1
13 Remove posteriors, replace with Bennett Fossas. Set Bennett at 5˚. Check for any lateral movement. If there is more than .002”, return unit to factory for repair. 2 Insert posteriors into upper. CHECK: Posteriors should slip into upper freely. 3 Align yellow line of posterior to 30˚ mark of upper. Tighten thumbscrew. Repeat for other side. 4 CHECK: Area under gauge should be smooth. Mount gauge block onto lower. 5 Mount adjustable incisal guide. 6 CHECK: Check that there is free rotation of condylar balls. 7 Mount upper assembly to lower assembly. Balls should move freely in posterior tracks without lateral movement. 8 CHECK to see that cylindrical gauge blocks are aligned properly. VERIFY by visual inspection and by using finger- tips to check alignment of cylinders. Alignment can also be checked with .002” shim. Overlap of cylinders should not exceed .002” in either direction. If aligned within tolerance, go to step 12. If not, go to step 9. 9 If articulator is out of alignment, loosen both condylar shafts’ set screws. Set screws are 5/64”. 12 VERIFY alignment in anterior/posterior direction within .002” using same procedure as step 8. If not within .002”, return unit to factory for repair. 14 VERIFY: Verify incisal pin lines up correctly. Loosen thumbnut. Drop incisal pin. When pin is on guide, dark line should line up. Calibration is complete. CALIBRATION INSTRUCTIONS: Hanau Modular Articulator with adjustable Bennett Instructions for Modular Articulator with no Bennett or 0 Immediate side shift on reverse side. 1 CHECK: Area under gauge should be smooth. Mount upper half of gauge and tighten. Raise incisal pin 3mm. The pin should not touch the adjustable incisal guide. 11 Lock set screws after lateral alignment is achieved. 10 With thumbs, gently push on condyle shafts, moving upper laterally until upper and lower gauges line up within .002”. CHECK: .002” alignment can be verified with shim stock. Whip Mix Corporation - West 1730 East Prospect Rd., Suite 101 Fort Collins, CO 80525 Toll-Free: 800-201-7286 Fax: 970-472-1793 www.whipmix.com

Upload: trinhbao

Post on 09-Aug-2018

303 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CALIBRATION INSTRUCTIONS: Hanau Modular Articulator …whipmix.com/wp-content/uploads/calibration-modular_w_bennett1.pdf · 14VERIFY: Verify incisal pin lines up correctly. Loosen

13 Remove posteriors,replace with BennettFossas. Set Bennett at 5˚.Check for any lateralmovement. If there ismore than .002”, returnunit to factory for repair.

2 Insert posteriors into upper.CHECK: Posteriors shouldslip into upper freely.

3 Align yellow line of posteriorto 30˚ mark of upper. Tightenthumbscrew. Repeat forother side.

4 CHECK: Area under gaugeshould be smooth.Mount gauge block ontolower.

5 Mount adjustable incisalguide. 6 CHECK: Check that there is

free rotation of condylarballs.

7 Mount upper assembly tolower assembly. Balls shouldmove freely in posteriortracks without lateralmovement.

8 CHECK to see that cylindricalgauge blocks are alignedproperly. VERIFY by visualinspection and by using finger-tips to check alignment ofcylinders. Alignment can also bechecked with .002” shim.Overlap of cylinders should notexceed .002” in either direction.If aligned within tolerance, goto step 12. If not, go to step 9.

9 If articulator is out ofalignment, loosen bothcondylar shafts’ set screws.Set screws are 5/64”.

12 VERIFY alignment inanterior/posterior directionwithin .002” using sameprocedure as step 8. If notwithin .002”, return unit tofactory for repair.

14VERIFY: Verify incisal pinlines up correctly. Loosenthumbnut. Drop incisalpin. When pin is on guide,dark line should line up.Calibration is complete.

CALIBRATIONINSTRUCTIONS:

Hanau™ModularArticulator

with adjustable BennettInstructions for Modular

Articulator with no Bennettor 0 Immediate side shift

on reverse side.

1 CHECK: Area under gaugeshould be smooth. Mountupper half of gauge andtighten. Raise incisal pin 3mm.The pin should not touch theadjustable incisal guide.

11 Lock set screws afterlateral alignment isachieved.

10 With thumbs, gently pushon condyle shafts, movingupper laterally until upperand lower gauges line upwithin .002”. CHECK: .002”alignment can be verifiedwith shim stock.

Whip Mix Corporation - West1730 East Prospect Rd., Suite 101Fort Collins, CO 80525Toll-Free: 800-201-7286Fax: 970-472-1793www.whipmix.com