abnormal intra-qrs potentials associated with percutaneous transluminal coronary...
Post on 14-Nov-2023
0 Views
Preview:
TRANSCRIPT
www.elsevier.com/locate/jelectrocard
Journal of Electrocard
Abnormal intra-QRS potentials associated with percutaneous transluminal
coronary angiography–induced transient myocardial ischemia
Paul Lander, PhDa, Pedro Gomis, PhDb,4, Stafford Warren, MDc, Gary Hartman, MDd,
Kathy Shuping, BScd, Ralph Lazzara, MDd, Galen Wagner, MDd
aFlow Metrix, Maynard, MA 01754, USAbCentre de Referencia en Bioenginyeria de Catalunya, Technical University of Catalonia, 08028 Barcelona, Spain
cCharleston Area Medical Center, Charleston, WV 25304, USAdDuke University Medical Center, Durham, NC 27705, USA
Abstract This article introduces a novel concept of abnormal intra-QRS potentials (AIQPs) associated with
0022-0736/$ – see fro
doi:10.1016/j.jelectroc
4 Corresponding
E-mail address: p
myocardial ischemia. AIQPs are microvolt-level potentials—subtle alterations in the QRS of the
high-resolution electrocardiogram (ECG)—isolated from the unfiltered signal-averaged ECG
(SAECG) by a method of mathematical modeling. The aims of the study were (1) to determine
the characteristics of potentials in the SAECG related to ischemically altered activation during
percutaneous transluminal coronary angiography (PTCA), (2) to determine their relationship with
standard 12-lead ECG variables, and (3) to investigate whether AIQPs have a specific
pathophysiologic basis in myocardial ischemia. Continuous high-resolution ECG data were
acquired from 12 patients before, during, and after PTCA. SAECGs were computed every
60 seconds using an enhanced method of signal averaging. AIQP, ST-segment deviation, and
changes in standard ECG QRS duration were measured in each 1-minute SAECG. AIQP
amplitudes increased significantly during balloon inflation, compared with the preinflation state.
AIQPs exhibited a greater prevalence (12 of 12 patients) than ST-segment deviation changes of
more than 100 lV (7 of 12 patients), or measurable changes in standard QRS duration (4 of
12 patients). In patients with significant changes in 12-lead ECG variables during balloon
inflation, AIQPs were strongly correlated with both ST-segment and QRS-duration changes. AIQP
timing was correlated with the artery occluded, suggesting a specific, ischemia-influenced origin
of the signal. AIQPs show promise as a time-localized, sensitive new ECG marker of ischemically
altered ventricular activation.
D 2006 Elsevier Inc. All rights reserved.
Keywords: Myocardial ischemia; Potentials; Angioplasty; Signal-averaged ECG; QRS
Introduction
Presently, no single noninvasive test is well suited to
continuous monitoring of myocardial ischemia, detecting
presence, extent, and severity. Emergency care of patients
with acute chest pain suggestive of myocardial infarction is
based on decisions regarding initiation and administration
of thrombolytic therapy. Standard 12-lead electrocardiogram
(ECG) remains the only readily available clinical method
to guide these decisions. The general objective of this
work was to compare the capabilities of standard and
high-resolution ECG (HRECG) to detect acute myocar-
dial ischemia.
nt matter D 2006 Elsevier Inc. All rights reserved.
ard.2006.02.006
author. Tel.: +34 934010763; fax: +34 93 4017045.
edro.gomis@upc.edu (P. Gomis).
Conventional large-scale notches and slurs in the QRS
complex of the standard ECG have long been associated
with altered conduction due to myocardial scarring.1-4 We
have defined abnormal intra-QRS potentials (AIQPs) as
low-amplitude notches and slurs (1-50 lV), or subtle
alterations, in the QRS complex of the HRECG. AIQPs
are quantified by using a computerized modeling technique
described previously.5,6 We have recently shown that AIQP
may be a better pathophysiologic signal representation of a
reentrant mechanism for ventricular tachycardia than ven-
tricular late potentials.7 AIQP and late potentials were
measured by using epicardial maps recorded during
experimental myocardial infarction.7 In one study, in the
presence of an arc of block, late potentials and AIQPs were
both present. Late potentials were distal, whereas AIQPs
were coincident with the localized site of reentry. This
iology 39 (2006) 282–289
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289 283
points to AIQP as a marker for altered conduction in the
border zone of the myocardial infarction.
In both myocardial infarction and myocardial ischemia,
abrupt, irregular, and localized changes in ventricular
conduction will occur. We hypothesize that in both cases,
AIQPs will result from altered activation around an area
of functional block. Changes in magnitude and direction
of conduction velocity would be transient, resulting in
low-level alterations in QRS morphology in the body
surface HRECG.
Previous studies8-11 have documented changes in high-
frequency QRS energy during periods of complete acute
coronary occlusion in man, which were reversed when
blood flow to the ischemic myocardium was reestablished.
