GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 1 CAL calib at SLAC SVAC
Stability of the CAL CalibrationsStability of the CAL Calibrations
• Online script “suites” acquire charge injection and muon data.
• Offline, calibGenCAL package processes that data for 3 clients:
1. fle, fhe, lac settings for online (that online can also make itself)
2. “calib”.xml files for offline reconstruction, via the rdb metadata database.
3. The “calib”.xml files to be used for Monte Carlo detector simulation (see Julie McEnery’s talk at Monday’s analysis
meeting).
• Test the whole chain via the final single crystal energy & positions.
• This talk: test the stability of the results for different data-taking configurations.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 2 CAL calib at SLAC SVAC
The MethodThe Method
• Extrapolate TKR track to CAL and predict which crystals get hit.
• Compare energy deposits, position resolution with Monte Carlo.
• Use Tkr1EndPos, Dir
• Require:TkrNumTracks == 1Tkr1KalThetaMs < 0.03Tkr1NumHits > 15
<2 MeV in adjacent crystalsTrack must traverse top & bottom of crystal (“V == 1”)
• Correct energy for cos
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 3 CAL calib at SLAC SVAC
Expect rms <27.35/12 = 7.9 mm
For even layers, CAL asymmetry gives position along the X direction. Y position is center of crystal face.
(Have cut 30 mm away from ends.)
Example: Bay 1 (=FM103)Example: Bay 1 (=FM103)Layer 4, Column 2 ( Layer 4, Column 2 ( All 2 tower B/2 data))
Width of CsI cells:
27.35 mm
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 4 CAL calib at SLAC SVAC
Six towersSix towers -- energy -- energy
( FIFO configuration goof for 5th, 6th towers -- after these 2 slides, show only 2
tower data. )
FM104 FM105 FM102 FM103 FM106 FM107
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 5 CAL calib at SLAC SVAC
Six towersSix towers -- position ( -- position (x,y) for odd layersx,y) for odd layers
calibrations stable across 3/8 of LAT.
No glaring issues of e.g. even vs odd bays.
Bigger, better calibration data samples needed.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 6 CAL calib at SLAC SVAC
How odd & even layers differHow odd & even layers differBird’s eye view of CAL AFEE boards
TPS
top
The four X rows areone layer ‘higher’than the Y rows.
X Y
TEM
X Y0 0 1 01 2 3 12 4 5 23 6 7 3
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 7 CAL calib at SLAC SVAC
2 Towers -- zoom on energy.2 Towers -- zoom on energy. Blue=data, Blue=data, Green=MCGreen=MC
FM104 FM105Monte Carlo width is ~150 keV narrower than for real data. Reflects a known misunderstanding of light yield in CsI. See e.g. Benoit Lott’s “quenching” studies.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 8 CAL calib at SLAC SVAC
2 Towers -- zoom on 2 Towers -- zoom on x,y and compare with x,y and compare with MCMC
FM104 FM105FM104 FM105
Odd layers: “Y” is the longitudinal measurement.
“X” structure an artifact of muon
angular distribution.
Widths as in MC.
“Y” means:
work-in-progress.
Less striking in “X”.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 9 CAL calib at SLAC SVAC
Glasnost…Glasnost…
• The preceding slides make it look like our <11.2 MeV> energy comes out perfectly.
• But… after the 2 tower data, a minor geometry bug was found and fixed in calibGenCAL.
• the bug fix raises reconstructed energy to ~11.4 MeV.
• We may have a CAL trigger induced shift in muon energy peak for the CAL-trigger calibration data, absent in the TKR-triggered data we are analysing.
• Several valiant attempts to confirm that were foiled by various technical hitches… we’re working on it … it’s only a 2% effect and will get straightened out.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 10 CAL calib at SLAC SVAC
2 Towers -- some data sets2 Towers -- some data sets
(from Eduardo’s May 13 “End to end” slides)
• B/2 -- (baseline) ( preceding slides )
• 2/1 and 2/2 -- 27 and 29 volts, (nominal spacecraft supply is 28 volt).
• 4/4 -- overlay rate of 20 kHz.
• 8/6 -- overlay rate of 1kHz, non-zero suppressed 4-range readout.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 11 CAL calib at SLAC SVAC
27 (red)27 (red),, 29 (yellow)29 (yellow) versusversus 28 volts (B/2, blue) 28 volts (B/2, blue)
27 volts 29 volts-28 keV -23 keV
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 12 CAL calib at SLAC SVAC
Hi rate (20 kHz, red),Hi rate (20 kHz, red), Big events (yellow) Big events (yellow) versus versus B/2 (blue)B/2 (blue)
20 kHz Big Events-43 keV -5 keV
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 13 CAL calib at SLAC SVAC
ConclusionsConclusions
• First look at 3/8 of LAT CAL suggests uniformity across towers.No glaring issues of e.g. even vs odd bays.
• A 2% shift between calibration muons and reconstructed muons exists, and is being tracked down.
• Spacecraft supply voltage ; data rate ; event size have no obvious effect on measured muon energy (or landau width, or x,y either)
• Recent v3 “parallelized” online acquisition, combined with forthcoming v4 “parallelized & streamlined” calibGenCAL are simplifying the calibration process.
• v3 also corrected ancient “badcalibGen” that made HE range continuity study harder to do.
• We’re in fairly good shape, and steadily getting better.
GLAST LAT Project Instrument Analysis Workshop #4, 14 July 2005
David Smith 14 CAL calib at SLAC SVAC
AFEE = Analog Front End ElectronicsAFEE = Analog Front End Electronics
This is “Y+ 43” (and “Y- 40”), at outlet of the data highway.(row 3=layer 7, column 7 or 5)
0 to 11 on Minus side11 to 0 on Plus side
bottom
top cable controllers
top related