Page 1: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Performance Maintenance Pro-

tocol

2535/2545/2555 Quaternary Gradient Module

Serial Number/Firmware Version:

     

Date:      

Performer:      

© 2014 WATERS CORPORATION. ALL RIGHTS RESERVED.

*715001946* *Rev.E*

Page 2: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Performance maintenance

GoalThis performance maintenance protocol is a guide to performing maintenance tasks on your system. Packaged as part of the performance maintenance kit, it is intended to ensure optimum performance and an uninterrupted period of trouble-free operation.

ScheduleWaters recommends that performance maintenance visits be performed once a year. Systems under heavy use or using aggressive solvents will need more frequent maintenance. For detailed maintenance schedules, refer to the appropriate operator’s guide(s) located on the Waters web site. Failure to follow the maintenance schedules as described in the operator’s guides may result in unreliable operation and inaccurate laboratory results. For some protocols, optional tests can be performed by Waters field service engineers for added confidence and reliability.

Safety considerations

Caution: Always observe Good Laboratory Practice and, when handling Waters Critical Clean™ parts, refer to Controlling Contamination in UPLC/MS and HPLC/MS Systems, P/N 715001307.

Warning: To prevent personal contamination with biologically hazardous or toxic materials, carefully dispose of the replaced component in the appropriate waste container.

Warning: When performing maintenance, remove power to the unit by powering down and disconnecting the power cord.

Warning: Always use proper static protection when handling any printed circuit boards.

Warning: Dispose of components in accordance with local or customer regulations. If the component comes with a Waters Corporation return sheet, follow its instructions.

Warning: Avoid contact with solvents by wearing suitable gloves and safety glasses.

Some protocols may involve removing instrument covers, exposing the performer to high voltages, high currents, and moving parts. This presents risks of electric shock and/or damage to the instrument. To reduce these risks, Waters recommends that these protocols be conducted by certified Waters field service engineers, who have access to the latest information on Waters products. Waters is not responsible for an unqualified performer initiating actions that they do not understand. If there is any doubt on the part of the performer, a Waters certified field service engineer should conduct the performance maintenance activity.

Waters Quality PartsWaters designs and manufactures Waters Quality Parts® to the same strict regulatory standards as your high-performance Waters systems. Waters recommends using Waters Quality Parts for all

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 2 OF 14

Page 3: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

maintenance activities. Failure to use Waters Quality Parts may compromise laboratory results and reliability.

Customer educationWaters offers several operation and maintenance courses for ACQUITY® UPLC®, Alliance® HPLC and MS systems as well as courses designed to help you maintain professional competency and further your knowledge in chromatography, UPLC, informatics and mass spectrometry. For more information on available training courses from Waters Educational Services please visit the Waters Web site.

Compliance servicesWaters Compliance Services help you achieve and maintain the regulatory compliance of your LC and MS instruments and software with a combination of expertise, compliant-ready systems and software, and a portfolio of services to ensure that you have quality data, consistently. To learn more about Compliance Services, please visit the Waters Web site.

General tasks

Use the following guidelines when conducting performance maintenance visits. This information is intended to be a guide only. For example:

If a particular module has been recently serviced, it may need only testing and replacement of fewer parts.

Not all procedures apply to all instruments (specify NA in the Comments section if the procedure does not apply).

If the instrument fails to perform adequately during pre-performance maintenance checks, further service may be required beyond the scope of this protocol. Such service may incur additional parts and labor charges.

The following table lists the general tasks to be performed during a performance maintenance visit. Perform and confirm the steps listed for each procedure and record any comments that you have about the results at the end of each section.For detailed instructions and a list of the necessary tools, refer to the operator’s guide located on www.waters.com.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 3 OF 14

Page 4: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Inspect Maintenance tasks

Post-perfor-mance main-

tenancechecks

Optional calibration tests1

Firmware version2 Filters Initialization di-agnostics Flow rate accuracy

Fan Pump head com-ponents Static leak test Gradient proportioning

Helium air regulator Vent valve rotor seal assembly

Create log en-tries

Tubing and fittings Battery3

Error logs

1 Optional tests performed by Waters field service engineers as part of a Total Assurance with Calibration Plan. For those without the plan, the tests can be performed for an additional charge. Some of these tests may require addi-tional chemistry kits that are not supplied in the performance maintenance kit.