Large-scale QRS changes in the standard ECG have also
been documented during the transient acute transmural
ischemia produced by angioplasty balloon occlusion.12 The
present study introduces a new concept—that low-level
abnormal intra-QRS potentials not seen in the standard ECG
may be a direct, sensitive marker of ischemically altered
ventricular activation. Percutaneous transluminal coronary
angioplasty (PTCA) provides an excellent model in which
to investigate the electrophysiologic changes of transmural
ischemia resulting from the spontaneous thrombotic occlu-
sion responsible for acute myocardial infarction. The PTCA
laboratory selected for the present study is unique in that it
uses a single, prolonged balloon inflation of up to 5 minutes
when clinically feasible. AIQP in the SAECG, and changes
in ST-segment deviation and QRS duration from the
standard 12-lead ECG, were measured every minute before,
during, and after PTCA. This permitted comparison of serial
changes during each minute of balloon inflation in both
standard and high-resolution ECG signals. Signal-averaged
ECGs with enhanced signal-to-noise ratio were obtained
every 60 seconds when the time-frequency plane Wiener
filter was used.13-15
The present study had 3 specific aims: (1) to determine
the prevalence and natural progression of AIQPs (abnor-
mal intra-QRS potentials not seen in the standard ECG)
during PTCA, (2) to determine whether correlations exist
among AIQP and ST-segment deviation and QRS prolon-
gation in the standard 12-lead ECG, and (3) to investigate
whether AIQPs are specific to acute PTCA-induced
myocardial ischemia by relating their timing to the
particular artery occluded.
Methods
Data acquisition
Continuous high-resolution ECGs were recorded in 12
patients undergoing elective PTCA in the catheterization
laboratory at the Charleston Area Medical Center, WVa.
Recording commenced at least 50 seconds before balloon
inflation. Balloon inflation periods ranged from 3 to
5 minutes, with the exact times of inflation and deflation
noted. HRECG data were recorded continuously by use of
orthogonal XYZ leads with the Predictor system (Corazonix
Corp., Oklahoma City, Okla), at a sampling rate of 1000 Hz.
The resolution of the analog to digital converter was 16 bits,
resulting in a 0.3-lV quantization step. Particular attention
was paid to obtaining a high-fidelity recording by careful
skin preparation and electrode application, and by minimal
use of fluoroscopy and other devices that are significant
sources of electromagnetic interference.
Patients who met the following criteria were enrolled in
the study: no history of coronary bypass surgery; no
previous myocardial infarction; absence of ST-segment
deviation, ventricular hypertrophy, bundle branch block,
and extrasystoles on the baseline standard 12-lead ECG.
Perfusion balloons or atherectomy devices were not used.
Of the 12 patients studied, 4 had occlusion of the left
anterior descending coronary artery (LAD), 4 of the left
circumflex coronary artery (LCX), and 4 of the right
coronary artery (RCA). One subject had ECG evidence of
a prior myocardial infarction.
In addition to the HRECG, a standard, 12-lead ECG
(Hewlett-Packard 4765A, Palo Alto, CA) was recorded at
1-minute intervals during PTCA. ST-segment deviations
(either depression or elevation) were measured 60 milli-
seconds after the J point and defined by the maximum value
present in any lead. During and after the period of balloon
inflation, ST-segment deviations were compared with a
reference ECG recorded just before balloon inflation.
Off-line processing of the continuous HRECG record
The continuous HRECG from each study was stored
digitally and subsequently transferred to a recordable
compact disc for later processing. An annotation process
was performed for each subject’s HRECG file before signal
averaging and analysis. Custom-designed computer algo-
rithms were used, which provided a visual overview of the
automatic annotation process. Initially, a normal sinus
rhythm QRS complex was selected from within the
preinflation segment of the HRECG record. This beat
served as a template for both classification (sinus or
nonsinus beat) and alignment for enhanced signal averag-
ing. The QRS onset and offset were marked visually on the
template beat. All other beats in the HRECG record were
then compared with the template. Nonsinus, fusion, or
excessively noisy beats were detected by separately
comparing the integrals (ie, areas) of each half of the QRS
with those of the template. Sinus rhythm beats that passed
this test were then aligned by using a correlation method
described previously.16 The correlation region was the first
half of the QRS, with a threshold value for the correlation
coefficient of .99 set for beat acceptance. Noise in each beat
was measured by the ensemble signal variance method
described previously.16 The result of the annotation process
was a list of all beats, with their alignment fiducial point,
classification, and noise measurement.
