Microscope
-
8/10/2019 Best Microscope Ppt Prepared ...
2/134
2
-
8/10/2019 Best Microscope Ppt Prepared ...
3/134
3
DEFINITION
- A microscope (Greek: micron= small and scopos= to look)
- MICROSCOPE:Is an instrument for viewing objects that
are too small to be seen by the naked eye- Microscope must accomplish three tasks:
*Magnification*Resolution
*Contrast
-
8/10/2019 Best Microscope Ppt Prepared ...
4/134
4
MAGNIFICATION
Degree of enlargement
No of times the length, breadth or diameter, of an
object is multiplied.
Which of these images
would be viewed at ahigher power of
magnification?
-
8/10/2019 Best Microscope Ppt Prepared ...
5/134
5
RESOLUTION
Ability to reveal closely adjacent structural
details as separate and distinct
LIMIT OF RESOLUTION (LR): The min distance between two visible bodies at
which they can be seen as separate and not in contact with each other
Types of microscope Resolving power
Compound Microscope 200 nanometers
Scanning Electron Microscope 10 nanometers
Transmission Electron Microscope 0.2 nanometers
-
8/10/2019 Best Microscope Ppt Prepared ...
6/134
Contrast
- Giving the details visible to the eye, camera, or other imagingdevice
-
8/10/2019 Best Microscope Ppt Prepared ...
7/134
7
HISTORICALBACKGROUND
2ndCentury BC: Ptolemy describes a stick that appears to bend in water
-
8/10/2019 Best Microscope Ppt Prepared ...
8/134
8
1stCentury: Romans experiment with glass and observe objects larger whenviewed through glass
12thCentury: Salvino DArmante made first eye glass
1590: Hans Janssen and his son Zacharias Janssen,mount two lenses in atube to create first compound microscope
-
8/10/2019 Best Microscope Ppt Prepared ...
9/134
9
1609: Galileo Galilee develops compound microscope with concave and convexlens
1665: Robert Hookes book Micrographia officially documents observationsthrough the microscope
-
8/10/2019 Best Microscope Ppt Prepared ...
10/134
10
1674: Anton van Leeuwenhoek achieves greater magnification allowing him toobserve bacteria. Generally credited with bringing the microscope to the
attention of biologists.
-
8/10/2019 Best Microscope Ppt Prepared ...
11/134
IS FIRST
Safety Biologica l co ntamination: Gloves
Grounded electrical connection (ideal)
Immersion oil in eyes
Location & Storage Do not use under vibration
Do not store in high humidity
Avoid intense temperatures
Keep clean
-
8/10/2019 Best Microscope Ppt Prepared ...
12/134
Wear Personal Protective Equipment
Always check PPE for defects or tears before using If PPE becomes torn or defective remove it and replace with new
equipment
Do not reuse disposable equipment
Dispose of contaminated PPE properly
Wash hands immediately after removing PPE Treat all blood and bodily fluids as if they are contaminated
Wash thoroughly during cleanup and decontamination
Properly dispose of all contaminated material
-
8/10/2019 Best Microscope Ppt Prepared ...
13/134
The following symbols are found on the microscope. Study the meaning of the symbols
and always use the equipment in the safest possible manner
-
8/10/2019 Best Microscope Ppt Prepared ...
14/134
TYPES OF MICROSCOPE
- Simple microscope
- Compound microscope
- Electron Microscope
- Phase Contrast
Microscope
- Dark Ground Microscope
- Fluorescent Microscope
- Others
-
8/10/2019 Best Microscope Ppt Prepared ...
15/134
SIMPLE MICROSCOPE
-
8/10/2019 Best Microscope Ppt Prepared ...
16/134
More than five hundred years ago, simple glass magnifiers weredeveloped. These were convex lenses (thicker in the center than the
periphery). The specimen or object could be focused by use of the magnifier
placed between the object and the eye. These simple microscopes, alongwith the cornea and eye lens, could spread the image on the retina by
magnification through increasing the visual angle on the retina.