2 Firmware upgrades, if required, should be performed by a Waters field service engineer prior to beginning perfor-mance maintenance. Firmware upgrades may require re-qualification of the upgraded instrument if it is in a regu-lated environment.

3 Batteries replaced by Waters field service engineers only.

Inspect

Firmware version

Confirm what firmware versions are installed. Update if required.

NOTE: Refer to Downloads under the Services & Support tab on the waters.com home page for the latest firmware and driver pack release notes. These release notes detail software and data system requirements, and installation notes.

To obtain the firmware version:

1. Power-on the module.2. Start the chromatographic data system software, and then navigate to the Quaternary

Gradient Module (QGM) console.3. From the console, click Configure > View module information.4. Check the firmware version and note it on the cover page.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 4 OF 14

Page 5: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Firmware check complete. Yes No N/A

Comments:      

     

     

Fan

To inspect the fans:

1. Observe the fan on the rear of the instrument.2. Ensure that the fan grate is clear of debris and dust. The fan should be blowing outward.3. Clean if necessary.

Fan inspection complete. Yes No N/A

Comments:      

     

     

Helium air regulator

NOTE: The helium air regulator pressure must read between 50-150 psi (345-1034 kPa; 3.45-10.34 bar). Adjust if necessary.

Helium air regulator pressure and minimum specifications for helium sparge check complete. Yes No N/A

Comments:      

     

     

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 5 OF 14

Page 6: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Tubing and fittingsCheck tubing and fittings for leaks.

Tubing and fittings inspection complete. Yes No N/A

Comments:      

     

     

Error logs

To review the error logs:

1. Power-on the module.2. Start the chromatographic data system software.3. In the console select the QGM from the system tree.4. Click on Logs to review the error logs.

Review error logs complete. Yes No N/A

Comments:      

     

     

Maintenance tasks

The following components are replaced/rebuilt as part of the performance maintenance. Wa-ters Quality Parts have been used to replace/rebuild all parts.

Filters

Pump head components

Vent valve rotor seal assembly

Battery

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 6 OF 14

Page 7: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Yes No N/A

Comments:      

     

     

Post-performance maintenance checks

Perform the post-performance maintenance checks only in the order listed.

Initialization diagnostics

Turn on power, allow instrument to complete and pass on-board software checks.

Initialization diagnostics check complete. Yes No N/A

Comments:      

     

     

Static leak test

1. Prime the system.2. Energize the seals.3. Start the leak test.4. Record the results.

N A leak rate of 300 psi/min or less for each pump head indicates a successful or passed test.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 7 OF 14

Pump head Leak rate (psi/min) Pass/Fail

Left       Pass FailRight       Pass Fail

Page 8: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Static leak test complete. Yes No N/A

Comments:      

     

     

Create log entriesAfter completing performance maintenance, create a log entry.

1. In the console, click on the QGM in the system tree.2. From the menu bar, click Maintain > Create log entry. The Create Log Entry dialog box

appears.3. Enter the operator’s name, the date of the performance maintenance, and any other com-

ments.Create log entries complete.

Yes No N/A

Comments:      

     

     

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 8 OF 14

Page 9: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Performance maintenance confirmation

I confirm on behalf of      

Company Name

that this performance maintenance procedure has been satisfactorily completed on:      

Serial No.

Signed: Date:      Customer

     Type Name

I certify that the maintenance procedure has been successfully completed.

Signed: Date:      Waters Engineer

     Type Name

Comments:     

     

Waters Corporation34 Maple Street Milford, Massachusetts01757 USA

Waters, MassLynx, The Science of What’s Possible and Waters Quality Parts are registered trademarks of Waters Corporation. All other trademarks are property of their respective owners. ©2014 Waters Corporation. Printed in the U.S.A.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 9 OF 14

Page 10: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Optional calibration tests

Optional calibration tests are performed by Waters field service engineers as part of a Total Assur-ance with Calibration Plan. For customers without this plan, these tests can be performed for an additional charge. Some of these tests may require additional chemistry kits that are not supplied in the performance maintenance kit.