The time-frequency plane Wiener–filtered SAECG
Signal averaging of several hundred beats is convention-
ally required to detect high-resolution ECG signals, such as
ventricular late potentials.17,18 In this study, the hypothe-
sized dynamic nature of abnormal intra-QRS potentials
during PTCA necessitated forming the signal-averaged ECG
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289284
(SAECG) with a significantly reduced number of beats, on
the order of 50 to 100 beats. The time-frequency plane
Wiener (TFPW)–filtered SAECG was used to accelerate
noise reduction. This is a new method of signal averaging,
which uses processing in the time-frequency plane to
enhance the signal-to-noise ratio of the average. The method
has previously been used successfully to detect late
potentials, with a reduced number of beats and averaged
noise levels less than 0.3 lV root-mean-square (rms).19 The
TFPW filtering method is summarized in the following. (A
full description of the method is given in References 13-15).
First, the ensemble (or signal) average is given by
xP tð Þ ¼XRi¼1
xi tð Þ=R ¼ s tð Þ þ nP tð Þ=ffiffiffiR
pð1Þ
where xi(t) are the R beats of the ensemble, s(t) is the
deterministic, cardiac signal component, and n(t) is the
assumed additive and Gaussianly distributed noise of a
typical beat. The quadratic time-frequency representation
(TFRQ, or spectrotemporal map) of the signal average is
given by
TFRQ xP tð Þ½ � ¼ XPt; fð Þ ¼O
ZxP sð Þw s � tð Þ e�j2pf s dsO
2 ð2Þ
where w(t) is the analysis window used to compute the
spectrotemporal map.20 Ideally, the quantity TFRQ[x(t)] can
be expressed mathematically as
E TFRQ xP tð Þ½ �� �
¼ S t; fð Þ þ NP
t; fð Þ=R ð3Þ
where S(t,f) and N(t,f)/R represent the cardiac signal and
averaged noise components in the time-frequency plane.
Similarly, the average of the TFRs computed for each beat
in the ensemble can be ideally expressed as
EXRi¼1
TFRQ xi½ �=R#¼ TFRQ xi tð Þ½ �P
¼ S t; fð Þ þ NPt; fð Þ
"
ð4Þ
In practice, the 2 quantities in Eqs. (3) and (4) are
smoothed with a 2-dimensional filter15 to reduce interfer-
ence and variance terms before being combined into a
weighting function, given by
w t; fð Þ ¼ S t; fð ÞS t; fð Þ þ N
Pt; fð Þ=R
ð5Þ
This is the TFPW filter, which is then applied to the
linear TFR (TFRL) of the ensemble average, x(t,f), by a
multiplicative operation. The resulting TFPW-filtered
SAECG, s(t), is recovered by an inverse transformation
from the time-frequency plane to the time domain, that is15
ss tð Þ ¼ TFRL�1
xP t; fð Þw t; fð Þ� �
ð6Þ
Subensemble averaging
The ensemble of all beats in the continuous HRECG
record before, during and after PTCA, was split into
subensembles by using the list of annotated beats described
here. A series of sequential, TFPW-filtered, subensemble
averages was computed as follows:
(1) A preinflation subensemble average, composed of
up to 128 beats available immediately before
balloon inflation.
(2) to (6) Between 3 and 5 subensemble averages,
each of 60 seconds duration, computed sequentially
during the balloon inflation period. All available
beats within each 1-minute epoch that passed the
alignment, classification, and noise criteria, were
included.
(7) to (10) Up to 4 sequential, postdeflation sub-
ensemble averages, each of 60 seconds duration.
Each subensemble average was TFPW filtered and
then analyzed.
Analysis of subensemble SAECGs
Abnormal intra-QRS potentials were measured in each
subensemble average with the computerized modeling
technique developed in References 5 and 6. This approach
is summarized as follows. First, QRS onset, offset, and
duration (QRSD) were measured automatically and then
manually overread, as has been described previously.16 A
discrete cosine transform (DCT) of the QRS waveform was
computed and then mathematically represented by using an
ARX model.21 This is given by
y nð Þ ¼ �Xnai¼1
aiy n� 1ð Þ þXnbj¼0
bju n� jð Þ þ e nð Þ
¼ yy nð Þ þ e nð Þð7Þ
where y(n) is the DCT of the QRS complex, y(n) is the
modeled signal DCT, u(n) is an impulse, and e(n) is the
residual, following standard terminology for an ARX
model.21 The model, h = [a1a2. . .ana b0b1. . .bnb], is found
by a nonlinear iterative search for the minimum mean
squared error between y(n) and y(n), the original and
modeled signals, respectively. A model order of [na = 7,
nb = 8]; that is, 7 autoregressive and 8 moving average
terms, was used for all subjects and all leads. This model
order was selected on the basis of previous experience with
SAECG waveforms.5,6,22,23
The residual signal, e(n) = y(n) � y(n), is the AIQP
waveform, which was quantified by calculating its rms
value within the duration of the high-resolution QRS,
that is
AIQPrms ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiXT
OAIQP tð ÞO2=T
sð8Þ
The timing of the abnormal intra-QRS potential was also
computed as the bcenter of massQ of AIQP(t), defined by
AIQPt0¼
Pt0QRSOn OAIQP tð ÞO
2
P QRSOffQRSOn OAIQP tð ÞO
2c0:5 ð9Þ
Fig. 1. Progression of AIQP amplitudes before, during, and after the PTCA
procedure for each of the 12 patients.
