![Page 1: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/1.jpg)
By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist Hospital & Research Centre
![Page 2: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/2.jpg)
About the competition Contouring stage Arc geometry Optimization Results Conclusions
![Page 3: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/3.jpg)
Case selection Peripheral targets
Inhomogeneous shape
Proximity to organs at risk The criteria In general, the criteria is well chosen
Constraints on the right side of the patient were difficult to achieve
I found that some objectives were easily “over” achieved
![Page 4: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/4.jpg)
Structure Points
Target (CI , HI, others) 45
Heart 20
Left lung 19
Right breast 6
Right lung 5
Spinal cord 5
Total: 100
![Page 5: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/5.jpg)
Intensity modulators:
MLC Speed (Max speed 2.5 cm/sec)
Dose Rate (0-1400 MU/min)
Gantry Speed (~4.8 deg/sec)
![Page 6: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/6.jpg)
Made the structures “high resolution” Used partially segmented OARs Used conformation structures (rings)
![Page 7: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/7.jpg)
![Page 8: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/8.jpg)
![Page 9: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/9.jpg)
![Page 10: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/10.jpg)
![Page 11: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/11.jpg)
Eclipse Ver 13.6 (AAA 11.0.31, TrueBeam) Limitations:
Target inhomogeneity multiple field sizes per target
Sparing ipsilateral OARs isocentre positioning
Sparing contralateral OAR’s limiting gantry rotation
MLC over-carriage max travel of 14.5 cm
Therefore, I used all available assets (nine coplanar partial arcs) @ 6 MV FFF
![Page 12: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/12.jpg)
![Page 13: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/13.jpg)
![Page 14: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/14.jpg)
Gantry 290 130 CW
Collimator 5 deg
![Page 15: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/15.jpg)
Gantry 119 330 CCW
Collimator 355 deg
![Page 16: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/16.jpg)
![Page 17: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/17.jpg)
Gantry 290 45 CW
Collimator 5 deg
![Page 18: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/18.jpg)
Gantry 179 70 CCW
Collimator 355 deg
![Page 19: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/19.jpg)
![Page 20: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/20.jpg)
Gantry 295 75 CW
Collimator 5 deg
![Page 21: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/21.jpg)
Gantry 75 295 CCW
Collimator 5 deg
![Page 22: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/22.jpg)
Gantry 82 170 CW
Collimator 355 deg
![Page 23: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/23.jpg)
Gantry 170 82 CCW
Collimator 355 deg
![Page 24: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/24.jpg)
![Page 25: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/25.jpg)
Gantry 179 95 CCW
Collimator 355 deg
![Page 26: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/26.jpg)
Start with targets and rings only:
SC + SC ring + overall ring
add axilla
add breast
add OARs
Try to be patient!
![Page 27: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/27.jpg)
![Page 28: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/28.jpg)
![Page 29: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/29.jpg)
Pause the optimizer until the cost function “plateaus” At the “odd number” steps
At the beginning of each phase (calc resolution)
Re-optimize: using the current plan dose as an intermediate
dose for optimization
![Page 30: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/30.jpg)
If your target priorities are P: 100%
Lt Lung: ▪ 95 % 15 % vol max: 2000 cGy
▪ 80% mean: 700 cGy
Rt Lung: ▪ 80 % 3 % vol max: 500 cGy
▪ 50% mean: 100 cGy
Rt Breast: ▪ 90 % 5 % vol max: 200 cGy
▪ 50% mean: 75 cGy
Heart: ▪ 35% mean: 400 cGy
Spinal cord: ▪ 40% 0% vol max: 600 cGy
▪ 40% mean: 150 cGy
![Page 31: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/31.jpg)
![Page 32: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/32.jpg)
![Page 33: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/33.jpg)
![Page 34: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/34.jpg)
![Page 35: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/35.jpg)
PTV
L Lung
Heart
R Brst
R Lung
Spinal Cord
![Page 36: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/36.jpg)
![Page 37: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/37.jpg)
![Page 38: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/38.jpg)
![Page 39: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/39.jpg)
Structure Min dose (cGy) Max dose (cGy) Mean dose (cGy)
PTV_TOTAL_EVAL 3868.8 5579.6 5118.7
Heart 48.9 4931.7 388.5
Left lung 69.2 5228.1 973.5
Right breast 0.0 847.9 47
Right lung 5.0 2363.0 103.7
Spinal cord 22.2 895.6 139.1
![Page 40: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/40.jpg)
Structure Available Points My score
Targets (coverage, conformation,
homogeneity, hotspot) 45 42.43
Heart 20 20
Left lung 19 17.62
Right breast 6 3.54
Right lung 5 5
Spinal cord 5 5
Total: 100 93.6
![Page 41: By Saad Aldelaijan - radmed.org - RapidArc Eclipse Webinar Prese… · By Saad Aldelaijan Medical Physicist, Radiation Physics Section Biomedical Physics Department King Faisal Specialist](https://reader033.vdocuments.mx/reader033/viewer/2022052010/60212be6eef8e106213d9e7a/html5/thumbnails/41.jpg)
General: Understand the criteria to score more
Make a quick table for points/organ
Contouring: Use high resolution structures
Use conformation structures (rings)
Use partially segmented structures for increased OARs sparing
Geometry: Nine coplanar partial arcs @ 6 MV FFF
SC (2), Axilla (2), Breast (4) and (1) “overall”
Optimization: Start with SC+rings , add axilla, add breast, then the rest
Try to be patient!