Page 1: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Copolymer Characterization

by

Multi-Detector

Size-Exclusion Chromatography

André M. StriegelNational Institute of Standards & Technology

Page 2: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Objectives

• Introduction to macromolecular distributions.

• Determine chemical heterogeneity in copolymer using

quadruple-detector SEC.

• Examine and correct effect of n/c heterogeneity on

calculated M averages, distribution, etc.

• Extend method to “dual-absorbing” copolymers.

• Apply quintuple-detector SEC to show

physicochemical basis for conformational changes

across MMD.

Page 3: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Styrene Polystyrene

n can take on a range

of values

Therefore, so can M

Page 4: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

10 20 30 40 50 60 70 80 900

20

40

60

80

100

120

Dif

fere

nti

al

We

igh

t F

rac

tio

n

Molar Mass

Molar Mass

(MM)

Short-Chain

Branching

(SCB)

Long-Chain

Branching

(LCB)

Polyelectrolyte

Charge

(+/-)

+

+

++

+

++

+

++

+

Chemical

Heterogeneity

(CH)

Chemical

Composition

Distribution

(CCD)

MM

SCB

LCB

+/-

CH

Ch

em

ica

l Co

mp

os

ition

Differential Weight Fraction

CCD

Striegel, Yau, Kirkland, & Bly, Modern Size-Exclusion Liquid Chromatography, 2nd ed. (2009)

Page 5: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Macromolecular distributions: Measurement and End-Use Effects

Macromolecular Representative end-use Separation method used for

Property properties affected determination

Molar mass Elongation, tensile strength, adhesion SEC, FFF, HDC, TGIC, CEC, SFC, MALDI-MS,

rheology

Long-chain branching Shear strength, tack, peel, crystallinity SEC-MALS, SEC-VISC, rheology, enzymology

Short-chain branching Haze, stress-crack resistance, crystallinity SEC-IR, SEC-NMR, TREF, CRYSTAF, enzymes

Crosslinking Gelation, vulcanization, surface roughness SEC-MALS, SEC-VISC, rheology

“Architecture” Flow modification, diffusion, encapsulation SEC-MALS-QELS-VISC

Tacticity Crystallinity, anisotropy, solubility SEC-NMR, TGIC, LCCC,

Chemical composition Morphology, miscibility, solubility GPEC, TGIC

Chemical heterogeneity Toughness, brittleness, biodegradability SEC-spectroscopy/spectrometry, LCCC, PFC

Chemical composition Mechanical properties, blending, plasticization 2D-LC (e.g., SEC-GPEC)

vs. molar mass

Block sequence Dielectric properties, reactivity, miscibility SEC-spectroscopy, 2D-LC (e.g., PFC-SEC)

Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS

Polyelectrolyte charge Flocculation, transport, binding of metals SEC-conductivity

Particle size Packing, drag, friction, mixing FFF, HDC, PSDA, sieving

Striegel, Yau, Kirkland, & Bly, Modern Size-Exclusion Liquid Chromatography, 2nd ed. (2009)

Page 6: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Measuring

Chemical Heterogeneity

in a

Random Copolymer

Page 7: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Macromolecular Distributions:Measurement & End-Use Effects

• Property: Chemical Heterogeneity

0

20

40

60

80

100

120

Dif

fere

nti

al W

eig

ht

Fra

cti

on

Molar Mass

Molar Mass

(MM) MM

CHChemical

Heterogeneity

(CH)

Relative ab

un

dan

ce

Page 8: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Macromolecular Distributions:Measurement & End-Use Effects

• Property: Chemical Heterogeneity

• Definition: Change in average chemical

composition of a copolymer as f(M),

uncorrected for local polydispersity.

Page 9: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Macromolecular Distributions:Measurement & End-Use Effects

• Property: Chemical Heterogeneity

• Affects: Melting point, gas permeation,

conductivity, interfacial fracture energy

• Measured using: SEC with physico-

chemical detection methods

Page 10: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Measuring Chemical Heterogeneity

• Random copolymer of styrene (St) and methyl methacrylate (MMA):

Is relative ratio of St:MMA constant as a function of molar mass?

• Styrene absorbs at 260 nm, but methyl methacrylate does not.

• UV detector @ 260 nm will respond only to styrene content in polymer.

• Differential refractometer will respond to both styrene and methyl methacrylate

in polymer.