The image produced by such a magnifier, held close to the observers eye,appears as if it were on the same side of the lens as the object itself. Such an
image, seen as if it were ten inches from the eye, is known as a virtual image.
These magnifiers had severe limitations in specimen positioning, illumination,
lens aberrations, and construction.
Convex Lenses are curved glass used to make microscopes (and glasses etc.)
Convex Lenses bend light and focus it inone spot
-
8/10/2019 Best Microscope Ppt Prepared ...
17/134
Simple magnifier. A simple magnifier uses a single lens system to enlarge the object in
one step
-
8/10/2019 Best Microscope Ppt Prepared ...
18/134
18
COMPOUND
MICROSCOPE
-
8/10/2019 Best Microscope Ppt Prepared ...
19/134
Around the beginning of the 1600s, through workattributed to the Janssen brothers in the Netherlands and
Galileo in Italy, the compound microscope was developed. Inits basic form, it consisted of two convex lenses aligned in
series: an object glass (objective) closer to the object or
specimen, and an eyepiece (ocular) closer to the observers
eyewith means of adjusting the position of the specimen
and the microscope lenses. The compound microscope
achieves a two-stage magnification. The objective projects a
magnified image into the body tube of the microscope and
the eyepiece further magnifies the image projected by the
objectiveFor example
The total visual magnification using a 10Xobjective and a 15Xeyepiece
is 150X
-
8/10/2019 Best Microscope Ppt Prepared ...
20/134
Compound magnifier. In the compound microscope, the intermediate image formed by
the objective and tube lens is enlarged by the eyepiece.
-
8/10/2019 Best Microscope Ppt Prepared ...
21/134
Parts of Microscope
-
8/10/2019 Best Microscope Ppt Prepared ...
22/134
i. Optical component
ii.Mechanical componentiii.Electrical component
-
8/10/2019 Best Microscope Ppt Prepared ...
23/134
Optical component
1. Its basic function is to gather the light passing through the specimenand then to project an accurate, real, inverted IMAGE of the specimen
up into the body of the microscope.
2. Other related functions of the objective are simply to magnified the
specimen with a given magnification rate i.e 4x,10x ,40x and 100x
magnification power
OBJECTIVE LENS
-
8/10/2019 Best Microscope Ppt Prepared ...
24/134
Scan - 4X
Low Power - 10X
High Power - 40X
Oil immersion - 100X
TYPES
http://hjs.geol.uib.no/marinemicro/images/image70.gif -
8/10/2019 Best Microscope Ppt Prepared ...
25/134
OIL IMMERSION OBJECTIVE
- Highest magnification
- Oil prevents refraction of light outwards and
allows it to pass straight in to objective
25
GLASS
OIL
A
B
C
D
E
G
F
FBEG - OILABCD - AIR
-
8/10/2019 Best Microscope Ppt Prepared ...
26/134
Using the 100x immersion objective
1. Focus on the specimen using all objectives,
starting from the lowest-power to the
highest-power objective
2. Before engaging immersion objective, place
a drop of oil onto the specimen.3. Engage the immersion objective
Make sure oil is free of bubbles
4. Use the fine adjustment knob to bring the
specimen into focus
5. After use, remove oil from objective front
lens by wiping with gauze moistened with
70% alcohol
-
8/10/2019 Best Microscope Ppt Prepared ...
27/134
Working distance
Objective
Magnification
Working
Distance4X 18.5
10X 10.6
40X 0.6
100X 0.13
-
8/10/2019 Best Microscope Ppt Prepared ...
28/134
-
8/10/2019 Best Microscope Ppt Prepared ...
29/134
-
8/10/2019 Best Microscope Ppt Prepared ...
30/134
-
8/10/2019 Best Microscope Ppt Prepared ...
31/134
31
EYE PIECE
1. Its basic function is to look at the
focused, magnified real image
projected by the objective.
