7/22/2019 Alcolyzer Plus Beer
1/129
Reference Manual
Alcolyzer Plus Beer
Beer Analyzing SystemFirmware Version: V5.02.09
-
7/22/2019 Alcolyzer Plus Beer
2/129
-
7/22/2019 Alcolyzer Plus Beer
3/129
Reference Manual
Alcolyzer Plus Beer
Beer Analyzing SystemFirmware Version: V5.02.09
-
7/22/2019 Alcolyzer Plus Beer
4/129
Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.
Nor is any liability assumed for damages resulting from information contained in the document.
Anton Paar GmbH reserves the right to content changes. This also extends to changes to delivery
volumes or any features of delivered parts.
All rights reserved (including translation). This document, or any part of it, may not be reproduced,
changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm
or any other process) without prior written permission by Anton Paar GmbH.
Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked
as such. They are the property of their respective owner.
Published by Anton Paar
Printed: Anton Paar, Austria
Copyright 2009 Anton Paar GmbH, Graz, Austria
Address: Anton Paar GmbH
Anton-Paar-Str. 20
A-8054 Graz / Austria Europe
Tel: +43 316 257-0
Fax: +43 316 257-257
E-Mail: [email protected]
Web: www.anton-paar.com
Date: 06.03.2009
Document number C27IB02M.fm
-
7/22/2019 Alcolyzer Plus Beer
5/129
C27IB02 5
Contents
1 About the Reference Manual .......................................................................................................... 9
2 Safety Instructions........................................................................................................................... 9
2.1 General Safety Instructions ..................................................................................................... 9
3 About Alcolyzer Plus Beer............................................................................................................ 11
4 Testing the Supplied Parts............................................................................................................ 11
5 Putting into Operation................................................................................................................... 14
5.1 Installing the Instruments....................................................................................................... 14
5.2 Mounting the pH Electrode .................................................................................................... 16
5.2.1 Preparing and Mounting the pH Electrode "Hamilton Polilyte Lab"........................ 16
5.2.2 Preparing and Mounting the pH Electrode "Orion model 91-63"............................ 175.3 Mounting the Hoses without Option pH/Color........................................................................ 19
5.4 Mounting the Hoses with Option pH/Color............................................................................. 20
5.5 Completing the Installation .................................................................................................... 21
5.6 Connecting a Printer or PC.................................................................................................... 24
5.6.1 Connecting Alcolyzer Plus Beer to a Printer .......................................................... 24
5.6.2 Connecting Alcolyzer Plus Beer to a PC................................................................ 24
6 Functional Components................................................................................................................ 25
6.1 Display and Keypad............................................................................................................... 25
6.1.1 Display in the Measuring Mode.............................................................................. 25
6.1.2 Soft Keys in the Measuring Mode .......................................................................... 27
6.2 Rear View .............................................................................................................................. 28
7 General Instrument Settings......................................................................................................... 29
7.1 Setting the Display Contrast .................................................................................................. 29
7.2 Setting Date and Time ........................................................................................................... 29
7.3 Configuring the Measuring System........................................................................................ 29
7.3.1 Setting the General Parameters............................................................................. 29
7.3.2 Defining the Peripheral Devices ............................................................................. 30
7.3.3 Defining the Interface Settings ............................................................................... 30
7.4 Setting a Custom Function .................................................................................................... 30
7.4.1 Setting a User-Defined Ethanol Table (in %V/V).................................................... 30
7.4.2 Setting a User-Defined Color Unit .......................................................................... 31
7.5 Selecting the Output Data...................................................................................................... 32
7.5.1 Selecting the Output Data for the Display .............................................................. 33
7.5.2 Selecting the Output Data for the Printer ............................................................... 33
7.5.3 Selecting the Output Data for the Data Memory..................................................... 33
7.6 Settings for the Sample Changer........................................................................................... 33
8 Setting the Measuring Methods.................................................................................................... 34
8.1 Description of the Measuring Methods .................................................................................. 348.2 Editing a Measuring Method.................................................................................................. 35
-
7/22/2019 Alcolyzer Plus Beer
6/129
6 C27IB02
8.2.1 Editing a Predefined Method.................................................................................. 35
8.2.2 Editing a New Method ............................................................................................ 35
8.3 Selecting the Measuring Method........................................................................................... 37
9 SOP (Standard Operating Procedure).......................................................................................... 38
10 Checking the Adjustment Validity................................................................................................ 41
10.1 Checking Procedure without Option pH/Color....................................................................... 41
10.1.1 Performing a Checking Procedure with Water (0%V/V)......................................... 41
10.1.2 Performing a Checking Procedure with a Water/Ethanol Solution of Known
Concentration ........................................................................................................................ 43
10.2 Checking Procedure with Option pH/Color............................................................................ 44
10.2.1 Checking Procedure for the Measuring Parameter "alcohol" ................................. 44
10.2.2 Checking Procedure for the Measuring Parameter "pH"........................................ 44
10.2.3 Checking Procedure for the Measuring Parameter "color"..................................... 45
11 Adjusting the Instrument .............................................................................................................. 4511.1 Adjustment without Option pH/Color...................................................................................... 45
11.1.1 Adjustment with Water (Zero Point)........................................................................ 46
11.1.2 Adjustment with a Water/Ethanol Solution of Known Alcohol Concentration ......... 48
11.2 Adjustment with Option pH/Color........................................................................................... 49
11.2.1 Adjustment Alcohol................................................................................................. 49
11.2.2 Adjustment Color.................................................................................................... 49
11.2.2.1 Adjustment with Water (Zero) ................................................................. 49
11.2.2.2 Adjustment with a Color Standard (Tartrazine) of Known Concentration 51
11.2.3 Adjustment pH........................................................................................................ 5211.2.3.1 Automatic ................................................................................................ 53
11.2.3.2 Manual .................................................................................................... 54
11.2.3.3 Edit Buffer Data....................................................................................... 55
11.2.3.4 Sort Buffer Data ...................................................................................... 56
11.3 Density Adjustment................................................................................................................ 56
12 Measurement.................................................................................................................................. 58
12.1 Identifying the Sample and Assigning the Method................................................................. 59
12.1.1 Types of Sample Identification ............................................................................... 59
12.1.2 Sample Identification for "text + number" and "text"............................................... 60
12.1.3 Sample Identification for "list" and "# + list"............................................................ 60
12.1.3.1 Entering a Sample List ............................................................................ 60
12.1.3.2 Editing the Sample List during Operation................................................ 62
12.1.3.3 Deleting a Sample List ............................................................................ 62
12.2 Performing a Measurement without Option pH/Color ............................................................ 63
12.3 Performing a Measurement with Option pH/Color................................................................. 64
12.4 Correcting Calculated Analysis Results................................................................................. 64
13 Data Memory .................................................................................................................................. 66
14 Cleaning Routines ......................................................................................................................... 67
14.1 Cleaning the Measuring Cells without Option pH/Color......................................................... 67
-
7/22/2019 Alcolyzer Plus Beer
7/129
C27IB02 7
14.1.1 Between Series of Measurements on the Same Day............................................. 67
14.1.2 At the End of the Day ............................................................................................. 67
14.1.3 Once a Week.......................................................................................................... 68
14.2 Cleaning the Measuring Cells with Option pH/Color.............................................................. 69
14.2.1 Storing the pH Electrode ........................................................................................ 69
14.2.2 Between Series of Measurements on the Same Day............................................. 70
14.2.3 At the End of the Day ............................................................................................. 70
14.2.4 Once a Week.......................................................................................................... 71
14.3 Cleaning after Wort Measurements....................................................................................... 72
14.4 Maintaining the pH Electrode................................................................................................. 72
14.4.1 Maintaining the Hamilton pH Electrode.................................................................. 72
14.4.2 Maintaining the Orion pH Electrode ....................................................................... 73
15 Menu Operation and Description ................................................................................................. 74
15.1 Menu Operation..................................................................................................................... 7415.1.1 Operating the Soft Keys on the Instrument ............................................................ 74
15.1.2 Using an External PC Keyboard............................................................................. 75
15.1.3 Using a Bar Code Reader ...................................................................................... 75
15.2 Menu Description and Function............................................................................................. 76
15.2.1 "adjustment" ........................................................................................................... 76
15.2.1.1 "adjust alcohol"........................................................................................ 76
15.2.1.2 "adjust pH" .............................................................................................. 77
15.2.1.3 "adjust color" ........................................................................................... 77
15.2.2 "instrument settings"............................................................................................... 7815.2.3 "method settings" ................................................................................................... 82
15.2.4 "user functions" ...................................................................................................... 83
15.2.5 "data memory"........................................................................................................ 83
15.2.6 "utilities".................................................................................................................. 84
15.2.7 "parameter calculator" ............................................................................................ 90
15.2.8 "service mode" ....................................................................................................... 91
15.2.9 "auto sampler settings"........................................................................................... 91
15.3 Entering and Changing a Password ...................................................................................... 91
15.4 Installing a New Firmware Version ........................................................................................ 92
16 Calculations.................................................................................................................................... 93
17 Filling Samples with the SF-1 Sample Filling Unit...................................................................... 97
Appendix A: Degassing Beer Samples.......................................................................................... 98
Appendix B: Available Parameters ................................................................................................ 99
B.1 Method Parameters............................................................................................. 99
B.2 Printer Protocol Parameters.............................................................................. 101
Appendix C: Status and Error Messages .................................................................................... 102
C.1 Status/Error Messages for the Alcohol Measurement....................................... 102
C.2 Status/Error Messages for the Color Measurement .......................................... 103
C.3 Status/Error Messages for the pH Measurement .............................................. 103
-
7/22/2019 Alcolyzer Plus Beer
8/129
8 C27IB02
C.4 System Error Messages.................................................................................... 104
C.5 Other Messages................................................................................................ 104
Appendix D: Technical Data ......................................................................................................... 105
Appendix E: Interface Commands for Communication PC and Alcolyzer Plus Beer ............. 106
Appendix F: Configuration for Communication between DMA and Alcolyzer Plus Beer ...... 108
Appendix G: Firmware Versions .................................................................................................. 110
Appendix H: Wetted Parts............................................................................................................. 112
Appendix I: Material Safety Data Sheet for Tartrazine.............................................................. 113
Appendix J: Density/Concentration Table for Binary Aqueous Ethanol Solutions ................ 115
Appendix K: Menu Tree................................................................................................................. 126
Index................................................................................................................................................... 128
-
7/22/2019 Alcolyzer Plus Beer
9/129
1 About the Reference Manual
C27IB02 9
1 About the Reference Manual
This reference manual informs you about the installation and the safe handling
and use of the instrument. Make sure that the reference manual is easily
accessible to all personnel involved with the instrument. Pay special attention to
the safety instructions and warnings in the manual and on the instrument.
