bt-750 - yahoo · 2019. 12. 18. · 5.4 ecg1, ecg2 setup ... bt-750 operator manual 7 p/n :...

66
P/N : OPM(BT-750)EN, Rev.02 BT-750 BT-750 PATIENT MONITOR

Upload: others

Post on 02-Feb-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

  • P/N : OPM(BT-750)EN, Rev.02

    BT-750

    BT-750

    PATIENT MONITOR

  • BT-750 Operator Manual 1

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Proprietary Materials Copyright © 2013 Bistos Co., Ltd. All rights reserved.

    Information and descriptions contained in this manual are the property of Bistos

    Corporation and may not be copied, reproduced, disseminated, or distributed without

    express written permission from Bistos Corporation.

    The materials in this document can contain typographic errors and be subject

    to change without notice.

    Printed in Korea

    Prior to beginning

    In this document, all information for correct using and operation is included.

    If you need a help for instruction or questions, please contact Bistos Co.,Ltd.

    Bistos Co.,Ltd. and the authorized agency do not take responsibility for the

    harmful cases that are happened by user’s obvious carelessness or incorrect uses.

    In this document, the product, system and equipment indicate BT-750. And the

    company and we indicate Bistos Co.,Ltd.

    We emphasize the safety and matters that require attention while using and

    operating this product by the following terms. User must know these terms – Warning, Caution, Note - before using this product.

    In the event that the product is damaged due to misuse or negligence by a user, the

    manufacturer or the authorized agents shall not be responsible for any damage or loss

    to the product.

    Warning

    The title “Warning” is used to inform the user of possible causes that could

    inflict the injury, death or property damage to the patients.

    Caution

    The title of “Caution” is used to inform the users of possible causes that

    could inflict the injury to the patients although it might not be severe enough

    to cause deaths.

    Note

    The title of “Note” is used to inform the users of items that are of importance

    in terms of installation, operation or maintenance of the equipment although

    the failure does not for the physical harm to the patients.

  • BT-750 Operator Manual 2

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Table Contents

    Proprietary Materials………………………………………………...1

    Prior to beginning……………………………………………………..1

    1. SAFETY ............................................................................................ 4 1.1 Intended Use ....................................................................................................... 4

    1.2 Instruction for the safe operation and use of the BT-750 .................................... 4

    1.3 General Precautions on Environment ................................................................. 4

    1.4 General Precautions on Electric safety ............................................................... 5

    1.5 Definition of Symbols ......................................................................................... 7

    2. INTRODUCTION ............................................................................ 8 2.1 About this Manual ............................................................................................... 8

    2.2 BT-750 Monitor Configuration ........................................................................... 8

    3. INSTALLATION ............................................................................ 10 3.1 Description of the BT-750 Front Panel ............................................................. 10

    3.2 Description of the BT-750 Rear Panel .............................................................. 11 3.3 Description of the BT-750 Right Panel ............................................................. 12 3.4 Description of the BT-750 Left Panel ............................................................... 12 3.5 Settings ............................................................................................................. 13

    4. BT-750 OPERATION ..................................................................... 17 4.1 Operation .......................................................................................................... 17

    4.2 Turn off the monitor .......................................................................................... 17 4.3 Display .............................................................................................................. 18 4.4 Basic Operation ................................................................................................. 19

    5. SYSTEM & ALARM SETUP ....................................................... 20 5.1 System Setup..................................................................................................... 20

    5.2 The modification of system setup(2nd

    display window) .................................... 24 5.3 Alarm Setup ...................................................................................................... 25 5.4 ECG1, ECG2 Setup .......................................................................................... 29

    5.5 SpO2 Setup ....................................................................................................... 32

    5.6 NIBP Setup ....................................................................................................... 33

    5.7 IBP Setup .......................................................................................................... 34

    5.8 Temperature Setup ............................................................................................ 36

    5.9 CO2 Setup ......................................................................................................... 37

    5.10 Trend Setup ..................................................................................................... 38

    5.11 Printer Setup .................................................................................................... 39

    6. MEASUREMENT OF ECG AND RESPIRATION .................... 40 6.1 Setup Connections ............................................................................................ 40

    6.2 Respiration Menu Display ................................................................................ 42

  • BT-750 Operator Manual 3

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    7. MEASUREMENT OF SPO2 ......................................................... 43 7.1 Preparation for SpO2 Measurement .................................................................. 43

    8. MEASUREMENT OF NIBP ......................................................... 45 8.1 Setting NIBP ..................................................................................................... 45

    9. MEASUREMENT OF TEMPERATURE .................................... 46 9.1 Temperature Measurement ................................................................................ 46

    10. MEASUREMENT OF IBP .......................................................... 47 10.1 IBP Measurement ............................................................................................ 47 10.2 Setup Connections .......................................................................................... 47 10.3 Adjustment of Zero point of the Converter ..................................................... 50

    11. MEASUREMENT OF ETCO2.................................................... 51 11.1 EtCO2 Measurement ....................................................................................... 51 11.1 How to measure Sidestream CO2 ................................................................... 52

    12. MAINTENANCE ......................................................................... 53 12.1 Maintenance and Cleaning .............................................................................. 53

    12.2 How to clean Accessories ............................................................................... 53

    13. TROUBLESHOOTING ............................................................... 55

    14. SPECIFICATION......................................................................... 57 14.1 Mechnical, Electric and Performance Specifications : Main Body ................. 57

    14.2 Mechnical, Electric and Performance Specifications : Accessories ................ 60 14.3 Compliances.................................................................................................... 61

    14.4 Environmental Conditions .............................................................................. 61 14.5 List of Standards ............................................................................................. 62

    11. WARRANTY ................................................................................. 64

  • BT-750 Operator Manual 4

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 1

    Safety

    1.1 Intended Use

    BT-750 is a patient monitor which displays state of patient’s data with wave form. Basic functions are measuring ECG 3lead select and respiration. Also it is possible to

    measuring arrhythmia, analysis ST level and detects pacemaker function. Other basic

    functions are measuring SpO2, pulse rate, NIBP, 2 IBP and 2 body temperatures.

    Optional functions are ECG 5 lead select and EtCO2 measurement.

    Note : This device is for use only by trained medical personnel located in a treatment

    room which is located in hospital, an operating room, a recovery room and an

    intensive care unit.

    Note : Because the incorrect use of this device can do patient’s life harm, this device

    must be installed, used and operated by trained medical personnel who can

    measure and analyze physiological signals.

    1.2 Instructions for the safe operation and use of the BT-750

    Examine the monitor and any accessories periodically to ensure that the cables, line cords,

    transducers, and instruments do not have visible evidence of damage that may affect patient

    safety or monitoring performance. The recommended inspection interval is once per week or

    less. Do not use the monitor if there is any visible sign of damage. Do not place the monitor

    to incline place to avoid falling of product.

    1.3 General Precautions on Environment Do not keep or operate the equipment under the environment listed below.

    Avoid placing in an area exposed to moisture. Do not touch the equipment with wet hand.

    Avoid exposure to direct sunlight

    Avoid placing in an area where there is a high variation of temperature. Operating temperature

    ranges from 10C to 40C.

    Avoid in the vicinity of Electric heater

  • BT-750 Operator Manual 5

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    1.4 General Precautions on Electric safety

    Prior to operation of the equipment, make sure to check followings

    Caution

    Whether power supply is appropriate. (100 – 240VAC)

    Caution

    Whether connections (power line or selected equipment) are properly

    made to the equipment.

    Caution

    Whether the correct measuring accessory for required parameter is

    connected to the equipment prior to turning on the equipment.

    Caution

    BT-750 can be used alongside with defibrillator while the accuracy

    of the product may be fallen in a short time

    Caution

    Use of this product is prohibited during MRI photographing. A fire

    may breakout from the induced current, and the accuracy of the

    product and the MRI may be influenced by cross interference.

    Caution

    BT-750 is intended only to measure and assess patient state. It must

    be used in conjunction with Clinical signs and symptoms when

    diagnosis is made.

    Caution

    The computerized ECG analysis should be judged by qualified

    medical professionals, and should not be the sole basis in

    determining necessity of treatment for the patient.

    Caution

    Leakage, heating, firing or disruption of the battery may cause fire or

    injury.

    Caution

    Do not use batteries beyond those specified in this manual.

    Caution

    Do not short-circuit the battery and does not heat, integrate, or

    dispose of it in a fire.

    Avoid placing in an area where there is an excessive humidity rise or ventilation problem. Operating humidity ranges

    to 80% RH max @40C.

    Avoid placing in an area where there is an excessive shock or vibration.

    Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage.

    Avoid dust and especially metal material into the equipment.

