bt-750 - yahoo · 2019. 12. 18. · 5.4 ecg1, ecg2 setup ... bt-750 operator manual 7 p/n :...
TRANSCRIPT
-
P/N : OPM(BT-750)EN, Rev.02
BT-750
BT-750
PATIENT MONITOR
-
BT-750 Operator Manual 1
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Proprietary Materials Copyright © 2013 Bistos Co., Ltd. All rights reserved.
Information and descriptions contained in this manual are the property of Bistos
Corporation and may not be copied, reproduced, disseminated, or distributed without
express written permission from Bistos Corporation.
The materials in this document can contain typographic errors and be subject
to change without notice.
Printed in Korea
Prior to beginning
In this document, all information for correct using and operation is included.
If you need a help for instruction or questions, please contact Bistos Co.,Ltd.
Bistos Co.,Ltd. and the authorized agency do not take responsibility for the
harmful cases that are happened by user’s obvious carelessness or incorrect uses.
In this document, the product, system and equipment indicate BT-750. And the
company and we indicate Bistos Co.,Ltd.
We emphasize the safety and matters that require attention while using and
operating this product by the following terms. User must know these terms – Warning, Caution, Note - before using this product.
In the event that the product is damaged due to misuse or negligence by a user, the
manufacturer or the authorized agents shall not be responsible for any damage or loss
to the product.
Warning
The title “Warning” is used to inform the user of possible causes that could
inflict the injury, death or property damage to the patients.
Caution
The title of “Caution” is used to inform the users of possible causes that
could inflict the injury to the patients although it might not be severe enough
to cause deaths.
Note
The title of “Note” is used to inform the users of items that are of importance
in terms of installation, operation or maintenance of the equipment although
the failure does not for the physical harm to the patients.
-
BT-750 Operator Manual 2
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Table Contents
Proprietary Materials………………………………………………...1
Prior to beginning……………………………………………………..1
1. SAFETY ............................................................................................ 4 1.1 Intended Use ....................................................................................................... 4
1.2 Instruction for the safe operation and use of the BT-750 .................................... 4
1.3 General Precautions on Environment ................................................................. 4
1.4 General Precautions on Electric safety ............................................................... 5
1.5 Definition of Symbols ......................................................................................... 7
2. INTRODUCTION ............................................................................ 8 2.1 About this Manual ............................................................................................... 8
2.2 BT-750 Monitor Configuration ........................................................................... 8
3. INSTALLATION ............................................................................ 10 3.1 Description of the BT-750 Front Panel ............................................................. 10
3.2 Description of the BT-750 Rear Panel .............................................................. 11 3.3 Description of the BT-750 Right Panel ............................................................. 12 3.4 Description of the BT-750 Left Panel ............................................................... 12 3.5 Settings ............................................................................................................. 13
4. BT-750 OPERATION ..................................................................... 17 4.1 Operation .......................................................................................................... 17
4.2 Turn off the monitor .......................................................................................... 17 4.3 Display .............................................................................................................. 18 4.4 Basic Operation ................................................................................................. 19
5. SYSTEM & ALARM SETUP ....................................................... 20 5.1 System Setup..................................................................................................... 20
5.2 The modification of system setup(2nd
display window) .................................... 24 5.3 Alarm Setup ...................................................................................................... 25 5.4 ECG1, ECG2 Setup .......................................................................................... 29
5.5 SpO2 Setup ....................................................................................................... 32
5.6 NIBP Setup ....................................................................................................... 33
5.7 IBP Setup .......................................................................................................... 34
5.8 Temperature Setup ............................................................................................ 36
5.9 CO2 Setup ......................................................................................................... 37
5.10 Trend Setup ..................................................................................................... 38
5.11 Printer Setup .................................................................................................... 39
6. MEASUREMENT OF ECG AND RESPIRATION .................... 40 6.1 Setup Connections ............................................................................................ 40
6.2 Respiration Menu Display ................................................................................ 42
-
BT-750 Operator Manual 3
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
7. MEASUREMENT OF SPO2 ......................................................... 43 7.1 Preparation for SpO2 Measurement .................................................................. 43
8. MEASUREMENT OF NIBP ......................................................... 45 8.1 Setting NIBP ..................................................................................................... 45
9. MEASUREMENT OF TEMPERATURE .................................... 46 9.1 Temperature Measurement ................................................................................ 46
10. MEASUREMENT OF IBP .......................................................... 47 10.1 IBP Measurement ............................................................................................ 47 10.2 Setup Connections .......................................................................................... 47 10.3 Adjustment of Zero point of the Converter ..................................................... 50
11. MEASUREMENT OF ETCO2.................................................... 51 11.1 EtCO2 Measurement ....................................................................................... 51 11.1 How to measure Sidestream CO2 ................................................................... 52
12. MAINTENANCE ......................................................................... 53 12.1 Maintenance and Cleaning .............................................................................. 53
12.2 How to clean Accessories ............................................................................... 53
13. TROUBLESHOOTING ............................................................... 55
14. SPECIFICATION......................................................................... 57 14.1 Mechnical, Electric and Performance Specifications : Main Body ................. 57
14.2 Mechnical, Electric and Performance Specifications : Accessories ................ 60 14.3 Compliances.................................................................................................... 61
14.4 Environmental Conditions .............................................................................. 61 14.5 List of Standards ............................................................................................. 62
11. WARRANTY ................................................................................. 64
-
BT-750 Operator Manual 4
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 1
Safety
1.1 Intended Use
BT-750 is a patient monitor which displays state of patient’s data with wave form. Basic functions are measuring ECG 3lead select and respiration. Also it is possible to
measuring arrhythmia, analysis ST level and detects pacemaker function. Other basic
functions are measuring SpO2, pulse rate, NIBP, 2 IBP and 2 body temperatures.
Optional functions are ECG 5 lead select and EtCO2 measurement.
Note : This device is for use only by trained medical personnel located in a treatment
room which is located in hospital, an operating room, a recovery room and an
intensive care unit.
Note : Because the incorrect use of this device can do patient’s life harm, this device
must be installed, used and operated by trained medical personnel who can
measure and analyze physiological signals.
1.2 Instructions for the safe operation and use of the BT-750
Examine the monitor and any accessories periodically to ensure that the cables, line cords,
transducers, and instruments do not have visible evidence of damage that may affect patient
safety or monitoring performance. The recommended inspection interval is once per week or
less. Do not use the monitor if there is any visible sign of damage. Do not place the monitor
to incline place to avoid falling of product.
1.3 General Precautions on Environment Do not keep or operate the equipment under the environment listed below.
Avoid placing in an area exposed to moisture. Do not touch the equipment with wet hand.
Avoid exposure to direct sunlight
Avoid placing in an area where there is a high variation of temperature. Operating temperature
ranges from 10C to 40C.
Avoid in the vicinity of Electric heater
-
BT-750 Operator Manual 5
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
1.4 General Precautions on Electric safety
Prior to operation of the equipment, make sure to check followings
Caution
Whether power supply is appropriate. (100 – 240VAC)
Caution
Whether connections (power line or selected equipment) are properly
made to the equipment.
Caution
Whether the correct measuring accessory for required parameter is
connected to the equipment prior to turning on the equipment.
Caution
BT-750 can be used alongside with defibrillator while the accuracy
of the product may be fallen in a short time
Caution
Use of this product is prohibited during MRI photographing. A fire
may breakout from the induced current, and the accuracy of the
product and the MRI may be influenced by cross interference.
Caution
BT-750 is intended only to measure and assess patient state. It must
be used in conjunction with Clinical signs and symptoms when
diagnosis is made.
Caution
The computerized ECG analysis should be judged by qualified
medical professionals, and should not be the sole basis in
determining necessity of treatment for the patient.
Caution
Leakage, heating, firing or disruption of the battery may cause fire or
injury.
Caution
Do not use batteries beyond those specified in this manual.
Caution
Do not short-circuit the battery and does not heat, integrate, or
dispose of it in a fire.
Avoid placing in an area where there is an excessive humidity rise or ventilation problem. Operating humidity ranges
to 80% RH max @40C.
Avoid placing in an area where there is an excessive shock or vibration.
Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage.
Avoid dust and especially metal material into the equipment.