Symbols in this reference manual
This reference manual features the following symbols:
2 Safety Instructions
Read the reference manual at hand before using Alcolyzer Plus Beer.
Follow all hints and instructions contained in this reference manual to ensure
the correct use and safe functioning of Alcolyzer Plus Beer.
2.1 General Safety Instructions
Liability
The reference manual at hand does not claim to address all safety issues
associated with the use of the instrument and samples. It is your
responsibility to establish health and safety practices and determine the
applicability of regulatory limitations.
Anton Paar GmbH only warrants the proper functioning of Alcolyzer Plus
Beer if no adjustments have been made to the mechanics, electronics, and
firmware.
Only use Alcolyzer Plus Beer for the purpose described in this reference
manual. Anton Paar GmbH is not liable for damages caused by incorrect use
of Alcolyzer Plus Beer.
Caution:
This symbol calls attention to the risk of instrument damage or
measurement errors. Do not proceed until the indicated conditions for
averting this threat are fully understood and met.
Information:
This symbol calls attention to any additional information of use to the
operator.
-
7/22/2019 Alcolyzer Plus Beer
10/129
2 Safety Instructions
10 C27IB02
Installation and use
Alcolyzer Plus Beer is not an explosion-proof instrument and therefore must
not be operated in areas with risk of explosion.
The installation procedure shall only be carried out by authorized personnel
who are familiar with the installation instructions.
Do not use any accessories or wearing parts other than those supplied or
approved by Anton Paar GmbH.
Make sure all operators are trained to use the instrument safely and correctly
before starting any applicable operations.
In case of damage or malfunction, do not continue operating Alcolyzer Plus
Beer. Do not operate the instrument under conditions which could result in
damage to goods and/or injuries and loss of life.
Check Alcolyzer Plus Beer for chemical resistance to the samples and
cleaning agents.
Maintenance and service
The results delivered by Alcolyzer Plus Beer not only depend on the correct
function of the instrument, but also on various other factors. We therefore
recommend to have the results checked (e.g. plausibility tested) by skilled
personnel before consequential actions are taken based on the results.
Service and repair procedures may only be carried out by authorizedpersonnel or by Anton Paar GmbH.
Disposal
Concerning the disposal of Alcolyzer Plus Beer, observe the legal
requirements in your country.
Precautions for highly flammable samples and cleaning agents
Observe and adhere to your national safety regulations for handling the
measured samples (e.g. use of safety goggles, gloves, respiratory protection
etc.).
Only store the minimum required amount of sample, cleaning agents and
other inflammable materials near the Alcolyzer Plus Beer.
Do not spill sample/cleaning agents or leave their containers uncovered.
Immediately remove spilled sample/cleaning agents.
Make sure that the setup location is sufficiently ventilated. The environment
of Alcolyzer Plus Beer must be kept free of inflammable gases and vapors.
-
7/22/2019 Alcolyzer Plus Beer
11/129
3 About Alcolyzer Plus Beer
C27IB02 11
Connect Alcolyzer Plus Beer to the mains via a safety switch located at a
safe distance from the instrument. In an emergency, turn off the power using
this switch instead of the power switch on Alcolyzer Plus Beer.
Supply a fire extinction unit.
Ensure the sufficient supervision of Alcolyzer Plus Beer during operation.
3 About Alcolyzer Plus Beer
Alcolyzer Plus Beer is based on a highly accurate and patented NIR
spectrometer, which measures the alcohol content of beer in the concentration
range of 0 - 12%V/V.
Optionally, Alcolyzer Plus Beer can also be equipped with a color measurement
device and a pH measurement device. The color value (EBC or ASBC value) is
determined by the measurement of light absorption. For the pH measurement, an
electrode Hamilton (Polilyte Lab) or a measuring cell Orion (model 91-63) is used.
In combination with an Anton Paar density meter DMA 4500/5000 and a sample
changer SP-1m, the automatic determination of relevant beer parameters can be
achieved.
4 Testing the Supplied PartsAlcolyzer Plus Beer was tested and packed carefully before shipment. However,
damage may occur during transport.
1. Keep the packaging material (box, foam piece, transport protection) for
possible returns and further questions from the transport and insurance
company.
2. Check the delivery for completion by comparing the supplied parts to those
noted in Table 4.1.
3. If a part is missing, contact your Anton Paar representative.
4. If a part is damaged, contact the transport company and your Anton Paar
representative.
http://-/?-http://-/?- -
7/22/2019 Alcolyzer Plus Beer
12/129
4 Testing the Supplied Parts
12 C27IB02
Table 4.1: Supplied Parts
Symbol Pcs. Item/Spare Part Mat. No.
1 Alcolyzer Plus Beer (without option pH/color)
or
Alcolyzer Plus Beer
with option color + pH Hamilton or
with option color + pH Orion
8921
8921
9918
12167
1 Instruction manualEnglish or
German
including a CD with the reference manual
12665
12666
1 Power cable
USA:
UK:
Europe:
52656
61865
65146
1 Accessory kit 9860
2m Hose 3 x 5 silicone / 1m
Hose 1.6 x 3.2 Tygon / 1m
50814
67540
5 Syringe 5 mL / 1pc. 63130
2 Filling nozzle Luer Alcolyzer Plus / 1pc. 7046
2 Adapter Luer " UNF / 1pc. 64792
2 Male Luer plug / 1pc. 63865
-
7/22/2019 Alcolyzer Plus Beer
13/129
4 Testing the Supplied Parts
C27IB02 13
3 Adapter Luer cone / 1pc. 63863
2 Filling nozzle fitting Alcolyzer Plus / 1pc. 10759
2 Filling nozzle UNF " Alcolyzer Plus / 1pc. 10758
2 Tube d=2.1, D=2.4, L=20 1.4571 / 1pc. 67982
1 Needle (only delivered with SF-1) / 1pc. 67394
3m Hose 2 x 4 silicone /1m
(only delivered with SF-1)
51273
1 Cable male 9 Pin / male 9 Pin / 1pc. 2359
1
(option)
Backup Card DMA 4500/5000 Alcolyzer for
firmware versions DMA V4.xxx or higher;
Alcolyzer V5.xx / 1pc.
23913
1 Waste vessel 500 mL 6210
1 Option color + pH Hamilton 9918
1 Electrode Hamilton pH = 0-14 0-80 9466
1 Option color + pH Orion 12167
1 Electrode Orion model 91-63 12555
Table 4.1: Supplied Parts
Symbol Pcs. Item/Spare Part Mat. No.
-
7/22/2019 Alcolyzer Plus Beer
14/129
5 Putting into Operation
14 C27IB02
5 Putting into Operation
5.1 Installing the Instruments
1. Put SP-1m into operation as described in the SP-1m instruction manual.
2. Connect the power cord of SP-1m to a power terminal with ground
connection.
3. Place the density meter on SP-1m.
4. Connect the connecting power cable (SP-1m accessory, Mat. No. 66178) to
the power outlet of SP-1m and the power outlet of the DMA.
5. Mount the filling nozzle fittings as described in the SP-1m instruction manual.
6. Place Alcolyzer Plus Beer on top of the density meter.
7. Connect the power cord of Alcolyzer Plus Beer to a power terminal with
ground connection.
