TechnicalInformationTI 207C/07/enNo. 51501593
Smartec S is a compact measuringsystem for inductive conductivitymeasurement, specially designed formeasurements in highly conductiveliquids. The various process connectionsof the Smartec S CLD 132 allow anindividual adaption of the transmitter todifferent mounting situations.
Areas of application• Alkali/acid concentration control• Phase separation of product/water
and product/product mixtures in pipesystems
• Monitoring and control of bottlecleaning systems
• Product monitoring in breweries,dairies and the beverage industry
• CIP system control.
Benefits at a glance• Transmitter housing made of stainless
steel (IP 67)• Sensor made of highly resistant
synthetic material (PEEK)• Sterilisable sensor (140 °C)• Insensitive to polarisation and soiling• Various mounting possibilities• Sensor versions for all industrial
process connections• Compact transmitter• Simple installation• Easy operation due to logical menu
structure• Large two-line display allows
simultaneous display of measuredvalue and temperature
• Direct calibration access via CAL key• Safe operation due to overvoltage
protection according to EN 61000-4-5• Alarm contact for error signalling• Communication interface HART® or
PROFIBUS• Standard version extendable by:
– Measuring range remote switching(MRS)
– Temperature coefficientdetermination
– Alarm contact switchable as limitcontactor.
Endress HauserThe Power of Know How
Conductivity Measurementsmartec S CLD 132
Transmitter for measurement of conductivityand concentration with inductive sensor
Quality made byEndress+Hauser
ISO 9001
Measuring system A complete measuring systemcomprises:• the transmitter Smartec S CLD 132• conductivity sensor CLS 52 with
integrated temperature sensor andfixed cable or
• the compact instrument Smartec SCLD 132 with integrated conductivitysensor CLS 52
Optional for the separate version:
• extension cable CLK 5• junction box VBM• Mounting kit for pipe mounting.
Basic version andadditional functions
ENDRESS+HAUSERSMARTEC SENDRESS+HAUSERSMARTEC S
ALARM
CAL
E
+
–
Smartec S CLD 132
Conductivity sensorCLS 52
Smartec S CLD 132with integrated
conductivity sensor
LD132SYS.CDR
Completemeasuring systems withSmartec S CLD 132
top:Separate version andconductivity sensorCLS 52
bottom:Compact version withintegrated conductivitysensor CLS 52
Fuctions of the basic version Options
MEASUREMENT
CALIBRATION of installation factorCALIBRATION of cell constantCALIBRATION of residual coupling
Read instrument DATA
Linear CURRENT OUTPUTCURRENT OUTPUT simulation
CHECK of measuring system by PCS alarm(live check)
TEMPERATURE COMPENSATION selectable(1 free table)
CONCENTRATION table selectable(4 defined tables, 1 free table)
RELAY as alarm contact
SERVICE functions
2nd current output for temperature
MEASURING RANGE REMOTE SWITCHING:
REMOTE SWITCHING of max. 4 measuring ranges
TEMPERATURE COMPENSATION selectable(4 free tables)
CONCENTRATION table selectable(4 defined tables, 4 free tables)
RELAY configurable as alarm or limit contactor
TEMPERATURE COEFFICIENTS can be determined
2
Measurement principle In inductive conductivity measurement,a transmitting coil generates a magneticalternating field that induces an electricvoltage in a liquid. The ions present inthe liquid enable a current flow whichincreases with increasing ionconcentrations.
The current in the liquid generatesa magnetic alternating field in thereceiving coil. The resulting current inthe receiving coil is measured and usedto determine the conductivity value. Theconductivity serves as a measure of ionconcentration.
This measurement principle has thefollowing advantages:• No electrodes, therefore no
polarisation• Error-free measurement in media with
a tendency to sedimentate• Complete galvanic isolation of
measurement from medium.
Measuring technology according tothe purity law
The sensor, injection-moulded for highlychemically, mechanically and thermallyresistant PEEK (polyether ether ketone),does not have joints or crevices and istherefore biologically safe.
The temperature sensor Pt 100 is indirect thermal contact with the mediumvia the stainless steel thermalconductivity socket. This assuresextremely fast temperature response(t90 < 5 s).
The use of special components andmaterials makes the sensor suitable forcontinuous exposure to temperaturesof +125 °C and brief exposures(max. 30 min) to +140 °C for sterilization.
ENDRESS+HAUSERSMARTEC SENDRESS+HAUSERSMARTEC S
ALARM
CAL
E
+
–
LD132FCT.CDR
1
3
4
6
2
5
7
8
9
Measurement andfunction principleof Smartec S
LS51-T12.EPS
Detail view of theconical measuringchannel
LS52-03.EPS
Sensor CLS 52 withtemperature sensorPt 100 with flow lines
3
Dimensions
LD132-01.CDR
207
242
(180
)
142
35.5
∅ 49
34
Dimensions in parentheses denote thetotal length of the compact device withsensor Varivent, APV or 2 “ Perlick.
