FUNDAMENTALS OF IMAGING
RAD 206
Control of Scatter Radiation
What is focal spot?What is effective focal spot ?Which side of the x-ray tube has better resolution ?Which focal spot size produces better resolution?Which OID produces better resolution?Which Object Thickness produces better resolution?Which Object shape produces better resolution?
OBJECTIVESBegin discussing factors that influence image detail or visibility of detail
• Scatter Radiation, factors affecting it, and there effect on Spatial & Contrast resolution
• Ways to reduce scatter & improve image quality• Primary beam restriction & Grids• Technique adjustments when using grids
WHAT IS CONTRAST
What is scatterWhat is scatter
RELATIONSHIP BETWEEN CONTRAST AND SCATTER
The relationship is INVERSED
IMAGE-FORMING X-RAYSTwo kinds of photons are responsible for the OD and contrast on an image: • Photons that pass through without interacting
and
• Photons that are scattered though Compton.
EFFECT OF SCATTER ON THE IMAGE
Which interaction produces scatter radiation ?
FACTORS CONTRIBUTING TO AN INCREASE IN SCATTER
• Increased kVpkVp (Beam Energy, Compton)
• Increased partpart thickness thickness (e.g. Dabbeh)
• Increased x-ray field sizefield size (opening Collimation to Larger Fields)
Q: what can you do to reduce scatter ?Q: what can you do to reduce scatter ?
1- KVPAt 50 kVp
7979% photoelectric, 21%21% Compton<1% transmission
At 80 kVp 4646% photoelectric, 52% 52% Compton2% transmission
RELATIONSHIP BETWEEN KVP AND CONTRAST
How does increasing kVp affect patient dose?
2- PATIENT THICKNESS
Imaging thick parts of the body results in more scatter radiation than thin parts
IMAGE TEST TOOL
Is patient thickness something the radiographer can control?
IS PATIENT THICKNESS SOMETHING THE IS PATIENT THICKNESS SOMETHING THE RADIOGRAPHER CAN CONTROL? RADIOGRAPHER CAN CONTROL?
Normally, No
But Compression Devices can be used in certain situations which reduces patient thickness and bringing the object closer to the IR, and therefore results in better spatial resolution and contrast resolution and also lesser exposure factors and patient dose
COMPRESSION
• Reduces OID• Improves spatial
resolution• Improves contrast
resolution (reducing fog or noise)
• Reduces patient dose
COMPRESSION
3- FIELD SIZEAs field size increases, intensity of scatter radiation also increases rapidly. Especially during fluoroscopy
Larger radiation field produces more scatter radiation
COMPARE IMAGES: WHAT DO YOU THINK ABOUT RADIOGRAPHIC CONTRAST & IMAGE NOISE?
INCREASE IN SCATTER
• kVpkVp
• PartPart ThicknessThickness
• Field SizeField Size
CONTROL OF SCATTER RADIATION
Technologists routinely use two types of devices to reduce the amount of scatter radiation reaching the IR
Beam RestrictorsCompression
Grids
Before Patient
On patient
After patient
BEAM-RESTRICTING DEVICESAKA LBD
1. Aperture Diaphragm
1. Cones or Cylinders
2. Variable aperture collimator
APERTURE DIAPHRAGMThe simplest of all beam-restricting devices
Lead or lead-lined metal diaphragm attached to the x-ray tube head
The opening in the diaphragm is usually designed to cover just less than the IR used
APERTURE DIAPHRAGMFixed lead opening
Fixed image receptor size Constant SID
Source-to-diaphragm distance = SDD
CONES & CYLINDERSAre modifications of the aperture diaphragmAlignmentis one difficultywhen usingconesNow mostly used with spinesteeth & heads
IMPROVED CONTRAST RESOLUTION OF THE L5 - S1
IMPROVED CONTRAST RESOLUTION OF THE FRONTAL SINUSES
VARIABLE APERTURE COLLIMATOR
The most common beam-restricting device is the light-localizing variable aperture collimator
The first part of the collimator serves to control off-focus radiation. What is off-focus radiation?
OFF - FOCUS RADIATIONALSO CALLED EXTRA-FOCAL RADIATION
X-ray tubes are designed so that the projectile e- interacts with the target. However, some of the e- bounce off the target and land on other areas
This caused x-rays to be produced out side the focal spot, they are undesirable because it extends the size of the focal spot, increases patient skin dose & reduces image contrast
Note: Using a grid does not reduce off-focal radiation
FIRST-STAGE ENTRANCE SHUTTERING DEVICEHas multiple collimator blades protruding from the top of the collimator into the tube housing
THE SECOND-STAGE COLLIMATOR SHUTTERSPb leaves are at least 3 mm thickThey work in pairs and are independently controlled
THE COLLIMATOR LAMP & MIRRORMust be adjusted so that the projected light field coincides with the x-ray beamMisalignment of the light field and beam can result in collimator cutoff of anatomic structures
Always keep the collimated area smaller than the size of the cassette
What is a PBL BLD
GRIDS
QUESTIONS …?
Technique adjustments problems