Delivered with both
options color + pH
1 Connection cable type K1 BNC 1m 62521
1 Buffer solution pH=4.01 Hamilton (500 mL) 125541 Buffer solution pH=7.00 Hamilton (500 mL) 12556
1 pH storage solution 500 mL Hamilton 12632
1 Color adjustment set
Color standard approx. 50 EBC 2x50 mL
Ultra pure water 0.0 EBC 3x50 mL
12627
Table 4.1: Supplied Parts
Symbol Pcs. Item/Spare Part Mat. No.
Information:
Alcolyzer Plus Beer does not require any special installation conditions.
Conditions in the installation area should be those in a typical laboratory.
To guarantee temperature stability, do not place Alcolyzer Plus Beer
- near a heater
- near an air conditioner
- in direct sunlight.
For air humidity restrictions seeAppendix D.
Caution:
To ensure the correct functioning of the built-in ventilator, make sure the
airflow at the bottom and the rear of the instrument is not blocked.
-
7/22/2019 Alcolyzer Plus Beer
15/129
5 Putting into Operation
C27IB02 15
8. Screw the filling nozzle fittings Alcolyzer Plus (accessory box) into the
openings of the measuring cell.
9. If the option pH/color is integrated, check that 2# Filling nozzle Luer
Alcolyzer Plus adapters are screwed into the openings of the pH measuring
cell (see Fig. 5 - 1).
Fig. 5 - 1 Mounting the adapters onto the pH measuring cell
10. Connect COM2 at the rear of Alcolyzer Plus Beer to COM1 at the rear of
DMA 4500/5000 using the connection cable (Mat. No. 2359).
11. If a printer is used, connect the printer interface cable to COM1 at the rear of
Alcolyzer Plus Beer.
12. Connect the interface cable of SP-1m to the S-Bus outlet of Alcolyzer Plus
Beer.
Caution:
Screw the adapters into the openings of the measuring cell hand-tight(without using any tools). Otherwise, the connections between the adapters
and the measuring cell are not leak-tight and liquid may enter the electronics
and affect operation.
Caution:
Before switching on Alcolyzer Plus Beer, check that the given voltage and
frequency (AC100 to 240 V, 50 to 60 Hz) corresponds with the local mains
voltage and mains frequency. If large voltage fluctuations are to be
expected, the use of a constant voltage source (UPS) is recommended.
The non-fused earth conductor of the power cord must be connected to
earth.
-
7/22/2019 Alcolyzer Plus Beer
16/129
5 Putting into Operation
16 C27IB02
5.2 Mounting the pH Electrode
5.2.1 Preparing and Mounting the pH Electrode "Hamilton
Polilyte Lab"
Preparing
1. Leave the pH electrode in an upright position for at least one hour.
2. Carefully remove the watering cap.
3. Flush the electrode with distilled water.
4. Check the inside of the electrode for unwanted air bubbles. Shaking the
electrode like a clinical thermometer will cause any bubbles to rise to the top.
Mounting
1. Loosen the protective tube by twisting it counterclockwise (see Fig. 5 - 2).
Fig. 5 - 2 Loosening the protective tube
2. Screw the connecting cable (Mat. No. 62521) to the rear end of the pH
electrode.
3. Connect the other end of the cable to the rear of Alcolyzer Plus Beer (see
Chapter 6.2).
4. Insert the pH electrode into the protective tube carefully until the end of the
electrode touches the bottom of the pH measuring cell.
5. Tighten the protective tube by turning it clockwise.
The pH electrode is fixed.
-
7/22/2019 Alcolyzer Plus Beer
17/129
5 Putting into Operation
C27IB02 17
Fig. 5 - 3 pH electrode Hamilton
6. Check that both Filling nozzle Luer Alcolyzer Plus are screwed into the
openings of the pH measuring cell (see Fig. 5 - 1).
5.2.2 Preparing and Mounting the pH Electrode "Orion
model 91-63"
Preparing
1. Remove the protective shipping cap from the sensing element and save for
storage.
2. Clean any salt deposits from the exterior by rinsing with distilled water.
3. Uncover the filling hole and add filling solution to the electrode.
4. Place the pH electrode in the electrode holder and suspend in air for 15
minutes to thoroughly wet the reference junction.
5. Shake the electrode (like a clinical thermometer) to remove air bubbles.
6. Soak the electrode in pH electrode storage solution for one hour.
Information:
To maintain an adequate flow rate, the level of filling solution must always
be above the reference junction.
The filling hole must be open whenever the electrode is in use.
Caution:
Once the junction is wet, do not allow the electrode to dry out.
-
7/22/2019 Alcolyzer Plus Beer
18/129
5 Putting into Operation
18 C27IB02
Mounting
1. Screw the connecting cable (Mat. No. 62521) to the rear end of the pH
electrode.
2. Connect the other end of the cable to the rear of Alcolyzer Plus Beer (see
Chapter 6.2).
3. Insert the pH electrode into the pH measuring cell carefully until the end of
the electrode touches the bottom of the pH measuring cell.
Fig. 5 - 4 pH electrode Orion
4. Tighten the knurled nut (see Fig. 5 - 4). The pH electrode is fixed.
5. Check that both Filling nozzle Luer Alcolyzer Plus are screwed into the
openings of the pH measuring cell (see Fig. 5 - 1).
Knurled nut
-
7/22/2019 Alcolyzer Plus Beer
19/129
5 Putting into Operation
C27IB02 19
5.3 Mounting the Hoses without Option pH/Color
Fig. 5 - 5 Hose connections without option pH/color
1. Cut a piece of the silicone hose 3 x 5 mm (1) approx. 24 cm in length.
2. Connect one end of the hose to the air pump outlet of the DMA.
3. Connect the Adapter Luer cone to the other end of the hose.
4. Connect a Tygon hose (2) approx. 26 cm in length to the upper end of the
SP-1m needle and to the front adapter of the density meter.
5. Connect a Tygon hose (3) approx. 22 cm in length to the rear adapter of the
density meter and to the lower adapter of Alcolyzer Plus Beer.
6. Connect a Tygon hose (4) approx. 22 cm in length to the upper adapter of
Alcolyzer Plus Beer.
7. Connect the tube, Mat. No. 67982, (5) to the other end of the hose.
8. Connect a silicone hose 2x4 mm (6) to the tube.
9. Lead the silicone hose via the peristaltic pump of SP-1m to a waste vessel.
10. The length of this hose can be selected according to the distance of the
waste vessel.
4 (T)
3 (T)
2 (T)
5 (Tube)
6 (S) to waste
6
1 (S)
(T) .... Tygon
(S) .... Silicone
-
7/22/2019 Alcolyzer Plus Beer
20/129
5 Putting into Operation
20 C27IB02
5.4 Mounting the Hoses with Option pH/Color
Fig. 5 - 6 Hose connections with option pH/color
1. Connect an adapter Luer cone (1) to each opening of the pH measuring cell.
2. Cut a piece of the silicone hose 3 x 5 mm (2) approx. 24 cm in length.
3. Connect one end of the hose to the air pump outlet of the DMA.
4. Connect the adapter Luer cone to the other end of the hose.
5. Connect a Tygon hose (3) approx. 26 cm in length to the upper end of the
SP-1m needle and to the front adapter of the density meter.
6. Connect a Tygon hose (4) approx. 22 cm in length to the rear adapter of the
density meter and to the lower adapter of Alcolyzer Plus Beer.
7. Connect a Tygon hose (5) approx. 22 cm in length to the upper adapter of
Alcolyzer Plus Beer.
8. Connect the tube, Mat. No. 67982, (6) to the other end of the hose.
9. Connect a silicone hose 2x4 mm (7) approx. 50 cm in length to the tube.
10. Lead the silicone hose via the peristaltic pump of SP-1m to the front adapterof the pH measuring cell.
5 (T)
8 (S) to waste
3 (T)
6 Tube
7 (S)
4 (T)
2 (S)
1
(T) .... Tygon
(S) .... Silicone
to airpump
-
7/22/2019 Alcolyzer Plus Beer
21/129
5 Putting into Operation
C27IB02 21
11. Connect a 2x4 mm silicone hose (8) to the rear adapter of the pH measuring
cell.
12. Lead this hose to a waste vessel.
13. Select the length of this hose according to the distance of the waste vessel.
5.5 Completing the Installation
1. Turn on the SP-1m and the density meter and wait for the initialization of
both instruments to be completed.
2. Configure the DMA 4500/5000 according toAppendix F.
3. Turn on Alcolyzer Plus Beer using the power switch at the rear of the
instrument.
4. Select the menu instrument settings > device listand configure the device
list as follows (for details about operating the menu, see Chapter 15.1):
Caution:
Before switching on Alcolyzer Plus Beer, check that the given voltage and
frequency (AC 100 to 240 V, 50 to 60 Hz) corresponds with the local
mains voltage and mains frequency. If large voltage fluctuations are to be
expected, the use of a constant voltage source (UPS) is recommended.