2. They are usually 10x,15x,40x and45x
- TYPES
(a) Monocular(Huygenian)
(b) Binocular (Ramsden)
(c) Trinocular (Compensating)
http://hjs.geol.uib.no/marinemicro/images/image71.gif -
8/10/2019 Best Microscope Ppt Prepared ...
32/134
Whats the power of this lens?
To calculate the power of magnification, multiply the power of the
ocular lens by the power of the objective, e..g.: 10x40=400 times
What are the powers of
magnification for eachof
the objectives we have
on our microscopes?
Fill in the table onyour worksheet.
-
8/10/2019 Best Microscope Ppt Prepared ...
33/134
-
8/10/2019 Best Microscope Ppt Prepared ...
34/134
Comparing Powers of Magnification
Which of these images
would be viewed at a
higher power ofmagnification?
We can see better details with higher the powers of
magnification, but we cannot see as much of the image.
10x 40x
-
8/10/2019 Best Microscope Ppt Prepared ...
35/134
.
SUBSTAGE CONDENSER
1. Its basic function is to gather the light coming from the light source and to
concentrate(focus) that light in a collection of parallel beams onto the
specimen.
-
8/10/2019 Best Microscope Ppt Prepared ...
36/134
COLLECTOR LENS.
This lens is placed in front of the light source. Its function is to project an
image of the light source onto the plane of the condensers aperturediaphragm. In some instruments a diffusion or frosted filter is placed just
after the collector lens (side closer to the specimen) in order to provide
more even illumination.
FIRST SURFACE MIRROR
Its function is to reflect the light coming from the lamp up into the
substage condenser. Just before that mirror (closer to the lamp side)
is another variable diaphragm known as the field diaphragm
-
8/10/2019 Best Microscope Ppt Prepared ...
37/134
TUBE LENS
Its function is to gather the parallel rays of light projected by the objective
and bring those rays to focus at the plane of the fixed diaphragm of the
eyepiece.
FILTERS -Blue, Green,
Heat absorbing filters,Barrier filters.
-
8/10/2019 Best Microscope Ppt Prepared ...
38/134
ii. Mechanical component
Body TubeThe body tube holds the objective lenses
and the ocular lens at the proper distance
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
39/134
Nose Piece
The Nose Piece holds the objective lenses
and can be turned to increase the
magnification
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
40/134
Stage ClipsThese 2 clips hold the slide/specimen in
place on the stage.
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
41/134
CONDENSER HOLDER.
This holder receives any one of several types of condenser
-
8/10/2019 Best Microscope Ppt Prepared ...
42/134
Diaphragm
The Diaphragm controls the amount of light
on the slide/specimen
Turn to let more light in or to
make dimmer.
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
43/134
Arm
Used to support the microscope when
carried. Holds the body tube, nose piece
and objective lenses
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
44/134
Stage
Supports the slide/specimen
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
45/134
Coarse Adjustment Knob
Moves the stage up and down (quickly) forfocusing your image
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
46/134
Fine Adjustment Knob
This knob moves the stage SLIGHTLY to
sharpen the image
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
47/134
Base
Supports the microscope
Diagram
-
8/10/2019 Best Microscope Ppt Prepared ...
48/134
48
PARTS OF COMPOUND MICROSCOPE
-
8/10/2019 Best Microscope Ppt Prepared ...
49/134
-
8/10/2019 Best Microscope Ppt Prepared ...
50/134
-
8/10/2019 Best Microscope Ppt Prepared ...
51/134
iii. Electrical component
ILLUMINATION -
Lamp, sunlight, battery
operated lamp, 60 W bulb,
Quartz halogen light.
It is light source
* step-down transformer
* switch* Rheostat
* lamp (halogen or incandescent)
* fuse
* Electronic sub system
-
8/10/2019 Best Microscope Ppt Prepared ...
52/134
Light Source
Projects light upwards through thediaphragm, the specimen and the lenses
Some have lights, others have mirrorswhere you must move the mirror to reflect
light
Diagram
Operational Principle
-
8/10/2019 Best Microscope Ppt Prepared ...
