AWAKE Electron Spectrometer

Design

Simon Jolly4th September 2012

Page 2: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Spectrometer Specifications• Wakefield accelerated electrons ejected

colinear with proton beam: need to separate the 2 and measure energy of electron beam only.

• Must be able to resolve energy spread as well as energy: spectrometer must accept a range of energies, probably 0-5 GeV.

• Simple conceptual layout:– Dipole mounted at plasma exit separates

proton and electron beams.– Scintillator screen 4 m downstream of dipole

intercepts electron beam ONLY.– Dispersion gives energy-dependent position

spread on screen.– Scintillator imaged by intensified CCD camera

viewing upstream face of scintillator screen.04/09/12 Simon Jolly, University College

London2

Page 3: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Spectrometer Layout

04/09/12 Simon Jolly, University College London

3

Dipole separates protons from electrons AND puts energy-dependent spread in beam.

Electrons intercepted by scintillator screen, imaged by high speed iCCD camera.

Screen nominally 4 m from dipole.

Dipole

Proton beam

Electron beam

Screen

iCCD Image

Low energy

High energy

iCCD Camera

Page 4: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Browne-Beuchner Spectrometer (Alexey)

04/09/12 Simon Jolly, University College London

4

C. P. Browne and W. W. Buechner, Rev. Sci. Instrum. 27, 899 (1956).

Removes angular convolution from energy measurement, but needs ~3 m magnet for 5 GeV at 1.5 T…

Focusing in broad energy range is achieved via pole face curvature.

Page 5: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Spectrometer Beam Trajectories (Alexey)

04/09/12 Simon Jolly, University College London

5

Page 6: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

With Quadrupole Doublet (Alexey)

04/09/12 Simon Jolly, University College London

6

Page 7: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

With Fringe Fields (Alexey)

04/09/12 Simon Jolly, University College London

7

Page 8: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Proton Microbunches After Plasma (Alexey)

04/09/12 Simon Jolly, University College London

8

0 m after plasma

2 m after plasma

6 m after plasma

Transverse Distributions

-60 mm < s < -40 mm

Page 9: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

CERN 1 m Dipole (Edda, Alexey)

04/09/12 Simon Jolly, University College London

9

Page 10: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

AWAKE Spectrometer Layout• Edge of scintillator screen is aligned

with dipole coils (position will depend on resolution).

• Screen mounted at 45 degrees to beam axis.

• Camera is 4m from centre of screen, mounted at 90 degrees to beam axis.

• Camera shown in horizontal bending plane.

• Camera can also be mounted vertically, directly above screen, with screen tilted at 45 degrees to vertical as well as 45 degrees to beam axis. Dipole to screen distance remains unchanged (independent of screen-camera orientation).

04/09/12 Simon Jolly, University College London

10

4 m

1.6 m

1 m

Camera

Dipole

Screen

Page 11: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

AWAKE Spectrometer Layout 2

Simon Jolly, University College London

2 m

1.6 m

1 m

Camera

Dipole

Screen

04/09/12 11

Page 12: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Layout Questions• Do we need 2 dipoles?

– Difficult to get good low/high field from single dipole.– Weak dipole (Browne-Buechner?) upstream of strong dipole analyses 1-

20 MeV electrons.– Reduces available longitudinal space, increases complexity: need 2nd

screen & camera + larger vacuum vessel.• Can we include quadrupole doublet?

– Alexey’s simulations show improvement in energy range on screen with doublet.

– Is there enough longitudinal space?• How close should screen be to dipole?

– Closer gives better range, less effect of emittance, higher light output.– Greater distance improves resolution, requires smaller dipole.– Probably needs to be closer due to low beam intensity, large energy

spread.• Optimum camera location?

– Camera not radiation hard (waiting on specs from manufacturer), so needs to be well shielded ie. far from beamline.

– Intensifier enhances low light output from screen (single photon detection) but more light into camera gives better signal-to-noise ie. close to beamline.

– Mount camera above beamline? Increases distance from beamline while reducing distance to screen.

04/09/12 Simon Jolly, University College London

12

Page 13: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Spectrometer Camera

04/09/12 Simon Jolly, University College London

13

• Camera already purchased.• Andor iStar 340T iCCD

camera:– 2048 x 512 total pixels

(1850 x 512 active with 25 mm intensifier).

– 13.5 um pixels.– Gen-2 W-AGT P43 intensifier,

gated at 7 ns.– Nikon F-mount lens mount.– 16-bit readout, 150 ke- pixel

full well.– http://www.andor.com/scientifi

c_cameras/istar_iccd_camera/340-intensified-sensor/

• Camera setup:– DAQ and control software.– Data analysis software to give

online measurement of witness beam energy spectrum.

– Mechanical design of camera support system for tests and installation in SPS tunnel (comes later).

Page 14: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Scintillator Choices• Scintillator requirements:

– High light output (very low intensity electron beam).– Radiation hard (survive proton beam?).– Vacuum safe (screen inside vacuum vessel).

• Scintillator choices from 3 different fields:– Accelerators (survive higher current):

• Al2O3 (“ruby” powder, used in LHC BTV).• YAG:Ce (high light output).• Quartz (very rad hard, low light output).

– Detectors (high light output):• CdWO4/PbWO4/ZnWO4 (CMS ECal).• CsI(Tl) (very high light output, not rad hard).• BGO/LYSO (more rad hard for lower light).• Plastics (less rad hard, cheap).

– Plasma Wakefield (large area):• Lanex (large screens, vacuum safe?)• P43/P46/P47 phosphors (x-ray phosphors).

• Need highest number of photons per solid angle to give maximum screen-to-camera distance.

• Manual lens probably means no zoom (prime lens), so resolution fixed by mechanical layout.

04/09/12 Simon Jolly, University College London

14

Page 15: AWAKE Electron Spectrometer Design Simon Jolly 4 th September 2012

Conclusions• Baseline design is evolving rapidly: this is good!• Looks like optimum will be 45 degree screen close to

dipole:– Alexey’s simulations show energies already well

separated at dipole exit.– Get as close to dipole exit as possible while still

imaging whole screen (camera view not obscured by dipole yoke).

– Light path needs to be absolutely light tight (light from plasma exit?) but not necessarily vacuum tight.

• Need input on scintillators to know what possibilities are:– Why is Lanex never used in HEP?– Need specs on light output to set maximum camera-to-

screen distance.• Spectrometer design will probably evolve from all

constraints ie. beamline space, radiation, vacuum vessel, then optimised within those constraints.