current in force compensation coil, direct attraction to/from surrounding structure Motion-dependent restoring forces
Constructional steelwork, ferromagnetic weight (s) electrical induction earth’s magnetic field, moving-coil meters
Magnetic Environmental
Bangladesh BEST Programme
20
RepeatabilityLoad cell hysteresis
RepeatabilityElectrical Interference
ZeroMains fluctuation
Sensitivity, Zero, RepeatabilityIncorrect horizontal support
RepeatabilityMechanical errors
Type of errorSource of error
Bangladesh BEST Programme
21
ACCURACY CLASS (OIML R111-1 :2004)
0.200.060.0200.0061 mg
0.200.060.0200.0060.0035 mg
0.250.080.0250.0080.00310 mg
0.40.120.040.0120.00450 mg
1.60.50.160.050.0160.005100 mg
103.01.00.30.100.030.0101 g
124.01.20.40.120.040.0122 g
165.01.60.50.160.050.0165 g
206.02.00.60.200.060.02010 g
301031.00.30.100.03050 g
501651.60.50.160.05100 g
100301031. 00.30.10200 g
50016050165.01.60.51 kg
100030010030103.01.02 kg
250080025080258.02.55 kg
5000160050016050165.010 kg
Class M3Class M2Class M1Class F2Class F1Class E2Class E1
Tolerance in mgNominal value
0.003
Bangladesh BEST Programme
Lowest class of weights required for different types of weighing machine
22
E2E2E2F2M1M2M21 kg to 30kg
E2E2F1/E2F2M1M1200g to 1 kg
E2F1F2M1M230kg to 100kg
E2F1F2M1M1Up to 200g
E2F1M1/F2M2Above 100 kg
0.01mg or less
0.1mg1mg10mg100mg1g10g100g
ResolutionCapacity
Bangladesh BEST Programme
Calibration of Top Loading and Analytical Balances
23
Bangladesh BEST Programme
Before use a Balance…………• Keep the Balance power on appropriate time to warm up.• Clean the balance as best as no contamination on the balance pan.• If it is applicable, level the balance before use.• Don’t move the balance after the calibration.• Lint or any other soft gloves should be worn by the user.• A pair of forceps or a pair tweezers with thermal insulating ends should
be used to lift and place weights.
24
Bangladesh BEST Pregramme
• Make sure with the soft placement of weights on the balance pan as level best of possible.
• Readings should be taken after the balance is stabilized well. • The place should be free from vibration.• Any other relevant precaution practiced by the user.
25
Before use a Balance Bangladesh BEST Programme
Balance Calibration
The calibration of the electromagnetic force compensation balance consists of the following tests:
• Test 1 - Scale Value • Test 2 - Repeatability of Reading• Test 3 - Departure from Nominal Value• Test 4 - Off centre Loading• Test 5 – Hysteresis
26
Bangladesh BEST Programme
Scale Value
Internal Scale Adjustment –Internal Mass
External Scale Adjustment – External Mass
Procedure – Balance Operating Manual
27
Bangladesh BEST Programme
Repeatability of Reading Objective of Measurement : The repeatability test is carried out to determine the ability of the
balance to give under defined conditions of use closely similar responses for repeated measurements
The balance Standard Deviation is the “repeatability of measurement”
28
Bangladesh BEST Programme
Repeatability of Reading :• M 1 -Set the reading to zero without any mass on the pan• M 2 - Record the zero reading of the balance (z)• M 3 - Place the mass on the pan and record the reading
(r)• M 4 - Remove the mass from the pan• M 5 – Repeat M2 to M4 ten times• during the process the balance should not be zeroed
29
Bangladesh BEST Programme
Repeatability of Reading:Performed at three points : Near zero
Near half load
Near full load
Bangladesh BEST Programme
DATA RECORDING Mass 1 : Mass 2: Mass 3:
31
No Zero reading (z)
Mass reading (r)
Corr. reading (r-z)
Zero reading (z)
Mass reading (r)
Corr. reading (r-z)
Zero reading (z)
Mass reading (r )
Corr. reading (r-z)
1
2
3
4
5
6
7
8
9
10
Bangladesh BEST Programme
Repeatability of Reading
Mass 1: Mass 2: Mass 3 : Mean value :Std. Dev. ( σ ) : Max. Diff. : Uncer. Type A :
32
Bangladesh BEST Programme
Repeatability of Reading
0.20.10.1
0.0670.0520.052
0.05100200
Maximum Difference Between Successive
Readings (mg)
Standard Deviation of Reading (mg)
Load on Pan(g)
Repeat three different values of M ; near zero, half maximum and maximum load.