Dimensions ofSmartec S CLD 132,compact version
LD132-02.CDR
142
225
175
109
170
95
Dimensions ofSmartec S CLD 132,separate transmitterwith post mountingaccessories
225
4
Mounting
160
225
LD132-10.CDR
∅ 7
Wall mounting of theseparate transmitterSmartec S CLD 132
LD132-11.CDR
Back plateincluded in delivery
Post mounting kit(accessories)
LD132-13.CDR
∅ max. 60
Post mounting of theseparate transmitterSmartec S CLD 132
5
Mounting
The minimum pipe socket diameterdepends on the sensor used.
The pipe socket and the wall distanceof the built-in sensor determine theinstallation factor.
Installation factor The electrical conductivity of the liquidprimarily depends on the ionconcentration. However, installation andsensor geometry must also be takeninto account for measurement. The cellconstant completely describes thesensor geometry.
If the distance from the wall is sufficient(a > 15 mm), then it is not necessary toconsider the installation factor (f = 1.00).If the distance from the wall is smaller,then the installation factor increases inthe case of electrically insulating pipes(f > 1) and decreases in the case ofelectrically conductive pipes (f < 1).
LD132-08.CDR
Horizontal flow Vertical flow
b
b
a
Flowdirection
Flow direction
a = wall distanceb = immersion depth
min. DN 40
min. DN 40
Mounting of thecompact version ofSmartec S CLD 132in flow pipes
DIAGR.CDR
2
1
1.40
1.20
1.00
0.80
Inst
alla
tion
fact
orf
0 5 10 15 20 25Wall distance a in mm
Installation factor fis dependent ondistance a from pipe wall1 Electrically
conductive pipe2 Insulating pipe
Sensor Diameter pipe socket Installation factor
Varivent, 2 “ Perlick , APV ≥ DN 40 f > 1
Dairy fitting, 2" SMS,G 1 ½ thread, 2" Clamp
≥ DN 50 f > 1
no influence of diameter ≥ DN 65 f = 1
6
Process connections
LD132-06.CDR
Dairy fitting DN 50(DIN 11851)
2" Clamp connection(ISO 2852)
G 1 ½ internal thread
83
60 AF
83 102
2 “ ClampG 1½
Varivent connectionDN 40 ... DN 125
2" Perlick connection
∅ 84 ∅ 90
46 33
∅ 92
83
∅ 84
2 “ SMS
APV connectionDN 40 ... DN 100
∅ 100
46
Process connectionsof Smartec S CLD 132,compact version
LD132-07.CDR
Dairy fitting DN 50(DIN 11851)
G 1 ½ internal thread 2" Clamp connection(ISO 2582)
60 AF
2 “ Clamp
83
G 1½
102
∅ 84
8383
Varivent connectionDN 40 ... DN 125
2" Perlick connection
46 33
∅ 89∅ 84
2 “ SMS
∅ 92
∅ 100
APV connectionDN 40 ... DN 100
46
Process connectionsof conductivity sensorsCLS 52
7
Electrical connection
Pressure andtemperature ranges
15 V85
86
10-50 V
81
82
10-50 V
93
94
41
42
43
Cd
mA31
32
mA33
3411
12
13
S
84
83
15
16
∼–∼–
Temperaturesensor Pt 100
Power supply
LD13212E.CDR
Auxiliary voltageoutput
Digital input 1(HOLD,MRS 3+4*)
Signal output 1:conductivity
Alarm(contact status:no current)
Digital input 2(MRS 1+2*)
Sensor
optionalSignal output 2:temperature(optional)
PE
* MRS = measuring range remote switching (optional)Electrical connectionof Smartec S CLD 132
60 80 100 120 14020 40125
0
10
0
–10–5
20
30
40
5055
55
35
60
70
[° C]
[° C]
brief exposurefor sterilization
(< 30 min)
T-DIAGR.CDRMedium temperature
Compact version
Sensor CLS 52 with separate version
Am
bien
ttem
pera
ture
Permissible pressureand temperature rangesof Smartec S CLD 132
60 80 100 120 14020 4012590
0
42
–5
8
12
16
20
[bar]
[° C]
Med
ium
pres
sure
Medium temperature
LS52-DTB.CDR
brief exposurefor sterilization
(< 30 min)
Permissible pressureand temperatureranges of the sensorCLS 52
8
Operation Everything at a glanceThe display simultaneously shows thecurrent measured value and thetemperature – the essential processdata. Brief informational texts in theconfiguration menu provide assistancewith parameter configuration.
