Types of Sample Matrix Interferences in Ion Chromatography
• Samples with components that will damage the column.
• Samples that are too acidic for cation analysis.
• Samples that are too basic for anion analysis.
• Samples that contain high concentrations of interfering ions.
a Grace company
Problems with Common Sample Preparation Techniques
Dilution and Filtration are not suitable when:
• The analytes are present at trace levels.
• The analytes differ greatly in concentration.
• Interfering compounds and analytes are both soluble.
a Grace company
Solid Phase Extraction (SPE) Cartridges for Ion Chromatography
•Stable from pH 2 to 13.
•Low IC interference levels.
•Requires only 2 - 3mL of samples.
a Grace company
The Alltech Maxi-Clean™ IC Cartridge Construction
Polypropylene Housing
Female LuerInlet
Polyethylene Frits
Packing Material
Male LuerOutlet
a Grace company
Types of Maxi-Clean™ IC SPE Cartridges:
•Maxi-Clean™ IC-RP
•Maxi-Clean™ IC-Ag
•Maxi-Clean™ IC-Ba
•Maxi-Clean™ IC-OH
•Maxi-Clean™ IC-H
•Maxi-Clean™ IC-Na
•Maxi-Clean™ IC-Chelate
a Grace company
General Procedure for Using Maxi-Clean™ IC Cartridges
1. Precondition the cartridge by passing 5-10mL of IC grade water (use 5mL of methanol followed by 5mL of Water for Maxi-Clean™ IC-RP).
2. Load the sample at a rate of 1mL/min or less. (Adjust the sample size to use 50% or less of the device’s capacity).
3. Discard the first 1mL eluate.
4. Collect and analyze the remaining eluate.
a Grace company
Product Specifications
Maxi-Clean™ IC-RP
Packing: PolystyreneParticle Size: 300µmRetains: Hydrophobic components by a
reversed-phase mechanismApplications: Removal of hydrophobic organic components
a Grace company
Maxi-Clean™ IC-RP Removes Organic Contaminants
Trace Anions in Aniline
Original Sample After Cleanup with Maxi-Clean™ IC-RP
Column: Anion/S, 250 x 4.1mm Mobile Phase: 4mM KHP pH 4.5Flowrate: 2.0mL/minDetector: Alltech Conductivity
1. Chloride2. Nitrate3. Sulfate
5351
a Grace company
Average % Recovery of Ions with Maxi-Clean™ IC-RP Cartridge
90929496
98100102104
Ions
a Grace company
Sample Preparation by Precipitation
Maxi-CleanTM IC-Ba• For removal of sulfate and phosphate prior to
analysisResin-2-Ba+2 + M+2SO4
-2 Resin-2-M+2 + Ba+2SO4-2
Maxi-CleanTM IC-Ag• For removal of halides prior to analysis
Resin--Ag+ + M+X- Resin--M+ + Ag+X-
M=Sample cations X= Cl-, Br-, I-
a Grace company
The Solubility of Barium and Silver Salts in Aqueous Solutions
Compound Solubility (g/100mL) Temperature (°C)
BaF2 0.12 25BaCl2 37.5 26Ba(NO2)2 67.5 20Ba(NO3)2 8.7 20BaHPO4 0.01 - 0.02 N/ABa2SO4 2.2 x 10-4 25AgF 182 15.5AgI 2.8 x 10-7 25AgCl 8.9 x 10-5 10AgNO2 0.155 0AgBr 3.7 x 10-4 100AgNO3 122 0Ag2SO4 0.57 100
a Grace company
Maxi-Clean™ IC-Ba
Product Specifications
Packing: Strong cation exchanger in the Ba2+ formParticle Size: 106µmCapacity: 0.8/2.4 meq.Retains: Sulfate by precipitation Applications: Removal of sulfate before the analysis of
other anions
a Grace company
Maxi-Clean™ IC-Ba Removes Sulfate
Original Sample After Cleanup with Maxi-Clean™ IC-BA
Column: Anion/R, 250 x 4.1mm Mobile Phase: 5mM p HBA, pH 8.5Flow rate: 1.5mL/min.Detector: Alltech Conductivity
5348
Anions in Battery Acid
1. Fluoride2. Chloride3. Nitrite4. Bromide5. Nitrate6. Phosphate7. Sulfate
a Grace company
Average % Recovery of Ions with Maxi-Clean™ IC-Ba Cartridge
0102030405060708090
100110
Ions
a Grace company
Maxi-Clean™ IC-Ag
Product Specifications
Packing: Strong cation exchanger in Ag+ formParticle Size: 106µmCapacity: 0.8/2.4 meq.