Bangladesh BEST Programme
Departure from Nominal Value:
Objective of Measurement :
“The departure from nominal value measurements are carried out to determine the deviation of the balance reading from the expected value”
This is the Correction to Balance Reading
34
Bangladesh BEST Programme
Departure from Nominal Value:
1 Divide the range into at least 10 equally spaced steps 2 The masses corresponding to M, 2M, 3M, 4M,
5M, 6M etc. until the capacity of balance is reached 3 In addition choose four more test points
corresponding to 20 %, 40 %,60%and 80 % of the first step for single range.
35
Bangladesh BEST Programme
Departure from Nominal Value: 1 - Read the zero z1
2 - Place the weight on the pan &record reading r1
3 – Remove the weight momentarily and place it back on the pan and note the new reading r2
4 - Remove the weight and read the zero z2
5 – Repeat 1 to 4 for weights of masses 2M,3M,4M,5M etc until the capacity of balance is reached
36
Bangladesh BEST Programme
Departure from Nominal Value: • Average zero reading zi = (z1 + z2 )/ 2
• Average mass reading ri = (r1 + r2 )/ 2
• Difference di = ri – zi
• Correction ci = Mi – di
• Mi is the conventional value of the mass37
Bangladesh BEST Programme
Correction to Balance Reading
Pan Load Conv. Value Mi
Reading Mean Difference ri - zi
Correction Mi - di
M- 100 100.02 0.00100.000.00100.02
0.00 100.01
100.01 0.01
2M
3M
4M
5M
6M
7M
8M9M
10M
Bangladesh BEST Programme
Departure from Nominal Value:
39
0.00.0-0.10.1
-0.00.00.0-0.1-0.1-0.1-0.1-0.2-0.3-0.3-0.3
After the Calibration
Before the Calibration
0.0960.0970.0970.098
0.0990.0980.0980.0990.0990.0990.100.100.100.110.11
--------
0.2----
0.4--
0.6--
0.4--
0.5--
481216
20406080
100120140160180200220
Expanded Uncertainty k=2+(mg)
Correction(mg) Reading(g)
Bangladesh BEST Programme
Off Centre Loading:
Objective of Measurement : “The off center loading is carried out to find the
variation in the displayed weighted as the object is moved to various positions on the weighing pan”
40
Bangladesh Best Programme
Off Centre Loading
41
This test to find the variation of the reading for several place of the balance pan. Test should be done for rectangular type balance pans and as well as for circular type balance pans.
A single weight should be used
43
1
52
3
2
4
5
1
Bangladesh BEST Programme
Off Centre Loading:
42
Center (g)100.0000-100.0000=0.0000Left (g)
100.0001-100.0000=0.0001
Back (g)100.0000-100.0000=0.0000
Right (g) 99.9999-100.0000= -0.0001
Front (g)100.0001-100.0000=0.0001
Max Difference =0.0002 g
Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time
Bangladesh BEST Programme
Off Centre Loading:
43
0.0002- 0.00010.00010.00000.00010.0000
Maximum Difference (g)
Right(g)Left(g)Back(g)Front(g)Centre(g)
Mass of 100g was moved to various positions on the pan. The balance readings obtained relative to the centre reading
Place a mass on the centre of the pan and move it successively to the front rear, left and right positions ( position being 2/3 of the distance between the centre and the lip of the pan) on the pan recording the reading each time
Bangladesh BEST Programme
Hysteresis:
Objective of Measurement : “The hysteresis measurements are performed to determine the
difference in the indication at the same load , depending upon the direction of approach to that load ie increasing or decreasing the load”
44
Bangladesh BEST Programme
Hysteresis
Perform the test at one point about mid range• Zero the balance and record the reading z1 • Place a mass M1 on the plan and record the reading r1
• Add extra mass M2 to bring the balance close to full range• Remove the extra mass M2 and record the reading with the
mass M1 still on the pan• Remove the mass M1 and record the reading r2
• Repeat above steps three times• Take the average of three readings
45
Bangladesh BEST Programme
Hysteresis Mass M = g
46
Bangladesh BEST Programme
Pan Load Reading 1 Reading 2 Reading 3
Zero ( z1 )
M ( m1 )
M + M1
M ( m2 )
Zero ( z2 )
m1 - m2
z1 - z2
Hysteresis:
47
0.1100
0.00
Hysteresis Error(mg)Pan Load (g)
Bangladesh BEST Programme
Hysteresis:
48
Mass Value VS Display Value
0
10
20
30
40
50
60
70
80
90
100
0 20 40 60 80 100
Mass Value
Disp
lay
Valu
e
