biological indicators and sterility assurance level

64
Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY [email protected] Biological Monitoring

Upload: others

Post on 05-Oct-2021

31 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Biological Indicators and Sterility Assurance Level

Prof. Duygu PERÇİN, MD Department of Clinical Microbiology

Erciyes University Faculty of Medicine, Kayseri-TURKEY

[email protected]

Biological Monitoring

Page 2: Biological Indicators and Sterility Assurance Level

Learning Objectives

The attandee will be able to Differantiate the biological indicators Describe the use and interpretation of BI Critique the reliability of the short incubation time

of rapid readout indicators Describe and debate the Sterility Assurance Level

concept Recognise the impact of excessive condensation

on sterilization efficacy Analyse the factors effecting sterilization efficacy

Page 3: Biological Indicators and Sterility Assurance Level

Routine control of sterilization: Why?

To detect PROBLEMS of a process IN A RELIABLE WAY and IN TIME by controlling one or more variables of a sterilization process

Page 4: Biological Indicators and Sterility Assurance Level

Problems A sterilizer malfunction that could account for

failure to achieve the endpoint A change in the product and/or sterile barrier

system A change in loading density A change in container/configuration Delayed or inappropriate sterilizer calibration

and/or routine maintenance Wrong sterilizer process Inappropriate handling of chemical indicator Changes in the utilities supplied to the sterilizer that

could materially affect cycle execution (pressure, flow rate, non-condensable gases in the steam supply, etc.)

Page 5: Biological Indicators and Sterility Assurance Level

ISO 17665-Routine control

Page 6: Biological Indicators and Sterility Assurance Level

Routine monitoring and control

Page 7: Biological Indicators and Sterility Assurance Level

ISO 17665

Page 8: Biological Indicators and Sterility Assurance Level

Biological qualification of a sterilization process

A sterility assurance level (SAL) of at least 10−6 should be demonstrated.

When sterilizing heat-stable materials an overkill approach is normally utilized

Acceptable results for three consecutive cycles of either half cycle or full cycle approach are required for each type of load

The biological indicators used in testing should contain heat resistant spores, such as Geobacillus stearothermophilus spores and should comply with applicable standards

ISO 17665

Page 9: Biological Indicators and Sterility Assurance Level

Biological challenge

ISO 17665

Page 10: Biological Indicators and Sterility Assurance Level

When to use BI

In the first cycle of sterilizer after installation

After every big repair of sterilizer

Routinely once in a week In each cycle including implant

Page 11: Biological Indicators and Sterility Assurance Level

Biological monitoring of LTS methods

According to ISO 11135 chemical indicators can not be used as the sole means of establishing sterilization process using ETO

The only way to prove efficiency of gas concentration and exposure time on biological death is biological monitoring

Every cycle of LTS methods must be monitored with BI and load must be released according to the BI result

Parametric release with physical and chemical parameters is limited

Page 12: Biological Indicators and Sterility Assurance Level

New biological indicators

3M Attest Rapid Readout BI 1295 & Autoreader 490H

Routine monitoring of vaporized hydrogen peroxide sterilization processes in STERRAD® NX and 100NX systems

New FDA approval

Page 13: Biological Indicators and Sterility Assurance Level

If a BI is positive

Check the indicator with negative and positive controls

Check the cycle parameters Check the load Check the packaging material (esp. For ETO

and H2O2) Check the sterilizer Inform IC team for the items if they are

used on the patients

Page 14: Biological Indicators and Sterility Assurance Level

First generation indicators

Spore suspensions inoculated on paper strips (log5-6)

These strips are placed within a sterile liquid medium under aseptic conditions following the sterilization cycle

Microbial growth = any turbidity in the medium

Negative result requires an incubation period of 24-168 hours

Page 15: Biological Indicators and Sterility Assurance Level

Second generation indicators

Contain the medium required for the growth of the spores together with the paper strips containing spores (log5-6)

No risk of contamination during the inoculation of the medium

Growth is seen within 24-48 hours by way of the visible change of color of the medium due to the change of pH as a result of growth

Page 16: Biological Indicators and Sterility Assurance Level

Third generation indicators

Rapid readout indicators Log 5-6 Capable of giving the results within a short

period of incubation like 3-10 hours??? Give CSSDs the opportunity of seeing the

results before the sterilized items are used on patients ???

Page 17: Biological Indicators and Sterility Assurance Level

Rapid readout indicators that are widely used

Attest 1292 Rapid Readout Biologic Indicator (3M, USA) based on showing the alfa-glucosidase enzyme

produced by the active spore with fluorescent radiation

within 3 hours

Bright-Check Rapid Readout Indicators (Etigam, The Netherlands) capable of showing the change of color related to

pH change within 10 hours

Page 18: Biological Indicators and Sterility Assurance Level

Super Rapid Readout Indicators

Page 19: Biological Indicators and Sterility Assurance Level

Question 1:

Is there a relationship between the number of surviving spores on a

biological indicator and

the overall grow-out time or detection time of indicator?