Optional calibration tests• Flow rate accuracy• Gradient proportioning

Tools/materials required• Graduated cylinder, 100 mL (flow rate accuracy test for 2535 QGM)• Graduated cylinder, 250 mL (flow rate accuracy test for 2545 QGM)• (2) Graduated cylinders, 500 mL (flow rate accuracy test for 2555 QGM and gradient propor-

tioning testing)• (2) Graduated cylinders, 1 L (gradient proportioning testing for 2555 QGM)• Timer (calibrated)• 100% HPLC–grade methanol• 100% HPLC-grade water• pressure-restriction device capable of producing a backpressure of 600 to 1000 psi (4136 to

6894 kPa; 41 to 69 bar)• Powder-free nitrile gloves and safety glasses

Caution: Never change directly between immiscible solvents, or between buffered solu-tions and organic solvents. Immiscible solvents form emulsions in the fluid pathways. Buffered solutions and organic solvents in combination can result in salt precipitation in the gradient proportioning valves, pump heads, check valves, and other parts of the system. Confirm that all fluids in the system are miscible with methanol.

Flow rate accuracy

1. Choose the appropriate graduated cylinder for the instrument being tested.

Caution: Ensure that the graduated cylinder is thoroughly dry.

QGM Graduated cylinder size (mL)

2535 1002545 2502555 500

2. Prime the system with fully degassed methanol or water.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 10 OF 14

Page 11: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Notice: If using the internal sparge, set the initial rate duty cycle to 100%, and the initial rate duration to a time dependent on the volume of your solvent (10 minutes per liter of mobile phase). Reduce the rate duty cycle to 20% before performing the test.

3. Prepare the timer to begin timing.4. Enter the test flow rate for the instrument being tested.

QGM Test flow rate (mL/min)

2535 202545 502555 100

5. Adjust the pressure-restriction device to create a backpressure of 600 to 1000 psi (4136

to 6894 kPa; 41 to 69 bar)6. When the flow and pressure are stable, insert the outlet tubing into the graduated cylin-

der.7. When the bottom of the meniscus reaches the specified level on the graduated cylinder,

start the stopwatch.

QGM Start level2535 202545 502555 100

8. Stop the timer when the bottom of the meniscus reaches the specified mark on the gradu-ated cylinder.

QGM Stop level2535 1002545 2502555 500

9. Record the time required to reach the final mark.N If the time is not within the acceptable range, other service adjustments may be

necessary for an additional charge.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 11 OF 14

Page 12: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

TimerSerial/ID number      Calibration date      

Flow rate accuracy test complete. Yes No N/A

Comments:      

     

     

Gradient proportioning

1. If you have not already done so, prime the system with fully degassed methanol.

Notice: If using the internal sparge, set the initial rate duty cycle to 100%, and the initial rate duration to a time dependent on the volume of your solvent (10 minutes per liter of mobile phase). Reduce the rate duty cycle to 20% before performing the test.

2. Place the primed solvent A and B lines from the separations module into two identical graduated cylinders of the appropriate size.

QGM Graduated cylinder size (mL)

2535 2502545 5002555 1000

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 12 OF 14

Actual value Expected result

      minutes 3:48 – 4:12 minutes

Page 13: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

3. Fill the cylinders to the appropriate meniscus level with 100% degassed methanol.

QGM Level2535 2002545 5002555 1000

4. Set the proper flow rate for the instrument you are using with 50% solvent A, and 50% solvent B.

QGM Flow rate (mL/min)2535 202545 502555 100

5. When the meniscus of the solvent in the A cylinder reaches the appropriate level for the instrument you are testing, stop the flow.

QGM Stop level2535 1002545 2502555 500

6. Record the values for the volumes in each cylinder.

NOTE: Expected Result: B cylinder volume = A cylinder volume ± 10 mL.

N If the volume is not within the acceptable range, other service adjustments may be necessary for an additional charge.

Solvent Actual vol-ume

A       mLB       mL

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 13 OF 14

Page 14: Blank Procedure - Waters Corporation · Web viewEnsure that the fan grate is clear of debris and dust. The fan should be blowing outward. Clean if necessary. Fan inspection complete

715001946 REV. E

Calibration confirmation

I confirm on behalf of      

Company Name

that the calibration has been satisfactorily completed on:      

Serial No.

Signed: Date:      Customer

     Type Name

I certify that the calibration has been successfully completed.

Signed: Date:      Waters Engineer

     Type Name

Comments:      

     

Waters Corporation34 Maple Street Milford, Massachusetts01757 USA

Waters, MassLynx, The Science of What’s Possible and Waters Quality Parts are registered trademarks of Waters Corporation. All other trademarks are property of their respective owners. ©2014 Waters Corporation. Printed in the U.S.A.

2535/2545/2555 QGM PERFORMANCE MAINTENANCE PROTOCOL 14 OF 14