    Do not disjoint or disassemble the equipment. BISTOS Co., Ltd. does not take responsibility of it. Disassembling of equipment should be done only by the authorized personnel.

    Power off when the equipment is not fully installed. Otherwise, the equipment could be damaged.

  • BT-750 Operator Manual 6

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Caution

    Classification of BT-750 :

    - BT-750’s class is “Class IIb” equipment.

    - It is type “CF grade” for ECG, Resp, IBP and “BF grade: for

    EtCO2, SpO2, NIBP and Temp.

    - The noise level is“ B” Class according to the IEC/EN 60601-1

    (Safety of Electric Medical Product), and the noise redemption is

    “B” Level according to the IEC/EN 60601-1-2 (Electromagnetic

    Compatibility Requirements)

    - Protection grade against water is“ IPX 0”

    - Do not use this product around a flammable anesthetic and solvent.

    Caution

    Although BT-750 has been value of measurement in compliance

    with existing EMI/EMC requirements, use of this system in the

    presence of an electromagnetic field can cause momentary

    degradation of the SpO2 waveform. If this occurs often, Bistos

    suggests a review of the environment in which the system is being used, to

    identify possible sources of electromagnetic waves. These waves could be

    from other electrical devices used within the same room or an adjacent

    room. Communication devices such as cellular phones and pagers can

    cause these waves. The existence of radio, TV or microwave transmission

    equipment located nearby can cause electromagnetic waves. If EMI causes

    disturbances, it may be necessary to relocate your system.

    Electrostatic discharge (ESD), commonly referred to as a static

    shock, is a naturally occurring phenomenon. ESD is the most

    prevalent during conditions of low humidity, which can be caused by

    heating or air conditioning. During low humidity conditions,

    electrical charges naturally build up on individuals and can create

    static shocks. An ESD condition occurs when an individual with an

    electrical energy build-up comes in contact with objects such as

    metal door knobs, file cabinets, computer equipment, and even other

    individuals. The static shock or ESD is a discharge of the electrical

    energy build-up from a charged individual to a lesser or non-charged

    individual or object. The level of electrical energy discharged from a

    system user or patient to the BT-750 can be significant enough to

    cause damage to the system or probes. The following precautions can

    help to reduce ESD: anti-static spray on carpets; anti-static spray on

    linoleum; anti-static mats.

  • BT-750 Operator Manual 7

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    1.5 Description of product Symbols

    Symbol Description Symbol Description

    Type CF Product,

    Defibrillation

    protected

    Power Switch

    Type BF Product,

    Defibrillation

    protected

    Alarm On/Off

    Noninvasive Blood

    Pressure

    Manual

    Measurement(NIBP)

    TREND

    Warning Sign

    Period of Auto Blood

    Pressure Measurement

    Liquid Protection

    Grade Liquid Protection

    Grade

    T1~2 Measuring Temp. Start Printing (Printer)

    CO2 In/Output

    EtCO2

    Gas Input/Output Menu ON/OFF

    SPO2 SPO2 Measurement Ground

    P 1~2 Invasive Blood

    Pressure RS232 port and nurse

    call

    ECG Electrocardiograph /

    Respiration Screen Transfer

    NIBP Non Invasive Blood

    Pressure

    Use for Central

    Monitoring

    Table 1-1. Definitions of symbols

    1.6 Description of equipment Symbols

    The symbols in the bottom of LCD indicates the present status of BT-750 and use

    bottoms for each function.

    Indicate that Audio alarm generates

    Display for current charging state of charger. So, make sure of the

    charging state when you use only with charger while the AC power is on.

    :Indicate that AC power is on

  • BT-750 Operator Manual 8

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 2

    Introduction

    2.1 Feature BT-750

    BT-750 is a patient monitor which display state of patient’s data with waveform.

    ※ Basic functions of Patient measuring

    - ECG 3 lead (I, II, III,) select and Respiration measurement

    - Arrhythmia measurement, ST Level analysis and Pacemaker Detect function

    - SpO2 and Pulse Rate measurement

    - NIBP, 2 IBP measurements

    - Body Temperature (2 Temp)

    ※ Optional functions of Patient measuring

    - ECG 5 lead (I, II, III, aVR, aVL, aVF, V) select

    - End-tidal partial pressure respiration (EtCO2)

    ※ Etc

    - Trend Data Saving for 72 hours

    - VGA (Display Output) Function

    - Clone Function (This function support patient checking thru Internet)

    - Built-in 3 channel Printer. (Option)

    2.2 List of Components

    Standard Accessories

    Items Qty

    BT-750 body 1 ea

    SpO2 Finger Probe(Adult) 1M 1 ea

    SpO2 Extension CABLE 2M 1 ea

    NIBP cuff (Adult) 1 ea

    NIBP HOSE 1 ea

    ECG Cable 3 lead type 1 ea

    ECG electrode (3M) 3 ea

    220V Power cord 1ea

    Operation Manual 1ea

    Table 2-2. Standard Accessories

  • BT-750 Operator Manual 9

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Optional Accessories

    Items

    SpO2 Finger Probe(Child, Neonate) DISPOSABLE SPO2 SENSOR

    Disposable SpO2 Finger

    Probe(Neonate) Press Measuring Kit

    NIBP Cuff (BIG Adult, Child,

    Pediatric, Infant) ECG Cable 5 lead Type

    Printer Module EtCO2 Kit set

    Roll Paper Temperature Sensor

    IBP KIT IBP Extension cable 2M

    Temperature sensor ( Rectal type)

    Table 2-3. Optional Accessories

    Caution

    The different results can come out according to the status of patients.

    If the results are doubtful, please check the patient’s status using other measuring equipment which is using in hospital.

    Caution

    Use only authorized accessories supplied with BT-750. Use of any

    other accessories may result in out of specification performance and

    possible safety hazards.

  • BT-750 Operator Manual 10

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 3

    Installation

    3.1 Description of the BT-750 Front Panel

    ▶ Function and Explanation

    3.2 Description of the BT-750 Rear Panel

    No Switch / LED Function Explanation

    1 AC Indicator When AC power is connected and main power switch is

    on, green light is on

    2 NIBP Switch It is used to measure blood pressure manually.

    3 Power Switch It supplies power to the monitor

    4 Printer Switch Start/Stop Printer

    5 Battery

    Charge Indicator

    On Battery charged, orange LED is on and while AC

    power is disconnected, LED is off.

    6 Alarm Switch Used for alarm or alarm silence

    7 Trend Switch The equipment stores SpO2 and pulse rate for 72 hours in

    every 1 minute and used to check it

    8 Alarm Indicator When the alarm is on it will show on led

    9 Menu Switch It shows menu on the monitor

    10 Speaker Sound of Alarm and Heart rate.

    11 Selection trim knob It shows entire menu selection.

  • BT-750 Operator Manual 11

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Figure 3-2. Rear panel

    No. Item Function Explanation

    1 AC Power Input

    Port

    Port for AC adopter to connect exterior power to the

    monitor.

    2 VGA Port It is used for when VGA port is connecting to other

    Monitor

    3 RS232 Port The communication port for connecting monitor and

    computer.

    4 LAN Port The communication port for connecting monitor and

    external devices.

    5 Label Display

    6 Ground Port Port of Ground Cable

    7 Battery Store Port of Battery

    8 Portable Handle Handle of carry

    3.3 Description of the BT-750 Right Panel

  • BT-750 Operator Manual 12

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Figure 3-3. Right panel

    No. Item Function Explanation

    1 Printer Power LED When Printer Power is connected ,green light is on

    2 Printer Error LED When Printer is in malfunction ,red light is on

    3 FEED Switch Printer Paper selection

    4 Printer Cover Inserting the printer paper and close the cover

    3.4 Description of the BT-750 Left Panel

    Figure 3-4. Left panel

    3.5 Settings

    No. Item Function Explanation

    1 ECG/RESP Connection Terminal It is connector of Measuring ECG/RESP

    2 SpO2 Connection Terminal Connector of SpO2 Sensor

    3 IBP 1 Connector of IBP 1

    4 IBP 2 Connector of IBP 2

    5 TEMP 1 Connection Terminal Connector of Temp 1 Sensor

    6 TEMP 2 Connection Terminal Connector of Temp 2 Sensor

    7 Cuff Connection Terminal Connector of NIBP Cuff Sensor

    8 CO2 Input, Output Connection

    Terminal Connector of CO2 Input and Output

  • BT-750 Operator Manual 13

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Precautions for Settings

    Pay attention when setting BT-750 for the following points

    ● Use BT-750 under conditions of 10℃∼40℃ of circumstantial

    temperature and within 80% humidity.

    ● Check the connection status of the power cord.