Do not disjoint or disassemble the equipment. BISTOS Co., Ltd. does not take responsibility of it. Disassembling of equipment should be done only by the authorized personnel.
Power off when the equipment is not fully installed. Otherwise, the equipment could be damaged.
-
BT-750 Operator Manual 6
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Caution
Classification of BT-750 :
- BT-750’s class is “Class IIb” equipment.
- It is type “CF grade” for ECG, Resp, IBP and “BF grade: for
EtCO2, SpO2, NIBP and Temp.
- The noise level is“ B” Class according to the IEC/EN 60601-1
(Safety of Electric Medical Product), and the noise redemption is
“B” Level according to the IEC/EN 60601-1-2 (Electromagnetic
Compatibility Requirements)
- Protection grade against water is“ IPX 0”
- Do not use this product around a flammable anesthetic and solvent.
Caution
Although BT-750 has been value of measurement in compliance
with existing EMI/EMC requirements, use of this system in the
presence of an electromagnetic field can cause momentary
degradation of the SpO2 waveform. If this occurs often, Bistos
suggests a review of the environment in which the system is being used, to
identify possible sources of electromagnetic waves. These waves could be
from other electrical devices used within the same room or an adjacent
room. Communication devices such as cellular phones and pagers can
cause these waves. The existence of radio, TV or microwave transmission
equipment located nearby can cause electromagnetic waves. If EMI causes
disturbances, it may be necessary to relocate your system.
Electrostatic discharge (ESD), commonly referred to as a static
shock, is a naturally occurring phenomenon. ESD is the most
prevalent during conditions of low humidity, which can be caused by
heating or air conditioning. During low humidity conditions,
electrical charges naturally build up on individuals and can create
static shocks. An ESD condition occurs when an individual with an
electrical energy build-up comes in contact with objects such as
metal door knobs, file cabinets, computer equipment, and even other
individuals. The static shock or ESD is a discharge of the electrical
energy build-up from a charged individual to a lesser or non-charged
individual or object. The level of electrical energy discharged from a
system user or patient to the BT-750 can be significant enough to
cause damage to the system or probes. The following precautions can
help to reduce ESD: anti-static spray on carpets; anti-static spray on
linoleum; anti-static mats.
-
BT-750 Operator Manual 7
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
1.5 Description of product Symbols
Symbol Description Symbol Description
Type CF Product,
Defibrillation
protected
Power Switch
Type BF Product,
Defibrillation
protected
Alarm On/Off
Noninvasive Blood
Pressure
Manual
Measurement(NIBP)
TREND
Warning Sign
Period of Auto Blood
Pressure Measurement
Liquid Protection
Grade Liquid Protection
Grade
T1~2 Measuring Temp. Start Printing (Printer)
CO2 In/Output
EtCO2
Gas Input/Output Menu ON/OFF
SPO2 SPO2 Measurement Ground
P 1~2 Invasive Blood
Pressure RS232 port and nurse
call
ECG Electrocardiograph /
Respiration Screen Transfer
NIBP Non Invasive Blood
Pressure
Use for Central
Monitoring
Table 1-1. Definitions of symbols
1.6 Description of equipment Symbols
The symbols in the bottom of LCD indicates the present status of BT-750 and use
bottoms for each function.
Indicate that Audio alarm generates
Display for current charging state of charger. So, make sure of the
charging state when you use only with charger while the AC power is on.
:Indicate that AC power is on
-
BT-750 Operator Manual 8
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 2
Introduction
2.1 Feature BT-750
BT-750 is a patient monitor which display state of patient’s data with waveform.
※ Basic functions of Patient measuring
- ECG 3 lead (I, II, III,) select and Respiration measurement
- Arrhythmia measurement, ST Level analysis and Pacemaker Detect function
- SpO2 and Pulse Rate measurement
- NIBP, 2 IBP measurements
- Body Temperature (2 Temp)
※ Optional functions of Patient measuring
- ECG 5 lead (I, II, III, aVR, aVL, aVF, V) select
- End-tidal partial pressure respiration (EtCO2)
※ Etc
- Trend Data Saving for 72 hours
- VGA (Display Output) Function
- Clone Function (This function support patient checking thru Internet)
- Built-in 3 channel Printer. (Option)
2.2 List of Components
Standard Accessories
Items Qty
BT-750 body 1 ea
SpO2 Finger Probe(Adult) 1M 1 ea
SpO2 Extension CABLE 2M 1 ea
NIBP cuff (Adult) 1 ea
NIBP HOSE 1 ea
ECG Cable 3 lead type 1 ea
ECG electrode (3M) 3 ea
220V Power cord 1ea
Operation Manual 1ea
Table 2-2. Standard Accessories
-
BT-750 Operator Manual 9
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Optional Accessories
Items
SpO2 Finger Probe(Child, Neonate) DISPOSABLE SPO2 SENSOR
Disposable SpO2 Finger
Probe(Neonate) Press Measuring Kit
NIBP Cuff (BIG Adult, Child,
Pediatric, Infant) ECG Cable 5 lead Type
Printer Module EtCO2 Kit set
Roll Paper Temperature Sensor
IBP KIT IBP Extension cable 2M
Temperature sensor ( Rectal type)
Table 2-3. Optional Accessories
Caution
The different results can come out according to the status of patients.
If the results are doubtful, please check the patient’s status using other measuring equipment which is using in hospital.
Caution
Use only authorized accessories supplied with BT-750. Use of any
other accessories may result in out of specification performance and
possible safety hazards.
-
BT-750 Operator Manual 10
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 3
Installation
3.1 Description of the BT-750 Front Panel
▶ Function and Explanation
3.2 Description of the BT-750 Rear Panel
No Switch / LED Function Explanation
1 AC Indicator When AC power is connected and main power switch is
on, green light is on
2 NIBP Switch It is used to measure blood pressure manually.
3 Power Switch It supplies power to the monitor
4 Printer Switch Start/Stop Printer
5 Battery
Charge Indicator
On Battery charged, orange LED is on and while AC
power is disconnected, LED is off.
6 Alarm Switch Used for alarm or alarm silence
7 Trend Switch The equipment stores SpO2 and pulse rate for 72 hours in
every 1 minute and used to check it
8 Alarm Indicator When the alarm is on it will show on led
9 Menu Switch It shows menu on the monitor
10 Speaker Sound of Alarm and Heart rate.
11 Selection trim knob It shows entire menu selection.
-
BT-750 Operator Manual 11
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Figure 3-2. Rear panel
No. Item Function Explanation
1 AC Power Input
Port
Port for AC adopter to connect exterior power to the
monitor.
2 VGA Port It is used for when VGA port is connecting to other
Monitor
3 RS232 Port The communication port for connecting monitor and
computer.
4 LAN Port The communication port for connecting monitor and
external devices.
5 Label Display
6 Ground Port Port of Ground Cable
7 Battery Store Port of Battery
8 Portable Handle Handle of carry
3.3 Description of the BT-750 Right Panel
-
BT-750 Operator Manual 12
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Figure 3-3. Right panel
No. Item Function Explanation
1 Printer Power LED When Printer Power is connected ,green light is on
2 Printer Error LED When Printer is in malfunction ,red light is on
3 FEED Switch Printer Paper selection
4 Printer Cover Inserting the printer paper and close the cover
3.4 Description of the BT-750 Left Panel
Figure 3-4. Left panel
3.5 Settings
No. Item Function Explanation
1 ECG/RESP Connection Terminal It is connector of Measuring ECG/RESP
2 SpO2 Connection Terminal Connector of SpO2 Sensor
3 IBP 1 Connector of IBP 1
4 IBP 2 Connector of IBP 2
5 TEMP 1 Connection Terminal Connector of Temp 1 Sensor
6 TEMP 2 Connection Terminal Connector of Temp 2 Sensor
7 Cuff Connection Terminal Connector of NIBP Cuff Sensor
8 CO2 Input, Output Connection
Terminal Connector of CO2 Input and Output
-
BT-750 Operator Manual 13
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Precautions for Settings
Pay attention when setting BT-750 for the following points
● Use BT-750 under conditions of 10℃∼40℃ of circumstantial
temperature and within 80% humidity.
● Check the connection status of the power cord.
