for xcellenc E e Form No. CAT-MX-50-023-R03 Specifications Options: AD-8121 B Dot Matrix Compact Printer AX-MX-31 Sample Pan (0 85mm x 100pcs) AX-MX-41 Calibration Mass (20g, equivalent to OIML class F1) AX-MX-43 Certified Temperature Calibrator MARKETED BY: ...Clearly a Better Value The contents of this catalogue are subject to change due to modifications and/or other reasons. Specifications are subject to change for improvement without notice. Measurement Method Max Sample Weight Capacity Weight Resolution Moisture Content Display Moisture Content Accuracy over 1g (Standard Deviation) over 5g Heating Technology Drying Temperature (1°Cincrement) Memory of Measurement Programs Measurement Programs Measurement Mode Heating Mode Display Type Interface Data Memory Function Operating Temperature GLP/GMP/ISO Self Check Function Communication Software Sample Pan Size Power Physical Dimention/Weight Standard Accessories Sample Pans - 20, Pan Handles -2, Spoon, Test Sample (30g of Sodium Tartrate Dihydrate), CD-ROM (WinCT-Moisture), Glass Fiber Sheet, RS-232C Cable, Display Cover, Dust Cover, Instruction Manual, Quick Reference Card, Power Cable, Fuse MX-50 0.001g 0.01% / 0.1% 0.10% 0.02% Halogen lamp (Straight type, 400 Watt max, 5000 hours) 50-200°C 20 sets Standard Mode/Automatic Mode/Quick Mode/Timer Mode/Manual Mode Moisture content(Wet or Dry base) /Dry content/Ratio/Weight Standard / Quick / Step /Ramp Large VFD RS-232C standard 100 5-40°C (41-104°F) less than 85%RH Available Standard WinCT-Moisture standard Ø85mm AC 200V to 240V (1.5A), 50/Hz, Approx. 400W 215(W) x 320(D) x 173(H) / Approx. 6kg 51g 400W straight halogen lamp heating system with SRA filter and SHS weighing technology AD-8121B Dot Matrix Compact Printer WinCT-Moisture consists of RsTemp software to determine the heating temperature and RsFig software for graphics. 1. Example of measurement using Rs Temp software to determine the heating temperature Automatically changes the heating temperature (by your setting increments and interval time) within a range of 50 C - 200 C. From the rate of change in moisture over time measurement it can determine the optimum heating temperature for the sample. 0 0 2. Example of measurement using RsFig graphics software Can be moisture rate aganist time when the heating temperature is changed as well as showing the result of repeated measurement at a certain temperature. From the graph, the highest possible heating range for the sample and the quickest measuring time can be determined. Soybean flour Sodium tartrate dihydrate Corn starch Sodium tartrate dihydrate

Upload: phammien

Post on 14-May-2018

215 views

Category:

Documents


1 download

TRANSCRIPT

for xcellencE e

For

m N

o. C

AT-M

X-5

0-02

3-R

03

Specifications

Options:AD-8121 B Dot Matrix Compact Printer

AX-MX-31 Sample Pan (0 85mm x 100pcs)

AX-MX-41 Calibration Mass (20g, equivalent to OIML class F1)

AX-MX-43 Certified Temperature Calibrator

MARKETED BY:

...Clearly a Better Value

The contents of this catalogue are subject to change due to modifications and/or other reasons.

Specifications are subject to change for improvement without notice.

Measurement Method

Max Sample Weight Capacity

Weight Resolution

Moisture Content Display

Moisture Content Accuracy over 1g

(Standard Deviation) over 5g

Heating Technology

Drying Temperature (1°Cincrement)

Memory of Measurement Programs

Measurement Programs

Measurement Mode

Heating Mode

Display Type

Interface

Data Memory Function

Operating Temperature

GLP/GMP/ISO

Self Check Function

Communication Software

Sample Pan Size

Power

Physical Dimention/Weight

Standard Accessories Sample Pans - 20, Pan Handles -2, Spoon, Test Sample (30g of Sodium Tartrate

Dihydrate), CD-ROM (WinCT-Moisture), Glass Fiber Sheet, RS-232C Cable, Display Cover,

Dust Cover, Instruction Manual, Quick Reference Card, Power Cable, Fuse

MX-50

0.001g

0.01% / 0.1%

0.10%

0.02%

Halogen lamp (Straight type, 400 Watt max, 5000 hours)

50-200°C

20 sets

Standard Mode/Automatic Mode/Quick Mode/Timer Mode/Manual Mode

Moisture content(Wet or Dry base) /Dry content/Ratio/Weight

Standard / Quick / Step /Ramp

Large VFD

RS-232C standard

100

5-40°C (41-104°F) less than 85%RH

Available

Standard

WinCT-Moisture standard

Ø85mm

AC 200V to 240V (1.5A), 50/Hz, Approx. 400W

215(W) x 320(D) x 173(H) / Approx. 6kg

51g

400W straight halogen lamp heating system with SRA filterand SHS weighing technology

AD-8121BDot Matrix Compact Printer

WinCT-Moisture consists of RsTemp software to determine the heating temperature and RsFig software for graphics.