    ● Do not connect multiple cords to the power supply.

    Place the main body on a flat area.

    If noise occurs, ground the equipment further.

    All information is stored in the internal memory, even in the case of power on and off.

    Do no use any electric cords generating connecting noise.

    Take care since the product can be broken by mechanical shock.

    Get rid of dust or inflammable material near the product when setting it up.

    3.5.1. How to Use when power is on.

    1) Insert AC Adaptor to AC-IN Connector at the rear of BT-750

    2) Turn on the On/Off switch at the rear of BT-750

    ( If the power is in Stand-by , battery starts to charge. )

    3) If AC lamp of the front side (Green light) is on and you press the power switch

    [AC Adaptor Ports]

    [the front side of the product]

    Note

    Though AC power cord is connected, charging lamp continuously

    lights on and after completing charging, the lamp turns off.

    ※ AC Power

    If AC power is connected to the monitor and turn on the AC power switch, Green LED is on to

    indicate Power-on. It is automatically converted to a recharging mode. (Input 100~240V, 50/60Hz)

    Caution

    Noise may occur in waveform, while using AC power. In this case,

    ground the product.

    It may occur the fatal damage to the product ,while connecting AC

    power with wet hand.

    3.5.2 How to Use Battery

    If you continuously press the sub power switch for

    about 1 second, the power is on.

    Lamp is on during battery charging. (Orange light)

    If you turn on the main power switch, the lamp

    lights up while AC adaptor is connected.

  • BT-750 Operator Manual 14

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    In case of sudden blackout or potable purpose, automatically AC Adaptor power transfers to

    battery power for continuous use.

    How to use

    1. Battery charge lamp is off, while battery used. 2. It is durable for approximately 4 hours but depends on battery charged state. ( Full

    charged time: 4 hours)

    3. Battery charging starts once main power switch of rear side is on after AC Adaptor connection.

    4. Battery charged state displays in 5 steps.

    Full charged 70% charged 50% charged 20% charged Discharged

    5. When battery is low charged, battery state display on the screen is turned off, beep sound generates and after about 5 minutes, power is disconnected automatically in

    order to protect battery.

    Note

    When battery electronic pressure is in low state, it may cause

    inaccurately measured value. Once battery state is low, connect AC

    Adaptor to the appliance at once.

    Note

    For battery replacement, make sure of polarity.

    If the battery polarity is placed incorrectly, it may cause fatal

    problem on equipment.

    Contact to Bistos head quarter or authorized agent, when you need

    battery replacement.

    If a side of power cord is put to power port of BT-750, power is supplied to the equipment. Make sure

    that each cord is connected to input / output terminal correctly. If there is incorrect connection,

    it may cause problem.

    ※ Battery Charge

    If AC power is disconnected, power is supplied thru battery. When battery power is not

    sufficient, you should charge it by connecting AC adaptor.

    Recharging time : More than 5 hours.

    Duration time: Under 3 hours.

    In the case of discharge battery.

    Caution

    Noise may occur in waveform, while using AC power. In this case,

    ground the product.

    It may occur the fatal damage to the product ,while connecting AC

    power with wet hand.

    Caution

    While the monitor is working, battery charging lamp is on. When

    sub-power switch is turned off and battery charging completes, lamp

    is out.

    ※ Battery type

    BT-750 uses Lithium-lon battery.

  • BT-750 Operator Manual 15

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    The table below shows the battery life while not in use.

    Temperature Shelf Life

    21 °C to 30 °C 12 months

    31 °C to 40 °C 6 months

    ※ Battery Replacement

    Replace the battery with a same type of battery. [The product supplied by Bistos Co.,Ltd.]

    Type: Rechargeable Lithium-Ion Battery

    Company : LG Chem

    ※ Battery Replacement Period

    If you start to use the system in battery-operated mode, it is necessary to keep it in use

    without stop, meanwhile, in care of discharge, the battery life will be shortened. You may

    refer the table below for battery replacement period.

    Rate of charge/discharge Times

    100 % charge/discharge 100 times

    50 % charge/discharge 200 times

    30 % charge/discharge 500 times

    You should replace the battery when the duration time gets shorten or charging is not done

    properly. When the monitor is connected to the power, battery is charged automatically.

    You can not use the battery, with separated from the monitor.

    ※ Battery Storage Maintenance

    Storage temperature (Humidity : 20 ~ 85 %RH without condensation)

    -20 ~ 20 ℃ : less than 1 year

    -20 ~ 45 ℃: less than 3 months

    -20 ~ 60 ℃ : less than 1 month

    ※ How to replace Battery

    You must contact the headquarter or agent of Bistos Co.,Ltd.

    Caution

    Dispose of the battery into a proper place for the protection of

    environment. Keep the battery dry.

    Caution

    Do not disassemble or reorganize the battery.

    3.5.3 Connection of Accessories

  • BT-750 Operator Manual 16

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Connect the accessories for measurement to each mating port of the equipment.

    3.5.4 Replacing Roll paper

    ① Verify if the printer is receiving the data, otherwise it will lose the receiving data.

    ② Open the printer door by pushing the button at the both side of the printer slightly as shown the figure below.

    ③ insert a new paper roll oriented properly like above figure, after verifying the direction of the paper. Remove the empty paper core and insert a new paper roll oriented properly.

    ④ Pull the paper out towards you until approximately 2 inches of paper and align the paper. Close the printer door.

    Caution

    ● When the printer door is not perfectly closed, the printing could

    not be operated or printed clearly.

    ● When printing more than 1 minute, the overheating of the printer

    could be faulty. In this case, stop the printing at least 3 minutes.

    ● Please use the right paper and do not use the core adhesive roll. If

    you use it, it may occur the malfunction as the printer fails to find the

    end of the paper.

  • BT-750 Operator Manual 17

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 4

    BT-750 Operation

    Prior to the Use of BT-750

    Make sure of followings prior to check the state of patient.

    Check if there is any factor to occur mechanical problem Review the cosmetic state of probe Check if there is any error in fiber-optic probe and AC power cable. Connect the probe to the probe port. Turn on the monitor and identify that the state of fiber-optic probe is normal and the cable is connected.

    Identify that fiber-optic probe has any problem on connection.

    4.1 Operation

    1) Turn on the power.

    2) Adjust the setting with exterior button when change of the value required.

    3) Figure and waveform are created on LCD screen once ECG probe connected to the

    patient.

    4) If you apply the SpO2 probe to the patient, data of the patient is displayed on LCD.

    5) Apply NIBP cuff to patient. Avoid the arm with SpO2 probe.

    6) Apply IBP Kit to patient. Avoid the arm with NIBP cuff.

    7) Connect the temperature sensor to the proper position of patient at the type of the sensor.

    8) Apply CO2 kit. (optional function)

    ※ Operation Notice - Only licensed person must use BT-750. - Read carefully followings prior to use the equipment. - Check the state of battery charging. - Make sure that power and probe are connected correctly to the monitor. - Avoid sudden impact when the machine is transported or each unit is connected. - Read carefully followings prior to use the equipment. - Identify that any error does not occur in entire equipment and patient. - Care for the patient safely and preferentially if any error is found on equipment and take

    possible action such as stopping the device operation.

    - Do not give any impact to the equipment. - Keep the probe not to be taken off from the port for exact measurement.

    4.2 Turn off the monitor Turn off the sub-power and verify if the LCD screen turns off.

  • BT-750 Operator Manual 18

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    4.3 Displays

    1. Basic screen composition

    Press the button in front of BT-750 for power-on.

    Display window shows company logo for 2 seconds and is changed to default setting.

    ▶ Menu 1 Sweep Speed 13 IBP2 Diastolic 25 Battery Status

    2 Status of Filter 14 NIBP Systolic 26 Date and Year

    Display

    3 Detection of

    Arrhythmia 15 NIBP Diastolic 27 Time Display

    4 Heart Rate

    Measurement 16 NIBP Mean 28

    Respiration Alarm

    Limit

    5 Heart Rate Alarm

    Limit 17 EtCO2 Measurement 29

    Respiration

    Measurement

    6 PVC Detect

    Measurement 18 InCO2 Measurement 30 ECG 1 Wave

    7 ST Segment Data 19 Temp. Measurement 31 Wave 2

    8 SpO2 Measurement 20 Temperature Alarm

    Limit 32 Wave 3

    9 SpO2 Bar Graph 21 Bed Number 33 Wave 4

    10 SpO2 Alarm Limit 22 Patient Mode 34 Wave 5

    11 IBP1 Measurement 23 Symbol of Alarm 35 LEAD Display

    12 IBP2 Systolic 24 Symbol of AC

    Power

    ※ When you want to exit from main display or return to basic display by pressing menu button.