● Do not connect multiple cords to the power supply.
Place the main body on a flat area.
If noise occurs, ground the equipment further.
All information is stored in the internal memory, even in the case of power on and off.
Do no use any electric cords generating connecting noise.
Take care since the product can be broken by mechanical shock.
Get rid of dust or inflammable material near the product when setting it up.
3.5.1. How to Use when power is on.
1) Insert AC Adaptor to AC-IN Connector at the rear of BT-750
2) Turn on the On/Off switch at the rear of BT-750
( If the power is in Stand-by , battery starts to charge. )
3) If AC lamp of the front side (Green light) is on and you press the power switch
[AC Adaptor Ports]
[the front side of the product]
Note
Though AC power cord is connected, charging lamp continuously
lights on and after completing charging, the lamp turns off.
※ AC Power
If AC power is connected to the monitor and turn on the AC power switch, Green LED is on to
indicate Power-on. It is automatically converted to a recharging mode. (Input 100~240V, 50/60Hz)
Caution
Noise may occur in waveform, while using AC power. In this case,
ground the product.
It may occur the fatal damage to the product ,while connecting AC
power with wet hand.
3.5.2 How to Use Battery
If you continuously press the sub power switch for
about 1 second, the power is on.
Lamp is on during battery charging. (Orange light)
If you turn on the main power switch, the lamp
lights up while AC adaptor is connected.
-
BT-750 Operator Manual 14
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
In case of sudden blackout or potable purpose, automatically AC Adaptor power transfers to
battery power for continuous use.
How to use
1. Battery charge lamp is off, while battery used. 2. It is durable for approximately 4 hours but depends on battery charged state. ( Full
charged time: 4 hours)
3. Battery charging starts once main power switch of rear side is on after AC Adaptor connection.
4. Battery charged state displays in 5 steps.
Full charged 70% charged 50% charged 20% charged Discharged
5. When battery is low charged, battery state display on the screen is turned off, beep sound generates and after about 5 minutes, power is disconnected automatically in
order to protect battery.
Note
When battery electronic pressure is in low state, it may cause
inaccurately measured value. Once battery state is low, connect AC
Adaptor to the appliance at once.
Note
For battery replacement, make sure of polarity.
If the battery polarity is placed incorrectly, it may cause fatal
problem on equipment.
Contact to Bistos head quarter or authorized agent, when you need
battery replacement.
If a side of power cord is put to power port of BT-750, power is supplied to the equipment. Make sure
that each cord is connected to input / output terminal correctly. If there is incorrect connection,
it may cause problem.
※ Battery Charge
If AC power is disconnected, power is supplied thru battery. When battery power is not
sufficient, you should charge it by connecting AC adaptor.
Recharging time : More than 5 hours.
Duration time: Under 3 hours.
In the case of discharge battery.
Caution
Noise may occur in waveform, while using AC power. In this case,
ground the product.
It may occur the fatal damage to the product ,while connecting AC
power with wet hand.
Caution
While the monitor is working, battery charging lamp is on. When
sub-power switch is turned off and battery charging completes, lamp
is out.
※ Battery type
BT-750 uses Lithium-lon battery.
-
BT-750 Operator Manual 15
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
The table below shows the battery life while not in use.
Temperature Shelf Life
21 °C to 30 °C 12 months
31 °C to 40 °C 6 months
※ Battery Replacement
Replace the battery with a same type of battery. [The product supplied by Bistos Co.,Ltd.]
Type: Rechargeable Lithium-Ion Battery
Company : LG Chem
※ Battery Replacement Period
If you start to use the system in battery-operated mode, it is necessary to keep it in use
without stop, meanwhile, in care of discharge, the battery life will be shortened. You may
refer the table below for battery replacement period.
Rate of charge/discharge Times
100 % charge/discharge 100 times
50 % charge/discharge 200 times
30 % charge/discharge 500 times
You should replace the battery when the duration time gets shorten or charging is not done
properly. When the monitor is connected to the power, battery is charged automatically.
You can not use the battery, with separated from the monitor.
※ Battery Storage Maintenance
Storage temperature (Humidity : 20 ~ 85 %RH without condensation)
-20 ~ 20 ℃ : less than 1 year
-20 ~ 45 ℃: less than 3 months
-20 ~ 60 ℃ : less than 1 month
※ How to replace Battery
You must contact the headquarter or agent of Bistos Co.,Ltd.
Caution
Dispose of the battery into a proper place for the protection of
environment. Keep the battery dry.
Caution
Do not disassemble or reorganize the battery.
3.5.3 Connection of Accessories
-
BT-750 Operator Manual 16
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Connect the accessories for measurement to each mating port of the equipment.
3.5.4 Replacing Roll paper
① Verify if the printer is receiving the data, otherwise it will lose the receiving data.
② Open the printer door by pushing the button at the both side of the printer slightly as shown the figure below.
③ insert a new paper roll oriented properly like above figure, after verifying the direction of the paper. Remove the empty paper core and insert a new paper roll oriented properly.
④ Pull the paper out towards you until approximately 2 inches of paper and align the paper. Close the printer door.
Caution
● When the printer door is not perfectly closed, the printing could
not be operated or printed clearly.
● When printing more than 1 minute, the overheating of the printer
could be faulty. In this case, stop the printing at least 3 minutes.
● Please use the right paper and do not use the core adhesive roll. If
you use it, it may occur the malfunction as the printer fails to find the
end of the paper.
-
BT-750 Operator Manual 17
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 4
BT-750 Operation
Prior to the Use of BT-750
Make sure of followings prior to check the state of patient.
Check if there is any factor to occur mechanical problem Review the cosmetic state of probe Check if there is any error in fiber-optic probe and AC power cable. Connect the probe to the probe port. Turn on the monitor and identify that the state of fiber-optic probe is normal and the cable is connected.
Identify that fiber-optic probe has any problem on connection.
4.1 Operation
1) Turn on the power.
2) Adjust the setting with exterior button when change of the value required.
3) Figure and waveform are created on LCD screen once ECG probe connected to the
patient.
4) If you apply the SpO2 probe to the patient, data of the patient is displayed on LCD.
5) Apply NIBP cuff to patient. Avoid the arm with SpO2 probe.
6) Apply IBP Kit to patient. Avoid the arm with NIBP cuff.
7) Connect the temperature sensor to the proper position of patient at the type of the sensor.
8) Apply CO2 kit. (optional function)
※ Operation Notice - Only licensed person must use BT-750. - Read carefully followings prior to use the equipment. - Check the state of battery charging. - Make sure that power and probe are connected correctly to the monitor. - Avoid sudden impact when the machine is transported or each unit is connected. - Read carefully followings prior to use the equipment. - Identify that any error does not occur in entire equipment and patient. - Care for the patient safely and preferentially if any error is found on equipment and take
possible action such as stopping the device operation.
- Do not give any impact to the equipment. - Keep the probe not to be taken off from the port for exact measurement.
4.2 Turn off the monitor Turn off the sub-power and verify if the LCD screen turns off.
-
BT-750 Operator Manual 18
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
4.3 Displays
1. Basic screen composition
Press the button in front of BT-750 for power-on.
Display window shows company logo for 2 seconds and is changed to default setting.
▶ Menu 1 Sweep Speed 13 IBP2 Diastolic 25 Battery Status
2 Status of Filter 14 NIBP Systolic 26 Date and Year
Display
3 Detection of
Arrhythmia 15 NIBP Diastolic 27 Time Display
4 Heart Rate
Measurement 16 NIBP Mean 28
Respiration Alarm
Limit
5 Heart Rate Alarm
Limit 17 EtCO2 Measurement 29
Respiration
Measurement
6 PVC Detect
Measurement 18 InCO2 Measurement 30 ECG 1 Wave
7 ST Segment Data 19 Temp. Measurement 31 Wave 2
8 SpO2 Measurement 20 Temperature Alarm
Limit 32 Wave 3
9 SpO2 Bar Graph 21 Bed Number 33 Wave 4
10 SpO2 Alarm Limit 22 Patient Mode 34 Wave 5
11 IBP1 Measurement 23 Symbol of Alarm 35 LEAD Display
12 IBP2 Systolic 24 Symbol of AC
Power
※ When you want to exit from main display or return to basic display by pressing menu button.