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289 285
Statistical analysis
Comparison of means between SAECG parameters
recorded in a series of subensemble averages (epochs) was
analyzed by repeated measures analysis of variance with
1-minute epoch measurements as within-subject factors.
Post hoc comparisons of each epoch with each of the others
were performed with the Bonferroni adjustment for multiple
comparisons. Correlation was computed by use of the
Pearson correlation coefficient. The probability that no
correlation was present (q = 0) was also calculated. All
analyses were performed with the SPSS for Windows
release 11.0 (SPSS Inc., Chicago, Ill).
Results
Changes in AIQP amplitudes during PTCA
Fig. 1 shows the progression of measured AIQP
amplitudes (mean of the 3 XYZ leads) in each of the
SAECG subensembles before, during, and after the PTCA
procedure for each of the 12 subjects. If an inflation lasted
only 4 or 3 minutes (3 cases), no data are plotted in the latter
Fig. 2. Illustration of HRECG waveforms showing (A) the preinflation SAECG,
waveforms recorded in the preinflation, inflation, and postdeflation periods; and
1 or 2 minutes, respectively, of the inflation period. That is,
all postdeflation data commence at minute/subensemble
number 7. AIQP amplitudes rise sharply during the first
minute of inflation (AIQPrms = 51.8 F 27.3 lV rms)
compared with the preinflation recording (AIQPrms =
9.3F3.2 lV rms). The increase typically is maintained
throughout the inflation period, and AIQP values slightly
decrease after balloon deflation. An analysis of variance
showed that changes in mean AIQP amplitudes between
some epochs were significant (F = 6.27, P b .0001).
Bonferroni corrected pairwise comparisons showed signif-
icant differences between preinflation epoch and any minute
of inflation values. Pairwise comparison also found signif-
icant increase from the second to the third minute of
inflation (P = .02). A marked decrease is observed during
the second minute after balloon deflation with respect to the
fourth minute of inflation but without statistical significance
(P = .07). No other differences between epochs were
statistically significant.
Fig. 2A illustrates, for 1 subject, the unfiltered, prein-
flation SAECG (0.05-300 Hz), the modeled preinflation
SAECG (dotted trace), and the sequence of AIQP wave-
forms recorded in the preinflation, five 1-minute inflations,
and first postdeflation epochs. The AIQP analysis interval
(ie, the SAECG QRS duration) is denoted by the vertical
dashed lines. Fig. 2B shows the progression of AIQP
amplitudes relative to the preinflation state, increasing
sharply in the first minute of inflation and decreasing after
balloon deflation.
Relationship between AIQP timing and artery occluded
Mean AIQP timing in the 3 XYZ leads is shown in
Fig. 3 for each artery. Timing expresses AIQPt0as a
percentage of the QRS duration. The vertical limit bars
show the minimum and maximum values of AIQPt0in the
4 patients for each of the LAD, LCX, and RCA arteries.
No statistical analysis was performed because of the small
number of subjects.
the modeled preinflation SAECG (dotted trace), and the sequence of AIQP
(B) the progression of AIQP amplitudes in the same SAECGs.
Fig. 3. Mean AIQP timing for each artery, expressed as a percentage of
QRS duration.
Table 1
Correlations and level of significance between AIQP and the standard ECG
variables, ST-segment deviation, and sQRSD measured during the
preinflation and inflation periods
ST segment sQRSD
AIQP 0.51 ( P = .0006)4 0.37 ( P = .015)4
ST segment 0.37 ( P = .017)4
4 P b .05.
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289286
Relationship between changes in AIQP and ST-segment
deviations and QRS duration measured from the standard
ECG
Seven (58.3%) of the 12 subjects studied exhibited a
clinically significant change in ST-segment level (greater
than 100 lV in any lead), measured in the standard 12-lead
ECG, during the course of PTCA (range, 0 to 4 mm, ie, 0 to
0.4 mV). Considering the transient ST-segment change
(=0.05 mV) criterion used in acute coronary syndrome,24
8 (66.7%) of 12 patients presented ST-segment changes.
Using a more sensitive criterion (= 25 lV), we found that
10 (83.3%) of 12 patients showed ST-segment changes.
Four subjects (33%) exhibited a measurable change in QRS
duration, as seen in the standard 12-lead ECG (range, 0-
80 milliseconds), during the PTCA procedure. In contrast,
all subjects (100%) showed a significant increase in AIQP
amplitudes between the preinflation SAECG and the first
subensemble average recorded during inflation. AIQP
amplitudes increased by 5.99 F 3.75 times (range, 2.28-
14.8 times) between these 2 recordings. AIQP amplitudes in
the preinflation recording ranged from 3.83 to 14.8 lV rms.
