Body Imaging at 7T?: Perspectives in 2012
Neil M. Rofsky, MD, FACR, FSCBTMR, FISMRM
In 1999, did you think you would use a 3T scanner for body imaging?
• 1) I was confident of this
• 2) I was unsure
• 3) I was very skeptical
• 4) I thought it was impossible
How would you characterize your current use of 3T for body imaging?
• 1) Interchangeably with 1.5T
• 2) Sparingly
• 3) Not at all
• 4) No access to a 3T in my practice
Starfruit at 7T
7T
• Unlike conventional 1.5T and 3T MR systems using a large volume built-in transmit body RF coil, 7 Tesla MRI at this point uses local transmit/receive RF coils
Umutlu,et al. Eur Radiol 2011; 21:841–849
Why 7T?
• SNR advantage
• Chemical separation
• Spectro
• Multipoint Dixon
• Na-23, O-17, P-31
• Unique contrast
• Susceptibility – brain, DWI?
• Fxnal blood flow
1H NMR spectra of subcutaneous fat (left) and tibial bone marrow (right) from a 26 year-old healthy male at 7 Tesla (7T).
Ren J et al. J. Lipid Res. 2008;49:2055-2062
So what?
• The ability to resolve protons adjacent to double bonds allows noninvasive estimation of the fatty acid composition of adipose tissue.
• Potential as an internal standard to quantify intramuscular lipids
• New insights into insulin sensitivity, diabetes, and obesity
Challenges: Images
• Artifacts
• Tissue susceptibility
• Chemical shift
• Signal heterogeneities
• Radiofrequency (RF) wavelength
• SAR
• Flip angle adjustments
• Echo train imaging
7T, 1H frequency = 298 MHz λ
• the radio frequency (RF) ��s in body
tissues are ≤ the object of interest resulting in non-uniform excitation and receive patterns due to complex RF field interferences.
300 cm
T2WI (TSE) at 7T
Umutlu,et al. Eur Radiol (2011) 21:841–849
16 channel transmit/receive array
Ladd et al. 2010 IEEE International Symposium on Biomedical Imaging Pages: 572-572
What # is optimal???? $$$$
7T Abdomen: T2WI
T2W, echo-train imaging: FSE, TSE
• Mult refocusing pulses
• SAR intensive
• Mitigate?
�� FAs
• Novel sequence design
T2W, echo-train imaging: FSE, TSE
1/100th reduction in SAR
Sarkar, et al. Radiology 2011; 259: 550-557.
T2W, echo-train imaging: FSE, TSE
1/100th reduction in SAR
Sarkar, et al. Radiology 2011; 259: 550-557.
*
Renal imaging at 7T (flat contrast)
T1WI GRE in phase T1WI GRE opposed phase
Umutlu,et al. Eur Radiol 2011; 21:841–849
Renal imaging at 7T (flat contrast)
T1WI GRE in phase T1WI GRE opposed phase
Umutlu,et al. Eur Radiol 2011; 21:841–849
add-on system for RF shimming integrated on the small signal side of the MR system
Renal imaging at 7T 2D FLASH
Umutlu,et al. Eur Radiol; 2011:841–849
L S
Prostate at 7T: Radiative Surface Coil
0.5 mm x 0.5 mm x 3 mm Raaijmakers et al. Mag Res in Medicine 66:1488–1497 (2011)
3T ERC 0.7x 0.7 x 2mm
7T: Patient acceptance
Mean grades < 3 (“very well tolerable”) except for exam duration at 7T.
Theyson, et al. Magn Reson Mater Phy (2008) 21:63–72
Conclusions
• Not useful for clinical in 2012
• Impossible??
• Niche applications??
• Stay tuned…..