  • BT-750 Operator Manual 19

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    4.4 Basic Operation

    4.4.1 Simple operation

    Order Operation Method

    1 Push the power switch for 1~2 second

    2 If you want to change the setting value etc. during use, push the Menu button

    to change the setting values per each measurement parameter.

    3 When the measurement sensor is connected to the patient, the information on

    the patient is displayed as values and waveforms on the LCD screen.

    4 Refer to the table below for the details of how to measure each parameter..

  • BT-750 Operator Manual 20

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 5

    System & Alarm Setup You can set up alarm range, volume, date, time of BT-750 in this menu.

    5.1 System Setup

    1) Press the button “ ” to select setup menu.

    2) Use the trim knob to go to required location and press the trim knob

    to set the location to be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    Value and save it.

    4) Press the button “ ” to return to main display.

    Setup Mode Range of Change Description

    CHANNEL

    SETUP

    3~6 waveforms

    Set up the location at each

    waveform.

    1~8

    Input the number of channels for

    waveform Set up the location at each

    waveform.

    Set up the color at each waveform

    BED NUMBER 01 ~ 99 Input BED numbers

    PATIENT ADULT/Pediatric/Neonate Set the age of patient

    DATE/ TIME YEAR : 2000 ~ 2099

    MONTH : 1 ~ 12

    DATE : 1 ~ 31

    HOUR: 0 ~ 23

    MINUTE: 0 ~ 59

    Date/Time Input

    PRINTER Refer to the setup of printer at 16th

    chapter

    KEY/BEEP

    VOLUME OFF/ 1 ~ 7(Change at 1 unit ) KEY, Beep Sound setup

    ALARM

    VOLUME OFF/ 1 ~ 7 (Change at 1 unit ) Set up the volume of ALARM Sound

  • BT-750 Operator Manual 21

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.1.1 Factory Setting

    Parameter Item Name Adult Pediatric Neonate

    ECG

    LEAD 1st CH LEAD II LEAD II LEAD II

    2st CH LEAD III LEAD III LEAD III

    GAIN 1st CH

    ⅹ1 ⅹ1 ⅹ1 2st CH

    SPEED 1st CH

    25 mm/sec 25 mm/sec 25 mm/sec 2st CH

    CONFIG

    MODE MONITOR MONITOR MONITOR

    LEAD 3 LEAD 3 LEAD 3 LEAD

    ARRHYTHMIA OFF OFF OFF

    PACEMAKER OFF OFF OFF

    HR ALARM ON ON ON

    ST P-R 96mS 96mS 96mS

    J 92mS 92mS 92mS

    ST ALARM HIGH Limit +3.6mm +3.6mm +3.6mm

    LOW Limit −3.6mm −3.6mm −3.6mm

    ALARM HIGH Limit 120 bpm 160 bpm 200 bpm

    LOW Limit 50 bpm 75 bpm 100 bpm

    SpO2

    WAVE SPEED 25 mm/sec 25 mm/sec 25 mm/sec

    PR ON ON ON

    ALARM

    ALARM ON ON ON

    HIGH Limit OFF OFF OFF

    LOW Limit 90 % 90 % 90 %

    TEMP

    UNIT ℃ / ℉ ℃ ℃ ℃

    ALARM

    1st CH ON ON ON

    2st CH ON ON ON

    1st HIGH 39.0℃/102.2℉ 39.0℃/102.2℉ 39.0℃/102.2

    2st LOW 36.0℃/96.8℉ 36.0℃/96.8℉ 36.0℃/96.8℉

    1st HIGH 39.0℃/102.2℉ 39.0℃/102.2℉ 39.0℃/102.2

    2st LOW 36.0℃/96.8℉ 36.0℃/96.8℉ 36.0℃/96.8℉

    RESP

    SPEED 12.5 mm/sec 12.5 mm/sec 12.5 mm/sec

    LEAD RA-LL RA-LL RA-LL

    GAIN ⅹ2 ⅹ2 ⅹ2

    APNEA ALARM OFF OFF OFF

    APNES TIME[SEC] 30 sec 30 sec 30 sec

  • BT-750 Operator Manual 22

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    ALARM

    ON / OFF OFF OFF OFF

    HIGH 30rpm 30rpm 100 rpm

    LOW 8rpm 8rpm 30 rpm

    NIBP

    MODE MANUAL MANUAL MANUAL

    INTERVAL 60 MIN 60 MIN 60 MIN

    UNIT mmHg mmHg mmHg

    ALARM

    ON / OFF ON ON ON

    HIGH 160/90(110)mmHg

    21.3/12.0(14.7)kPa

    120/70(90)mmHg

    16.0/9.3(12.0) kPa

    90/40(70)mmHg

    12.0/8.0(9.3)

    kPa

    LOW 90/50(60)mmHg

    12.0/6.7(8.0) kPa

    70/40(50)mmHg

    9.3/5.3(6.7) kPa

    60/20(30)mmHg

    5.3/2.7(4.0) kPa

    EtCO2

    WAVE SPEED 12.5 mm/sec 12.5 mm/sec 12.5 mm/sec

    UNIT mmHg mmHg mmHg

    SCALE ~25mmHg ~25mmHg ~25mmHg

    AVERAGE 10 SEC 10 SEC 10 SEC

    GAS COM. NON NON NON

    DISABLE/ENABLE DISABLE DISABLE DISABLE

    ALARM

    ON / OFF OFF OFF OFF

    EtCO2

    HIGH

    80mmHg

    10.7 kPa

    10.5%

    80mmHg

    10.7 kPa

    10.5%

    80mmHg

    10.7 kPa

    10.5%

    LOW

    00mmHg

    0 kPa

    0%

    00mmHg

    0 kPa

    0%

    00mmHg

    0 kPa

    0%

    INCO2

    HIGH

    20mmHg

    2.7 kPa

    2.6%

    20mmHg

    2.7 kPa

    2.6%

    20mmHg

    2.7 kPa

    2.6%

    LOW

    00mmHg

    0 kPa

    0%

    00mmHg

    0 kPa

    0%

    00mmHg

    0 kPa

    0%

    IBP

    SPEED 1st CH 25 mm/sec 25 mm/sec 25 mm/sec

    2st CH 25 mm/sec 25 mm/sec 25 mm/sec

    LABEL 1st CH P1 P1 P1

    2st CH P2 P2 P2

    SCALE 1st CH ~200mmHg ~200mmHg ~200mmHg

  • BT-750 Operator Manual 23

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    2st CH -50mmHg -50mmHg 50mmHg

    ALARM

    1st CH 0N 0N 0N

    2st CH 0N 0N 0N

    1st CH

    HIGH 160/90(110)mmHg

    21.3/12.0(14.7)kPa

    120/70(90)mmHg

    16.0/9.3(12.0) kPa

    90/40(70)mmHg

    12.0/8.0(9.3) kPa

    LOW 90/50(60)mmHg

    12.0/6.7(8.0) kPa

    70/40(50)mmHg

    9.3/5.3(6.7) kPa

    60/20(30)mmHg

    5.3/2.7(4.0) kPa

    2st CH

    HIGH 160/90(110)mmHg

    21.3/12.0(14.7)kPa

    120/70(90)mmHg

    16.0/9.3(12.0) kPa

    90/40(70)mmHg

    12.0/8.0(9.3) kPa

    LOW 90/50(60)mmHg

    12.0/6.7(8.0) kPa

    70/40(50)mmHg

    9.3/5.3(6.7) kPa

    60/20(30)mm

    Hg

    5.3/2.7(4.0)

    kPa

    IBP DISABLE/ENABLE ENABLE ENABLE ENABLE

    The other setup values

    SETU

    P

    CHANNEL

    SETUP

    DISPLAY 5-CH WAVE

    Wave ECG1,PLETH,BP1,BP2,PR

    BED NUMBER 01

    PATIENT ADULT

    KEY/BEEP VOLUME 02

    ALARM VOLUME 02

  • BT-750 Operator Manual 24

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.2 The modification of system setup (2nd

    display window)

    On operating the service mode such as CHANNEL SET, Date /Time, Printer setup, additional

    active window is opened.

    5.2.1 Selection of CHANNEL

    1) Go to the required location with trim knob and select the location, then the display

    window becomes active.

    2) Set up the values, turning the trim knob to the right / left and save the values after

    pressing the trim knob.

    3) Press “ok” button to revert to the main display window

    4) On selection of ‘exit”, it turns to 1st display window.

    5) Press “ ” button to revert to the main display window.