Intelligent and simpleAll instrument control functions arearranged in a logical menu structure.The individual parameters can only bemodified by following the access codeentry. The current position within themenu structure is displayed (as shownin figure below – O213).
ENDRESS+HAUSERSMARTEC S
ALARM
CAL
E
+
–
2000mS/cm
O213
20 mA
DISPLAY1.CDRUser interface:Display and keys
SETUP 1A
SETUP 2B
ALARMF
CHECKC
RELAYSR
ALPHA TABLET
CONCENTRATIONK
SERVICES
INTERFACEI
TEMP. COEFF.D
MEAS. RANGEREMOTE SWITCHING
M
CURRENT OUTPUTO
Selection of measurement unit, entry of measured value damping
Adjustment of current output assignment, curve and simulation for verification
Selection of contact type; adjustment of switching behaviour;adjustment of error current
Adjustment of live-check alarm
Configuration of contact as limit contactor or alarm contact
Definition of table for temperature compensation (up to 10 value pairs)α
Selection and editing of concentration curves(4 defined, up to 4 free curves)
Selection of language version; configuration of HOLD function;entry of release codes
Entry of device address for communication interface HART or PROFIBUS
Determination of temperature coefficients
Selection of digital inputs and parameter sets for remote switching ofmeasuring range
Temperature measurement and temperature compensation
CALIBRATIONC
Calibration of cell constant; calibration of residual coupling;entry of installation factor; adaptive calibration
Current temperature display can be suppressed by pressing the PLUS key
Display of active errors (sequence according to priority)
MEAS. VALUE
TEMPERATURE
Display
LD132BEE.CDR
Overview of Smartec SCLD 132 menu.This overview coversall options that can beinstalled.
9
Technical data
General specifications Manufacturer Endress+Hauser
Product designation Smartec S CLD 132
Physical data / design Dimensions of separate transmitter with mounting rail(L × W × D)
225 × 142 × 109 mm
Weight approx. 2.5 kg
Dimensions of compact deviceversions MV1, CS1, GE1, SMS (L × W × D)
225 × 142 × 242 mm
Dimensions of compact deviceversions VA1, AP1, PER (L × W × D)
225 × 142 × 180 mm
Weight with sensor CLS 52 approx. 3 kg
Measured value display LC display, two-line, five and nine digits for text,current status display
Materials Housing stainless steel SS 304 S
Front window polycarbonate
Sensor CLS 52 PEEK
Input Measured quantities conductivity, concentration, temperature
Minimum distance for 0 / 4 ... 20 mA signal
For conductivity measurement meas. value 0 ... 19.99 µS/cm: 2 µS/cmmeas. value 20 ... 199.9 µS/cm: 20 µS/cmmeas. value 200 ... 1999 µS/cm: 200 µS/cmmeas. value 2 ... 19.99 mS/cm: 2 mS/cmmeas. value 20 ... 200 mS/cm: 20 mS/cmmeas. value 200 ... 2000 mS/cm: 200 mS/cm
For concentration measurement no minimum distance
Inductive conductivity measurement
Measuring range 10 µS ... 2000 mS/cm
Cell constant k = 5.9 cm–1
Maximum cable length 55 m (CLK 5)
Measuring frequency 2 kHz
Temperature measurement
Temperature sensor Pt 100, class A acc. to IEC 60751
Measuring range –10 ... +150 °C
Temperature offset range ±5.0 °C
Temperature compensation
Compensation types none (α = 0), linear, table, NaCl
Range –10 ... +150 °C
Minimum distance 1 K
Reference temperature 25 °C
Digital inputs 1 and 2
Voltage 10 ... 50 V
Current consumption max. 10 mA at 50 V
Output Conductivity signal output (not with PROFIBUS)
Current range 0 / 4 ... 20 mA, galvanically isolated; error current 2.4 / 22 mA
Load max. 500 ΩMaximum resolution 700 digits/mA
Transmission range adjustable
Separation voltage max. 350 Veff / 500 V DC
Overvoltage (lightning) protection acc. to EN 61000-4-5:1995
Temperature signal output (optional, not with PROFIBUS)
Current range 0 / 4 ... 20 mA, galvanically isolated
Load max. 500 ΩMaximum resolution 700 digits/mA
Transmission range adjustable, ∆ 10 ... ∆ 100 % of measuring range
Separation voltage max. 350 Veff / 500 V DC
Overvoltage protection acc. to EN 61000-4-5:1995
10
Technical dataAuxiliary voltage output
Output voltage 15 V ± 0.6 V