-
BT-750 Operator Manual 19
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
4.4 Basic Operation
4.4.1 Simple operation
Order Operation Method
1 Push the power switch for 1~2 second
2 If you want to change the setting value etc. during use, push the Menu button
to change the setting values per each measurement parameter.
3 When the measurement sensor is connected to the patient, the information on
the patient is displayed as values and waveforms on the LCD screen.
4 Refer to the table below for the details of how to measure each parameter..
-
BT-750 Operator Manual 20
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 5
System & Alarm Setup You can set up alarm range, volume, date, time of BT-750 in this menu.
5.1 System Setup
1) Press the button “ ” to select setup menu.
2) Use the trim knob to go to required location and press the trim knob
to set the location to be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
Value and save it.
4) Press the button “ ” to return to main display.
Setup Mode Range of Change Description
CHANNEL
SETUP
3~6 waveforms
Set up the location at each
waveform.
1~8
Input the number of channels for
waveform Set up the location at each
waveform.
Set up the color at each waveform
BED NUMBER 01 ~ 99 Input BED numbers
PATIENT ADULT/Pediatric/Neonate Set the age of patient
DATE/ TIME YEAR : 2000 ~ 2099
MONTH : 1 ~ 12
DATE : 1 ~ 31
HOUR: 0 ~ 23
MINUTE: 0 ~ 59
Date/Time Input
PRINTER Refer to the setup of printer at 16th
chapter
KEY/BEEP
VOLUME OFF/ 1 ~ 7(Change at 1 unit ) KEY, Beep Sound setup
ALARM
VOLUME OFF/ 1 ~ 7 (Change at 1 unit ) Set up the volume of ALARM Sound
-
BT-750 Operator Manual 21
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.1.1 Factory Setting
Parameter Item Name Adult Pediatric Neonate
ECG
LEAD 1st CH LEAD II LEAD II LEAD II
2st CH LEAD III LEAD III LEAD III
GAIN 1st CH
ⅹ1 ⅹ1 ⅹ1 2st CH
SPEED 1st CH
25 mm/sec 25 mm/sec 25 mm/sec 2st CH
CONFIG
MODE MONITOR MONITOR MONITOR
LEAD 3 LEAD 3 LEAD 3 LEAD
ARRHYTHMIA OFF OFF OFF
PACEMAKER OFF OFF OFF
HR ALARM ON ON ON
ST P-R 96mS 96mS 96mS
J 92mS 92mS 92mS
ST ALARM HIGH Limit +3.6mm +3.6mm +3.6mm
LOW Limit −3.6mm −3.6mm −3.6mm
ALARM HIGH Limit 120 bpm 160 bpm 200 bpm
LOW Limit 50 bpm 75 bpm 100 bpm
SpO2
WAVE SPEED 25 mm/sec 25 mm/sec 25 mm/sec
PR ON ON ON
ALARM
ALARM ON ON ON
HIGH Limit OFF OFF OFF
LOW Limit 90 % 90 % 90 %
TEMP
UNIT ℃ / ℉ ℃ ℃ ℃
ALARM
1st CH ON ON ON
2st CH ON ON ON
1st HIGH 39.0℃/102.2℉ 39.0℃/102.2℉ 39.0℃/102.2
℉
2st LOW 36.0℃/96.8℉ 36.0℃/96.8℉ 36.0℃/96.8℉
1st HIGH 39.0℃/102.2℉ 39.0℃/102.2℉ 39.0℃/102.2
℉
2st LOW 36.0℃/96.8℉ 36.0℃/96.8℉ 36.0℃/96.8℉
RESP
SPEED 12.5 mm/sec 12.5 mm/sec 12.5 mm/sec
LEAD RA-LL RA-LL RA-LL
GAIN ⅹ2 ⅹ2 ⅹ2
APNEA ALARM OFF OFF OFF
APNES TIME[SEC] 30 sec 30 sec 30 sec
-
BT-750 Operator Manual 22
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
ALARM
ON / OFF OFF OFF OFF
HIGH 30rpm 30rpm 100 rpm
LOW 8rpm 8rpm 30 rpm
NIBP
MODE MANUAL MANUAL MANUAL
INTERVAL 60 MIN 60 MIN 60 MIN
UNIT mmHg mmHg mmHg
ALARM
ON / OFF ON ON ON
HIGH 160/90(110)mmHg
21.3/12.0(14.7)kPa
120/70(90)mmHg
16.0/9.3(12.0) kPa
90/40(70)mmHg
12.0/8.0(9.3)
kPa
LOW 90/50(60)mmHg
12.0/6.7(8.0) kPa
70/40(50)mmHg
9.3/5.3(6.7) kPa
60/20(30)mmHg
5.3/2.7(4.0) kPa
EtCO2
WAVE SPEED 12.5 mm/sec 12.5 mm/sec 12.5 mm/sec
UNIT mmHg mmHg mmHg
SCALE ~25mmHg ~25mmHg ~25mmHg
AVERAGE 10 SEC 10 SEC 10 SEC
GAS COM. NON NON NON
DISABLE/ENABLE DISABLE DISABLE DISABLE
ALARM
ON / OFF OFF OFF OFF
EtCO2
HIGH
80mmHg
10.7 kPa
10.5%
80mmHg
10.7 kPa
10.5%
80mmHg
10.7 kPa
10.5%
LOW
00mmHg
0 kPa
0%
00mmHg
0 kPa
0%
00mmHg
0 kPa
0%
INCO2
HIGH
20mmHg
2.7 kPa
2.6%
20mmHg
2.7 kPa
2.6%
20mmHg
2.7 kPa
2.6%
LOW
00mmHg
0 kPa
0%
00mmHg
0 kPa
0%
00mmHg
0 kPa
0%
IBP
SPEED 1st CH 25 mm/sec 25 mm/sec 25 mm/sec
2st CH 25 mm/sec 25 mm/sec 25 mm/sec
LABEL 1st CH P1 P1 P1
2st CH P2 P2 P2
SCALE 1st CH ~200mmHg ~200mmHg ~200mmHg
-
BT-750 Operator Manual 23
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
2st CH -50mmHg -50mmHg 50mmHg
ALARM
1st CH 0N 0N 0N
2st CH 0N 0N 0N
1st CH
HIGH 160/90(110)mmHg
21.3/12.0(14.7)kPa
120/70(90)mmHg
16.0/9.3(12.0) kPa
90/40(70)mmHg
12.0/8.0(9.3) kPa
LOW 90/50(60)mmHg
12.0/6.7(8.0) kPa
70/40(50)mmHg
9.3/5.3(6.7) kPa
60/20(30)mmHg
5.3/2.7(4.0) kPa
2st CH
HIGH 160/90(110)mmHg
21.3/12.0(14.7)kPa
120/70(90)mmHg
16.0/9.3(12.0) kPa
90/40(70)mmHg
12.0/8.0(9.3) kPa
LOW 90/50(60)mmHg
12.0/6.7(8.0) kPa
70/40(50)mmHg
9.3/5.3(6.7) kPa
60/20(30)mm
Hg
5.3/2.7(4.0)
kPa
IBP DISABLE/ENABLE ENABLE ENABLE ENABLE
The other setup values
SETU
P
CHANNEL
SETUP
DISPLAY 5-CH WAVE
Wave ECG1,PLETH,BP1,BP2,PR
BED NUMBER 01
PATIENT ADULT
KEY/BEEP VOLUME 02
ALARM VOLUME 02
-
BT-750 Operator Manual 24
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.2 The modification of system setup (2nd
display window)
On operating the service mode such as CHANNEL SET, Date /Time, Printer setup, additional
active window is opened.
5.2.1 Selection of CHANNEL
1) Go to the required location with trim knob and select the location, then the display
window becomes active.
2) Set up the values, turning the trim knob to the right / left and save the values after
pressing the trim knob.
3) Press “ok” button to revert to the main display window
4) On selection of ‘exit”, it turns to 1st display window.
5) Press “ ” button to revert to the main display window.