    Setup

    Mode

    Range of Change Description

    DISPLAY 3~6 waveforms

    Input the number of

    channels for waveform

    display

    Selection of

    waveform

    Selection of

    location

    1CH : ECG 1 (fixed)

    2CH : ECG2, PELTH, BP1, BP2 (4)

    3CH : PELTH, BP1, BP2, RR, CAPNO (5)

    4CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)

    5CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)

    6CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)

    Set up the location at

    each waveform

    Color at

    each

    waveform

    8 Set up the color at each

    waveform.

    5.2.2 Setup date /time

    1) Go to the required location with trim knob and select the location, then the display

    window becomes active.

    2) Set up the values, turning the trim knob to the right / left and save the values after

    pressing the trim knob.

    3) Press “RETURN” button to go to 1st display window.

  • BT-750 Operator Manual 25

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.3 Alarm Setup

    5.3.1 Modification Display Window for 1st

    Setup (SpO2,Respiration and Temp.)

    1) Press button “ ” to go to adjustment display of alarm

    2) Select the menu to change with trim knob and press the trim knob to set the location to be

    modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric value

    and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

  • BT-750 Operator Manual 26

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.3.2 Modification Display Window for the 2nd

    Setup (IBP, NIBP, ECG HR, EtCO2, ST)

    1) Press the button “ ” to go to adjustment display

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Activate the display window by pressing trim knob

    4) Use the trim knob to go to required location and press the trim knob

    to set the location to be modified. Then, selected menu is displayed.

    5) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it..

    6) Press “Return” button to go back to 1st display window.

    7) Press the button “ ” to return to main display.

  • BT-750 Operator Manual 27

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.3.3 Alarm properties

    BT-750 offers Alarm function to user. User should have full knowledge about this chapter

    and take right action on alarming. BT-750 provides alarm service in two ways. One is

    audio Alarm and the other is alarm message.

    1) AUDIO ALARM

    The high level alarm beeps 5 times and it beeps 5 times. The high level alarm has a

    specific rhythm that meets international standard. The medium level alarm beeps 3 times

    at 15 second intervals. Low-level alarm beeps to inform the occurrence of following

    state. Low level alarm beeps once at 30 seconds

    Alarm type Tone Pitch Beep Rate

    IEC60601-1‐8 IEC60601-1‐8

    High Level Alarm ~976 Hz 10 beeps in 15 sec

    Medium Level Alarm ~697 Hz 3 beeps in 15 sec

    Low Level Alarm ~488 Hz 1 beeps in 30 sec

    High Level Alarm

    It operates only when the value is extremely in abnormal state. For example, if pulse rate indicates

    0, high level alarm generates regardless of alarm limit.

    Medium Level Alarm

    It operates only when the value is beyond alarm limit.

    Low level Alarm

    It beeps when abnormal result brings up or there is anything that the user needs to be

    informed.

    Note

    When you use battery and alarm generates , the alarm volume is

    small.

    Visual alarm

    To help you detect the cause of an alarm, the numeric values of the relevant items keep

    blinking.

  • BT-750 Operator Manual 28

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    2) Alarm icon

    The alarm icon appears on the information window of the lower screen..

    Function Message Description

    SpO2 STAND BY Probe is connected but the finger is out of probe.

    CHECK PROBE Finger probe is out of probes

    ECG

    CHECK SENSOR Sensor is out of body or cable is out of the monitor IBP

    Temperature

    EtCO2

    NIBP

    CUFF HOSE Failure of pressurization for definite time

    NO PULSE Fail to detect pulse rate

    REDA FAIL Error in measurement values for moving

    ALARM

    LOW

    Refer to Alarm Generating Chart MIDIULM

    HIGH

    [Alarm Generating Chart]

    Message Generating Cause

    HIGH When pulse rate is not detected (“0” BPM)

    MEDIUN Out of numeric value

    LOW

    1. SP02 PROBE ERROR 2. LOW BATTERY

    3. Fail to measure NIBP or Over Pressure

    4. Lead Fault during ECG measurement

    5. Finger is out of SpO2 probe during SpO2 measurement

    6. Fault during Temp measurement 7. Fault during IBP measurement

    8. PACEMAKER Detect

    Alarming display is illuminated in every alarming. It is changed to the below chart,

    in every moment pressing alarming button ,when alarm generates.

    NO DISPLAY SOUND REMARK

    1

    Normal alarming Default

    2

    Holding alarm for 60 sec. Pressing alarm button once

    3

    Alarm sound off for 120 sec. Pressing alarm button twice

    4

    Alarm off Refer to the chapter 14

    Please press the alarm button ‘once’ in low level alarm to cancel it.

    Caution

    Messages generate and indicators are illuminated, regardless of

    alarming or not.

    3) Setting Up Alarm Hold

    After alarm generates, user can hold audio alarm and set up the time to 1min/2min/normal.

    If each event which generates alarming within setting time is not cancelled, audio alarm will

    generate again.

  • BT-750 Operator Manual 29

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.4 ECG1, ECG2 Setup

    5.4.1 ECG Menu Display

    1) “Press the button “ ” to go to adjustment display

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric value

    and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    Setup Mode Range of modification Description

    ECG 3 LEAD LEAD I, II, III

    Set ECG Lead ECG 5 LEAD LEAD I, II, III, aVR, aVL,

    aVF, V(Chest)

    ECG GAIN 1/2,1,2 times Set the size of ECG wave

    WAVE SPEED 6.5, 12.5, 25, 50 mm/sec Set the velocity of ECG wave

    ECG MODE Monitor/Operation/Diagnosis

    Set Analog and Digital filter in ECG

    wave

    LEAD MODE 3 LEAD, 5 LEAD Set ECG Cable

    ARRHYTHMIA ON/OFF Set the usage of Arrhythmia analysis

    PACEMAKER ON/OFF Set the detection of the pacemaker

    PVC ALARM ON/1~30 Select whether the alarm generates

    or no on detection of PVC .

    HR ALARM Refer to Ch.8 Alarm setup

    ST

    SEGMENT

    P-R /J + 4~252 ms ,Alarm

    ON/OFF Set P-R, J Point

  • BT-750 Operator Manual 30

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.4.2 ECG2 Menu Display

    1) “Press the button “ ” to go to adjustment display

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric value

    and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    Setup Mode Range of modification Description

    ECG 3 LEAD LEAD I, II, III

    Set ECG Lead ECG 5 LEAD

    LEAD I, II, III, aVR, aVL, aVF,

    V(Chest)

    ECG GAIN 1/2,1,2 times Set the size of ECG wave

    WAVE

    SPEED 6.5, 12.5, 25, 50 mm/sec Set the velocity of ECG wave

  • BT-750 Operator Manual 31

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.4.2.1 ST Analysis (When ST Segment is on)

    ST Level is measured by horizontal comparison between PR-segment and ST-segment.

    User can adjust J point and P-R point at 1~250 4m/sec interval. These measured values

    are available only in normal heart rate. In case of abnormal and pacemaker, for example,

    Premature Ventricular Contraction, Ventricular Tachycardia etc., it is except ST analysis

    because it is possible to cause distortion of ST form.

    1) “Press the button “ ” to go to adjustment display

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric value

    and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    ● Arrhythmia Analysis can judge problem of Ventricle.

    Caution Our ST measurement is only on the basis of ECG1.

    5.4.2.2 PACEMAKER Detection (When PACEMAKER is on)

    ● In case Pacemaker is Detection ON, red bar indicate it on the ECG wave form periodically

    from input of pacemaker detection

  • BT-750 Operator Manual 32

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.5 SpO2 Setup

    [SpO2 Menu Display]

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    Range of SpO2 setup

    Setup Mode Range of change Description

    Wave Speed 6.5,12.5,25,50 mm/sec SpO2 Waveform Speed

    SpO2 PR Pulse rate ON/OFF Display Pulse Rate ON/OFF

    SpO2 Alarm Refer to Ch.8 Alarm setup

  • BT-750 Operator Manual 33

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.6 NIBP Setup

    [NIBP Menu Display]

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it..

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display

    Range of NIBP Setup

    Setup Mode Range of change Description

    NIBP Mode AUTO , Manual, Stat Setup Automatic/Manual/Stat

    measurement

    Interval 1, 3, 5, 10, 30, 60, 90, 120, 240

    min

    Set the interval of the automatic

    NIBP measurement time

    NIBP Unit mmHg, kPa Set the unit of blood pressure

    NIBP Alarm Refer to Ch. 8 Alarm Setup

    *STST Mode is Setup mode to measure the blood pressure for 5 minutes continuously.

    Note

    When patient needs to use Stat Mode, it must press button at

    the bottom of the monitor after adjustment of Stat Mode.