In the SAECG recorded during the first minute of inflation,
AIQP amplitudes ranged from 17.2 to 116.5 lV rms. There
is no overlap between these 2 ranges. Note that all
measurements throughout the PTCA procedure were made
relative to the model of the preinflation SAECG.
Fig. 4 shows the mean progressions of AIQP, ST
segment, and standard QRS duration (sQRSD) changes,
Fig. 4. Progression of AIQP, ST-segment, and sQRSD changes, averaged
over the 12 subjects.
averaged over the 12 subjects. The traces show similarities
in form. Table 1 gives the significant correlations present
between any 2 of the 3 variables during the first 6 minutes
(preinflation plus balloon inflation epochs). The post-
deflation AIQP measurements were not correlated with
the standard ECG variables of ST deviation and QRS
duration change.
Discussion and conclusions
Significance and pathophysiologic basis of AIQP during
acute transmural myocardial ischemia
The results of the present study suggest that AIQP may
be a time-localized representation of ischemically altered
activation. The evidence for this is the strong correlation
present between AIQP amplitudes and ST-segment devia-
tion and between AIQP timing and the coronary artery
occluded. We hypothesize that AIQPs result from disrupted
activation around an area of functional block, and not from
altered properties of the myocardial action potential. This
seems likely because AIQPs appear to be more sensitive to
the presence of ischemia than ST-segment deviation, yet
ischemic injury currents occur before changes to the
upstroke of the action potential during ischemia. AIQPs
are also present when there is no prolongation of the QRS,
further suggesting that the phenomenon is based on changed
activation patterns contained within the normal period of
ventricular activation. The significance of ST depression
and elevation with acute transmural ischemia is a subject of
ongoing research.25,26 Some ECG leads may show elevation
while others exhibit depression. In the present study, the
intent was to maximize the sensitivity of ST measures by
considering the greatest change in any lead, whether
depression or elevation. With this comparison, AIQPs
emerge as the most sensitive ECG index of ischemia.
As measured by AIQP timing, ischemically altered
activation of the myocardium occurred earliest in the time
course of ventricular activation for LAD occlusions, later for
circumflex, and latest for RCA occlusions. The LAD
supplies 60% of the left ventricle, including the portions
likely to be activated earliest. In contrast, the RCA, as
evidenced by ECG changes during right bundle branch
block, supplies the ventricular myocardium activated latest.
The circumflex artery with its less distinct anatomical path
could be expected to be intermediate and to supply areas
overlapping those supplied by the LAD and RCA. These
projections are supported by the data of Fig. 3. AIQP
timing, that is, the point of disruption in ventricular
activation, occurs first with occlusion of the LAD, next in
the circumflex artery, and latest with occlusion of the RCA.
Fig. 5. Illustration of the differences in AIQP caused by large-scale QRS
morphologic changes and low-amplitude notches and slurs. A, Shows a
patient where large-scale QRS changes during PTCA created AIQPs (cf
Fig. 1). From top to bottom, 4-second ECG rhythm strips are shown for the
preinflation epoch, the five 1-minute inflation epochs, and the postdeflation
epoch. The corresponding subensemble-averaged AIQP waveforms are
shown to the right. B, Shows a patient with no large-scale QRS
morphologic changes. The ECG rhythm strips are invariant and AIQP
waveforms result from low-level QRS changes. Note: AIQP are not shown
to scale with the ECG rhythm strips, or between the 2 patients. The duration
of AIQP shown is the high-resolution QRS duration.
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289 287
The variability in AIQP timing with occlusion of the
circumflex artery is also greater (see error bars) than with
the LAD and RCA, as anticipated. These data support the
assertion that the timing of AIQP is consistent with AIQP
representing localized, ischemically altered activation of the
ventricular myocardium. Possibly owing to the small
number of subjects (n = 12), no relationship between AIQP
timing and location of the occlusion site within a specific
artery (proximal, mid, or distal) could be discerned. In
addition, AIQP timing did not appear to be lead specific for
particular arteries in this study. Future work will consider
standard precordial and limb leads measurements.
AIQP amplitudes showed a statistically significant
increase between the preinflation and first minute of balloon
inflation states in all 12 patients studied. The tendency in the
second to fourth minutes of inflation was a further, gradual
increase in AIQP amplitude. Comparison between paired
epochs showed a significant AIQP amplitude augmentation
between the second and third minute of inflation. However,
during the fifth minute, AIQP amplitudes declined in some
cases. One possible explanation may be that the myocardi-
um in and around the region of functional block had become
stunned after 4 minutes of occlusion. Postdeflation, AIQP
amplitudes typically decreased upon reperfusion and then
increased slightly or remained constant. A marked AIQP
reduction was observed during the second minute after
balloon deflation compared with the fourth minute of
occlusion. Further studies are needed to determine if and
when AIQP amplitudes return to baseline (preinflation)
values, and whether there may be any significance, relating
to reperfusion, in the postdeflation AIQP time course.