Setup
Mode
Range of Change Description
DISPLAY 3~6 waveforms
Input the number of
channels for waveform
display
Selection of
waveform
Selection of
location
1CH : ECG 1 (fixed)
2CH : ECG2, PELTH, BP1, BP2 (4)
3CH : PELTH, BP1, BP2, RR, CAPNO (5)
4CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)
5CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)
6CH : PELTH, BP1, BP2, RR, CAPNO, --- (6)
Set up the location at
each waveform
Color at
each
waveform
8 Set up the color at each
waveform.
5.2.2 Setup date /time
1) Go to the required location with trim knob and select the location, then the display
window becomes active.
2) Set up the values, turning the trim knob to the right / left and save the values after
pressing the trim knob.
3) Press “RETURN” button to go to 1st display window.
-
BT-750 Operator Manual 25
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.3 Alarm Setup
5.3.1 Modification Display Window for 1st
Setup (SpO2,Respiration and Temp.)
1) Press button “ ” to go to adjustment display of alarm
2) Select the menu to change with trim knob and press the trim knob to set the location to be
modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
-
BT-750 Operator Manual 26
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.3.2 Modification Display Window for the 2nd
Setup (IBP, NIBP, ECG HR, EtCO2, ST)
1) Press the button “ ” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Activate the display window by pressing trim knob
4) Use the trim knob to go to required location and press the trim knob
to set the location to be modified. Then, selected menu is displayed.
5) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it..
6) Press “Return” button to go back to 1st display window.
7) Press the button “ ” to return to main display.
-
BT-750 Operator Manual 27
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.3.3 Alarm properties
BT-750 offers Alarm function to user. User should have full knowledge about this chapter
and take right action on alarming. BT-750 provides alarm service in two ways. One is
audio Alarm and the other is alarm message.
1) AUDIO ALARM
The high level alarm beeps 5 times and it beeps 5 times. The high level alarm has a
specific rhythm that meets international standard. The medium level alarm beeps 3 times
at 15 second intervals. Low-level alarm beeps to inform the occurrence of following
state. Low level alarm beeps once at 30 seconds
Alarm type Tone Pitch Beep Rate
IEC60601-1‐8 IEC60601-1‐8
High Level Alarm ~976 Hz 10 beeps in 15 sec
Medium Level Alarm ~697 Hz 3 beeps in 15 sec
Low Level Alarm ~488 Hz 1 beeps in 30 sec
High Level Alarm
It operates only when the value is extremely in abnormal state. For example, if pulse rate indicates
0, high level alarm generates regardless of alarm limit.
Medium Level Alarm
It operates only when the value is beyond alarm limit.
Low level Alarm
It beeps when abnormal result brings up or there is anything that the user needs to be
informed.
Note
When you use battery and alarm generates , the alarm volume is
small.
Visual alarm
To help you detect the cause of an alarm, the numeric values of the relevant items keep
blinking.
-
BT-750 Operator Manual 28
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
2) Alarm icon
The alarm icon appears on the information window of the lower screen..
Function Message Description
SpO2 STAND BY Probe is connected but the finger is out of probe.
CHECK PROBE Finger probe is out of probes
ECG
CHECK SENSOR Sensor is out of body or cable is out of the monitor IBP
Temperature
EtCO2
NIBP
CUFF HOSE Failure of pressurization for definite time
NO PULSE Fail to detect pulse rate
REDA FAIL Error in measurement values for moving
ALARM
LOW
Refer to Alarm Generating Chart MIDIULM
HIGH
[Alarm Generating Chart]
Message Generating Cause
HIGH When pulse rate is not detected (“0” BPM)
MEDIUN Out of numeric value
LOW
1. SP02 PROBE ERROR 2. LOW BATTERY
3. Fail to measure NIBP or Over Pressure
4. Lead Fault during ECG measurement
5. Finger is out of SpO2 probe during SpO2 measurement
6. Fault during Temp measurement 7. Fault during IBP measurement
8. PACEMAKER Detect
Alarming display is illuminated in every alarming. It is changed to the below chart,
in every moment pressing alarming button ,when alarm generates.
NO DISPLAY SOUND REMARK
1
Normal alarming Default
2
Holding alarm for 60 sec. Pressing alarm button once
3
Alarm sound off for 120 sec. Pressing alarm button twice
4
Alarm off Refer to the chapter 14
Please press the alarm button ‘once’ in low level alarm to cancel it.
Caution
Messages generate and indicators are illuminated, regardless of
alarming or not.
3) Setting Up Alarm Hold
After alarm generates, user can hold audio alarm and set up the time to 1min/2min/normal.
If each event which generates alarming within setting time is not cancelled, audio alarm will
generate again.
-
BT-750 Operator Manual 29
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.4 ECG1, ECG2 Setup
5.4.1 ECG Menu Display
1) “Press the button “ ” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
Setup Mode Range of modification Description
ECG 3 LEAD LEAD I, II, III
Set ECG Lead ECG 5 LEAD LEAD I, II, III, aVR, aVL,
aVF, V(Chest)
ECG GAIN 1/2,1,2 times Set the size of ECG wave
WAVE SPEED 6.5, 12.5, 25, 50 mm/sec Set the velocity of ECG wave
ECG MODE Monitor/Operation/Diagnosis
Set Analog and Digital filter in ECG
wave
LEAD MODE 3 LEAD, 5 LEAD Set ECG Cable
ARRHYTHMIA ON/OFF Set the usage of Arrhythmia analysis
PACEMAKER ON/OFF Set the detection of the pacemaker
PVC ALARM ON/1~30 Select whether the alarm generates
or no on detection of PVC .
HR ALARM Refer to Ch.8 Alarm setup
ST
SEGMENT
P-R /J + 4~252 ms ,Alarm
ON/OFF Set P-R, J Point
-
BT-750 Operator Manual 30
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.4.2 ECG2 Menu Display
1) “Press the button “ ” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
Setup Mode Range of modification Description
ECG 3 LEAD LEAD I, II, III
Set ECG Lead ECG 5 LEAD
LEAD I, II, III, aVR, aVL, aVF,
V(Chest)
ECG GAIN 1/2,1,2 times Set the size of ECG wave
WAVE
SPEED 6.5, 12.5, 25, 50 mm/sec Set the velocity of ECG wave
-
BT-750 Operator Manual 31
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.4.2.1 ST Analysis (When ST Segment is on)
ST Level is measured by horizontal comparison between PR-segment and ST-segment.
User can adjust J point and P-R point at 1~250 4m/sec interval. These measured values
are available only in normal heart rate. In case of abnormal and pacemaker, for example,
Premature Ventricular Contraction, Ventricular Tachycardia etc., it is except ST analysis
because it is possible to cause distortion of ST form.
1) “Press the button “ ” to go to adjustment display
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric value
and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
● Arrhythmia Analysis can judge problem of Ventricle.
Caution Our ST measurement is only on the basis of ECG1.
5.4.2.2 PACEMAKER Detection (When PACEMAKER is on)
● In case Pacemaker is Detection ON, red bar indicate it on the ECG wave form periodically
from input of pacemaker detection
-
BT-750 Operator Manual 32
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.5 SpO2 Setup
[SpO2 Menu Display]
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
Range of SpO2 setup
Setup Mode Range of change Description
Wave Speed 6.5,12.5,25,50 mm/sec SpO2 Waveform Speed
SpO2 PR Pulse rate ON/OFF Display Pulse Rate ON/OFF
SpO2 Alarm Refer to Ch.8 Alarm setup
-
BT-750 Operator Manual 33
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.6 NIBP Setup
[NIBP Menu Display]
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it..
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display
Range of NIBP Setup
Setup Mode Range of change Description
NIBP Mode AUTO , Manual, Stat Setup Automatic/Manual/Stat
measurement
Interval 1, 3, 5, 10, 30, 60, 90, 120, 240
min
Set the interval of the automatic
NIBP measurement time
NIBP Unit mmHg, kPa Set the unit of blood pressure
NIBP Alarm Refer to Ch. 8 Alarm Setup
*STST Mode is Setup mode to measure the blood pressure for 5 minutes continuously.
Note
When patient needs to use Stat Mode, it must press button at
the bottom of the monitor after adjustment of Stat Mode.