  • BT-750 Operator Manual 34

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.7 IBP Menu Display

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location

    to be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    ● Table for labels according to the site of measurement for BP1

    Label Site Scale

    ABP Arterial blood pressure 0~200 mmHg

    P1 Standard labels 0~200 mmHg

    ● Table for labels according to the site of measurement for BP21

    Label Site Scale

    P2 Standard labels 0~25 mmHg

    CVP Central venous pressure 0~25 mmHg

    PAP Pulmonary arterial pressure 0~50 mmHg

    LAP Left arterial pressure 0~25 mmHg

  • BT-750 Operator Manual 35

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    ●IBP measurement range

    Setup Mode Range of change Description

    WAVE

    SPEED

    6.5, 12.5, 25,50 mm/sec Set the speed of IBP waveform

    BP1 LABEL P1 , ABP

    Set the level according to the

    measurement site of BP. Change the

    alarm setting range according to

    level.

    BP1 SCALE 50,100,200,300 mmHg Set the display of BP1 wave scale

    BP1 ZERO YES/NO Adjust the zero point of the

    transducer

    BP1 ALARM Refer to Ch.8 Alarm Setup

    BP2 LABEL

    P2,CVP,PAP,LAP

    Set the level according to the

    measurement site of BP. Change the

    alarm setting range according to

    level.

    BP2 SCALE 25,50,100,200,300 mmHg Set the display of BP2 wave scale

    BP2 ZERO YES/NO Adjust the zero point of the

    transducer

    BP2 ALARM Refer to Ch.8 Alarm Setup

    IBP Enable / DISABLE Select the IBP display

    IBP inactive display window (Show the unified NIBP display window.)

  • BT-750 Operator Manual 36

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.8 Temperature Setup

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    ● Temperature Menu Setup Range

    Setup Mode Range of change Description

    Temp Unit ℃, ℉ Set the unit of temperature

    Temp 1 Alarm Refer to Ch.8 Alarm Setup

    Temp 2 Alarm

  • BT-750 Operator Manual 37

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.9 CO2 Setup

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    ● Setup range of EtCO2

    Setup Mode Range of Change Description

    WAVE

    SPEED

    6.5, 12.5, 25 mm/sec Set the speed of EtCO2 waveform

    UNIT mmHg, kPa, % Set the unit of EtCO2

    measurement

    SCALE ~12,~25, ~50, ~70, ~100,~150

    mmHg Set the WAVE SCALE

    ZERO YES / NO Adjust the zero point of the

    transducer

    AVERAGE 1 BREATH, 10, 20 Sec Set DATA average time DATA

    Gas COMP NON,O2/N2O, N2O, O2 Set the Gas type

    EtCO2 Enable / DISABLE Set ON/OFF of EtCO2 module

    ALARM Refer to Ch.8 Alarm Setup

  • BT-750 Operator Manual 38

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.10 Trend Setup

    1) Press the button “ ”to convert to Trend display window.

    2) Adjust the tram knob and select SCAN Then, selected bar becomes active. Adjust the bar to

    the required location with tram knob.

    3) Select the PAGE with tram knob, to turn over display window.

    4) Press the button “ ” to return to main display.

    Mode Description

    SCAN Select the data of current page by adjusting Trim Knob.

    PAGE

    Could see 1~17 pages by adjusting Trim Knob

    Show the data of 174 cases each page.

    -- SEC Store interval data by selected interval second.

    Store the data of approximately 2960 cases

    Note

    Up to the selection of Second, it could store minimum 8 hours to

    maximum 72 hours trend data.

  • BT-750 Operator Manual 39

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5.11 Printer Setup

    [Pinter Menu Display]

    1) Press the menu button “ ” to go to Setup Printer Setup.

    2) Select the menu to change with trim knob and press the trim knob to set the location

    to be modified. Then, selected menu is displayed.

    3) Press the trim knob to make the display window active in second time.

    4) Go to the required location with tramp knob and press the tram knob to set thelocation

    to be modified ,then the setting location becomes active .

    5) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    6) Press “Return” button to go back to 1st display window.

    7) Press the button “ ” to return to main display.

    ● Select the each Printer Mode using same method.

    Setup Mode Range of change Description

    WAVE ECG 1+ PLETH

    ECG 1+ CAPNO Select the waveform to print out

    PRINT SPEED 15/30 mm/sec Select speed of printer

    PRINT

    ALARM ON / OFF

    ON: Print out in case of over alarm limit

    automatically.

    OFF: manual printing

  • BT-750 Operator Manual 40

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 6

    Measurement of ECG and Respiration 6.1 Setup Connections

    1) Connect the ECG cable to the ECG connection terminal of the measurement module.

    ● The process of depolarization and polarization of the myocardium generates electronic

    potentials that are sensed by ECG electrodes on the skin surface.

    ● Changes in the chest impedance resulting from the patient’s respiration is measured and

    displayed as a waveform and the values will appear on the screen.

    2) Attach the patient’s electrode to the sites shown below.

    R: Red

    L: Yellow

    C: White

    N: Black F: Green

    R: Red

    L: Yellow

    F: Green

    ECG Connector

  • BT-750 Operator Manual 41

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    3) Description of Lead IEC(AAMI)

    Lead Color of

    electrode

    Attaching site

    R(RA) Red (White) Just below the clavicle (scapula) near the right shoulder

    L(LA) Yellow(Black) Just below the clavicle (scapula) near the left shoulder

    F(LL) Green(Red) Lower area of the left side of the abdomen

    N(RL) Black(Green) Lower area of the right side of the abdomen

    C(V) White(Brown) Select the candidate lead for measurement among the

    chest leads

    Note

    ECG and SpO2 will display synchronized numeric when you stop ECG

    measurement while you are measuring ECG SpO2 at the same time

    The default setting of ECG Lead is ECG1: Lead II, ECG2: Lead III

    In case of using ECG 5 Lead cable, you need to set menu 5 Leads at ECG Cable.

    Respiration signals are relatively more sensitive to interference from radiated

    electromagnetic signals. .

    Do not rely entirely on the equipment.

    Respiration is measured with the ECG and HR etc., using the ECG cable

    To setup Apnea alarm, push Menu button RESP->APNEA Time

    Priority is EtCO2 measurement when measuring EtCO2.

    Respiration function is very sensitively effected by electromagnetic wave, so it

    should avoid from judging the status of patient by numeric value and waveform

    of respiration.

    Alarm cases of arrhythmia analysis

    Name of Arrhythmia Description

    Tachycardia: TAC Heart rate is over 140bpm for a minute.

    Bradycardia: BRD Heart rate is under 40bpm for a minute.

    Premature Ventricular

    Contraction: PVC Abnormal heart rate is over 1.

    Ventricular Tachycardia: VTAC Over 8 PVC is over 140bpm

    Asystole: ASY No electric action of heart.

    Ventricular Bigeminy: BGM After normal heart rate once, two times or more PVC

    Ventricular Trigeminy: TGM After normal heart rate at two times, two times or more PVC

    Ventricular RUN After normal heart rate at Over 3 (PVC)

    Venticule of over 100bpm

    Ventricular Fibrillation:

    VFIB

    Very fast ,chaotic heart rate in the lower chambers of

    the hearts, resulting from multiple areas of the ventricle

    is attempting to control the heart’s rhythm.

    Multifocal PVCs Over the last 15 beats two or more premature ventricular

    beats.

    Ventricular Couplet (COP) Abnormal heart rate is more than two times

    Missed Beat (MIB) ECG wave is missed intermediately

    R on T PVC PVC is detected before finishing T wave in the normal ECG Wave.

    Frequent PVCs PVC generates higher than alarm limit.

  • BT-750 Operator Manual 42

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Caution

    Inspection does not carry out the patients with arrhythmia, premature

    ventricular contraction or ventricular tachycardia.

    6.2 Respiration Menu Display

    1) Press the button “ ” to go to adjustment display.

    2) Select the menu to change with trim knob and press the trim knob to set the location to

    be modified. Then, selected menu is displayed.

    3) Turn to right and left with trim knob and press the trim knob to set up the numeric

    value and save it.

    4) Press “Return” button to go back to 1st display window.

    5) Press the button “ ” to return to main display.

    The range of respiration setup

    Setup Mode Range of modification Description

    WAVE SPEED 6.5/12.5/25 mm/s Set Wave Speed

    RESP LEAD RA-LL/RA-LA Set one of two

    RESP GAIN ⅹ1/2, ⅹ1, ⅹ2 Set the size of RESP waveform

    APNEA ALARM ON/OFF Set alarm for apnea

    APNEA TIME [SEC] 20~120 Sec. Set detection of Apnea

    RESP ALARM Refer to CH.8 Alarm Setup

  • BT-750 Operator Manual 43

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 7

    Measurement of SpO2

    7.1 Preparation for SpO2 Measurement

    1) Connect SpO2 probe to the probe port.