53/134
53
Operational Principle
Light
Knowledge of the behavior of light and the effects resulting when light passes from
air through a glass convex lens and out into air again is fundamental to the
understanding of image formation. When light passes from air into a convex lens,
the speed of light is slowed. The various colors, differing in wave length, are slowed
at different rates (dispersion). This bending (refraction) effect differs for different
colors. Those rays which strike the central area of the lens at a perpendicular
emerge unrefracted. Light passing through the other parts of the convex lens are
refracted or bent. The blue rays are bent more than thegreen rays, more than
the red rays.
When white light passes through convex lenses of objectives, eyepieces, or
condensers, two main kinds of aberrations may occur; chromatic aberration and/or
spherical aberration. These aberrations can be corrected in the design of the
lenses.0
-
8/10/2019 Best Microscope Ppt Prepared ...
54/134
-
8/10/2019 Best Microscope Ppt Prepared ...
55/134
-
8/10/2019 Best Microscope Ppt Prepared ...
56/134
-
8/10/2019 Best Microscope Ppt Prepared ...
57/134
nstallation and commission
-
8/10/2019 Best Microscope Ppt Prepared ...
58/134
-
8/10/2019 Best Microscope Ppt Prepared ...
59/134
-
8/10/2019 Best Microscope Ppt Prepared ...
60/134
-
8/10/2019 Best Microscope Ppt Prepared ...
61/134
DETAILED OBSERVATION PROCEDURE
-
8/10/2019 Best Microscope Ppt Prepared ...
62/134
DETAILED OBSERVATION PROCEDURE
-
8/10/2019 Best Microscope Ppt Prepared ...
63/134
-
8/10/2019 Best Microscope Ppt Prepared ...
64/134
-
8/10/2019 Best Microscope Ppt Prepared ...
65/134
-
8/10/2019 Best Microscope Ppt Prepared ...
66/134
-
8/10/2019 Best Microscope Ppt Prepared ...
67/134
-
8/10/2019 Best Microscope Ppt Prepared ...
68/134
-
8/10/2019 Best Microscope Ppt Prepared ...
69/134
-
8/10/2019 Best Microscope Ppt Prepared ...
70/134
-
8/10/2019 Best Microscope Ppt Prepared ...
71/134
-
8/10/2019 Best Microscope Ppt Prepared ...
72/134
-
8/10/2019 Best Microscope Ppt Prepared ...
73/134
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
74/134
Preventive Maintenance
Daily Preventative Maintenance
Objective (after using oil immersion lens)
Body
Yearly Preventative Maintenance
Microscope frame
Observation Tubeoptical axis
Observation Tubeleft/right axis
Observation Tuberevolving axis
Observation Tubeparfocality
As needed Grease replacement parts
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
75/134
General Maintenance
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
76/134
General Maintenance
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
77/134
Cleaning the eyepieces
Never rub the lens surface strongly
Do not use the same lens tissue to clean more than one lens
Do not moisten lens tissue with excessive amounts of cleaning solution
When cleaning with tweezers, be careful not to protrude its tip from the lens tissue
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
78/134
General Maintenance -
Cleaning the Objective Lens
*When cleaning oil immersion objective, first wipe objective with a dry lens tissue
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
79/134
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
80/134
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
81/134
Preparing for Inspection
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
82/134
Inspection Sheet
Use the CX21 inspection sheet
and check the electrical unit,
mechanical and optical
performance
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
83/134
Inspection: Checking for Dust
Close aperture iris diaphragm
When aperture iris diaphragm is closed
the depth of focus increases and the dirt
position is brought into focus
Observe a specimen through the
eyepiece
Image influence caused by dirt on each component
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
84/134
General Maintenance
Inspection: Checking for Dust
Note: If dirty portion
can not be
identified, it is
assume that the
internal lens andprism are
contaminated.
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
85/134
General Maintenance
Inspection: Checking Lenses
General Maintenance
-
8/10/2019 Best Microscope Ppt Prepared ...
86/134
General Maintenance
Inspection: Checking Lenses
General Maintenance:
-
8/10/2019 Best Microscope Ppt Prepared ...