Page 20: Biological Indicators and Sterility Assurance Level

When sterilization fails

Due to the adverse effects of sterilization on the spore structure, germination time can be delayed

Surviving but damaged spores can cause prolonged grow-out times

Rapid readout biological indicators should also be able to detect these cases to ensure the efficacy of sterilization process

Hurst A and Gould GW. The Bacterial Spore 2. 1983

Page 21: Biological Indicators and Sterility Assurance Level

Aim of the study

To evaluate the reliability of the short incubation time of these rapid readout indicators

by using a simulative model including different numbers of spores

Page 22: Biological Indicators and Sterility Assurance Level

Inoculation scheme

Sheep blood agar

Page 23: Biological Indicators and Sterility Assurance Level

Incubation

Page 24: Biological Indicators and Sterility Assurance Level

0

50

100

150

200

250

300

350

400

450

500

550

600

650

2000000 200000 20000 2000 200 20 2

CFU

min

ute

Bright-Cheq3M Attest

Advised incubation time for 3M Attest (180 min)

Advised incubation time for Bright-Cheq (600 min)

Page 25: Biological Indicators and Sterility Assurance Level

Conclusions from the study The less inoculum is in the indicator the longer

detection time is achieved for rapid readout indicators

Minimal incubation time to report negative result for both rapid readout indicators cannot be shorter than 9 hours

spore germination time (2 h)

+ min incubation time (7 h)

Page 26: Biological Indicators and Sterility Assurance Level

ISO 14161 ISO 17665 BIs are used to test

the effectiveness of a given sterilization process and the equipment used, by evaluating microbial lethality according to the concept of sterility assurance level

BIs do not prove if

the load is sterile!

A BI is a microbiological challenge of known resistance that is used to confirm sterilization process lethality at locations on or in product where it is placed.

The physical parameters measured during the sterilization process must be used to verify that the defined sterilization process has been carried out

Page 27: Biological Indicators and Sterility Assurance Level

SAL concept

Page 28: Biological Indicators and Sterility Assurance Level

Elimination of microorganisms

A time-dependent process Influenced by the intensity of treatment the initial microbial contamination level

Effect of some risks in CSSD non condensable gases improper cleaning excessive condensate

Page 29: Biological Indicators and Sterility Assurance Level

Tem

pera

ture

134 oC

121 oC

If we prolong sterilization cycles to

be sure to achieve SAL 10 – 6 PO

OR

cl

eann

ing

EXCE

SSIV

E co

nden

sate

NCG

Do we increase our mistakes with it ???

Question 2:

Page 30: Biological Indicators and Sterility Assurance Level

The heavier our sterilization packs are, the more condensate we are generating at heating up

If this condensate is trapped into sterilization pack it does not gain temperature as fast as surfaces that are not in the condensate

Excessive condensate

Page 31: Biological Indicators and Sterility Assurance Level

Difference in F value

Condensate (green)

Without condensate

(red)

Up to:

-60% …at short cycles

Effect of excessive condensate on sterilization efficacy

Page 32: Biological Indicators and Sterility Assurance Level

Materials and methods Preparation of Geobacillus

stearothermophilus (ATCC 7953) spores from 105 to 109

Inoculation of nuts Steam sterilization Device for generation of

condensate Culture and incubation Microbiological results Electron microscopic evaluation

Page 33: Biological Indicators and Sterility Assurance Level

Spore production (Writz-Conklin staining)

Photo by Dr. Duygu Perçin

Page 34: Biological Indicators and Sterility Assurance Level

Bolted nuts

Photo by Peter Kozin

Page 35: Biological Indicators and Sterility Assurance Level

Bolted nuts

Spore inoculation

bolt

washer

nut

Page 36: Biological Indicators and Sterility Assurance Level

Steam sterilization aparatus and cycle Steam sterilizer Getinge Ge336c Validated cycle -Temperature 135,5oC -3 transatmospheric pulses

for air removal -Different holding times -Short vacuum drying time

Page 37: Biological Indicators and Sterility Assurance Level

Device for production of condensate during sterilization cycle

Photo by Peter Kozin

Page 38: Biological Indicators and Sterility Assurance Level

Photo by Peter Kozin

Page 39: Biological Indicators and Sterility Assurance Level

Transfer into broth and incubation

Page 40: Biological Indicators and Sterility Assurance Level

Turbidity in broths in 72 hours Photo by Duygu Perçin

Page 41: Biological Indicators and Sterility Assurance Level

Gram staining of turbid broth

Photo by Duygu Perçin

Page 42: Biological Indicators and Sterility Assurance Level

STEP 1: Results of bolted nuts inoculated with 109 spores

Sterilization time

Sample size Cycle (134oC) Growth

3 min 6 correct +

6 condensate +

4 min 6 correct +

6 condensate +

5 min 6 correct +

6 condensate +

Page 43: Biological Indicators and Sterility Assurance Level

STEP 2: Results from bolted nuts with less load and metal plates (2cm2)

Sterilization time Cycle (1340C) Sample size / type / load Growth 3 min Correct 6 / Screws / 106 No

Condensate 6 / Screws / 106 No 3 min Correct 2 / Screws / 107 No

Condensate 4 / Screws / 107 No 4 min Condensate 4 / Screws / 107 No 3 min Correct 6 / Plates / 106 No