    7.1.1 Setting SpO2 Probe

    1) Clean the body part (a finger or a toe) to be inserted to probe for measurement with

    alcohol.

    2) Set the probe on the finger to measure. (Avoid using it with other medical devices to

    affect blood flow and setting it on the body part in medical treatment )

    3) Do not make any motion and be in pause as possible as you can for stable measurement

    and fix the probe cable with your fingers or plasters but be careful not to make interference in

    blood flow.

    SpO2 Connector

  • BT-750 Operator Manual 44

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    4) Make sure that the probe is set on the body part correctly in every 2 or 3 hours. If the state

    of your skin is different, change the part to be attached to other side.

    Note

    When probe is used incorrectly, inaccurate value or waveform can

    come out as below.

    ● When the patient uses the probes without certification of Bistos

    Technology.

    ● Hemoglobin dysfunction

    Caution

    Probe should be treated with attention. Careless use might cause

    damage to probe. Keep the probe cable away from acute objects.

    When the patient has a high fever or peripheral circulatory failure,

    the skin, temperature increases to 2~3 degree.

    In case of the patients with cold fingers, SpO2 data is not accurate.

    For the patients who have abnormally high oxyhemoglobin or

    methemoglobin , SpO2 data is not accurate.

    You can not use disposable probe to different patients. For same

    patient, it is possible to reuse and remeasure even in different

    portion. If you want to reuse disposable probe, disinfection is

    indispensable.

    Caution

    It might fail to measure pulse rate under occasions as follows;

    When probe is fastened tightly.

    When probe excessively exposes to Medical illumination, bilirubin

    illumination or sunlight.

    In case of being measured on arms or legs in cuffs or bandages.

    Warning

    In case of measuring with probe while MRI is in use, the patient may

    have burns. To minimize the danger, use non-inductive line, if there

    exists any danger even with proper use, remove the connector from

    the patient.

    ● Allergic patients should not use disposable probes.

    ● Make sure that the probes radiate light, faced each other correctly,

    and the light reaches to the tissue of a patient. When applying to a

    newborn baby, use disposable probes outside of the incubator.

    ● If used in an incubator, accurate results may not be obtained

    because the humid air can have an effect on the outcome.

    When probe is excessively fixed with tape or used in other incorrect

    way, the patient may suffer from injury.

    Warning

    If the probe are excessively exposed to medical illumination

    ( especially for xenon light ), bilirubin illumination, fluorescent

    lamp, IR heating appliances or direct ray of light , this may cause

    problem in function . To avoid excessive exposure to the light, use it

    by the general operation and be careful that the probe is not exposed

    to outside ray with opacity. If outdoor circumstance is under

    extremely high light, it may generate inaccurate value.

    Do not use probe at the body part where intravenous injection or

    artery catheters are placed.

    Do not use damaged SpO2 probe or optical appliances.

    Do not clean with oxygenated water, ultraviolet ray, direct ray of

    light and steam.

  • BT-750 Operator Manual 45

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 8

    Measurement of NIBP

    8.1 Setting NIBP

    1) Choose the proper cuff for the patient.

    2) Connect the cuff hose to cuff.

    3) Connect the cuff hose to NIBP font.

    4) Put the cuff on the patient.

    5) Use the proper cuff to the patient before measuring blood pressure.

    8.1.1 NIBP measurement

    1) It should be measured at the same label of the patient’s heart in the arms and legs of the

    patient.

    2) The cuff ties part under the position to be measured.

    It may cause errors in measurement values.

    3) NIBP measurement will be done with button

    and if you press the button one more time, it

    stops measurement.

    Note

    It is impossible to measure the blood pressure in upper arm, in case

    that infusion pump or catheter is inserted in the blood vessel. It may

    cause damages in the organ around catheter during the supply of

    blood pressure.

    Caution

    Use the proper cuff to the patient before measuring blood pressure.

    Inspect excess blood pressure or folding of cuff.

    Avoid the cuff from getting wetness out of any kinds of incidents,

    and use it after dry completely.

    NIBP hose connector

  • BT-750 Operator Manual 46

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 9

    Measurement of Temperature

    9.1 Temperature Measurement

    ● Changes in impedance according to the change in the patient’s body temperature is

    perceived by the temperature sensor, and then displayed numerically on the screen after a

    series of calculations. Measured Temp

    1) Connect the temperature sensor to the connection terminal of the measurement module

    and sensor to patient.

    Caution

    The temperature sensor should be sterilized before application to

    another patient.

    Temperature Connector

  • BT-750 Operator Manual 47

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 10

    Measurement of IBP

    10.1 IBP Measurement

    Arterial Blood Pressure is the pressure exerted by the blood against the walls of the arterial

    vessels. The cardiac cycle consists of a period of relaxation called Diastole, during which

    the heart fills with blood, followed by a period of contraction called Systole. While the heart

    is contracting and relaxing, the sensor gets the measurement of the pressure.

    10.2 Setup Connections 1) Connect the interface cable for the transducer to the IBP connector on the monitor’s front

    panel. An interface cable for the transducer has to be selected correctly as it depends on

    the each transducer type.

    Note

    The monitor is designed to accept signals from BT4812-3 disposable

    transducer, or equivalent (pressure range of 0~300 mmHg). Refer to

    the transducer directions for use for details.

    *This transducer is verified under the connection with to see if

    protective means is provided against hazards to patient when used

    with HF surgical equipment.

    IBP Connector

  • BT-750 Operator Manual 48

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    2) Set up the patient circuit according to the directions for use of the transducer, monitoring kit.

    3) “Hold the backside of the clear cover surrounding the connector, and connect the converter

    to the reusable monitor interface.

    4) In a clean environment, open the package to check if all parts are well connected and if the

    handle of the “stopcock” is positioned properly. All side ports of the stopcock are protected

    by the outlet plugs and may not be removed till the system is full and the foam has been

    removed. These outlet plugs should always be replaced with other outlet plugs. (A pouch is

    included in the kit.).

  • BT-750 Operator Manual 49

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    5) How to use Flush

    Hold both sides of the flush operator and pull upward gently. Do not revolve the operator and

    take care not to put the power on one side.

    Caution

    If a fast flush is performed on the patient, the user should check

    carefully for the presence of foam and particulate matter. If a large

    volume is flushed by force, a short flush with an increase rate less

    than 2cc is recommended to avoid central embolism.

    Blood pressure for a liquid infusion line over 300mmHg requires an

    infusion of more than 3cc per hour. In this case, the pressure must

    not exceed 775mmHg. A protective function is set in the flush

    device, which prevents overpressure of the converter by making the

    liquid bypass the device. If a more precise control of fluid volume is

    required, it is necessary to connect the infusion pump around the

    flush device.

    6-1) Twist the sterilized solution bag lightly using a drip chamber spike

    6-2) Open the roller clamp and remove the air inside the bag completely by squeezing

    using an infusion set or a needle inserted into the injection port in the back. By emptying

    the back this way, it can prevent air input into the patient’s body.

    6-3) Before inputting the solution into the infusion set, push both sides and fill the drip

    chamber partially.

    6-4) Operate the flush device gently. Since air rises from the bottom, make sure that the

    solution is in the bottom at all times.

    6-5) Prime the side ports and the plug of the zero point stopcock in the converter using a

    door-lock plug.

    6-6) Check if bubbles are present in the monitoring system. To check the presence of

    invisible bubbles, rap it lightly

    6-7) Add pressure to the system up to 300mmHg using C-fuse or the clear cuff. Flush the

    system for 2~3 seconds. Check for the presence of air bubbles which may result from a fast

    flush. Now, the system is ready for zero point setting and measuring. Place the stopcock in

    a 90 degree position to let it off state.

  • BT-750 Operator Manual 50

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Caution

    The stopcocks of the converter or the outlet should be locked before

    connecting the door-lock plug.

    Note

    >

    ● Priming should be done slowly!

    - Slow priming will lessen the effort to remove air bubbles afterward.

    ● Priming using gravity!

    - Pressing may result in leakage of liquid or bubbles by forcing the

    liquid to flow into the system. If small bubbles flow too slowly

    inside the system, lift up the supply bag. 1 inch corresponds to

    2mmHg, and the primed 4-feet line gives a pressure of about

    100mmHg if it is completely spread out The solution bag should

    be placed at a higher position than the converter and the pressing tube,

    in order for priming using gravity.

    Caution

    If air foam or bubbles are present inside the system, it may cause

    significant bias of the pressure waveform or air embolism.