Page 11: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Measuring Chemical Heterogeneity

DRIDifferential refractometer

UVUltraviolet

MALSMulti-angle light scattering

VISCDifferential Viscometer

Styrene (but not MMA)

absorbs @ 260 nm

Both St & MMA

elicit DRI response

Page 12: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Relative Detector Response:

DRI vs. UV

20 22 24 26

0.0

0.2

0.4

0.6

0.8

1.0

UV Signal Conc. of St only

Re

lati

ve

De

tec

tor

Sig

na

l (a

rbit

rary

un

its

)

Retention volume (mL)

DRI Signal Conc. of both St and MMA

iDRI

iUV

i

PSDRI

PSUV

PMMAPS

i

PS

PMMA

i

i

S

SZ and constant

S

SF where

c

n

c

nZ

c

nF

c

nZ

St

,

,

,

,

%100)(%

Haidar Ahmad & Striegel, Anal. Bioanal. Chem., 396 (2010) 1589

Page 13: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

5.0x104

1.0x105

1.5x105

2.0x105

2.5x105

3.0x105

3.5x105

0.5

1.0

1.5

2.0

2.5

3.0

18

20

22

24

26

28

30

Dif

fere

nti

al W

eig

ht

Fra

cti

on

(d

Wf /d

M)

Molar Mass, M (g/mol)

MMD

% Styrene

Weig

ht P

erc

en

t Sty

ren

e (%

St)

Measuring Chemical Heterogeneity

Haidar Ahmad & Striegel, Anal. Bioanal. Chem., 396 (2010) 1589

Page 14: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Chemical Heterogeneity

Translates Into

n/c Heterogeneity

c

nn

c

n

c

0

0lim

n: refractive index of solution

n0: refractive index of solvent

c: concentration of analyte in solution

Page 15: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

n/c versus %PS

y = 0.0011x + 0.0852

R2 = 0.9951

0.08

0.12

0.16

0.2

0 25 50 75 100% PS

dn

/dc

(mL

/g)

dn/dc vs % PS

Linear (dn/dc vs % PS)

75% PS Block

copolymer

100% PS

100%

PMMA

Alternating

copolymers

(50% PS)

20% PS random

copolymer

25% PS Block &

Random copolymers

Page 16: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

What Does This Mean?

• Molar mass M determined by multi-detector SEC is

influenced by chemical identity of analyte.

• This influence enters through the n/c-dependence

of the static light scattering (SLS) and differential

refractive index (DRI) detector signals:

• M averages and distribution determined from ratio

of these signals at each SEC elution slice i:

cc

nM SSLS

2

c

ncSDRI

i

i

i

i

i

i

i

i

i

c

nM

cc

n

cc

nM

S

S

2

DRI,

,SLS

Page 17: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

What Does This Mean?

• Therefore, chemical heterogeneity translates into

n/c heterogeneity.

• In turn, heterogeneity in n/c translates into a bias

in the molar mass averages and distribution

calculated via SEC/MALS/DRI .

• (This bias will also manifest itself in SEC using

LALS, peak-position calibrant-relative calibration, or

universal calibration).

Page 18: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

We can correct for this:

• First, correct n/c for chemical heterogeneity:

iMMA

PMMA

iSt

PSicorrected

wc

nw

c

n

c

n,,

,

Page 19: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

y = 0.0011x + 0.0852

R2 = 0.9951

y = 0.001x + 0.087

0.08

0.12

0.16

0.2

0 25 50 75 100% PS

dn

/dc

(mL

/g)

dn/dc vs % PS

Linear fit

Predicted

n/c = [(0.191 wSt) + [0.087 (1 - wSt)]

n/c versus %PS

Page 20: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

We can correct for this:

• First, correct n/c for chemical heterogeneity:

• Then, use this value to correct calculated M:

iMMA

PMMA

iSt

PSicorrected

wc

nw

c

n

c

n,,

,

icorrected

duncorrecteiduncorrecteicorrected

c

n

c

n

MM

,

,,

Page 21: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Chemical Heterogeneity Corrected MMD

5.0x104

1.0x105

1.5x105

2.0x105

2.5x105

3.0x105

3.5x105

0.5

1.0

1.5

2.0

2.5

3.0

18

20

22

24

26

28

30

Dif

fere

nti

al W

eig

ht

Fra

cti

on

(d

Wf /d

M)

Molar Mass, M (g/mol)

Uncorrected MMD

Corrected MMD

% Styrene

We

igh

t Pe

rce

nt S

tyre

ne

(%S

t)