87/134
General Maintenance:
Checking Left/Right Alignment Place the cross eyepiece into the right
sleeve
Place a slide with an image with a
recognizable reference point (e.g. center
of a spiral), secure it on stage
Align the reference point in the center of
the cross of the eyepiece (looking
through right eyepiece) by moving the
stage with the control knob
General Maintenance:
-
8/10/2019 Best Microscope Ppt Prepared ...
88/134
General Maintenance:
Checking Left/Right Alignment
Cont.
General Maintenance:
-
8/10/2019 Best Microscope Ppt Prepared ...
89/134
General Maintenance:
Checking Left/Right Alignment
Cont.
General Maintenance:
-
8/10/2019 Best Microscope Ppt Prepared ...
90/134
General Maintenance:
Checking Left/Right Alignment Switch the cross eyepiece to the left
sieve
If the crosshair is aligned with the same
reference point, the left/right alignment iscorrect
If the crosshair is not aligned, do NOT move
the stage. The left eyepiece sleeve needsadjusted.
General Maintenance: Fixing
-
8/10/2019 Best Microscope Ppt Prepared ...
91/134
General Maintenance: Fixing
Left/Right Alignment
General Maintenance: Fixing
-
8/10/2019 Best Microscope Ppt Prepared ...
92/134
General Maintenance: Fixing
Left/Right Alignment
General Maintenance: Fixing
-
8/10/2019 Best Microscope Ppt Prepared ...
93/134
General Maintenance: Fixing
Left/Right Alignment2`------------
---
Common Failures
-
8/10/2019 Best Microscope Ppt Prepared ...
94/134
Electrical/Power Failure Light does not turn on when power switch is turned on
Light bulb burnt out
Objective not placed in alignment
Dirty components
Eyepiece
Objectives
Filters
General User Error
Not in focus
Brightness not adjusted properly
Objective not rotated fully into secured position
T bl h ti
-
8/10/2019 Best Microscope Ppt Prepared ...
95/134
Troubleshooting
What is troubleshooting?
A logical, systematic process that is able to determine "what" happened, "why" ithappened and a method to develop effective fixes for the "why" it happened.
Why do we Troubleshoot?
To prevent the equipment and machinery from repeat incidents and failures
5 St f T bl h ti
-
8/10/2019 Best Microscope Ppt Prepared ...
96/134
5 Steps of Troubleshooting
oPreparation
oStep 1 Observation
oStep 2 Define Problem Area
oStep 3 Identify Possible CausesoStep 4 Determine Most Probable Cause
oStep 5 Test and Repair
oFollow-up
Preparation
-
8/10/2019 Best Microscope Ppt Prepared ...
97/134
Preparation
oBefore you begin to troubleshoot any piece ofequipment, you must be familiar with safety rules andprocedures for working on electrical equipment.
Be Safe!
Turn off power, ground yourself, and use gloves when
appropriate
oNext, gather information regarding the equipment andthe problem.
Be sure you understand how the equipment is designed to
operate it.Operation or equipment manuals and drawings are great
sources of information and are helpful to have available.
If there are equipment history records, you should review
them to see if there are any reoccurring problems
-
8/10/2019 Best Microscope Ppt Prepared ...
98/134
Step 1: Observation
oWhat Happened?oMost faults provide obvious clues as to their cause.
oThrough careful observation and a little bit of reasoning,most faults can be identified as to the actual componentwith very little testing.
oWhen observing malfunctioning equipment, look forvisual signs of mechanical damage such as indications
of impact, chafed wires, loose components or parts lying
in the bottom of the cabinet.
oLook for signs of overheating, especially on wiring, relay
coils, and printed circuit boards
oDo not forget to use your other senses
-
8/10/2019 Best Microscope Ppt Prepared ...
99/134
when inspecting equipment.
The smell of burnt insulation is something you would not
miss. Listening to the sound of the equipment operating may
give you a clue to where the problem is located.