    Care should be taken not to let air bubbles inside the 3-way outlet

    stopcock or the cannula reflush into the patient. To check, confirm if

    the monitoring line is fully filled with liquid before connecting the

    monitoring line and let a small volume of blood flow through the

    cannula.

    10.3 Adjustment of Zero point of the Converter

    1) Lock the zero point stopcock (toward OFF) and release the converter according to the air

    pressure.

    2) Set “Menu”-> “Zero”-> “Yes” in order. If final value is “0”, the measurement starts.

  • BT-750 Operator Manual 51

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 11

    Measurement of EtCO2

    11.1 EtCO2 Measurement (Optional)

    Sidestream CO2 Gas Measurement

    Sidestream CO2 module can be fitted onto BT-750. Sidestream CO2 provides the function of

    ETCO2 and the function to measure respiratory CO2 and respiration numbers using the small

    filterline based on a small lumen type filterline. In the Sidestream CO2 type, the level is

    measured through the cannula (non-tubing type) attached to the nose or a sampling line

    equipped in the respiratory circuit (tubing type).

    Caution

    The maximum sampling rate of the nose tube is 50mL/min. The

    product should not applied to patients who may experience

    respiratory distress by the vacuum flow level.

    To prevent infection of the medical staff by the patient’s respiration

    sample, the outlet of BT-750 should be connected to the air

    elimination system of the hospital.

    Caution

    The gas inspiring vacuum pressure (negative pressure) of the gas

    elimination system should not exceed the pump outlet standard of

    BT-750, 1mmHg. Excessive pressure will display the message of

    “OCCLUSION” and it may damage the BT-750 product pump.

    During zeroing, the air elimination system should be operating.

    Waste from Microstream and CO2 filter should be treated as fatal

    biological materials for the human body

    Note

    Connection of all tubes should be safe, and the nose cannula should

    be kept away from the CO2-present area during the self test period

    (including the outlet of the ventilator and the respiration of the user).

    Range of Respiration Measuring : 0 ~ 150bpm

    For the best accuracy, a warm-up of about 2 minutes is required.

    Note

    We do not use specific gas for calibration and manually operate “O”

    Calibration.

    When you adjust “O” calibration, connect it to the transducer and

    monitor in the air. The patient may not connect it to the other gas.

    Operate “ Menu → ZERO → YES” in order .

    Use it after verifying “O” value at the window screen, otherwise the

    measurement value would not be correct.

  • BT-750 Operator Manual 52

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    11.2 How to measure Sidestream CO2

    1) Attach the gas/tube to BT-750 Sidestream port and the other side attach to hospital gas

    device

    2) Respiration numeric is automatically changed when EtCO2 is measured.

    3) EtCO2 will be displayed on BT-750 screen after 20 seconds when it detect proper input.

    4) EtCO2 waveform and data is displayed. If waveform is not displayed on the screen, change

    setting menu “Menu → EtCO2 → EtCO2 → Enable

    5) You can change EtCO2 waveform when you need.

    Caution

    The watertrap is disposable and should only be used for a single

    patient. Do not reuse the watertrap for another patient.

    Caution

    The cannula is disposable and should only be used for a single

    patient. Do not reuse the cannula for another patient.

    Caution

    If oxygen is being delivered while using Sidestream CO2, be sure to

    use a CO2 Sampling and O2 Delivery cannula. Using a different type

    of cannula could obstruct oxygen delivery.

    Warning

    Assembly, extension, repair should be carried out by authorized

    person from Bistos Technology.

  • BT-750 Operator Manual 53

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 12

    Maintenance 12.1 Maintenance and Cleaning

    BT-750 and its accessories can be cleaned by using various methods. Please follow the

    methods below to avoid unnecessary damage or contamination to the equipment.

    Keep the equipment far from the place possibly water sprinkling or soaking into the equipment.

    Avoid the place possibly affecting to the correct measurement with air pressure, temperature, humidity, ventilation, sunlight, dust or air including with salt, ion.

    Storage Temperature

    Packaged storage (in box) : -20℃~80℃

    Unpackaged storage : -20℃~60℃

    Storage Humidity : 0~80% under 12℃~60℃ range Keep it safe from tilt, vibration or impact. Avoid the location exposed to chemical or explosive gas. Keep the equipment clean for the next operation. Stop operating and contact the right personnel immediately when malfunction

    occurs.

    Never let the unauthorized person disassemble, repair or alter. At least once a month, clean and wipe off the monitor by using the soft cloth after wetting it in lukewarm water or alcohol. Do not use lacquer, thinner, ethylene or oxides which could be harmful to the equipment. Make sure both cables and accessories are free of dust or contaminants and wipe them off with soft cloth wet by lukewarm water (40℃/104℉) and at least once a week clean them with clinic alcohol. Do not soak the accessories in any liquid or detergent. Also, don't let the liquid penetrate into the equipment or probe.

    12.2 How to clean Accessories

    For cleaning the SpO2 finger probes, be directed below.

    Chemicals How to clean

    Alcohol Isopropyl Alcohol Wash with wet gauze

    Liquid soap Benzalconium chloride 0.05 W/V%(200x diluted)/Wash with wet gauze

    0.01 W/V%(50x diluted)/Wash with wet gauze

    Iodine Povidone Iodine 0.02 W/V%(50x diluted)/Wash with wet gauze

    Glutaral C5H8O2 2 W/V%(experiment solution)/Wash with wet gauze

  • BT-750 Operator Manual 54

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Caution

    When harmful (unaccepted) materials are used for cleaning, we shall not

    offer the service without charge in regardless of warranty period.

    Check carefully both monitor and probe after cleaning the equipment.

    Do not use the equipment that is worn out or damaged.

    Caution

    You can get incorrect measurement figure under the following conditions.

    - The patients’ excessive motion

    - Using electrical operational instrument with high-frequency or the

    instrument for cardioversion.

    - Venous blood pulse

    - Appling to the monitor parts to measure blood pressure or use arterial

    catheter/bandage and intravascular parts

    - When the patients are on low blood pressure, severe blood vessel

    contract, sever anemia or low body temperature.

    - Arterial occlusion nearby probes

    - When the patients are on cardiac arrest or shock.

    Premature neonates and patients with chronic pulmonary disease should be checked for oxygenation levels before starting treatment.

    To prevent electric noise during use, the product should be installed apart from dynamo, X-ray equipment, broadcast equipment or portable

    cables. An inaccurate result may occur when these equipment are placed

    near the product.

    To ensure patient is electrically isolated, connect only to other

    equipment that provides patient electrical isolation. When the patient

    with pacemaker use the monitor, only qualified personnel should

    operate and control all functions of the monitor.

  • BT-750 Operator Manual 55

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    Chapter 13

    Troubleshooting

    This chapter explains the solutions to simple problems. One might feel difficulty to use while

    operating the monitor. At this moment, before regarding it as breakdown, check the following

    points.

    Problem Sign Solution

    Cannot turn

    the Power on

    Check whether AC power

    line is plugged in.

    Faulty power switch

    ☞ Connect to AC power properly. ☞ Contact our customer service center or repair center.

    Cannot produce

    the proper

    waveform

    Unstable connection. ☞ Check the connections of the power, connectors, and the

    accessories.

    ▣ Problem of ECG/Respiration measurement

    Problem Solution

    The message“ LEAD

    FAULT” is displayed.

    1. Check the ECG cable and ensure that the gel in the electrode is

    not dry.

    2. Try using another ECG cable.

    The ECG waveform

    is not displayed.

    1. Check if the cable connection is stable. 2. Check if the ECG cable is damaged.

    Severe noise on the

    waveform.

    1. Check if the electrode attached to the patient is stable. 2. Check if the disposal electrode is one provided by the

    company. (Noise may occur according to the kind of

    electrode.)

    ▣ Problems in SpO2 Measurement

    Problem Cause Solution

    Error display

    by ray interference

    Errors generated by excessive ray

    outside.

    Cover the probes with opaque

    substances.

    Error by

    electromagnetic

    wave.

    Power cable or connectors are placed

    closely with probes.

    Keep probes away from

    power cable and connector.

    Error display

    Use of improper probes or clinical

    problem Misplaced probes or patient

    moving.

    Probes do not attach to the patient.

    Use of proper probes. Place

    probes in right location Fix

    probe and cable with tape and

    use new probes as required.

    Output failure of

    SpO2 value

    Change of arterial blood capacity by

    shock. Apply to other body part

    SpO2 waveform

    displays in small

    or flat shape.

    Fingers in cold temperature or

    improper probe use , setup

    error for waveform size

    Apply to other body part or

    check if it is used in proper

    way and waveform size is set

    up correctly.

  • BT-750 Operator Manual 56

    P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)

    ▣ NIBP measurement

    Problem Solution

    NIBP measurement is not d