1. Example of measurement using Rs Temp software to determine the heating temperatureAutomatically changes the heating temperature (by your setting increments and interval time) within a range of 50 C - 200 C.

From the rate of change in moisture over time measurement it can determine the optimum heating temperature for the sample.

0 0

2. Example of measurement using RsFig graphics software

Can be moisture rate aganist time when the heating temperature is changed as well as showing the result of repeated

measurement at a certain temperature. From the graph, the highest possible heating range for the sample and the quickest

measuring time can be determined.

Soybean flour Sodium tartrate dihydrate

Corn starch Sodium tartrate dihydrate

Fast and uniform heating with halogen lamp and innovative SRA technology

High Repeatability

Standard win CT-Moisture for real-time graph displaying

Sodium Tartrate Dihydrate comes as standard for accuracy checking

Calibration of the heater temperature

Straight halogen lamp and uniquely designed SRA (Secondary Radiation Assist) filter gives shortermeasurement time, thanks to fast and uniform heating

with SHS (Super Hybrid Sensor) featured as the weight sensor, ultra accurate moisture contentdetermination is possible based on high precision weighing of even a small sample

Win CT-Moisture is an original software application designed to display a graph of moisture content rate change while measuring with a connected PC

Sodium Tartrate Dihydrate is a chemical material that has stable moisture content of +0.3 15.66%

and thus is best to use for accuracy check to maintain the reference value of the analyzer

With the temperature calibration (optional), calibration result can be output in the format that conforms to GLP, GMP, ISO

Memory functionAccording to sample up to 20 suitable measurement conditions can be stored and recalled , which saves timeand prevents the user from making a mistake when settingFor measurement result, up to 100 data can be stored and output at once

Five measurement programsFive choices of measurement programs, Standard, Automatic, Quick, Timer, and Manual Mode are providedStandard Mode : just measurement accuracy, HI, MID or LO needs to be setAutomatic Mode : Ends measurement when moisture content changes at a rate less than the set rate

0Quick Mode : Begins heating samples at 200 C for 3 minutes, then is the same as Automatic ModeTimer Mode : Continues measurement for a set duration of time (1~60mins.: by 1min, 60~480mins.: by 5mins.)Manual Mode : STOP button should be pressed to end measurement (Max. heating time: 480mins.)

Selectable Heating ModeChoose the heating mode from standard, quick, step and ramp heating modes for the most suitable measurement

Clear and esay-to-see, large VFD displayMeasurement, setting value, change in moisture content, action status, data number and other

-0.1,

Select the best moisture analyzerfor your application-Test with The Best

Progress Window for heating checkHeating process can be checked through translucent window

Self check functionDefect check function is available along with temperature control

Quick Reference cardA convenient operating guide is installed at the bottom of the analyzer

Standard RS-232CBi-directional communication with a PC or connects directly to a printer

Conformity to GLP, GMP, GCP, and ISO with date/time, ID, Calibration data and check records outputsData output for daily record management

With our WinCT-Moisture software, data measured by the Moisture Analyzers can

be easily displayed on your computer.

Effective for determine measurement conditions such as heating temperature,

and useful in reducing the time needed for measurement and improving accuracy.

Displays moisture rate changes over time in a graph (Rs Fig)

Measures moisture rate in a minimum time with excellent accuracy

Automatically determines the most suitable heating conditions in a short time (Rs Temp)

Shows sample data summary

Displays changes in moisture rate in real time

Heats at the highest temperature without change the physical properties of the sample and provides measurements with good repeatability

Automatically changes the heat applied by the set increments and interval time within a range of

0 050 C-200 C. From the moisture rate change over the time can determine the most suitable heating temperature in one time measurement (*patent pending)

Provides a data summary for the sample with the results of moisture rate change for the representative material's measurementCalculates measurement data statisticsSaves the recorded data as a CSV fileDetermine other changes to the sample material in addition to moisture rate

Moisture rate

Change in moisture rate