Checking the temperature of components can also help
find problems, but be careful while doing this, some
components may be alive or hot enough to burn you.
oPay particular attention to areas that wereidentified either by past history or by the person that
reported the problem.
-
8/10/2019 Best Microscope Ppt Prepared ...
100/134
Here is a note of caution! Do not let these mislead you, pastproblems are just thatpast problems, they are not
necessarily the problem you are looking for now.
Also, do not take reported problems as a fact, always checkfor yourself if possible. The person reporting the problem
may not have described it properly or may have made their
own incorrect assumptions.
Step 2: Define the Problem Area
-
8/10/2019 Best Microscope Ppt Prepared ...
101/134
Step 2: Define the Problem Area
oAt this stage you apply logic and reasoning to yourobservations to determine the problem area of the malfunctioning
equipment.
Often times when equipment malfunctions, certain parts of
the equipment will work properly while others not.
oThe key is to use your observations (from step 1) to rule outparts of the equipment or circuitry that are operating properly and
not contributing to the cause of the malfunction. You should continue to do this until you are left with only the
part(s). If faulty, it could cause the symptoms that the
equipment is experiencing.
o To help you define the problem area you should have a
-
8/10/2019 Best Microscope Ppt Prepared ...
102/134
oTo help you define the problem area you should have aschematic diagram of the circuit, in addition to your noted
observations.
oStarting with the whole circuit as the problem area, take eachnoted observation and ask yourself, "what does this tell me about
the circuit operation?".
If an observation indicates that a section of the circuitappears to be operating properly, you can then eliminate it
from the problem area.
As you eliminate each part of the circuit from the problem
area, make sure to identify them on your schematic. This will help you keep track of all your information.
Step 3: Identify Possible Causes
-
8/10/2019 Best Microscope Ppt Prepared ...
103/134
Step 3: Identify Possible Causes
oOnce the problem area(s) have been defined, it is necessaryto identify all the possible causes of the malfunction.
This typically involves every component in the problem
area(s).
oIt is necessary to list (write down) every fault which couldcause the problem no matter how remote the possibility of itoccurring.
Use your initial observations to help you do this.
During the next step you will eliminatethose which are
not likely to happen.
Step 4 Determine the Most Probably
-
8/10/2019 Best Microscope Ppt Prepared ...
104/134
Step 4 Determine the Most Probably
Cause
oOnce the list of possible causes has been made, it is thennecessary to prioritize each item as to the probability of it being the
cause of the malfunction.
The following are some rules of thumb when prioritizing
possible causes.
Although, it may seem to be possible for two components to
fail at the same time, it is not very likely. Start by looking for
one faulty component as the culprit.
oThe following list shows the order in which you should check
-
8/10/2019 Best Microscope Ppt Prepared ...
105/134
components based on the probability of them being defective:
1.Components that burn out or have a tendency to wear out: i.e.
mechanical switches, fuses , relay contacts, and light bulbs.(Remember, fuses burn out for a reason. You should find out why
before replacing them.)
2.Coils, motors, transformers, and other devices with windings:
These usually generate heat and with time, can malfunction.3.Connections: especially screw or bolted type. Over time these
can loosen and cause a high resistance. In some cases this
resistance will cause overheating and eventually will burn open.
Connections on equipment that is subject to vibration, are
especially prone to coming loose.4.Defective wiring: Pay particular attention to areas where the
wire insulation could be damaged causing short circuits. Do not
rule out incorrect wiring, especially on a new piece of equipment.
Step 5 Test and Repair
-
8/10/2019 Best Microscope Ppt Prepared ...
106/134
p p
oOnce you have determined the mostprobable cause, you must either prove it to bethe problem or rule it outby
inspection/observation or by using test
instruments
oTest Instruments: used to helpnarrow the problem area and
identify the problem component.
specialized instruments designed
to measure various behaviors ofspecific equipment
General instruments (i.e. multimeters)
oImportant Rule: when taking meter readings, predict what the
-
8/10/2019 Best Microscope Ppt Prepared ...