The non-fused earth conductor of the power cord must be connected to
earth.
Information:
Once switched on, Alcolyzer Plus Beer performs a system check and anattemperating procedure:
After transport, full stabilization of Alcolyzer Plus Beer might take up to
one day.
Usually Alcolyzer Plus Beer is ready for measurement after a start-up
period of one hour.
To guarantee an optimal temperature stability over a long time, always
leave the system switched on.
-
7/22/2019 Alcolyzer Plus Beer
22/129
5 Putting into Operation
22 C27IB02
Fig. 5 - 7 Menu: instrument settings/device list
Now Alcolyzer Plus Beer initializes the communication between all
connected peripherals.
The COM2 interface of Alcolyzer Plus Beer is automatically configured
(connected to: PIB, baud rate: 9600, format: 8 bit No) and cannot be
changed unless densityis set to not used.
If all devices are detected properly, the following message is displayed:
Fig. 5 - 8 Device check
5. If for a certain peripheral the message "offline" is displayed, check the
interface settings and/or connection of the corresponding connection cable
(seeAppendix F).
INSTRUMENT SETTINGS
filling devicedensityCO2O2
SP-1m
device list
::::
DMA xxxx series
not usednot used
DEVICE CHECK
filling unitdensity :: SP-1mpassed
-
7/22/2019 Alcolyzer Plus Beer
23/129
5 Putting into Operation
C27IB02 23
Fig. 5 - 9 Device check failed
To repeat the device check, select device check.
To edit the device list, select change device list.
6. Set the control parameters for SP-1m in the menu auto sampler settings
(see Chapter 15.2.9).
To make sure that the measured sample is fully replaced with the next
sample, we recommend using the control parameter 1100 (100 seconds
filling time, approx. 30 mL sample). For further information, see the SP-1m
instruction manual.
Fig. 5 - 10 Measuring system: Alcolyzer Plus Beer Analyzing System
BUS ERROR
filling unit : offline
device check 20 secchange device list
-
7/22/2019 Alcolyzer Plus Beer
24/129
5 Putting into Operation
24 C27IB02
5.6 Connecting a Printer or PC
5.6.1 Connecting Alcolyzer Plus Beer to a Printer
Connect the printer cable (Mat. No. 64433) to the COM1 on the rear side of
Alcolyzer Plus Beer.
Depending on the printer that you use, you can choose between 4 different printer
types for optimal communication between the connected printer and Alcolyzer
Plus.
Set the COM1 interface to the appropriate printer in the menu instrument
settings > COM1.
5.6.2 Connecting Alcolyzer Plus Beer to a PC
To transfer the measured results to a PC, connect the PC to Alcolyzer Plus Beer.
For further details, seeAppendix F.
1. Connect the interface cable DMA-PC (Mat. No. 72979) to the PC and to
COM1 on the rear side of Alcolyzer Plus Beer.
2. If you use the Anton Paar software VisioLab for Alcolyzer, set the COM1
interface to VisioLabin the menu instrument settings > COM1.
3. If you use the Anton Paar software AP-SoftPrint, set the COM1 interface to
AP-SoftPrintin the menu instrument settings > COM1.
Printer type Holding time
after 1 line
End-of-line
character
Remarks
printer_1 800 msec LF CR after 40
characters
printer_2 1500 msec LF
printer_3 800 msec LF
printer_4 10 msec LF+CR
AP-SoftPrint 10 msec LF
-
7/22/2019 Alcolyzer Plus Beer
25/129
6 Functional Components
C27IB02 25
6 Functional Components
6.1 Display and Keypad
After switching on, Alcolyzer Plus Beer performs a self-test and the initialization.
6.1.1 Display in the Measuring Mode
Fig. 6 - 1 Display after the end of the measurement
A ... MethodThe selected method is displayed.
B ... Measuring status of the alcohol measurementFor details, see the reference manual, Appendix B.1.
C ... Time
D ... Sample nameE ... Measuring status of the color measurement
For details, see the reference manual, Appendix B.2.
F ... Measuring status of the pH measurementFor details, see the reference manual, Appendix B.3.
G ... Measuring status of the density measurementThe current status of the density measurement is displayed.
H ... Display area for measured values
I ... Information about the sample changer
J ... Soft keys
Information:
It may be necessary to adjust the contrast of the display with the black knob
"Display contrast" at the rear of the instrument.
BEERvalid09:44:21sample #validvalidvalid
SP-1m
%V/V 20 CBeer3c 7.4 EBCpH 4.43
1.00521 g/cm3
: stopped
A
B
C
D
E
F
G
H
I
J
-
7/22/2019 Alcolyzer Plus Beer
26/129
6 Functional Components
26 C27IB02
Display of measured values:
The alcohol content is displayed in %V/V (2 trailing digits) at 20C (percent
volume per volume at the indicated temperature). If the measurement is not
finished, "....." is displayed. As long as the measurement is still in progress
("predetermined"), only one trailing digit is displayed. If the measuring resultis lower than -0.50%V/V or higher than 30.00%V/V, "low" or "high" is
displayed.
The color measurement is displayed in EBC or ASBC with one trailing digit
at 20C. If the measurement is not finished, "....." is displayed. As long as the
measurement is still in progress ("predetermined"), no trailing digit is
displayed. If the measuring result is lower than 0.5 EBC (-0.25 ASBC) or
higher than 130.0 EBC (66 ASBC), "low" or "high" is displayed.
The pH value is displayed in pH units (2 trailing digits) at 25C. If the
measurement is not finished, "....." is displayed.
Fig. 6 - 2 Display of further parameters
K ... Display area for calculated valuesSelectable, calculated parameters are displayed in this area after pressing
the key (see Chapter 7.5).
BEERvalid09:44:21sample #validvalidalcoholpErEaRDF
SP-1m
%V/V 20 CBeer3c 7.4 EBCpH 4.43 3.86 %m/m
11.12 %m/m 3.59 %m/m 1.80 %m/m 68.95 %
: stopped
KK
-
7/22/2019 Alcolyzer Plus Beer
27/129
6 Functional Components
C27IB02 27
6.1.2 Soft Keys in the Measuring Mode
Toggle key:
To toggle between displays of measured andcalculated values
To enter the menu (the key is
available)
Selects the measuring method.
Selects the main menu for various settings and
configurations. Access to the main menu can be
protected by a password.
Starts printing.
This soft key is only displayed when interface
COM1 or COM2 is configured for a printer in the
menu instrument settings.
Switches the air pump on.
Switches the air pump off.
Starts an automatic measuring process.
If a sample changer SP-1m is connected to
Alcolyzer Plus Beer and activated, the key
on Alcolyzer Plus Beer is not displayed. Start the
measurement with the key on the sample
changer.
Starts the next measurement.
To enter the sample identification.
Selects the list for the sample identification.
Confirms the changes.
Enters the SOP menu.
-
7/22/2019 Alcolyzer Plus Beer
28/129
6 Functional Components
28 C27IB02
6.2 Rear View
Fig. 6 - 3 Rear view
A ... Cable of the pH electrode
B ... pH electrode
C ... "air pump inlet": Connection for a desiccator
D ... "pH / color Module": Connection for the pH electrode
E ... "COM2": Serial interface RS 232 for connecting a density meter or a
printer
F ... "Keyboard" Connection for a keyboard and/or bar code reader
G ... Power inlet
H ... Power switch
I ... "COM1": Serial interface RS 232 for connecting a computer or printer
J ... "S-Bus": Connection for a sample changer
K ... "Display contrast"L ... Display contrast setting
M ... Cooling fins
A
B
C D E F G
I
J
KL
H
-
7/22/2019 Alcolyzer Plus Beer
29/129
7 General Instrument Settings
C27IB02 29
7 General Instrument Settings
7.1 Setting the Display Contrast
If necessary, adjust the display contrast with the "Display contrast" knob at the
rear of the instrument (see Chapter 6.2).
7.2 Setting Date and Time
Set date and time in the menu instrument settings > timeor dateusing the soft
keys , or . Press to confirm the settings.
7.3 Configuring the Measuring System
1. Press and to enter the main menu.
2. Select the menu instruments settings using the soft keys and .
Confirm with .
Fig. 7 - 4 Menu: instrument settings
7.3.1 Setting the General Parameters
Select the menu configurationto set general instrument settings.
Use the soft keys , and the key to select the parameter.
Press to confirm the settings.
Information:
For a detailed description of all setting options, see Chapter 15.2.2.
INSTRUMENT SETTINGS
device listCOM1COM2timedate
configuration
-
7/22/2019 Alcolyzer Plus Beer
30/129
7 General Instrument Settings
30 C27IB02
7.3.2 Defining the Peripheral Devices
Change to the menu instrument settings > device list to select the
connected peripherals like density meter, sample changer SP-1m or filling
device SF-1.
7.3.3 Defining the Interface Settings
Configure the two interfaces COM1or COM2for computer or printer in the
menu instrument settings.