AIQPs generally result from small-signal changes in QRS
shape. However, 1 subject exhibited a large-scale change in
AIQP amplitudes between the second and third minutes of
balloon inflation. In this case, AIQPs were the result of gross
changes in QRS morphology, as illustrated in Fig. 5A. The
traces from top to bottom show representative 4-second ECG
rhythm strips, and the corresponding AIQP waveforms, for
the preinflation, inflation, and postdeflation 1-minute
epochs. AIQPs emerge because large-scale changes in
QRS morphology occur with significant ST elevation. This
was an unusual phenomenon. Fig. 5B shows a more typical
case. The ECG rhythm strips did not show significant
variation in either QRS morphology or ST-segment deviation
throughout the PTCA procedure. The corresponding AIQP
waveforms reflect low-amplitude changes in QRS morphol-
ogy. (The AIQP waveforms are not shown to scale, either
between patients or with the ECG rhythm strips. Their
duration is that of the high resolution QRS.)
Relationship of the present work to previous studies
Abboud and colleagues8,27-29 have shown that the high-
resolution ECG complex exhibits reduced high-frequency
energy in the bandwidth 150 to 250 Hz with myocardial
ischemia. These changes are not apparent in the standard
ECG. The phenomenon has been demonstrated in both
animal27 and clinical8,9 settings. A computer simulation
suggested that low-level QRS morphology changes could be
attributable to slowing of conduction velocity in a region of
ischemia.29 Aversano et al10 and Berkalp et al11 have
demonstrated that similar changes in high-frequency
content of the QRS after myocardial infarction could be a
marker of successful reperfusion after thrombolytic therapy.
These studies lend weight to the hypothesis of this article,
that ischemically altered activation can often induce
electrical changes visible as microvolt-level potentials in
the HRECG.
However, the techniques used in these prior studies are
oriented toward the frequency domain, considering only the
150- to 250-Hz band. The fixed-frequency spectral window
filter used produces phase distortion in the QRS.30 By their
nature, there is no time resolution available with frequency
domain methods.20 Neither frequency domain methods nor
conventional filtering will allow an assessment of low-level,
time-localized activity in the QRS. The modeling process
used with AIQP filters the HRECG with an arbitrary
frequency characteristic to extract an abnormal signal
wholly in the time domain. In contrast to reduced high-
frequency energy, subtle, time-localized alterations in the
HRECG QRS complex (ie, AIQP) may be a direct marker of
ischemically altered activation.
A quantitative comparison of ST segment and tomo-
graphic scintigraphy changes during PTCA has recently
been reported.31 This study concluded that ST-segment
deviations, when present, yielded only a crude estimate of
the extent of myocardial ischemia. In our study, AIQPs had
a prevalence of 100%, as opposed to 60% for ST-segment
deviation (considering ST deviation greater than 100 lV)and up to 83.3% with a more sensitive criterion (= 25 lV).If AIQPs are verified as a time-localized marker of
ischemically altered activation, they may yield more
accurate electrocardiographic estimates of ischemia. Further
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289288
work is planned to characterize AIQP during and up to 24
hours after PTCA. Scintigraphic estimates of ischemia in
both the occluded and reperfused states will be compared
with abnormal HRECG signals in a large cohort of patients.
Advantages and limitations of the experimental design
Handling large amounts of continuous HRECG data
presented significant challenges, including quality control of
signal averaging (final noise levels and correct beat
classification) and developing reliable methods of automat-
ed analysis. The TFPW-filtered SAECG13-15 greatly en-
hanced noise reduction, enabling the computation of high
fidelity SAECGs every 60 seconds. This technique enabled
measurement of serial changes in AIQP amplitude and their
comparison with ST-segment deviation and standard QRS
duration. The modeling of the HRECG QRS complex to
extract microvolt-level AIQPs is a computational method
described in detail in References 5 and 6. The principal step
in the process is choosing the model order, which will affect
the computed AIQP waveform. The model order used (na =
7, nb = 8) was selected on the basis of prior experi-
ence.5,6,22,23 We expect the major findings of this study to
be substantially independent of the choice of model order.
Further studies are planned to determine the most appropri-
ate methods for AIQP calculation.
Balloon occlusions of exact timing and duration during
PTCA provide a model of acute transmural ischemia in a
controlled setting. The clinical environment imposed some
limitations. The most important of these was a limited
period of HRECG acquisition after balloon deflation. This
was typically 3 to 4 minutes. Interpretation of the HRECG
in the short term (up to 1 hour) after balloon deflation
would be problematic because of the expected electrophys-
iologic effects of reperfusion and coronary dilation. The
effects of reperfusion and the potentially altered substrate
for AIQP, due to changed myocardial blood flow, are
unpredictable. Therefore, the significance of AIQP changes
after balloon deflation is difficult to determine. Some
patients had a subsequent balloon inflation shortly after
the first PTCA procedure. However, reperfusion effects and
altered myocardial blood flow preclude using subsequent
inflations to test the reproducibility of the ECG changes
observed after the first balloon inflation. Further experi-
ments are planned to address the long-term (up to 24 hours)
behavior of AIQP.