107/134
p g g p
meter will read before taking the reading.
Use the circuit schematic to determine what the meter will
read if the circuit is operating normally.If the reading is anything other than your predicted value, you
know that this part of the circuit is being affected by the fault.
oDepending on the circuit and type of fault, the problem areaas defined by your observations, can include a large area of the
circuit. It creates a very large list of possible and probable
causes.
Use a divide and eliminate approach to eliminate parts of the
circuit from the problem area.The results of each test provides information to help you
reduce the size of the problem area, until the defective
component is identified.
-
8/10/2019 Best Microscope Ppt Prepared ...
108/134
oOnce you have determined the cause of the faulty operation
of the circuit, replace the defective component.
oAfter replacing the component, you must test operate allfeatures of the circuit, to be sure you have replaced the
proper component and that there are no other faults in
the circuit.
It can be very embarrassing to tell the customer that youhave repaired the problem only to have him find another
problem with the equipment just after you leave.
Follow-UpNot an official step of the troubleshooting process but it
-
8/10/2019 Best Microscope Ppt Prepared ...
109/134
Not an official step of the troubleshooting process, but it
should be done once the equipment has been repaired and
put back in service.
oTry to determine the reason for the malfunction. Did the component fail due to age?
Did the environment the equipment operates in cause
excessive corrosion? Are there wear points that caused the wiring to short out?
Did it fail due to improper use?
Is there a design flaw that causes the same component to
fail repeatedly?
oThrough this process further failures can be minimized.
oMany organizations have their own follow-up documentationand processes
o If you determine there was human performance (or user
-
8/10/2019 Best Microscope Ppt Prepared ...
110/134
y p (
error) involved in the equipment issue, the troubleshooter
needs to step out of the equipment analysis role and
interview people who have interacted with the equipment or
machinery in question.
o May need to properly train users on equipment usage.
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
111/134
-
8/10/2019 Best Microscope Ppt Prepared ...
112/134
-
8/10/2019 Best Microscope Ppt Prepared ...
113/134
-
8/10/2019 Best Microscope Ppt Prepared ...
114/134
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
115/134
Troubleshooting
Electrical Unit
If lamp does not turn on when power switch
is turned on, there is a problem with the
electrical unit, unless
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
116/134
g
Electrical Unit
After verifying the light bulb is not the issue,
check that the microscope is plugged in and
that outlet has power(use another deviceyou know to be working to validate this)
Verify the chord shows no obvious damage
Verify the fuse is intact
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
117/134
If the previous steps revealed no culprit
The problem is most likely the circuit board
itselfReplacing individual components is not feasible,
entire board needs replaced
If an identical microscope exists that is not
usable for another reason, harvesting its circuitboard is an option. See the manual for
instructions on removing the circuit board
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
118/134
g
Coarse/Fine adjustment knobs do not
turn smoothly
New grease needs applied to knobs
If using a different microscope, turn to
manual for knob-removal instructions
Instructions for CX-21 provided next
Troubleshooting:Focus Knobs
-
8/10/2019 Best Microscope Ppt Prepared ...
119/134
g
Dissassemble fine
adjustment knob
assembly (A) and fine
adjustment knob(B) by
turning them in arrow
directions
Remove the spring
washer (C) and washer
(D)
Remove the fine shaft
mount (E) with a
spanner (wrench) whileholding the coarse
adjustment knob (F)
Troubleshooting: Focus Knobs
-
8/10/2019 Best Microscope Ppt Prepared ...
120/134
Remove the coarse adj.
knob (G) by turning it
counterclockwise while
holding the coarse adj.
knob (F).
Remove the tension
knob (H) by turning it
counterclockwise.
(The washer (I) is
attached to tension
knob (H) with grease.)
Troubleshooting: Focus Knobs
-
8/10/2019 Best Microscope Ppt Prepared ...
121/134
Pull out the tension
ring (J) while holding
the convex part using
pliers
Place grease wherever
indicated in figure
Reassemble them in
the reverse order
Troubleshooting
-
8/10/2019 Best Microscope Ppt Prepared ...