7.4 Setting a Custom Function
You can define customized ethanol tables and color units.
7.4.1 Setting a User-Defined Ethanol Table (in %V/V)
If you want to use a different density/ethanol table than the OIML %V/V table, you
can edit a polynomial function (4th order) that calculates an ethanol value based
on your specific density/ethanol table.
Calculating the coefficients
The input value for the polynomial function is the ethanol %V/V value (c)
measured by Alcolyzer Plus Beer. The result of the calculation depends on the
entered coefficients according to the following formula.
Polynomial formula: Output (%V/V) = A0 + A1 x (c [%V/V]) + A2 x (c [%V/V]) 2+
A3 x (c [%V/V])3+ A4 x (c [%V/V])4
The default values are the polynomial coefficients valid for the OIML %V/V table(see Fig. 7 - 5).
Caution:
You can only use density/ethanol tables based on %V/V.
http://-/?-http://-/?- -
7/22/2019 Alcolyzer Plus Beer
31/129
7 General Instrument Settings
C27IB02 31
Fig. 7 - 5 Default setting of the polynomial coefficients
1. To calculate the coefficients, take the appropriate table (%V/V) that converts
the measured density into a corresponding ethanol value.
2. By using an appropriate mathematical program, you can calculate the
polynomial coefficients which are based on the table data.
Editing the user-defined ethanol table
1. Select the menu user functions > user defined ethanol table.
2. Enter the 5 polynomial coefficients (A0 to A4) as calculated.
Activating the user-defined ethanol table
1. At Alcolyzer Plus Beer, select the menu instrument settings >
configuration > calc. parameter.
2. Select CU V/Vand press the key.
7.4.2 Setting a User-Defined Color Unit
If the color measurement according to EBC or ASBC is not desired, you can
define a user-defined color unit. This is done by correcting the EBC value c [EBC]
according to the formula shown in Fig. 7 - 6.
Editing the user-defined color unit
1. To enter the correction factor for the slope of the color measurement and the
offset, select the menu user functions > user defined color unit.
USER FUNCTIONS
user defined ethanol table
A0 : 9.982000e-01A1 : - 1.520000e-03A2 : + 1.924620e-05A3 : - 1.461290e-07A4 : - 4.574060e-09
+
Caution:
To obtain a correct alcohol adjustment of Alcolyzer Plus Beer, use the same
ethanol table in the DMA that you used for calculating the polynomial
coefficients (seeAppendix F).
http://-/?-http://-/?- -
7/22/2019 Alcolyzer Plus Beer
32/129
7 General Instrument Settings
32 C27IB02
The following window is displayed:
Fig. 7 - 6 Menu: User functions/user defined color unit
2. Enter a slope value between 0 and 7.
Use the key to move the cursor. Edit the highlighted number using the and keys. Press to confirm the changes.
3. Enter an offset value between -10 and +10.
If not allowed values are entered, the slope is automatically set to 1.0 and
the offset to 00.0 after pressing .
Activating the user-defined color unit
1. Select the menu instrument settings > configuration > color unit.
2. Select CUand press the key.
7.5 Selecting the Output Data
When you select the parameters for display, printer and memory, you get a list
containing all parameters that are available for the predefined methods BEER,
CIDER, SOFT DRINK, WASH and MOLASSES (seeAppendix B.1).
If you then select your appropriate measuring method, you get only the
parameters that are of importance for this very method.
Example: If you select the parameter p (original extract) [%Plato], this parameter
is only available in the methods BEER, WASH and MOLASSES.
USER FUNCTIONS
user defined color unit
c [CU] = .000 * c [EBC] + 00.001
-
7/22/2019 Alcolyzer Plus Beer
33/129
7 General Instrument Settings
C27IB02 33
7.5.1 Selecting the Output Data for the Display
Change to the menu utilities > display configuration.
Select up to 5 parameters using the soft keys , and thekey.
Press to confirm the settings.
7.5.2 Selecting the Output Data for the Printer
You can select output data for the printer and save this selection in a printer
protocol. A printer protocol consists of a head and parameters. There are 4
factory-set printer protocols and 6 free printer protocols available.
Factory-set printer protocols:
- Printer protocol No. 6, used with BEER method
- Printer protocol No. 7, used with CIDER method
- Printer protocol No. 8, used with WASH and MOLASSES method
- Printer protocol No. 9, used with SOFT DRINK method.
For details of the settings, seeAppendix B.2.
If required, you can change these pre-settings according to your demands.
To define or change a printer protocol, change to the menu utilities > output
configuration > printer protocol:
To activate/deactivate the parameters, use the key.
Press to confirm the settings.
7.5.3 Selecting the Output Data for the Data Memory
Change to the menu utilities > output configuration > data memory.
You can activate up to 10 parameters to be stored in the data memory:
To activate/deactivate the parameters, use the key.
Press to confirm the settings.
7.6 Settings for the Sample Changer
Enter the control parameter for SP-1m in the menu auto sampler settings
as described in the SP-1m instruction manual.
Information:
If data should be sent to a LIMS system, activate additionally the parameters
alcohol cond.and color cond.
-
7/22/2019 Alcolyzer Plus Beer
34/129
8 Setting the Measuring Methods
34 C27IB02
This menu is only displayed if a sample changer is connected and configured
in the menu instrument settings > device list > filling device.
8 Setting the Measuring Methods
8.1 Description of the Measuring Methods
Alcolyzer Plus Beer provides 10 different methods:
Fig. 8 - 1 Menu: method settings
Method 1: BEER
Standard method for beers. This method can be used for almost all samples.
Method 2: CIDER
Method for ciders.
Method 3: SOFT DRINK
Method for alcohol-free beverages.
Method 4: WASH
Method for wash. Wash is an intermediate product of whisky production. It is an
alcohol containing liquid produced in the fermentation of worts by yeasts. The
alcohol content in wash varies between 7 and 10%vol.
Information:
We recommend setting the following filling times: Without option pH/color: 80 seconds (sample changer parameter "1080").
With option pH/color: 100 seconds (sample changer parameter "1100").
METHOD SETTINGS
no. 1no. 2no. 3no. 4no. 5no. 6no. 7no. 8
no. 9no. 10
::::::::
::
BEERCIDERSOFT DRINKWASH
new method
MOLASSES
-
7/22/2019 Alcolyzer Plus Beer
35/129
8 Setting the Measuring Methods
C27IB02 35
Method 5: MOLASSES
Method for molasses. Molasses is a honey-like, dark-brown sugar syrup which is
used as a fermentation raw material in the production of bioethanol. With this
method, you can determine the ethanol content of fermenting or ready-fermented
molasses.
Methods 6 to 10:
These methods are sample-specific and can be edited as described in Chapter
8.2.
8.2 Editing a Measuring Method
8.2.1 Editing a Predefined Method
You can edit the predefined methods "BEER", "CIDER", "SOFT DRINK", "WASH"
or "MOLASSES" to assign another printer protocol or enter a residual CO2content.
1. In the menu method settings, select one of the predefined methods and
confirm with .
2. In protocol no.:, you can select one of 10 printer protocols to assign it to the
selected method (For editing a printer protocol, see Chapter 7.5.2).
8.2.2 Editing a New Method
You can edit a new method to assign a special printer protocol (see Chapter 7.5.2)
or to enter a sample-specific alcohol correction factor (see Chapter 12.4). For the
types "BEER", "CIDER" and "SOFT DRINK", the residual CO2can be entered
(see Chapter 8.2.1).
1. Select a row between 6 and 10 in the menu method settings.
"new method" is displayed.
2. Confirm with .
3. Enter a name for the method and confirm the method name with .
Information:
If you measure fermenting molasses, we recommend to filtrate the sample
prior to measurement.
Information:The menu item residual co2 is only used with PBA-B and the methods
"BEER", "CIDER" and "SOFT DRINK".
-
7/22/2019 Alcolyzer Plus Beer
36/129
8 Setting the Measuring Methods
36 C27IB02
4. Select a type"BEER", "CIDER", "SOFT DRINK", "WASH" or "MOLASSES"
for the new method.
The type has to correspond with the sample to be measured.
5. If necessary, enter an offset between 0.99%vol and +0.99%vol (seeChapter 12.4).
6. If necessary, enter a pH-offset between -0.99 and +0.99 in pH-deg. offset.
7. In protocol no.:,you can assign a printer protocol from 1 to 10 to the new
edited method.
For details about editing a printer protocol, see Chapter 7.5.2.
8. Press to confirm the settings.
Fig. 8 - 2 Menu: method settings/new method/name
EDIT METHOD no. 6
typeoffsetpH-deg.offset
BEER:::
+0.50 %V/V
name special beer:
protocol no.residual co2
::
+0.50050.00 g/l
Information:
The menu item residual co2 is only used with PBA-B and the methods
"BEER", "CIDER" and "SOFT DRINK".
