aerosol particle counter models 310b, 310c, 350l, 5100€¦ · comprehensive validation manual...

2
BENEFITS APPLICATIONS ISO 21501 compliant with built-in Pulse Height Analyzer (PHA) – 100 LPM flow samples 1m³ in 10 minutes Ultra quiet pump improves work environment Calculates and creates reports for ISO 14644-1 and EC GMP (PIC/S GMP) Annex #1 Comprehensive validation manual makes it easy to meet pharmaceutical regulatory requirements Operates in cleanrooms via a web browser: Setup, sample, display, print, download data Choose from nine languages for display and printout Removable batteries with optional external charger for continuous mobile use Stores sampling recipes to reduce operator errors Cleanroom monitoring Facility certification for ISO or EC GMP Trend analysis Statistical process control Troubleshooting particle excursions Manifold compatible Portable or dedicated use 100 LPM, 50 LPM or 1 CFM flow rates available Large IR touch screen for simplified use Without measurement, there is no control. The Lasair III particle counter sets the new standard for aerosol particle counters. The on- board Pulse Height Analyzer (PHA) ensures ISO 21501 compliance. The large IR touch screen controls and intuitive user interface make the system very easy to use. The particle counter’s rugged enclosure is chemically resistant and lightweight for portability. The enclosure also has a streamlined design that minimizes particle traps and makes the unit easy to clean. Aerosol Particle Counter Models 310B, 310C, 350L, 5100 Exclusive Distributor and Service Centers: www.enviroterm.com

Upload: lamkhuong

Post on 09-Jun-2018

226 views

Category:

Documents


0 download

TRANSCRIPT

BENEFITS

APPLICATIONS

ISO 21501 compliant with built-in Pulse Height Analyzer (PHA)

– 100 LPM flow samples 1m³ in 10 minutes

Ultra quiet pump improves work environment

Calculates and creates reports for ISO 14644-1 and EC GMP (PIC/S GMP) Annex #1

Comprehensive validation manual makes it easy to meet pharmaceutical regulatory requirements

Operates in cleanrooms via a web browser: Setup, sample, display, print, download data

Choose from nine languages for display and printout

Removable batteries with optional external charger for continuous mobile use

Stores sampling recipes to reduce operator errors

Cleanroom monitoring

Facility certification for ISO or EC GMP

Trend analysis

Statistical process control

Troubleshooting particle excursions

Manifold compatible

Portable or dedicated use

100 LPM, 50 LPM or 1 CFM flow rates available

Large IR touch screen for simplified use

Without measurement, there is no control.

The Lasair III particle counter sets the new standard for aerosol particle counters. The on-board Pulse Height Analyzer (PHA) ensures ISO 21501 compliance. The large IR touch screen controls and intuitive user interface make the system very easy to use. The particle counter’s rugged enclosure is chemically resistant and lightweight for portability. The enclosure also has a streamlined design that minimizes particle traps and makes the unit easy to clean.

Aerosol Particle Counter Models 310B, 310C, 350L, 5100

Exclusive Distributor and Service Centers:

www.enviroterm.com

Aerosol Particle Counter

Specificationq/Modelu 310BChannels

Flow rate 1.0 CFM (28.3 LPM) ± 5%

aMaximum Concentration > 1,380,270/ft.³

Sample Tubing ID

Zero count

U.S. and foreign patents pending.

www.pmeasuring.com

0.3, 0.5, 1.0, 3.0, 5.0, 10µm

310C

1.0 CFM (28.3 LPM) ± 5%

> 1,380,270/ft.³

ID: 3/8"; Max. Length: 8 m

Data storage 3000 complete data sets. Data cannot be edited or altered meeting 21 CFR Part 11 compliance.