Uncorrected Corrected

Mn (g/mol) 170,000 11,900 175,000 11,900

Mw (g/mol) 200,000 800 209,000 800

Mz (g/mol) 234,000 2,100 250,000 2,100

Mw/Mn 1.18 0.08 1.19 0.07

Error in M depends both

on M and on

n/c difference among

monomers in copolymer

Haidar Ahmad & Striegel, Anal. Bioanal. Chem., 396 (2010) 1589

Page 22: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

22

As % composition

changes…

[(n/c)PS × wSt] +

[(n/c)PMMA × wMMA](n/c)copolymer =

(n/ c)copolymer

changes

SDRI = c (n/c)copolymer

Concentration

changes

ccnMS 2SLS )/(

Molar mass

changes

Intrinsic viscosity

changes c

sp

c

lim

0

][

Solution

conformation

changes

Chemical Heterogeneity:

Effect on Calculated Parameters

Page 23: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Chemical Heterogeneity Also Affects

Calculated Conformation

7x104

105

2x105

3x105

4x105

50

100

150

Intr

insic

Vis

co

sit

y,

[]

(mL

/g)

Molar Mass, M (g/mol)

Uncorrected

Corrected

0.71 0.01

0.60 0.01

105

2x105

3x105

10

15

20

25

Ra

diu

s o

f G

yra

tio

n,

RG (

nm

)

Molar Mass, M (g/mol)

Uncorrected

Corrected

0.64 0.01

0.59 0.01

Mark-Houwink Plot Conformation Plot

Polymer adopts a less extended conformation in solution than originally believed.

Haidar Ahmad & Striegel, Anal. Bioanal. Chem., 396 (2010) 1589

Page 24: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Determining

Chemical Heterogeneity

in Copolymers:

The “Dual-Absorption” Case

Page 25: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

“Dual-Absorption” Copolymers

• In previous case, only one monomer in copolymer absorbed at a given UV wavelength.

• What is both monomers absorb at the same wavelength, however?

• Also, what can adding QELS detection tell us about the copolymer?

• How does chemical heterogeneity affect solution structure?

• What is the underlying physicochemical basis?

Page 26: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Poly(AM-co-DMAM)

H2C CH CH2 CH

C O

NH2

C O

N

( () )x 1-x

Poly(acrylamide-co-N,N-dimethyl acrylamide)

Page 27: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Homopolymer Absorptivities

Absorptivity = 84 ± 1 mL ∙ mg-1 ∙ cm-1 Absorptivity = 1913 ± 1 mL ∙ mg-1 ∙ cm-1

@ 240 nm

Polyacrylamide

(PAM)

Poly(N,N-dimethyl acrylamide)

(PDMAM)

CH2 CH

C O

N

( )n

H2C CH

C O

NH2

( )n

Page 28: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

SEC/MALS/DLS/VISC/UV/DRI

© Wyatt Technology Corporation 2005 - All Rights Reserved

What’s Next…

• Remainder of day

– hands-on experience with hardware and software

– batch lab experiment, real data!

• Summary session bring questions!

Solvent

Reservoir

Pump Injector Thermostatted

SEC columns MALS/DLS

DRI VISC

Fluid connection

Electronic connection

UV

Page 29: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Wt% DMAM at each SEC slice i

%

c

n

c

nSaaZ

aZc

nS

DMAM

PDMAMPAM

r,iR,CopolymePAMPDMAMPDMAM

PAMPDMAM

PAM

r,iR,Copolyme

i 100%

iDRI

iUV

iCopolymerRS

SS

,

,

,,

PDMAM

PDMAMPDMAMR

PDMAM

PDMAM

PDMAMDRI

PDMAMUV

PDMAMa

c

n

Sa

c

n

S

SZ

,

,

,

where

a = absoptivity of constituent

homopolymers

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 30: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Chemical Heterogeneity of

P(AM-co-DMAM)

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

104

105

106

0.1

0.2

0.3

0.4

0.5

0.6

0.7

Molar Mass, M (g mol-1)

Dif

fere

nti

al

We

igh

t F

rac

tio

n

72

74

76

78

80

82

84

86

88

90

92%

N,N

-dim

eth

yla

cry

lam

ide

Page 31: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Polymer Size(s)

• Radius of gyration (a.k.a. rms radius): From MALS + DRI

• Hydrodynamic (a.k.a. Stokes) radius: DLS (a.k.a. QELS) + DRI

• Viscometric radius: MALS + VISC + DRI

2/1

2)(1

1

i

c miG Rrn

R

Ts

BH

D

TkR

6

3/1

10

3

AN

MR

Page 32: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

RG and RH Conformation Plots

Rigid Rod

Random Coil

Hard Sphere

0.60 ≥ α ≥ 0.50

α = 1

α = 0.33

105

106

1

10

100

Rad

ius (

nm

)

Molar Mass (g mol-1)