-
7/22/2019 Alcolyzer Plus Beer
37/129
8 Setting the Measuring Methods
C27IB02 37
8.3 Selecting the Measuring Method
1. Press the key in the measuring window.
The following window is displayed:
Fig. 8 - 3 Selecting a method
2. Use the key to toggle between the different methods.
3. Select the appropriate method (e.g. "BEER") with the key.
The selected method is displayed in the left upper corner.
BEER
BEERvalid09:44:21
validvalid
method :
%V/V 20 C
c 7.4 EBCpH 4.43
Information:
If you have assigned another method to a certain sample in the sample list,
this method instead of the selected method is used to measure the sample
(see Chapter 12.1.3.1).
-
7/22/2019 Alcolyzer Plus Beer
38/129
9 SOP (Standard Operating Procedure)
38 C27IB02
9 SOP (Standard Operating Procedure)
To ensure that Alcolyzer Plus Beer gives correct results, check the adjustments
of the measured parameters regularly.
The integrated SOP (Standard Operating Procedure) allows you to perform an
automated control measurement of all measuring parameters as well as an
adjustment afterwards.
Carrying out the SOP
1. Select the menu utilities > SOP configuration and choose SOP #1 or
SOP #2.
"SOP #1" is defined as the "routine" SOP and contains the parameters water,
alcohol, pH1 and pH2.
2. Place the appropriate control solutions in the correct order into the sample
changer magazine (for preparing the control solutions, see Chapter 10).
The following control solutions are used for the SOP:
- Bi-distilled water to check the zero point of the alcohol and the color
adjustment
- A binary ethanol/water solution with an alcohol content of 8 to 12%V/V to
check the "known concentration" of the alcohol adjustment
- pH buffer pH 7.00 and
- pH buffer pH 4.01 to check the pH adjustment
3. Press the key in the measuring window of Alcolyzer Plus Beer to
access the following menu:
Information:
The SOP can only be used in combination with a sample changer.
Information:
Carry out the "routine" SOP once a day. If desired, an additional SOP
(SOP #2) can be defined as an "extended" SOP with additional parameters
(e.g. color standard).
-
7/22/2019 Alcolyzer Plus Beer
39/129
9 SOP (Standard Operating Procedure)
C27IB02 39
Fig. 9 - 1
Use the key to toggle between "SOP #1" and "SOP #2".
4. Press the key to start the SOP.
5. Press the key on the sample changer.
The control solutions are measured automatically.
After the last measurement, the following window is displayed:
Fig. 9 - 2
6. If a measured value deviates significantly from the desired value, this value
can be readjusted now.
STANDARD OPERATING PROCEDURE #1
#1:2:
3:4:
#
##
wateralcohol
pH 1pH 2
adjust ?adjust ?
adjust ?
STANDARD OPERATING PROCEDURE #1
water
alcoholpH 1pH 2
SP-1m :
-0.01 %V/V 0.0710.01
EBC %V/V
4.02 pH 6.99 pH
stopped
adjust ?
10: 32: 54
-
7/22/2019 Alcolyzer Plus Beer
40/129
9 SOP (Standard Operating Procedure)
40 C27IB02
7. Highlight the respective row "adjust?" and press the key (see Fig. 9 - 2).
8. Enter the desired value and press .
Fig. 9 - 3
The following window is displayed (in this example, the alcohol value has
been readjusted):
Fig. 9 - 4
Information:
The adjustment of the known concentration for the alcohol can only be
performed successfully if an accurate density meter and an appropriate
density/alcohol concentration table are available.
If the known concentration for the alcohol shall be adjusted, the alcohol
value measured by the DMA is displayed in a line below the SOP
parameters after pressing the key. This value can be taken directlyfor the readjustment.
STANDARD OPERATING PROCEDURE #1
water
alcoholpH 1pH 2
adjust ?adjust ?
adjust ?
10: 34: 27
-0.01 %V/V 0.0710.01
EBC %V/V
4.02 pH 6.99 pH
1concentration: 0.05 %V/V
adjust ?adjustedadjust ?
STANDARD OPERATING PROCEDURE #1
water
alcoholpH 1pH 2
-0.01 %V/V 0.0710.05
EBC %V/V
4.02 pH 6.99 pH
adjust ?
10: 35: 16
-
7/22/2019 Alcolyzer Plus Beer
41/129
10 Checking the Adjustment Validity
C27IB02 41
9. Finish the SOP by pressing the key or the key when all
parameters are readjusted.
10 Checking the Adjustment Validity
If no SOP (see Chapter 9) is carried out, check the adjustments regularly as
described below.
10.1 Checking Procedure without Option pH/Color
Before each series of measurements, check the zero point using degassed,
bi-distilled water.
To check the validity of the slope of the adjustment, use a water/ethanol
solution of known concentration. Perform a slope check at least once everyweek.
10.1.1 Performing a Checking Procedure with Water (0%V/V)
1. Boil fresh, bi-distilled water for several minutes to remove dissolved gas.
2. Fill a clean glass flask full with the boiled water and cover it.
3. Wait until the water has cooled down to 20 2C.
4. Clean the measuring cell as described in Chapter 14.
Information:
If one (or both) pH buffer deviates too much from the correct value, the
measured pH value and the measured temperature will be displayed in a
line below the SOP parameters after confirming the question "adjust ?".
The pH value can then be corrected according to the pH/temperature
table of the corresponding buffer.
The readjustment for the pH measurement is carried out for both
adjustment points simultaneously. Therefore, only one "adjust? is
displayed for both buffers.
Information:
For the color adjustment, it is only necessary to adjust the zero point with
bi-distilled water regularly. If it is desired to check also the known concentration for the color
measurement (e.g. with a standard beer or color standard) this parameter
can be added to the SOP. If the known concentration for the color
measurement shall be adjusted, the color value of the color standard can
be entered in a line below the SOP parameters after pressing the key. Color standards are available at Anton Paar. Contact your local distributor.
-
7/22/2019 Alcolyzer Plus Beer
42/129
10 Checking the Adjustment Validity
42 C27IB02
5. Fill a sample vial with degassed, bi-distilled water and put it into the
magazine of the sample changer.
6. Press the key on the sample changer.
7. Wait until the measuring status displays "valid".
8. If the displayed value is 0 0.03%V/V and is stable for a few minutes,
measurements can be performed.
9. If the displayed value is outside this concentration range, clean the
measuring cell and perform another control measurement.
10. If the measuring result is still out of range (0 0.03%V/V) perform a zero
point adjustment (see Chapter 11.1.1).
11. If "no sample" is displayed, clean the measuring cells according to Chapter
14.1.3or Chapter 14.2.4(with option pH/color).
12. Then perform an adjustment for the zero point and for the known
concentration.
Checking the validity of the density adjustment
1. As soon as the DMA gives a valid result (flashing display), check the density
reading.
2. If the result deviates more than 0.0001 g/cm3(DMA 4500) or 0.00005 g/cm3
(DMA 5000) from the reference value (0.998203 g/cm3), perform a cleaning
procedure according to Chapter 14.
3. Repeat the checking procedure.
4. If the density result still exceeds the max. allowed deviation, perform an air/
water adjustment for the density meter (see Chapter 11.3).
Information:
If a density meter (DMA 4500, DMA 5000) is connected to Alcolyzer Plus
Beer, check also the validity of the density adjustment as described below.
-
7/22/2019 Alcolyzer Plus Beer
43/129
10 Checking the Adjustment Validity
C27IB02 43
10.1.2 Performing a Checking Procedure with a Water/
Ethanol Solution of Known Concentration
1. Prepare a water/ethanol solution with a concentration between 8 and
12%V/V.
2. Cover the flask and leave the solution for several hours at a temperature of
20 3C.
Before filling the measuring cell, the sample must be free of air bubbles.
3. Clean the measuring cell, if necessary (see Chapter 14).
4. Fill a sample vial with the water/ethanol solution and put it into the magazine
of the sample changer.
5. Press the key on the sample changer.
6. Wait until Alcolyzer Plus Beer and the DMA show "valid".
7. If the difference between the measuring values from Alcolyzer Plus Beer and
the DMA is less than 0.03%V/V and the results remain stable within this
range for a few minutes, the adjustment of Alcolyzer Plus Beer is correct.
8. If the displayed value is out of this range, clean the measuring cell andperform another control measurement.
9. If the measuring result is still out of range, perform an adjustment using a
water/ethanol solution (see Chapter 11.1.2).
Information:
This measurement can only be performed successfully if an accurate density
meter and an appropriate density / alcohol concentration table is available.
Caution:
Only use high-proof ethanol without denaturant!
Information:
If Alcolyzer Plus Beer is used in combination with a DMA 4500/5000, the
alcohol concentration is directly displayed on the density meter.
-
7/22/2019 Alcolyzer Plus Beer
44/129
10 Checking the Adjustment Validity
44 C27IB02
10.2 Checking Procedure with Option pH/Color
10.2.1 Checking Procedure for the Measuring Parameter
"alcohol"
The checking procedure for the measuring parameter "alcohol" is performed as
described in Chapter 10.1.