-
BT-750 Operator Manual 34
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.7 IBP Menu Display
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location
to be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
● Table for labels according to the site of measurement for BP1
Label Site Scale
ABP Arterial blood pressure 0~200 mmHg
P1 Standard labels 0~200 mmHg
● Table for labels according to the site of measurement for BP21
Label Site Scale
P2 Standard labels 0~25 mmHg
CVP Central venous pressure 0~25 mmHg
PAP Pulmonary arterial pressure 0~50 mmHg
LAP Left arterial pressure 0~25 mmHg
-
BT-750 Operator Manual 35
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
●IBP measurement range
Setup Mode Range of change Description
WAVE
SPEED
6.5, 12.5, 25,50 mm/sec Set the speed of IBP waveform
BP1 LABEL P1 , ABP
Set the level according to the
measurement site of BP. Change the
alarm setting range according to
level.
BP1 SCALE 50,100,200,300 mmHg Set the display of BP1 wave scale
BP1 ZERO YES/NO Adjust the zero point of the
transducer
BP1 ALARM Refer to Ch.8 Alarm Setup
BP2 LABEL
P2,CVP,PAP,LAP
Set the level according to the
measurement site of BP. Change the
alarm setting range according to
level.
BP2 SCALE 25,50,100,200,300 mmHg Set the display of BP2 wave scale
BP2 ZERO YES/NO Adjust the zero point of the
transducer
BP2 ALARM Refer to Ch.8 Alarm Setup
IBP Enable / DISABLE Select the IBP display
IBP inactive display window (Show the unified NIBP display window.)
-
BT-750 Operator Manual 36
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.8 Temperature Setup
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
● Temperature Menu Setup Range
Setup Mode Range of change Description
Temp Unit ℃, ℉ Set the unit of temperature
Temp 1 Alarm Refer to Ch.8 Alarm Setup
Temp 2 Alarm
-
BT-750 Operator Manual 37
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.9 CO2 Setup
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
● Setup range of EtCO2
Setup Mode Range of Change Description
WAVE
SPEED
6.5, 12.5, 25 mm/sec Set the speed of EtCO2 waveform
UNIT mmHg, kPa, % Set the unit of EtCO2
measurement
SCALE ~12,~25, ~50, ~70, ~100,~150
mmHg Set the WAVE SCALE
ZERO YES / NO Adjust the zero point of the
transducer
AVERAGE 1 BREATH, 10, 20 Sec Set DATA average time DATA
Gas COMP NON,O2/N2O, N2O, O2 Set the Gas type
EtCO2 Enable / DISABLE Set ON/OFF of EtCO2 module
ALARM Refer to Ch.8 Alarm Setup
-
BT-750 Operator Manual 38
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.10 Trend Setup
1) Press the button “ ”to convert to Trend display window.
2) Adjust the tram knob and select SCAN Then, selected bar becomes active. Adjust the bar to
the required location with tram knob.
3) Select the PAGE with tram knob, to turn over display window.
4) Press the button “ ” to return to main display.
Mode Description
SCAN Select the data of current page by adjusting Trim Knob.
PAGE
Could see 1~17 pages by adjusting Trim Knob
Show the data of 174 cases each page.
-- SEC Store interval data by selected interval second.
Store the data of approximately 2960 cases
Note
Up to the selection of Second, it could store minimum 8 hours to
maximum 72 hours trend data.
-
BT-750 Operator Manual 39
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5.11 Printer Setup
[Pinter Menu Display]
1) Press the menu button “ ” to go to Setup Printer Setup.
2) Select the menu to change with trim knob and press the trim knob to set the location
to be modified. Then, selected menu is displayed.
3) Press the trim knob to make the display window active in second time.
4) Go to the required location with tramp knob and press the tram knob to set thelocation
to be modified ,then the setting location becomes active .
5) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
6) Press “Return” button to go back to 1st display window.
7) Press the button “ ” to return to main display.
● Select the each Printer Mode using same method.
Setup Mode Range of change Description
WAVE ECG 1+ PLETH
ECG 1+ CAPNO Select the waveform to print out
PRINT SPEED 15/30 mm/sec Select speed of printer
PRINT
ALARM ON / OFF
ON: Print out in case of over alarm limit
automatically.
OFF: manual printing
-
BT-750 Operator Manual 40
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 6
Measurement of ECG and Respiration 6.1 Setup Connections
1) Connect the ECG cable to the ECG connection terminal of the measurement module.
● The process of depolarization and polarization of the myocardium generates electronic
potentials that are sensed by ECG electrodes on the skin surface.
● Changes in the chest impedance resulting from the patient’s respiration is measured and
displayed as a waveform and the values will appear on the screen.
2) Attach the patient’s electrode to the sites shown below.
R: Red
L: Yellow
C: White
N: Black F: Green
R: Red
L: Yellow
F: Green
ECG Connector
-
BT-750 Operator Manual 41
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
3) Description of Lead IEC(AAMI)
Lead Color of
electrode
Attaching site
R(RA) Red (White) Just below the clavicle (scapula) near the right shoulder
L(LA) Yellow(Black) Just below the clavicle (scapula) near the left shoulder
F(LL) Green(Red) Lower area of the left side of the abdomen
N(RL) Black(Green) Lower area of the right side of the abdomen
C(V) White(Brown) Select the candidate lead for measurement among the
chest leads
Note
ECG and SpO2 will display synchronized numeric when you stop ECG
measurement while you are measuring ECG SpO2 at the same time
The default setting of ECG Lead is ECG1: Lead II, ECG2: Lead III
In case of using ECG 5 Lead cable, you need to set menu 5 Leads at ECG Cable.
Respiration signals are relatively more sensitive to interference from radiated
electromagnetic signals. .
Do not rely entirely on the equipment.
Respiration is measured with the ECG and HR etc., using the ECG cable
To setup Apnea alarm, push Menu button RESP->APNEA Time
Priority is EtCO2 measurement when measuring EtCO2.
Respiration function is very sensitively effected by electromagnetic wave, so it
should avoid from judging the status of patient by numeric value and waveform
of respiration.
Alarm cases of arrhythmia analysis
Name of Arrhythmia Description
Tachycardia: TAC Heart rate is over 140bpm for a minute.
Bradycardia: BRD Heart rate is under 40bpm for a minute.
Premature Ventricular
Contraction: PVC Abnormal heart rate is over 1.
Ventricular Tachycardia: VTAC Over 8 PVC is over 140bpm
Asystole: ASY No electric action of heart.
Ventricular Bigeminy: BGM After normal heart rate once, two times or more PVC
Ventricular Trigeminy: TGM After normal heart rate at two times, two times or more PVC
Ventricular RUN After normal heart rate at Over 3 (PVC)
Venticule of over 100bpm
Ventricular Fibrillation:
VFIB
Very fast ,chaotic heart rate in the lower chambers of
the hearts, resulting from multiple areas of the ventricle
is attempting to control the heart’s rhythm.
Multifocal PVCs Over the last 15 beats two or more premature ventricular
beats.
Ventricular Couplet (COP) Abnormal heart rate is more than two times
Missed Beat (MIB) ECG wave is missed intermediately
R on T PVC PVC is detected before finishing T wave in the normal ECG Wave.
Frequent PVCs PVC generates higher than alarm limit.
-
BT-750 Operator Manual 42
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Caution
Inspection does not carry out the patients with arrhythmia, premature
ventricular contraction or ventricular tachycardia.
6.2 Respiration Menu Display
1) Press the button “ ” to go to adjustment display.
2) Select the menu to change with trim knob and press the trim knob to set the location to
be modified. Then, selected menu is displayed.
3) Turn to right and left with trim knob and press the trim knob to set up the numeric
value and save it.
4) Press “Return” button to go back to 1st display window.
5) Press the button “ ” to return to main display.
The range of respiration setup
Setup Mode Range of modification Description
WAVE SPEED 6.5/12.5/25 mm/s Set Wave Speed
RESP LEAD RA-LL/RA-LA Set one of two
RESP GAIN ⅹ1/2, ⅹ1, ⅹ2 Set the size of RESP waveform
APNEA ALARM ON/OFF Set alarm for apnea
APNEA TIME [SEC] 20~120 Sec. Set detection of Apnea
RESP ALARM Refer to CH.8 Alarm Setup
-
BT-750 Operator Manual 43
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 7
Measurement of SpO2
7.1 Preparation for SpO2 Measurement
1) Connect SpO2 probe to the probe port.