Increasing
Decreasing
Hysteresis error
Bangladesh BEST Programme
Calibration of Single Pan two Knife Edge Balances
49
Bangladesh BEST Programme
Balance Calibration
The calibration of the single pan two knife edge balance consists of the following tests:
• Test 1 - Repeatability of Reading• Test 2 - Departure from Nominal Value• Test 3 – Built-in-Mass• Test 4 - Off centre Loading• Test 5 – Hysterisis
50
Bangladesh BEST Programme
Balance Calibration
The calibration of the single pan two knife edge balance consists of the following tests:
• Test 1 - Repeatability of Reading- same as electronic balance
• Test 2 - Departure from Nominal Value- same as electronic balance
• Test 3 – Built-in-Mass• Test 4 - Off centre Loading- same as electronic balance
• Test 5 – Hysteresis- same as electronic balance
51
Bangladesh BEST Programme
Balance Calibration The procedure to determine the Built-in- Mass• With all dials set to the zero and no load on the pan,
release the balance and record the reading z1
• Set the "1" on the dial to be tested, place a calibrated mass, release the pan and record the reading m1a
• Remove the mass , replace it and record the reading m1b
• Return the dial to zero, remove the mass from the pan and record the zero z2
• Repeat the above steps for all dial settings of the balance (i.e. 2 , 3, 10 , 20 ,…………..90 , 100 , 200 )
• Follow the same procedure used for linearity test to determine the correction
52
Bangladesh BEST Programme
Calibration of Triple Beam Balances
53
Bangladesh BEST Programme
Balance Calibration
The calibration of triple beam balances consists of the following tests:
• Test 1 - Repeatability of Reading - same as electronic balance
• Test 2 - Departure from Nominal Value (Linearity Test) -
same as electronic balance
54
Bangladesh BEST Programme
Triple Beam Balance
Before the calibration• Clean the weighing pan and other parts of the balance
taking due care• Level the balance• Exercise the balance by making several weighing
55
Bangladesh Best Programme
Triple Beam BalanceRepeatability of readings:• Set the reading to zero without any mass on the pan• Record the zero reading zi of the balance without any mass
on the pan • Place the mass M on the pan and record the mass reading
mi
• Remove the mass M from the pan• Repeat the above steps ten times (The balance should not
be zeroed during the series of readings)• Repeat for this for near zero, half load and near maximum
56
Bangladesh BEST Programme
Triple Beam Balance
• Select a minimum of 10 equally spaced points ( M , 2M ,…..10M ) on the measuring range in such a way they cover the entire range or the part of the measuring range to be calibrated. In addition four test points corresponding to 20%,40%,60% and 80 % of the first step
• Record the zero reading z1
• Set the slider scale to the first step on the measuring range (M) , place the calibrating mass(es) on the pan until the balance comes to equilibrium and record the mass reading mi
• Remove the mass(es), place it back on the pan and record the mass reading mi2
• Remove the mass(es) and record the zero reading zi+1
• Repeat the above steps for the remaining points ( 2M , 3M ,….)
57
Bangladesh Best Programme
Weighing Scale (digital or analogue)
58
Bangladesh Best Programme
The calibration of weighing scales consists of the following tests:Test 1 - Repeatability of Reading - same as electronic balance
Test 2 - Departure from Nominal Value (Linearity Test) - same as electronic balance
Calibration Results of a Set of Weights –Class F1
59
Bangladesh BEST Programme
Nominal Value Conventional Value Uncertainty k=2 Max. Perm. Error
1 mg 1 mg – 0.001 mg 0.006 mg ±0.02 mg
2 mg 2 mg + 0.011 mg 0.006 mg ±0.02 mg
5 mg 5 mg+0.002 mg 0.006 mg ±0.02 mg
20 mg 20 mg – 0.005 mg 0.01 mg ±0.03 mg
50 mg 50 mg + 0.009 mg 0.012 mg ±0.04 mg
100 mg 100 mg + 0.011 mg 0.015 mg ±0.05 mg
500 mg 500 mg + 0.033 mg 0.025 mg ±0.08 mg
1 g 1 g – 0.028 mg 0.03 mg ±0.1 mg
2 g 2 g + 0.022 mg 0.04 mg ±0.12 mg
20 g 20 g + 0.009 mg 0.08 mg ±0.25 mg
50 g 50 g – 0.12 mg 0.1 mg ±0.3 mg
100 g 100 g – 0.22 mg 0.15 mg ±0.5 mg
200 g 200 g + 0.04 mg 0.3 mg ±1 mg
Issued by an accredited laboratory (DKD)
Thank you
60
Bangladesh BEST Programme