Defibrillator Implants:

Bruce Murdoch PhD CHP CLSOArgonne National LaboratoryArgonne, IL USAbtmurdoch@anl.gov

AIHce 2011, Portland OR Roundtable Session RT 201

Occupational Exposure to Electromagnetic Radiation

Typical ICD Device

Dimensions: 2” X 2.5” X 0.2”

AIHce 2011, Portland OR

2

Implant Placement

EM Vulnerabilities Leads Control Unit

Failure Modes False Trigger Stopping

AIHce 2011, Portland OR

3

Source: www.defibrillator-help.com

AIHce 2011, Portland OR

4

CASE STUDY

Where? APS (Advanced Photon Source)

Who? Accelerator Operator w/ICD

Why? Work Clearance

AIHce 2011, Portland OR

6

AIHce 2011, Portland OR

7

APS Control Room

Manufacturer Test Standard

ANSI/AAMA PC69 - 2007

AIHce 2011, Portland OR

8

Typical Manufacturer (Medtronic) Exposure Limit Recommendations Static B-Field B ≤ 0.5 mT (5G)

50/60 Hz B ≤ 0.1 mT (1G) E ≤ 6000 V/m

≥ 10 kHz B ≤ 1.25 mT (12.5 G)

R-F E-Field (≥ 150 kHz) E ≤ 100 V/m

AIHce 2011, Portland OR

9

ACGIH 2010 TLV Book Exposure Limit Recommendations

Static B-Field B ≤ 0.5 mT (5G)

ELF (5 Hz–2kHz) B ≤ 0.1 mT (1G)

VLF (2kHz–400 kHz) No recommendation

R-F (100 kHz–3GHZ) E ≤ 3-10 V/m

AIHce 2011, Portland OR

10

Static B-Fields

MetroLab Model ETM-1

3-Axis Magnetometer

AIHce 2011, Portland OR

11

ELF

Holaday HI-3627ELF Magnetic Field Meterw/ HI-3602 Probe

30 Hz – 2 KHz

AIHce 2011, Portland OR

12

VLF

Holaday HI-3637VLF Magnetic Field Meterw/ HI-3091 Probe

2kHz – 400 KHz

AIHce 2011, Portland OR

13

R-F

Narda NBM-520RF Survey Meterw/ EF3091 E-Field Probe

100 kHz – 3 GHz

AIHce 2011, Portland OR

14

Spectrum Coverage

AIHce 2011, Portland OR

15

RF

AIHce 2011, Portland OR

16

The Instrument Ensemble

Survey Plan

Daily Control Room work area

Weekly maintenance walkthroughs

ESH Coordinator guidance

84 specific survey locations plus area sweeps

Full spectrum survey at each location

AIHce 2011, Portland OR

17

Control Room Panels

AIHce 2011, Portland OR

19

Electrical Distribution RoomELF: B ≤ 0.4 G @ 6 inches from transformerRF: E ≤ 4 V/m

AIHce 2011, Portland OR

22

Mezzanine RF System Freq = 325 MHz E(RF) up to 3.5 V/m @ 6 in.

AIHce 2011, Portland OR

23

Power Control Systems

AIHce 2011, Portland OR

24

Pulser Power Transformers

AIHce 2011, Portland OR

25

RF Distribution Waveguides

Freq = 325 mHz

AIHce 2011, Portland OR

27

Linac KlystronsFreq = 2.8565 GHzE(RF) = 2 V/m @ 1 ft

AIHce 2011, Portland OR

29

Case Study Results

Employee barred from

– Fundamental RF Room: 7 V/m RF

– Electrical Distribution Room: 4 V/m RF

Cautions

– Some instrument racks: 4 V/m RF @ 12”

– Some waveguides: 3.5 V/m RF @ 12”

AIHce 2011, Portland OR