7.1.1 Setting SpO2 Probe
1) Clean the body part (a finger or a toe) to be inserted to probe for measurement with
alcohol.
2) Set the probe on the finger to measure. (Avoid using it with other medical devices to
affect blood flow and setting it on the body part in medical treatment )
3) Do not make any motion and be in pause as possible as you can for stable measurement
and fix the probe cable with your fingers or plasters but be careful not to make interference in
blood flow.
SpO2 Connector
-
BT-750 Operator Manual 44
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
4) Make sure that the probe is set on the body part correctly in every 2 or 3 hours. If the state
of your skin is different, change the part to be attached to other side.
Note
When probe is used incorrectly, inaccurate value or waveform can
come out as below.
● When the patient uses the probes without certification of Bistos
Technology.
● Hemoglobin dysfunction
Caution
Probe should be treated with attention. Careless use might cause
damage to probe. Keep the probe cable away from acute objects.
When the patient has a high fever or peripheral circulatory failure,
the skin, temperature increases to 2~3 degree.
In case of the patients with cold fingers, SpO2 data is not accurate.
For the patients who have abnormally high oxyhemoglobin or
methemoglobin , SpO2 data is not accurate.
You can not use disposable probe to different patients. For same
patient, it is possible to reuse and remeasure even in different
portion. If you want to reuse disposable probe, disinfection is
indispensable.
Caution
It might fail to measure pulse rate under occasions as follows;
When probe is fastened tightly.
When probe excessively exposes to Medical illumination, bilirubin
illumination or sunlight.
In case of being measured on arms or legs in cuffs or bandages.
Warning
In case of measuring with probe while MRI is in use, the patient may
have burns. To minimize the danger, use non-inductive line, if there
exists any danger even with proper use, remove the connector from
the patient.
● Allergic patients should not use disposable probes.
● Make sure that the probes radiate light, faced each other correctly,
and the light reaches to the tissue of a patient. When applying to a
newborn baby, use disposable probes outside of the incubator.
● If used in an incubator, accurate results may not be obtained
because the humid air can have an effect on the outcome.
When probe is excessively fixed with tape or used in other incorrect
way, the patient may suffer from injury.
Warning
If the probe are excessively exposed to medical illumination
( especially for xenon light ), bilirubin illumination, fluorescent
lamp, IR heating appliances or direct ray of light , this may cause
problem in function . To avoid excessive exposure to the light, use it
by the general operation and be careful that the probe is not exposed
to outside ray with opacity. If outdoor circumstance is under
extremely high light, it may generate inaccurate value.
Do not use probe at the body part where intravenous injection or
artery catheters are placed.
Do not use damaged SpO2 probe or optical appliances.
Do not clean with oxygenated water, ultraviolet ray, direct ray of
light and steam.
-
BT-750 Operator Manual 45
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 8
Measurement of NIBP
8.1 Setting NIBP
1) Choose the proper cuff for the patient.
2) Connect the cuff hose to cuff.
3) Connect the cuff hose to NIBP font.
4) Put the cuff on the patient.
5) Use the proper cuff to the patient before measuring blood pressure.
8.1.1 NIBP measurement
1) It should be measured at the same label of the patient’s heart in the arms and legs of the
patient.
2) The cuff ties part under the position to be measured.
It may cause errors in measurement values.
3) NIBP measurement will be done with button
and if you press the button one more time, it
stops measurement.
Note
It is impossible to measure the blood pressure in upper arm, in case
that infusion pump or catheter is inserted in the blood vessel. It may
cause damages in the organ around catheter during the supply of
blood pressure.
Caution
Use the proper cuff to the patient before measuring blood pressure.
Inspect excess blood pressure or folding of cuff.
Avoid the cuff from getting wetness out of any kinds of incidents,
and use it after dry completely.
NIBP hose connector
-
BT-750 Operator Manual 46
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 9
Measurement of Temperature
9.1 Temperature Measurement
● Changes in impedance according to the change in the patient’s body temperature is
perceived by the temperature sensor, and then displayed numerically on the screen after a
series of calculations. Measured Temp
1) Connect the temperature sensor to the connection terminal of the measurement module
and sensor to patient.
Caution
The temperature sensor should be sterilized before application to
another patient.
Temperature Connector
-
BT-750 Operator Manual 47
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 10
Measurement of IBP
10.1 IBP Measurement
Arterial Blood Pressure is the pressure exerted by the blood against the walls of the arterial
vessels. The cardiac cycle consists of a period of relaxation called Diastole, during which
the heart fills with blood, followed by a period of contraction called Systole. While the heart
is contracting and relaxing, the sensor gets the measurement of the pressure.
10.2 Setup Connections 1) Connect the interface cable for the transducer to the IBP connector on the monitor’s front
panel. An interface cable for the transducer has to be selected correctly as it depends on
the each transducer type.
Note
The monitor is designed to accept signals from BT4812-3 disposable
transducer, or equivalent (pressure range of 0~300 mmHg). Refer to
the transducer directions for use for details.
*This transducer is verified under the connection with to see if
protective means is provided against hazards to patient when used
with HF surgical equipment.
IBP Connector
-
BT-750 Operator Manual 48
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
2) Set up the patient circuit according to the directions for use of the transducer, monitoring kit.
3) “Hold the backside of the clear cover surrounding the connector, and connect the converter
to the reusable monitor interface.
4) In a clean environment, open the package to check if all parts are well connected and if the
handle of the “stopcock” is positioned properly. All side ports of the stopcock are protected
by the outlet plugs and may not be removed till the system is full and the foam has been
removed. These outlet plugs should always be replaced with other outlet plugs. (A pouch is
included in the kit.).
-
BT-750 Operator Manual 49
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
5) How to use Flush
Hold both sides of the flush operator and pull upward gently. Do not revolve the operator and
take care not to put the power on one side.
Caution
If a fast flush is performed on the patient, the user should check
carefully for the presence of foam and particulate matter. If a large
volume is flushed by force, a short flush with an increase rate less
than 2cc is recommended to avoid central embolism.
Blood pressure for a liquid infusion line over 300mmHg requires an
infusion of more than 3cc per hour. In this case, the pressure must
not exceed 775mmHg. A protective function is set in the flush
device, which prevents overpressure of the converter by making the
liquid bypass the device. If a more precise control of fluid volume is
required, it is necessary to connect the infusion pump around the
flush device.
6-1) Twist the sterilized solution bag lightly using a drip chamber spike
6-2) Open the roller clamp and remove the air inside the bag completely by squeezing
using an infusion set or a needle inserted into the injection port in the back. By emptying
the back this way, it can prevent air input into the patient’s body.
6-3) Before inputting the solution into the infusion set, push both sides and fill the drip
chamber partially.
6-4) Operate the flush device gently. Since air rises from the bottom, make sure that the
solution is in the bottom at all times.
6-5) Prime the side ports and the plug of the zero point stopcock in the converter using a
door-lock plug.
6-6) Check if bubbles are present in the monitoring system. To check the presence of
invisible bubbles, rap it lightly
6-7) Add pressure to the system up to 300mmHg using C-fuse or the clear cuff. Flush the
system for 2~3 seconds. Check for the presence of air bubbles which may result from a fast
flush. Now, the system is ready for zero point setting and measuring. Place the stopcock in
a 90 degree position to let it off state.
-
BT-750 Operator Manual 50
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Caution
The stopcocks of the converter or the outlet should be locked before
connecting the door-lock plug.
Note
>
● Priming should be done slowly!
- Slow priming will lessen the effort to remove air bubbles afterward.
● Priming using gravity!
- Pressing may result in leakage of liquid or bubbles by forcing the
liquid to flow into the system. If small bubbles flow too slowly
inside the system, lift up the supply bag. 1 inch corresponds to
2mmHg, and the primed 4-feet line gives a pressure of about
100mmHg if it is completely spread out The solution bag should
be placed at a higher position than the converter and the pressing tube,
in order for priming using gravity.
Caution
If air foam or bubbles are present inside the system, it may cause
significant bias of the pressure waveform or air embolism.
Care should be taken not to let air bubbles inside the 3-way outlet
stopcock or the cannula reflush into the patient. To check, confirm if
the monitoring line is fully filled with liquid before connecting the
monitoring line and let a small volume of blood flow through the
cannula.