10.2.2 Checking Procedure for the Measuring Parameter "pH"
Check the validity of the pH adjustment with buffers pH 7 and pH 4 once a day
(before starting the routine measurements).
Control measurement for buffer pH 7
1. Fill a sample vial with buffer pH 7 and place it in the sample changer
magazine.
2. Press the key on the sample changer.
After filling is completed, the measurement starts automatically.
3. Wait until the pH value is displayed in the measuring window ("valid").
This value is valid for 25C.
4. If the difference between the measured pH value and the set point is less
than 0.05 pH units, the pH adjustment is correct.
5. If the displayed value is out of this range, fill the pH measuring cell with the
same buffer and repeat the control measurement.
Caution:
Always use fresh pH buffers.
Information:
Instead of using the sample changer, one of the following filling procedures
can be used:
Disconnect the silicone hose that is plugged to the front adapter of the pH
measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to
the front adapter of the pH measuring cell and fill in the buffer.
Or:
Remove the pH electrode from the pH measuring cell and immerse the
electrode into a vessel filled with pH buffer. Ensure that the buffer
temperature does not deviate significantly from the temperature in the pH
measuring cell as the temperature sensor for the pH measurement is
located in the pH measuring cell.
-
7/22/2019 Alcolyzer Plus Beer
45/129
11 Adjusting the Instrument
C27IB02 45
6. If the displayed value is still out of range, perform an adjustment (see
Chapter 11.2.3).
Control measurement for buffer pH 4
The control measurement is performed in the same way as for buffer pH 7.
10.2.3 Checking Procedure for the Measuring Parameter
"color"
Checking procedure with water (0.0 EBC)
This procedure should preferably be performed together with the checking
procedure for 0% alcohol (see Chapter 10.1.1)
1. If the displayed value is 0 0.3 EBC (0 0.15 ASBC), measurements can beperformed.
2. If the displayed value is out of this range, clean the measuring cells and
repeat the control measurement.
3. If the measuring result is still out of range, perform a "zero" adjustment (seeChapter 11.2.2).
11 Adjusting the Instrument
11.1 Adjustment without Option pH/Color
Regular adjustments are performed with degassed bi-distilled water (0%V/V
alcohol, zero point).
Extended adjustments are performed with degassed bi-distilled water and a
water/ethanol solution of exactly known concentration in the range of 2 to
30 %V/V, preferably 8 to 12%V/V.
An extended adjustment is necessary if the checking procedure with a water/
ethanol solution of known concentration shows deviations of more than
0.03 %V/V from the reference value, even after correct zero point
adjustment.
Information:
During the first months, usually a drift of the color value of approx. 0.1 EBC
per day can be observed. This drift (in the positive direction) is a result of a
decreasing LED intensity.
-
7/22/2019 Alcolyzer Plus Beer
46/129
11 Adjusting the Instrument
46 C27IB02
If the adjustment of Alcolyzer Plus Beer is properly performed, it is sufficient
to check only the zero point before each series of measurements.
11.1.1 Adjustment with Water (Zero Point)
1. Boil fresh, bi-distilled water for several minutes to remove dissolved gas.
2. Fill a clean glass flask full with the boiled water and cover it.
3. Wait until the water has cooled down to approx. 20 2C.
4. Select Alcolyzer Plus Beer menu adjustment > adjust alcohol > zero.
5. Answer the question "Do you really want to adjust?" with .
If there is no bi-distilled water in the measuring system, press on
Alcolyzer Plus Beer. Then press the key on the sample changer.
The bi-distilled water is now filled into the measuring system by the sample
changer. After the filling procedure is completed, the adjustment procedure
starts automatically.
If bi-distilled water has already been filled into the measuring system, press
the key instead of the key to start the adjustment
procedure immediately.
6. Wait until the adjustment has been completed.
Information:
Before the adjustment, Alcolyzer Plus Beer has to be turned on for at leastone hour. However, after transport full stabilization of Alcolyzer Plus Beer
might take up to one day.
The ambient temperature should be stable to < 3C.
Caution:
Before starting the adjustment, ensure that the measuring cells are properly
cleaned (see Chapter 14).
-
7/22/2019 Alcolyzer Plus Beer
47/129
11 Adjusting the Instrument
C27IB02 47
Fig. 11 - 1 Successful adjustment
7. If the measurement value is stable to 0.02%V/V, the adjustment is
successful and "adjustment saved" is displayed (see Fig. 11 - 1).
Press to confirm the adjustment.
Press the key three times to return to the measuring window.
8. If "no stability - restart" is displayed after the adjustment, the required stability
of the measuring result could not be achieved. Problems during sample
filling (for example, air bubbles in the cell) could be the reason.
Clean the cells again and repeat the adjustment with bi-distilled water.
ADJUSTMENT ALCOHOL
S1 = 87.30 % of factory adjustment
adjustment 0.00 %V/V
adjustment saved
Information:
Additionally, the value S1 gives information about the measured signal
intensity of the current adjustment compared to the signal intensity at the
factory adjustment. This value decreases by-and-by due to ageing of the LED
and/or coating of the measuring cell.
If the S1 value is below 80%, clean the measuring cell thoroughly (see
Chapter 14) and repeat the zero adjustment. If the S1 value is still too low,
contact your Anton Paar representative.
Information:
After this adjustment we recommend performing the zero adjustment for the
optional color measurement because the bi-distilled water which is inside the
cells can be used for the zero adjustment of the color measurement.
http://-/?-http://-/?- -
7/22/2019 Alcolyzer Plus Beer
48/129
11 Adjusting the Instrument
48 C27IB02
11.1.2 Adjustment with a Water/Ethanol Solution of Known
Alcohol Concentration
1. Prepare a water/ethanol solution.
2. Cover the flask and leave the solution for several hours at 20 3C. Before
filling the measuring cell, the sample must be free of gas bubbles.
3. Select the Alcolyzer Plus Beer menu adjustment > adjust alcohol > known
concentration.
4. Answer the question "Do you really want to adjust?" with .
5. If there is no water/ethanol solution in the measuring system, press
on Alcolyzer Plus Beer.
6. Press the key on the sample changer.
The water/ethanol solution is now filled into the measuring system by the
sample changer. After the filling procedure is completed, the adjustment
procedure starts automatically.
If a DMA 4500/5000 is connected and configured properly (seeAppendix F),
"DMA: measuring" is displayed on Alcolyzer Plus Beer. As soon as "DMA:
valid" is displayed, the alcohol concentration that comes from the DMA is
displayed in the first line of Alcolyzer Plus Beer display and can be used
directly as the "known concentration" value for the adjustment.
7. Press the key to continue the adjustment.
8. Wait until the adjustment is finished.
Caution:
Only use high-proof ethanol without denaturant.
Information:
The concentration of the water/ethanol solution for the adjustment must be
between 2 and 30%V/V, preferably between 8 and 12%V/V alcohol.
Caution:
Before starting the adjustment, ensure that the measuring cells are properly
cleaned (see Chapter 14).
Information:
If a water/ethanol solution has already been filled into the measuring system,
press the key instead of the key.
-
7/22/2019 Alcolyzer Plus Beer
49/129
11 Adjusting the Instrument
C27IB02 49
9. If the measurement is stable to 0.02%V/V, the adjustment was successful
and "adjustment saved" is displayed.
Press to confirm the adjustment.
Press the key three times to return to the measuring window.
10. If "adjustment failed" is displayed after the adjustment, the required stability
of the measuring result could not be achieved. Problems during sample
filling (for example, air bubbles in the cell) could be the reason.
Clean the cells again and repeat the adjustment with a water/ethanol
solution.
11.2 Adjustment with Option pH/Color
11.2.1 Adjustment Alcohol
The adjustment for the measuring parameter "alcohol" is performed as described
in Chapter 11.1.
11.2.2 Adjustment Color
For the color adjustment, use the delivered color adjustment set.
11.2.2.1 Adjustment with Water (Zero)
1. Fill a clean sample vial with the delivered water standard (ultra pure water)
and place the vial in the sample magazine.
2. Select the menu adjustment > adjustment color > zero.
3. Answer the question "Do you really want to adjust?" with .
4. Press the key on Alcolyzer Plus Beer.
5. Press the key on the sample changer.
The ultra pure water is filled into the measuring system.
Information:
Alcohol solutions have a tendency to form gas bubbles. Handle the sample
with care.
Caution:
Before starting the adjustment, ensure that the measuring cells are properly
cleaned (see Chapter 14).
-
7/22/2019 Alcolyzer Plus Beer
50/129
11 Adjusting the Instrument
50 C27IB02
After the filling procedure is completed, the adjustment procedure starts
automatically.
6. Wait until the adjustment is finished.
Fig. 11 - 2 Successful adjustment
If the measurement value is stable, the adjustment is successful,
"adjustment saved" is displayed (see Fig. 11 - 2).
Press to confirm the adjustment.
Press the key three times to return to the measuring window.