Condensate 6 / Plates / 106 No

Page 44: Biological Indicators and Sterility Assurance Level

STEP 3: Effect of condensation and sterilization time on bolted nuts carrying 109 spores

Sterilization time Cycle (134oC) Growth

7 min Correct No

Condensate Growth +

10 min Correct No

Condensate Growth +

18 min Correct No

Condensate Growth +

Page 45: Biological Indicators and Sterility Assurance Level

Tem

pera

ture

134 oC

121 oC

POO

R

clea

nnin

g EX

CESS

IVE

cond

ensa

te

EVEN IF WE PROLONG THE

CYCLE WE ALSO INCREASE OUR

MISTAKES TOGETHER WITH

IT

STERILIZATION EFFICACY AT 134oC; WHAT IS GOING ON?

Page 46: Biological Indicators and Sterility Assurance Level

STEP 4: Effect of inoculum (sterilization in 134oC for 3 min)

Inoculum Cycle Result

24 h 48 h 72 h 105-106-107 Correct No No No

Condensate No No No

108 Correct No No No Condensate No No Yes

109 Correct No Yes Yes

Condensate Yes Yes Yes

Page 47: Biological Indicators and Sterility Assurance Level

G.stearothermophilus before sterilization

Photo by Duygu Perçin

Page 48: Biological Indicators and Sterility Assurance Level

109 condensate

Photo by Duygu Perçin

Page 49: Biological Indicators and Sterility Assurance Level

109 without

condensate

Photo by Duygu Perçin

Page 50: Biological Indicators and Sterility Assurance Level

108 condensate

Photo by Duygu Perçin

Page 51: Biological Indicators and Sterility Assurance Level

No growth

108 without

condensate

Photo by Duygu Perçin

Page 52: Biological Indicators and Sterility Assurance Level

G.stearothermophilus before sterilization

Page 53: Biological Indicators and Sterility Assurance Level

109 condensate

Page 54: Biological Indicators and Sterility Assurance Level

109 without

condensate

Page 55: Biological Indicators and Sterility Assurance Level

108 condensate

Page 56: Biological Indicators and Sterility Assurance Level

Log Reduction

3

Theoretical curve shows fast reduction with app. one decade (90%) every 6 seconds

As far as we could have confirmed, slope of reduction curve of biological indicators and items with simple to sterilize shape, is not affected by excessive condensate phenomena. It seems that slope is identical to theoretical curve.

If instruments with difficult structure are immersed in condensate, it seems that we are unable to sterilize them if bioburden is higher than 108 CFU

108 109

106

1030

6 9 18

Reduction at 134 oC

Page 57: Biological Indicators and Sterility Assurance Level

Conclusions from the study Inoculum has a big effect on sterilization

efficacy impresses the importance of cleaning

Condensation lowers the sterilization efficacy impresses the importance of proper loading of

packs and sterilizer Instrument shape has a big impact on

sterilization efficacy impresses the importance of challenging

structure of instruments and packaging

Page 58: Biological Indicators and Sterility Assurance Level
Page 59: Biological Indicators and Sterility Assurance Level

STERRAD NX (ASP, USA) 28 min standard cycle 1 – 2 – 3 – 4 injections

Page 60: Biological Indicators and Sterility Assurance Level

Results

Page 61: Biological Indicators and Sterility Assurance Level

In conclusion

108 spore concentration is a breakpoint for both steam and H2O2 gas plasma sterilization methods

Theoretical mathematical models are not applicable on high inoculum of microorganisms equal to or more than 108

Biological load difference is huge! 106 = 1.000.000 107 = 10.000.000 108 = 100.000.000 109 = 1.000.000.000

SAL concept is questionable…

Impact of BI (~106 spores) to approve SAL concept is even more questionable!

Page 62: Biological Indicators and Sterility Assurance Level

If you have positive BI it means you might be in trouble!

If you have negative BI it doesn’t mean that

you might be at ease!

It will be of value in sterility assurance only - if it is used and interpreted correctly,

- if the user takes appropriate action in response to the results

Page 63: Biological Indicators and Sterility Assurance Level

References Perkins JJ. Growth and death of microorganisms. In: Perkins JJ, ed. Principles and

Methods of Sterilization in Health Sciences. 2nd ed. Illinois: Charles C Thomas; 1969: 56-62.

NEN-EN-ISO 14161. Sterilization of health care products - Biological indicators - Guidance for the selection, use and interpretation of results; (2008).

ISO 17665. Sterilization of health care products - Moist heat- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices, (2006).

ISO 11138. Sterilization of health care products – Biological Indicators, (2006).

CDRH, FDA Guide for Validation of Biological Indicator Incubation Time. Do. 98984, (2007).

J.R. Gillis, G.A. Mosley, J.B. Kowalski, G. Krushefski, P.T. Nirgenau, K.J. McCauley. Understanding Biological Indicator Grow-Out Times. Pharmaceutical Technology, 34, (2010).

Percin D, Kozin P, Renders W. The impact of excessive condensate on the sterility assurance level. Central Service 2015; 1: 40-43.

Page 64: Biological Indicators and Sterility Assurance Level