10.3 Adjustment of Zero point of the Converter
1) Lock the zero point stopcock (toward OFF) and release the converter according to the air
pressure.
2) Set “Menu”-> “Zero”-> “Yes” in order. If final value is “0”, the measurement starts.
-
BT-750 Operator Manual 51
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 11
Measurement of EtCO2
11.1 EtCO2 Measurement (Optional)
Sidestream CO2 Gas Measurement
Sidestream CO2 module can be fitted onto BT-750. Sidestream CO2 provides the function of
ETCO2 and the function to measure respiratory CO2 and respiration numbers using the small
filterline based on a small lumen type filterline. In the Sidestream CO2 type, the level is
measured through the cannula (non-tubing type) attached to the nose or a sampling line
equipped in the respiratory circuit (tubing type).
Caution
The maximum sampling rate of the nose tube is 50mL/min. The
product should not applied to patients who may experience
respiratory distress by the vacuum flow level.
To prevent infection of the medical staff by the patient’s respiration
sample, the outlet of BT-750 should be connected to the air
elimination system of the hospital.
Caution
The gas inspiring vacuum pressure (negative pressure) of the gas
elimination system should not exceed the pump outlet standard of
BT-750, 1mmHg. Excessive pressure will display the message of
“OCCLUSION” and it may damage the BT-750 product pump.
During zeroing, the air elimination system should be operating.
Waste from Microstream and CO2 filter should be treated as fatal
biological materials for the human body
Note
Connection of all tubes should be safe, and the nose cannula should
be kept away from the CO2-present area during the self test period
(including the outlet of the ventilator and the respiration of the user).
Range of Respiration Measuring : 0 ~ 150bpm
For the best accuracy, a warm-up of about 2 minutes is required.
Note
We do not use specific gas for calibration and manually operate “O”
Calibration.
When you adjust “O” calibration, connect it to the transducer and
monitor in the air. The patient may not connect it to the other gas.
Operate “ Menu → ZERO → YES” in order .
Use it after verifying “O” value at the window screen, otherwise the
measurement value would not be correct.
-
BT-750 Operator Manual 52
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
11.2 How to measure Sidestream CO2
1) Attach the gas/tube to BT-750 Sidestream port and the other side attach to hospital gas
device
2) Respiration numeric is automatically changed when EtCO2 is measured.
3) EtCO2 will be displayed on BT-750 screen after 20 seconds when it detect proper input.
4) EtCO2 waveform and data is displayed. If waveform is not displayed on the screen, change
setting menu “Menu → EtCO2 → EtCO2 → Enable
5) You can change EtCO2 waveform when you need.
Caution
The watertrap is disposable and should only be used for a single
patient. Do not reuse the watertrap for another patient.
Caution
The cannula is disposable and should only be used for a single
patient. Do not reuse the cannula for another patient.
Caution
If oxygen is being delivered while using Sidestream CO2, be sure to
use a CO2 Sampling and O2 Delivery cannula. Using a different type
of cannula could obstruct oxygen delivery.
Warning
Assembly, extension, repair should be carried out by authorized
person from Bistos Technology.
-
BT-750 Operator Manual 53
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 12
Maintenance 12.1 Maintenance and Cleaning
BT-750 and its accessories can be cleaned by using various methods. Please follow the
methods below to avoid unnecessary damage or contamination to the equipment.
Keep the equipment far from the place possibly water sprinkling or soaking into the equipment.
Avoid the place possibly affecting to the correct measurement with air pressure, temperature, humidity, ventilation, sunlight, dust or air including with salt, ion.
Storage Temperature
Packaged storage (in box) : -20℃~80℃
Unpackaged storage : -20℃~60℃
Storage Humidity : 0~80% under 12℃~60℃ range Keep it safe from tilt, vibration or impact. Avoid the location exposed to chemical or explosive gas. Keep the equipment clean for the next operation. Stop operating and contact the right personnel immediately when malfunction
occurs.
Never let the unauthorized person disassemble, repair or alter. At least once a month, clean and wipe off the monitor by using the soft cloth after wetting it in lukewarm water or alcohol. Do not use lacquer, thinner, ethylene or oxides which could be harmful to the equipment. Make sure both cables and accessories are free of dust or contaminants and wipe them off with soft cloth wet by lukewarm water (40℃/104℉) and at least once a week clean them with clinic alcohol. Do not soak the accessories in any liquid or detergent. Also, don't let the liquid penetrate into the equipment or probe.
12.2 How to clean Accessories
For cleaning the SpO2 finger probes, be directed below.
Chemicals How to clean
Alcohol Isopropyl Alcohol Wash with wet gauze
Liquid soap Benzalconium chloride 0.05 W/V%(200x diluted)/Wash with wet gauze
0.01 W/V%(50x diluted)/Wash with wet gauze
Iodine Povidone Iodine 0.02 W/V%(50x diluted)/Wash with wet gauze
Glutaral C5H8O2 2 W/V%(experiment solution)/Wash with wet gauze
-
BT-750 Operator Manual 54
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Caution
When harmful (unaccepted) materials are used for cleaning, we shall not
offer the service without charge in regardless of warranty period.
Check carefully both monitor and probe after cleaning the equipment.
Do not use the equipment that is worn out or damaged.
Caution
You can get incorrect measurement figure under the following conditions.
- The patients’ excessive motion
- Using electrical operational instrument with high-frequency or the
instrument for cardioversion.
- Venous blood pulse
- Appling to the monitor parts to measure blood pressure or use arterial
catheter/bandage and intravascular parts
- When the patients are on low blood pressure, severe blood vessel
contract, sever anemia or low body temperature.
- Arterial occlusion nearby probes
- When the patients are on cardiac arrest or shock.
Premature neonates and patients with chronic pulmonary disease should be checked for oxygenation levels before starting treatment.
To prevent electric noise during use, the product should be installed apart from dynamo, X-ray equipment, broadcast equipment or portable
cables. An inaccurate result may occur when these equipment are placed
near the product.
To ensure patient is electrically isolated, connect only to other
equipment that provides patient electrical isolation. When the patient
with pacemaker use the monitor, only qualified personnel should
operate and control all functions of the monitor.
-
BT-750 Operator Manual 55
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
Chapter 13
Troubleshooting
This chapter explains the solutions to simple problems. One might feel difficulty to use while
operating the monitor. At this moment, before regarding it as breakdown, check the following
points.
Problem Sign Solution
Cannot turn
the Power on
Check whether AC power
line is plugged in.
Faulty power switch
☞ Connect to AC power properly. ☞ Contact our customer service center or repair center.
Cannot produce
the proper
waveform
Unstable connection. ☞ Check the connections of the power, connectors, and the
accessories.
▣ Problem of ECG/Respiration measurement
Problem Solution
The message“ LEAD
FAULT” is displayed.
1. Check the ECG cable and ensure that the gel in the electrode is
not dry.
2. Try using another ECG cable.
The ECG waveform
is not displayed.
1. Check if the cable connection is stable. 2. Check if the ECG cable is damaged.
Severe noise on the
waveform.
1. Check if the electrode attached to the patient is stable. 2. Check if the disposal electrode is one provided by the
company. (Noise may occur according to the kind of
electrode.)
▣ Problems in SpO2 Measurement
Problem Cause Solution
Error display
by ray interference
Errors generated by excessive ray
outside.
Cover the probes with opaque
substances.
Error by
electromagnetic
wave.
Power cable or connectors are placed
closely with probes.
Keep probes away from
power cable and connector.
Error display
Use of improper probes or clinical
problem Misplaced probes or patient
moving.
Probes do not attach to the patient.
Use of proper probes. Place
probes in right location Fix
probe and cable with tape and
use new probes as required.
Output failure of
SpO2 value
Change of arterial blood capacity by
shock. Apply to other body part
SpO2 waveform
displays in small
or flat shape.
Fingers in cold temperature or
improper probe use , setup
error for waveform size
Apply to other body part or
check if it is used in proper
way and waveform size is set
up correctly.
-
BT-750 Operator Manual 56
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
▣ NIBP measurement
Problem Solution
NIBP measurement is not d