Support for a local ischemic origin of AIQP comes
primarily from the association between AIQP timing and the
artery occluded. This was a cornerstone of the experimental
design. The number of subjects in the study (n = 12) did not
enable the consistency of AIQP measurements to be
thoroughly evaluated. An expanded study is currently in
progress that will address the issue of consistency by
concentrating on subjects with occlusions of, for example,
the proximal and mid LAD.
Because all subjects had AIQP and acute transmural
ischemia was present by definition during arterial occlusion,
the specificity of AIQP could not be tested. In previous
work, studying a canine model of infarction, we presented
evidence that AIQP are the pathophysiologic signal repre-
sentation of a reentrant substrate for ventricular tachycar-
dia.7 In studies of 2 clinical populations, we have
demonstrated that the amplitude of AIQP is progressively
greater in subjects with myocardial scarring and with a
future arrhythmic event post-MI, compared with subjects
with no prior myocardial infarction.5,22,23 This evidence
suggests that the characteristics of AIQP will depend on the
extent of myocardial ischemia.
Clinical relevance and perspective for future studies
Changes in AIQP that are observed only in the HRECG
appear to be more sensitive for detection of acute transmural
ischemia than QRS or ST-segment changes in the standard
ECG. Therefore, further investigation of their potential
clinical usefulness is indicated, and expanded studies of
HRECG signals during PTCA are in progress. QRS changes
apparent on the standard ECG are even less sensitive than
ST-segment deviations.32 They are incapable of providing
the clinician with the information that appears to be
available in the high-resolution ECG.
From a clinical perspective, the most interesting result of
the present study is the presence of AIQP in patients without
any ST-segment deviation during the period of prolonged
balloon inflation. The strong correlation between AIQP
amplitudes and ST-segment deviations suggests that these
methods might have future potential for a noninvasive
clinical indication of the extent or the severity of acute
ischemia. Future studies with an independent measurement
of ischemia, such as tomographic scintigraphy,33 are needed
to investigate whether the HRECG could provide a more
accurate electrocardiographic measurement of ischemically
altered electrophysiologic changes.
Acknowledgments
This work was supported in part by CICYT grant
TEC2004-02274 from the Spanish government.
References
1. Langner PH. The value of high fidelity electrocardiography using the
cathode ray oscillograph and an expanded time scale. Am Heart J
1952;5:429.
2. Langner PH, Geselowitz DB, Mansure FT. High-frequency compo-
nents in the electrocardiograms of normal subjects and of patients with
coronary heart disease. Am Heart J 1961;62:749.
3. Flowers NC, Horan LG, Thomas R, et al. The anatomic basis for high
frequency components in the electrocardiogram. Circulation
1969;39:531.
4. Flowers NC, Horan LG, Tolleson WG, et al. Localization of the site of
myocardial scarring in man by high-frequency components. Circulation
1969;40:927.
5. Gomis P, Jones DL, Caminal P, Berbari EJ, Lander P. Analysis of
abnormal signals with the QRS complex of the high resolution
electrocardiogram. IEEE Trans Biomed Eng 1997;44:681.
6. Gomis P, Lander P, Caminal P. Parametric linear and non-linear
modeling techniques for estimating abnormal intra-QRS potentials in
the high resolution ECG. In: Cerrolaza M, Jugo D, Brebbia C, editors.
Simulation Modeling in Bioengineering. Southampton (UK)7 Compu-
tational Mechanics Publications; 1996. p. 271.
7. Lander P, Gomis P, Caminal P, Berbari EJ, Ballouz G, Patterson E, et al.
Pathophysiological insights into abnormal intra-QRS signals in the
high resolution ECG. Comput Cardiol 1995;22:273.
P. Lander et al. / Journal of Electrocardiology 39 (2006) 282–289 289
8. Abboud S, Cohen RJ, Selwyn A, Ganz P, Sadeh D, Friedman PL.
Detection of transient myocardial ischemia by computer analysis of
standard and signal-averaged high frequency electrocardiograms in
patients undergoing percutaneous transluminal coronary angioplasty.
Circulation 1987;76:585.
9. Pettersson J, Pahlm O, Carro E, Edenbrandt L, Ringborn M, Sfrnmo L,
et al. Changes in high-frequency QRS components are more sensitive
than ST-segment deviation for detecting acute coronary artery
occlusion. J Am Coll Cardiol 2000;36:1827.
10. Aversano T, Rudikoff B, Washington A, Traill S, Combs V, Raqueno J.
High frequency QRS electrocardiography in the detection of reperfu-
sion following thrombolytic therapy. Clin Cardiol 1994;17:175.
11. Berkalp B, Oral D, Caglar N, Omurlu K, Pamir G, Alpman A, et al.
Effects of percutaneous transluminal coronary angioplasty on late
potentials and high frequency mid-QRS potentials. Cardiology
1994;85:216.