Output current max. 10 mA
Contact outputs
Switching current with ohmic load (cos ϕ = 1) max. 2 A
Switching current with inductive load (cos ϕ =0.4) max. 2 A
Switching voltage max. 250 V AC, 30 V DC
Switching power with ohmic load (cos ϕ = 1) max. 1250 VA AC, 150 W DC
Switching power with inductive load (cos ϕ = 0.4) max. 500 VA AC, 90 W DC
Limit contactor
Pickup / dropout delay 0 ... 2000 s
Alarm
Function (switchable) steady contact / fleeting contact
Alarm delay 0 ... 2000 s (min)
Accuracy Conductivity measurement
Deviation of indication1 max. 0.5 % of measured ± 4 digits
Reproducibility max. 0.2 % of measured ± 2 digits
Measurement deviation1, conductivity signal output 0.75 % of current output range
Temperature measurement
Resolution 0.1 °C
Deviation of indication1 max. 0.6 % of measuring range
Measurement deviation1, temperature signal output max. 0.75 % of current output range
Power requirements Supply voltage 100 / 115 / 230 V AC +10 / –15 %, 48 ... 62 Hz24 V AC/DC +20 / –15 %
Power consumption max. 7.5 VA
Fuse protection fine-wire fuse, medium time-lag, 250 V / 3.15 A
Operating data compact version Operating temperature max. 55 °C at 55 °C ambient temperaturemax. 140 °C (30 min) at 35 °C ambient temperature
Operating pressure max. 16 bar (90 °C)
Ambient conditions Storage and transport temperature –25 ... +70 °C
Relative humidity (nominal operating conditions) 5 ... 95 %, non-condensing
Ingress protection of housing IP 67
Electromagnetic compatibility interference emission and interference immunityacc. to EN 61326-1:1997
Vibrational resistanceacc. to IEC 770
Mounting place pipe
Vibration frequency 10 ... 60 Hz
Peak amplitude 0.21 mm
Impact test Impact strength of display window 9 J
Sensor data CLS 52 Conductivity measuring range 10 µS ... 2000 mS/cm
Cell constant k = 5,9 cm-1
Temperature sensor Pt 100, class A acc. to IEC 60751
Temperature measuring range Pt 100 –5 ... +140 °C
Temperature response time t90 < 5 s
Measuring value deviation ±10 µS/cm + 0.5% of meas. value at –5 ... 100 °C±30 µS/cm + 0.5% of meas. value at > 100 °C
max. cable length 55 m
Material in contact with medium PEEK, stainless steel SS 316 S, Chemraz
Ambient temperature –10 ... +70 °C
max. operating temperature 140 °C (max. 30 min)
max. operating pressure max. 16 bar (90 °C)
Ingress protection IP 67
Supplementary documentation Technical Information CLS 52 order no.: 50086110
1acc. to IEC 60746-1, for nominal operating conditions Subject to modifications.
11
Accessories
Product structure
Endress+Hauser GmbH+Co.- Instruments International -
P.O. Box 2222D-79574 Weil am Rhein
Tel. (07621) 975-02Fax (07621) 975-345
E-Mail: ehii@compuserve.com
TI 207C/07/en/12.99Printed in Germany / DT / CV5 (51501593)
Endress HauserThe Power of Know How
VersionP Compact versionW Separate transmitter, 5 m cable lengthX Separate transmitter, 10 m cable length
Process connectionMV1 Dairy fitting DN 50 (acc. to DIN 11851)CS1 2" Clamp connection (acc. to ISO 2852)GE1 G 1 ½ internal threadVA1 Varivent connection from DN 40 upwardAP1 APV connection from DN 40 upwardSMS 2" SMS connectionPER 2" Perlick connection
Cable entry1 Cable gland Pg 13.53 Cable gland M 20 × 1.55 Conduit adapter NPT ½"
Power supply0 230 V AC1 115 V AC5 100 V AC8 24 V AC / DC
Current output / communicationAA Current output conductivityAB Current output conductivity and temperatureHA Current output conductivity, HARTHB Current output conductivity and temperature, HARTPE PROFIBUS-PAPF PROFIBUS-PA, M 12 plugPP PROFIBUS-DP
Additional equipment1 Basic version2 Measuring range remote switching /
extended functionality
CLD 132- complete order code
Conductivity and resistance transmitter Smartec S CLD 132
Mounting accessories
Type Features Order number
Postmounting kit
Kit for mounting of field housing on horizontalor vertical pipes (∅ max. 60 mm)Material: stainless steel (SS 304 S)
50062121
Cable extension
Type Features Order number
CLK 5 Extension cable for inductive sensorsup to 55 m
50085473
VBM Junction box for inductive measuring cables 50003987