122/134
Stage falls down spontaneously or an in-
focus image quickly moves out of focus
Cause: tension on coarse adjustment knob
is too loose, needs to be adjusted
Troubleshooting:Coarse Knob Tension
-
8/10/2019 Best Microscope Ppt Prepared ...
123/134
On the CX-21, the tension of the
coarse adjustment knob can be
adjusted by inserting a flat head
screwdriver into (2). Rotating
clockwise increases tension,
rotating counterclockwise
reduces tension
If the stage is falling on its own,
increase the tension.
If the knob is hard to turn and
grease has already been
applied, reduce the tensionslightly
Troubleshooting: Stage Movement
-
8/10/2019 Best Microscope Ppt Prepared ...
124/134
If the stage does not move smoothly in
the X or Y directions, the tension of the
X/Y wires must be adjusted
Instructions for CX-21 are provided in
following slides, for other models see
manual
roubleshooting: Stage Movement
-
8/10/2019 Best Microscope Ppt Prepared ...
125/134
Troubleshooting:Stage Movement
-
8/10/2019 Best Microscope Ppt Prepared ...
126/134
X-tension adjustment
Troubleshooting:Stage Movement
-
8/10/2019 Best Microscope Ppt Prepared ...
127/134
g g Y-tension adjustment
Troubleshooting:Backlash
-
8/10/2019 Best Microscope Ppt Prepared ...
128/134
CARE OF THE MICROSCOPE
-
8/10/2019 Best Microscope Ppt Prepared ...
129/134
129
- Handling
- Proper storage
- Care of Lenses
- Care of mechanical part
- Care of electrical part and lamp
are and maintenance of your microscope
-
8/10/2019 Best Microscope Ppt Prepared ...
130/134
The microscope has 3 major enemies:
1. Dust
Protect the microscope from dust- make sure you use a dust
cover.
NEVER CLEAN THE OPTICS WITH A DRY TISSUE /SWAB/ COTTON BUD OR YOUR SHIRT SLEEVE! This willscratch the lens and cause irreparable damage. This isbecause the lens is usually coated with an anti-glare layer(bluish in colour). Try to remove dust from the optics byblowing gently on them, or use an air blower.
SOLUTION: make sure you use a dust cover when themicroscope is not in use or is being stored.
-
8/10/2019 Best Microscope Ppt Prepared ...
131/134
Source: www.funsci.com/fun3_en/dust/dust.htm
usty microscope image
2. Humidity
http://www.funsci.com/fun3_en/dust/dust.htmhttp://www.funsci.com/fun3_en/dust/dust.htm -
8/10/2019 Best Microscope Ppt Prepared ...
132/134
Humidity (warm and wet) offers the perfect environment forfungus to grow.
Fungal growth will occur primarily on the optics of themicroscope. Once you have a fungal infection on yourmicroscope it is difficult to remove and may permanentlydamage your optics.
Using a plastic dust jacket (like the one supplied with thescope) is a bad idea.
SOLUTION: Buy some thick calico and use the plastic cover as a
template to make dust covers that are washable and canbreathe. This prevents condensation inside the cover and thedevelopment of ideal conditions for fungal growth (warm andwet)
3. Water
-
8/10/2019 Best Microscope Ppt Prepared ...
133/134
Microscopes are electrical items. Water
and electricity are a dangerous mix!
Storing your microscope when it is wet
may lead to corrosion of parts and hygiene
problems. See also enemy number 2! If wet mounts are used, the stage is to be
cleaned and dried before returning the
scope to storage. The optics must be kept
dry to prevent fungal attack.
Summary
-
8/10/2019 Best Microscope Ppt Prepared ...
134/134
y
Safety is the first priority: microscopes are used to
examine biological specimens, gloves are imperative
Microscopes have many moving parts which must be
properly positioned for clear viewing
Always check that the microscope is being used
properly before assuming there is another problem
Dirt and dust can reduce image quality, when trying to
deduce debris location