7. If "no stability - restart" is displayed after the adjustment, the required stability
of the measuring result could not be achieved. Problems during sample
filling (for example, air bubbles in the cell) could be the reason.
Clean the cells again and repeat the adjustment with bi-distilled water.
ADJUSTMENT COLOR
S = 98.30 % of factory adjustment
adjustment 0.00 EBC adjustment saved
Information:
Additionally, the value S gives information about the measured signal intensity
of the current adjustment compared to the signal intensity at the factory
adjustment. This value decreases by-and-by due to ageing of the LED and/orcoating of the measuring cell.
If the S value is below 80%, clean the measuring cell thoroughly (see Chapter
14) and repeat the zero adjustment. If the S value is still too low, contact your
Anton Paar representative.
Information:
After the color adjustment has been done properly, it is only necessary to
check the zero point adjustment with bi-distilled water regularly.
http://-/?-http://-/?- -
7/22/2019 Alcolyzer Plus Beer
51/129
11 Adjusting the Instrument
C27IB02 51
11.2.2.2 Adjustment with a Color Standard (Tartrazine) of Known
Concentration
1. Fill a sample vial with the delivered color standard (tartrazine) and place it in
the sample changer magazine.
2. Select the menu adjustment > adjust color > known concentration.
3. Answer the question "Do you really want to adjust?" with .
4. Press the key on Alcolyzer Plus Beer.
5. Press the key on the sample changer.
The standard is filled into the measuring cells.
6. Enter the EBC value of the delivered standard in the line "color".
7. Press the key to continue with the adjustment.
8. Wait until the adjustment is finished.
9. If the measurement is stable, the adjustment was successful and
"adjustment saved" is displayed.
Press to confirm the adjustment.
Press the key three times to return to the measuring window.
10. If "adjustment failed" is displayed after the adjustment, the required stability
of the measuring result could not be achieved. Problems during sample
filling (for example, air bubbles in the cell) could be the reason.
Clean the cells again and repeat the adjustment with the color standard.
Information:
The concentration of the color standard for the adjustment must be between1.00 and 60.00 EBC (0.50 to 30.48 ASBC), preferably at approx. 50.00 EBC
(25.40 ASBC).
Information:
After the color adjustment has been done properly, it is only necessary to
check the zero point adjustment with bi-distilled water regularly.
-
7/22/2019 Alcolyzer Plus Beer
52/129
11 Adjusting the Instrument
52 C27IB02
11.2.3 Adjustment pH
The adjustment of the pH electrode is performed with two buffers (pH 7 and pH 4).
Check the validity of the pH adjustment with buffers pH 7 and pH 4 once a day
(before starting the routine measurements). For details, see Chapters 9 and
10.2.2.
For the pH buffers "Hamilton Duracal" the buffer data (pH value depending on the
temperature) are integrated in Alcolyzer Plus Beer.
If the pH electrode is adjusted with these buffers, the correct buffer data are
selected automatically.
If you use a different buffer to "Hamilton Duracal", the buffer data (pH value
depending on the temperature) can be entered in the menu edit buffer data. If
the pH electrode is then adjusted with these buffers, the correct buffer data areselected automatically.
Caution:
Always use fresh pH buffers.
Temperature (C) Buffer pH 7.00 Buffer pH 4.01
5 7.09 4.01
10 7.06 4.00
15 7.04 4.00
18 7.03 4.00
20 7.02 4.00
22 7.01 4.00
25 7.00 4.01
30 6.99 4.01
35 6.98 4.02
40 6.97 4.03
45 6.97 4.04
50 6.97 4.05
Information:
For the adjustment, the first buffer should be pH 7, the second buffer should
be pH 4.
-
7/22/2019 Alcolyzer Plus Beer
53/129
11 Adjusting the Instrument
C27IB02 53
11.2.3.1 Automatic
The adjustment is usually performed with "Hamilton Duracal" buffers. However,
the adjustment can be performed with other buffers. The data of these buffers
have to be entered in the menu edit buffer data. The edited buffers have to be
selected in the menu sort buffer data.
1. Select menu adjustment > adjust pH > automatic.
"clean pH sensor/expose to buffer 1" is displayed.
2. To fill buffer pH 7 in the measuring cells using the sample changer SP-1m,
press first the key on the Alcolyzer and then the key on
the sample changer.
3. After the filling procedure is completed, wait until the voltage and the pH
value is stable (also check if the pH value is plausible) before pressing the
key to start the adjustment.
"Standard 7.00" (or the name of your buffer) as well as a number (in row"adjust") is displayed.
4. Press the key.
The adjustment with the first buffer is finished.
"clean pH sensor/expose to buffer 2" is displayed.
5. Fill buffer pH 4 in the measuring cells and wait until the voltage and the pH
value is stable (also check if the pH value is plausible).
6. Then press the key to start the adjustment.
Information:
Instead of using the sample changer, one of the following filling procedurescan be used:
Disconnect the silicone hose that is plugged to the front adapter of the pH
measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to
the front adapter of the pH measuring cell and fill in the buffer. Use at least
15 mL buffer to make sure that the previous liquid in the pH measuring
cell is fully replaced.
Or:Remove the pH electrode from the pH measuring cell and immerse the
electrode into a vessel filled with pH buffer. Ensure that the buffer
temperature does not deviate significantly from the temperature in the pH
measuring cell as the temperature sensor for the pH measurement is
located in the pH measuring cell.
Information:
If "no buffer saved" is displayed, the measured voltage deviates from the
stored tables (Hamilton or your own tables).
-
7/22/2019 Alcolyzer Plus Beer
54/129
11 Adjusting the Instrument
54 C27IB02
"Standard 4.01" (or the name of your buffer) is displayed. The adjustment is
finished. The voltages of the adjustment data (for the offset and the slope)
are displayed.
7. Press the key to finish the adjustment.
11.2.3.2 Manual
The pH adjustment can also be performed manually. To do this, any buffer can be
used.
1. Select menu adjustment > adjust pH > manual.
"clean pH sensor/expose to buffer 1" is displayed.
2. To fill buffer pH 7 in the measuring cells using the sample changer SP-1m,
press first the key on the Alcolyzer and then the key onthe sample changer.
3. After the filling procedure is completed, wait until the voltage and the pH
value is stable (also check if the pH value is plausible) before pressing the
key to start the adjustment.
4. Enter the theoretical pH value of the buffer (pH value depending on the
current temperature) using either the , , keys or a keyboard.
5. Then press .
6. Wait until the adjustment with the first buffer is finished.
7. Then press the key.
"clean pH sensor/expose to buffer 2" is displayed.
8. Fill buffer pH 4 in the measuring cells and wait until the voltage and the pH
value is stable (also check if the pH value is plausible).
9. Then press the key to start the adjustment.
Information:
Instead of using the sample changer, one of the following filling procedures
can be used:
Disconnect the silicone hose that is plugged to the front adapter of the pH
measuring cell. Fill a 20 mL syringe with the buffer. Connect the syringe to
the front adapter of the pH measuring cell and fill in the buffer. Use at least
15 mL buffer to make sure that the previous liquid in the pH measuring
cell is fully replaced.
Or:Remove the pH electrode from the pH measuring cell and immerse theelectrode into a vessel filled with pH buffer. Ensure that the buffer
temperature does not deviate significantly from the temperature in the pH
measuring cell as the temperature sensor for the pH measurement is
located in the pH measuring cell.
-
7/22/2019 Alcolyzer Plus Beer
55/129
11 Adjusting the Instrument
C27IB02 55
10. Enter the theoretical pH value of the buffer (pH value depending on the
current temperature) using either the , , keys or a keyboard.
11. Then press .
12. Wait until the adjustment with the second buffer is finished.
The voltages of the adjustment data (offset and slope) are displayed.
13. Press the key to finish the adjustment.
11.2.3.3 Edit Buffer Data
If you use a different buffer to "Hamilton Duracal" for the adjustment, the buffer
data (pH value depending on the temperature) can be edited in this menu.
In the first two rows of the list the buffers "Hamilton Duracal" are predefined. Thedata of these buffers are already programmed and cannot be edited.
The rows 3 10 are individually editable. For any buffer the buffer data can be
entered and stored using a certain name for the new edited buffer.
Entering buffer data
1. Select a row between 3 and 10 using the keys and .
"new buffer" appears on the display.
2. Press to enter the designation (e.g. "buffer pH 4") of the new bufferusing the , , keys or a keyboard.
3. Press the key.
Now the list can be entered (temperature and corresponding pH value).
4. Use the key to enter the data. Use the key to move the cursor.Use the key to go to the next row. Press to save all entered data.
5. Answer the question "save new pH buffer?" with the key.
Deleting buffer data
1. To delete a newly edited buffer, press the key.
2. Answer the question "really clear buffer data?" with and press the
key.
Caution:At least two pairs of temperature and corresponding pH value that enclose the
temperature, at which the pH is later measured, have to be entered.
-
7/22/2019 Alcolyzer Plus Beer
56/129
11 Adjusting the Instrument
56 C27IB02