Retains: Halides by precipitationApplications: Removal of halides before analysis of
other anions
a Grace company
5347
Maxi-Clean™ IC-Ag Removes Chloride
Column: 250 x 4.1mm Packing: Anion/RMobile Phase: 5mM p HBA, pH 8.5Flow rate: 1.5mL/min.Detector: Alltech Conductivity
Original Sample After Cleanup with Maxi-Clean™
Nitrite in Brine
1. Chloride2. Nitrite
1
2
a Grace company
Average % Recovery of Ions with Maxi-Clean™ IC-Ag Cartridge
0102030405060708090
100110
Ions
a Grace company
Maxi-Clean™ Cartridges to Adjust Sample pH
•Baseline disturbances and/or system peaks result when sample pH differs from eluant pH.
•Strong anion exchange resin in the hydroxide form removes anions and neutralizes an acidic sample.
•Cation exchange resin in the hydrogen form removes cations and neutralizes an alkaline sample.
a Grace company
Maxi-Clean™ IC-OH
Product Specifications
Packing: Strong anion exchanger in OH- formParticle Size: 75µm
Capacity: 0.6/1.8 meq
Retains: AnionsApplications: Adjusting sample pH before
cation analysis
a Grace company
5352
Maxi-Clean™ IC-OH Neutralizes Acid
Column: Cation/R, 100 x 43.1mm Mobile Phase: 3.2mM nitric acidFlowrate: 1.0mL/minDetector: Alltech ConductivityOriginal Sample After pH
Neutralization with Maxi-Clean™ IC-OH
Monovalent Cations in 1M Nitric Acid
1. Sodium 2. Potassium
a Grace company
Recovery of Ions with Maxi-Clean™ IC-OH Cartridges
0102030405060708090
100110
Ions
a Grace company
Maxi-Clean™ IC-H
Product Specifications
Packing: Strong cation exchanger in H+ form Particle Size: 106µm
Capacity: 0.8/2.4 meq. Retains: CationsApplications: Adjusting sample pH before
anion analysis
a Grace company
5349
Maxi-Clean™ IC-H Neutralizes Hydroxide
Column: Anion/R, 150 x 4.1mm Mobile Phase: 5mM pHBA, pH 8.5Flowrate: 1.5mL/min.Detector: Alltech Conductivity
Original Sample After pH Neutralization with Maxi-Clean™ IC-H
Trace Anions In Caustic Solution
1. Fluoride2. Bicarbonate3. Chloride4. Nitrite5. Bromide6. Nitrate7. Phosphate8. Sulfate
a Grace company
Recovery of Ions with Maxi-Clean™ IC-H Cartridges
0102030405060708090
100110
Ions
a Grace company
Maxi-Clean™ IC-Na
Product Specifications
Packing: Strong cation exchanger in Na+ form Particle Size: 106µm
Capacity: 0.8/2.4 meq.
Retains: Cations (exchanges cations for sodium)
Applications: Removal of cations from sample without changing sample pH
a Grace company
5813
Maxi-Clean™ IC-Na+ Isolates Iron (III)
Column: Alltech Universal Anion, 150 x 4.6mm Mobile Phase: 2mM EDTA, disodium saltFlowrate: 1.0mL/min.Detector: UV at 330nm*Problem with IC-H+--> the precipitation of H4EDTA
Analysis of Iron (III)Procedures:1. Precondition IC-Na+ cartridge with 5mL water.2. Load the sample to preconcentrate iron (III) present.3. Wash with 10mL water.4. Elute the iron (III) with 5mL of 25mM EDTA, disodium salt.5. Analyze the eluant for iron (III).
1. Iron (III)
a Grace company
Maxi-Clean™ IC-Chelate
Product Specifications
Packing: Styrene-divinylbenzene copolymer containing iminodiacetate ionParticle Size: 225µmCapacity: 0.2/0.6 meq.Retains: Polyvalent metal ions (exchanges cations for sodium)Applications: Removal of polyvalent metal ions from sample
a Grace company
Benefits of Alltech Maxi-Clean™ IC Cartridges
• Remove interfering components from the sample matrix.
• Little or no loss of sample.
• Easy, quick, and convenient.
• Disposable.