RH

RG

slope = 0.58

= 0.56

df = 1.8

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 33: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

ρ ( ≡ RG /RH) Parameter

105

106

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

R

G/R

H

Molar Mass, M (g/mol)

Linear Random CoilTransition to More

Extended Coil

PDMAM

homopolymer

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 34: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Mark-Houwink & Conformation Plots

Rigid Rod

Random

Coil

Hard

Sphere

0.80 ≥ a ≥ 0.50

a = 2

a = 0

105

106

1

10

100

10

100

1000

Rad

ius (

nm

)

Molar Mass (g mol-1)

a = 1.07

df = 1.45

a = 0.64

df = 1.8

[]

RH

RG

slope = 0.58

= 0.56

df = 1.8 In

trinsic

Vis

cosity

, [n] (m

L g

-1)

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 35: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Rη/RG Plot

105

106

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

R/R

G

Molar Mass, M (g/mol)

Linear Random CoilTransition to More

Extended Coil

PDMAM

homopolymer

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 36: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

6x104

2x105

4x105

6x105

8x105

106

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

R

Molar mass, M (g/mol)

Poly(St-ran-MMA)

Poly(St-ran-MMA)

Poly(St-alt-MMA)

Poly(St-b-MMA), 75% St

Poly(St-b-MMA), 25% St

PS

PMMA

Hard-sphere limit

Appro

xim

ate

random

coil

ra

nge

RG

R/RG of Homo- and Copolymers

Haidar Ahmad, Striegel, & Striegel, Polymer, 52 (2011) 1268

Page 37: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Lack of Chemical Heterogeneity in

Random S-MMA Copolymers

Random Copolymer 126k

Random Copolymer 186k

Haidar Ahmad, Striegel, & Striegel, Polymer, 52 (2011) 1268

Page 39: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

R/R

G

R/RG

R/RG

Fraction A Fraction B Fraction C

Area under

1074 cm-1

Area under

1074 cm-1

Area under

1074 cm-1< <

Molar mass (g/mol)

Off-Line IR Analysis of SEC “Heart-Cuts”

Page 40: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Persistence Length, Lp

Lpr

Î1

2/1

2

GR

M

1/M

2/1)3(2

3LpL MLM

2/1

3

p

L

L

M

222

31

32/12/1

2

Lp

Lp

p

L

GML

M for ,

M

ML

L

M

R

M

Striegel, Yau, Kirkland, & Bly, Modern Size-Exclusion Liquid Chromatography, 2nd ed. (2009)

Page 41: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Persistence Length of Poly(AM-co-DMAM)

3.0x10-6

4.0x10-6

5.0x10-6

6.0x10-6

7.0x10-6

32

34

36

38

40

42

(M/2LPM

L) = 2.2

Lp = 12 1 nm

ML = 3120 6 g mol

-1 nm

-1

(M/R

G2)1

/2

1/M

R2 = 0.993

Poly(AM-co-DMAM)

Rowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 42: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

104

105

106

0.1

0.2

0.3

0.4

0.5

0.6

0.7

Molar Mass, M (g mol-1)

Dif

fere

nti

al W

eig

ht

Fra

cti

on

72

74

76

78

80

82

84

86

88

90

92

% N

,N-d

imeth

yla

cry

lam

ide

Transition in Polymer Conformation

Extended Coil

(“Alt”-like)

T

r

a

n

s

i

t

i

o

n

R

e

g

i

o

n

Random Coil

(“Homo”-like)

RG ≤ Lp Lp < RG ≤ 2Lp 2Lp < RG

Lp = 12 1 nmRowland & Striegel, Anal. Chem., 84 (2012) 4812

Page 43: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

Info From Multi-Detector SEC(Not a Comprehensive List)

MALS

QELS

VISC

DRI

UV

[]

CC

M-H plot,

R, R/RG

M aves, MMD,

Confor. plot, Lp

CH

CC: Chemical compostion, CH: Chemical heterogeneity, M-H: Mark-Houwink, RG,z/RH,z

Page 44: André M. Striegel · 2017. 2. 17. · quadruple-detector SEC. ... Base pair sequence Genetic code, heredity, mutations Automated DNA sequencing, MALDI-MS Polyelectrolyte charge Flocculation,

CONCLUSIONS

• Chemical heterogeneity in copolymers results in n/c heterogeneity.

• n/c heterogeneity can affect calculated M averages, distributions, etc.

• Appropriate choice of SEC detectors allows accurate characterization of copolymers.

• Can also inform knowledge of physicochemical basis for conformational changes as f(M).

• Method has also been extended to binary blends.