automatic programmable melting point apparatus model

4
Automatic Programmable Melting Point Apparatus Model : µThermoCal50 | Make : ANALAB | HSN Code : 90278090 Supplied with Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube, Sample Tapping Rod, Two Melting Standards 49ºC and 238 ºC, Forceps, Power Cord, Operational & Instruction Manual with SOP, Calibration Test Certicate (without Traceability Document) and Warranty Card. www.analab.co.in

Upload: others

Post on 18-Dec-2021

11 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Automatic Programmable Melting Point Apparatus Model

Automatic Programmable Melting Point Apparatus

Model : µThermoCal50 | Make : ANALAB | HSN Code : 90278090

Supplied with Melting Capillary One End Sealed (Pack of 100 Nos.), Boiling Tube,

Sample Tapping Rod, Two Melting Standards 49ºC and 238 ºC, Forceps, Power Cord,

Operational & Instruction Manual with SOP, Calibration Test Certicate (without

Traceability Document) and Warranty Card.

www.analab.co.in

Page 2: Automatic Programmable Melting Point Apparatus Model

Technical Specication:

www.analab.co.in

Temperature Controller : Microcontroller Based

Detection Principal of Melting Point : Automatic by Silicon Photodiode

Heating Media : By Aluminum Block

Temperature Range : +5°C above ambient to 350°C

Readability : 0.1°C

Accuracy of Temperature : A) +5°C above ambient to 200°C………… ± 0.2°C

B) 200°C to 350°C …………................…..± 0.5°C

Overheating Protection : In-built - Set by user in method parameter

Heating Before EMP* : Automatic heating depending upon temperature difference

Heating Rate : 0.2, 0.5, 1.0, 2.0, 3.0 & 5.0 ºC/min -Selectable

Display : 20 x 4 Line Alphanumeric Backlit LCD

Detection of Melting : A) Melting Point B) Melting Range

Melting Point Average Reading : Display of average reading for three same sample

Type of Detection : Auto / Manual

Detection of Boiling Point : Manually by viewing in TFT Display – One Boiling Tube

Magnication : 5X

Keypad : Membrane waterproof polycarbonate soft touch keypad compliance

to UL/ROHS

Viewing of Melting Capillary : Camera with TFT Display for viewing Melting Capillary and Boiling

Tube in CAPillaryVIEW

Method : 40 method for Melting & 10 for Boiling sample with View, Print &

Delete facility

Calibration : Seven Calibration Standards

(As per Pharmacopeia AND / OR Customer's known standards)

Melting Capillary Ports : Three - one sample at a time

Glass Melting Capillary : OD: 1.4 to 1.6mm | ID: 1.0mm| Length: 100mm | One End Sealed

Sample Filling Height : 3mm

Password Protection : Date & Time, Calibration Mode, Delete Stored Analysis Data and

Restore Factory Set Calibration Data

Maximum Heating Time : Around 6 minutes from 50°C to 350°C

Maximum Cooling Time : Around 6 minutes from 350°C to 50°C

Temperature Sensor : Duplex PT-100, One for internal used and One for calibration purpose

by external source

Page 3: Automatic Programmable Melting Point Apparatus Model

www.analab.co.in

Calibration Data Storage : Non Volatile Memory which can store, print & view calibration date &

time, company name, standard name with temperature, actual

reading of standard temperature, calibration done by, instrument

serial number, instrument ID, model number etc.

Searching of data : With record number, batch number, Date and by scrolling

Report Formation : As per GLP Compliance

Analysis Data Storage : More than 750 analysis data can be stored with corresponding

calibration data with batch number/ID, users name, sample name,

instrument serial number, instrument ID, model number, calibration

point, calibration date & time, analysis date & time, upper & lower

calibration standard based on sample temperature, heating rate,

melting range (MP1# & MP2#), signature etc. Secondly, analysis

data can print directly from analysis mode without saving analysis

data in instrument.

Printer Interface : Centronics Parallel Port / Serial (RS-232 C) Port

Printer : 40 / 80 Column Dot Matrix Printer

(Printer have to procure optionally)

Power Supply : 230 VAC ±10%, 50 Hz

Page 4: Automatic Programmable Melting Point Apparatus Model

Speci�cation & design are subject to change without prior notice for our continuous development.

www.analab.co.in

Plot No. 361, Sayajimarket,

N. H. No. 8 By-Pass,

At & Post : Sayajipura,

Vadodara - 390019, Gujarat, India.

+91 265 2514404, 9714503393

[email protected]

www.analab.co.in, www.analabscienti�c.com

CIN: U33110GJ2005PTC046555

AN ISO 9001 : 2015 CERTIFIED CO.

Optional

1. IQ / OQ / PQ / DQ Documents

Note:

a) Please provide Customer's Name & Address prior to dispatch.

b) To perform onsite documentation, visit charge will be extra depending upon place of Installation.

c) Traceability documents along with installation report.

d) Standards / Chemicals / Buffers required to perform documents are in customer's scope.

e) We don't send documents by email.

2. Traceability documents

3. Calibration Standard Kit for Melting Point Apparatus - (ANALAB Certied)

Consist of 07 Melting Standards as specied bellow, Melting Capillary One End Sealed 100 mm Long

(Pack of 100) – 02 Pack, Boiling Tube – 02 Nos., Sample Tapping Rod, Forceps, Glass for Heating Block

– 02 Nos., Glass for Camera – 02 Nos.

Melting Standards :

a) Benzophenone - Melting Range : 47 - 49 ºC, Capacity : 5gm

b) Vanillin - Melting Range : 81 - 83 ºC, Capacity : 5gm

c) Benzoic Acid - Melting Range : 121.5 - 123.5 ºC, Capacity : 5gm

d) Sulfanilamide - Melting Range : 164 - 167 ºC, Capacity : 5gm

e) Caffeine - Melting Range : 235 - 238 ºC, Capacity : 5gm

f) Phenolphthalein - Melting Range : 258 - 262 ºC, Capacity : 5gm

g) Potassium Nitrate - Melting Range : 334 – 336 ºC, Capacity : 5gm

Note: Melting Range / Melting Point specied on standard bottle is re-pack based on the values of batch

number specied by manufacturer and their COA.

4. GMP Enclosure made up of Stainless Steel

5. Keypad Screen Guard