12. Wagner GS, Selvester RH, Wagner NB, Krucoff MW. QRS changes
during acute ischemia induced by balloon occlusion of LAD artery.
J Electrocardiol 1988;21(Suppl):S18.
13. Lander P, Berbari EJ. Performance assessment of optimal filtering of
the high resolution electrocardiogram. Comput Cardiol 1994;21:693.
14. Lander P, Berbari EJ. Time-frequency plane Wiener filtering of the
high resolution ECG: Background and time-frequency representations.
IEEE Trans Biomed Eng 1997;44:247.
15. Lander P, Berbari EJ. Time-frequency plane Wiener filtering of the
high resolution ECG: Development and application. IEEE Trans
Biomed Eng 1997;44:256.
16. Lander P, Berbari EJ. Principles and signal processing techniques of
the high-resolution electrocardiogram. Prog Cardiovasc Dis 1992;
35:169.
17. Steinberg JS, Bigger Jr JT. Importance of the endpoint of noise
reduction in analysis of the signal-averaged electrocardiogram. Am J
Cardiol 1989;63:556.
18. Lander P, Berbari EJ, Rajagopalan CV, Vatterott P, Lazzara R. Critical
analysis of the signal-averaged electrocardiogram. Improved identifi-
cation of late potentials. Circulation 1993;87:105.
19. Lander P, Berbari EJ, Lazzara R. Optimal filtering and quality control
of the signal-averaged ECG: high-fidelity 1-minute recordings.
Circulation 1995;91:1495.
20. Lander P, Albert DE, Berbari EJ. Spectrotemporal analysis of
ventricular late potentials. J Electrocardiol 1990;23:95.
21. Ljung L. System identification: theory for the user, chapter 4 and 10.
Englewood Cliff (NJ)7 Prentice-Hall; 1987.
22. Lander P, Berbari EJ, Lazzara R, Gomis P, Steinberg JS. Analysis of
intra-QRS late potentials improves the predictive value of the signal-
averaged ECG. Pace 1995;18:796.
23. Lander P, Gomis P, Goyal RM, Berbari EJ, Caminal P, Lazzara R, et al.
Analysis of abnormal intra-QRS potentials: Improved predictive value
for arrhythmic events of the signal-averaged electrocardiogram.
Circulation 1997;95:1386.
24. Braunwald E, Antman EM, Beasley JW, et al. ACC/AHA guidelines for
the management of patients with unstable angina and non-ST-segment
elevation myocardial infarction. J Am Coll Cardiol 2000;36:970.
25. Atar S, Birnbaum Y. Ischemia-induced ST-segment elevation: classi-
fication, prognosis, and therapy. J Electrocardiol 2005;38(Suppl):1.
26. Peterson ED, Hathaway WR, Zabel KM, Pieper KS, Granger CB,
Wagner GS, et al. Prognostic significance of precordial ST segment
depression during inferior myocardial infarction in the thrombolytic
era: results in 16,521 patients. J Am Coll Cardiol 1996;28:305.
27. Mori-Avi V, Shargorodsky B, Abboud S, Laniado S, Akselrod S.
Effects of coronary occlusion on high frequency content of the
epicardial electrogram and body surface electrocardiogram. Circulation
1987;76:237.
28. Abboud S. High-frequency electrocardiogram analysis of the entire
QRS in the diagnosis and assessment of coronary artery disease. Prog
Cardiovasc Dis 1993;35:311.
29. Abboud S, Berenfeld O, Sadeh D. Simulation of high-resolution QRS
complex using a ventricular model with a fractal conduction system.
Effects of ischemia on high-frequency QRS potentials. Circ Res
1991;68:1751.
30. Lander P, Berbari EJ. Use of highpass filtering to detect late potentials
in the signal averaged ECG. J Electrocardiol 1989;22(Suppl):7.
31. Steg PG, Faraggi M, Himbert D, Juliard JM, Cohen-Solal A, Lebtahi
R, et al. Comparison using dynamic vector cardiography and MIBI
SPECT of ST-segment changes and myocardial MIBI uptake during
percutaneous transluminal coronary angioplasty of the left anterior
descending coronary artery. Am J Cardiol 1995;75:998.
32. Mager A, Sclarovsky S, Wurtzel M, Menkes H, Strasberg B, Rechavia
E. Ischemia and reperfusion during intermittent coronary occlusion in
man. Studies of electrocardiographic changes and CPK release. Chest
1991;99:386.
33. Gibbons RJ, Verani MS, Behrenbeck T, Pellikka PA, O’Connor MK,
Mahmarian JJ, et al. Feasibility of tomographic technetium-99m-
hexakis-2-methylpropyl-isonitrile imaging for the assessment of
myocardial area at risk and the effect of acute treatment in myocardial
infarction. Circulation 1989;80:1277.
top related