Communication Modes Ethernet or RS-232; USB data downloading; optional wireless Ethernet

Controlling Software DataAnalyst®, Facility Net, Pharmaceutical Net, FacilityPro, Microsoft® Internet Explorer 5.0+, Firefox®

Remote web browser operation; USB download, downloadable to Facilty Net/PharmaNet.Remote Operation

Environmental Sensors Four 4-20 mA inputs

Languages English, French, German, Italian, Japanese, Korean, Chinese (Traditional or Simplified), Spanish, Russian

8.4” color VGA display (640 x 480); IR touch screen (IP65 rated); Display Printer Built-in thermal printer

External Surface Polycarbonate (PC)

Enclosure Cleaning Materials Bleach, formaldehyde, ethyl/isopropyl alcohol, peroxide/quaternary ammonium solutions

Sample Output Internally filtered to HEPA standards (> 99.97% at 0.3 µm) filtering

Power and Battery 85-264 V, 50-60 Hz; Lithium battery: operation 3 hrs. (single) 6 hrs. (dual) for 1 CFM unit

12 x 13 x 10 inch (30x 33 x 26 cm) Dimensions (h, w, d)Weight 13 lbs. (5.9 kg) without battery, 16 lbs. (7.3 kg) with two optional batteries

Operating Environment Temperature: 0-30°C; Humidity:5-95% RH non-condensing

Lasair® is a registered trademark of Particle Measuring Systems, Inc.Microsoft® is a registered trademark of Microsoft Corporation.

Particle Measuring Systems, Inc. reserves the right to change specifications without notice.a. Less than 10% coincidence loss at maximum recommended concentration

Compliance to ISO-21501-4 requirementsCalibration

Reports USB reports; sample printouts; cleanroom certification reports for ISO and EC GMP (PIC/S GMP); averaging

ID: 3/8"; Max. Length: 8 m

5100

100 LPM ± 5%

> 688,495/ft.³

2 counts/m (100 LPM)³

ID: ¾"; Max. Length: 8 m

0.3, 0.5, 1.0, 5.0,10.0, 25µm

7.07 counts/m (1 CFM) ³4 counts/m³

7.07 counts/m (1 CFM) ³4 counts/m³

350L

50 LPM ± 5%

> 832,599/ft.³

ID: ½"; Max. Length: 8 m

2 counts/m (50 LPM)³

Counting Efficiency50% ± 20% for most sensitive threshold (100%±10% at 1.5 times the most sensitive sensitivity), fully complies with ISO 21501-4

Exclusive Distributor and Service Centers:

ENVIROTERM SDN. BHD. 369410-P

Particle Measuring Systems Headquarters

5475 Airport Blvd., Boulder, CO 80301

1-303-443-7100 1-800-238-1801 Fax: 1-303-546-7331

URL www.enviroterm.com Email : [email protected]

37-1, Jalan Setia Tropika 1/25,Taman Setia Tropika, 81200 Johor Baru,Johor, Malaysia.Tel : 07-2362768 Fax: 07-2362769

SOUTHERN OFFICE

0.3, 0.5, 1.0, 5.0,10.0, 25µm 0.3, 0.5, 1.0, 5.0,10.0, 25µm

H igh Pressure

• Monitor pressurized gases with air particle counter• Pressura range: 25 - 100 psi (60 - 100 psi for 100 LPM) • Compatible with CDA, Argon, N , or other non-toxic inert gases. 2

• Flow rate selection for 0.1, 1.0 CFM, 50, & 100 LPM

APPLICATION

Temperature/Relative Humidity handheld probe TRH-9008TR

Temperature/Relative Humidity Sensor TRH-100AND

Temperature Range:-40 to+80°C, ±0.3°C RH Range: 0-100%,±2% RH Cable Length: 2 meters, with handheld.

Temperature Range: 0 to+50°C, ±0.5°C RH Range: 0-100%,±3% RH Fixed Module on the rear panel of Lasair III

Differential Air Pressure Sensor DAP-100-AO Pressure Range:-50 to+50 Pascal, ±1.5% Tubing Length: 2 meters x 2 pcs

Filter Scanning Probe, For 1.0 CFM Flow rate FSP-ET-28

Many more options available to cater for your specific applications, includes, trace moisture, VOC, and etc. please contact one of our sales to discuss further on your application.

Instrument Service & Support: 1-800-557-6363

Customer Response Center: 1-877-475-3317

© 2012 Enviroterm Sdn. Bhd. All rights reserved.

OPTION: CAM

20, Lorong PS1, Bandar Perda,14000 Bukit Mertajam,Pulau Pinang, Malaysia.Tel : 04-5371371 Fax: 04-537 1373

NORTHERN OFFICELot 5035, Jalan 18/62, Taman Perindustrian Sri Serdang, 43300 Seri Kembangan, Selangor D.E., Malaysia.Tel : 03-89434558 Fax: 03-89436012

HEAD OFFICE

D iffuser

, Model : CAM

This CAM can be used in applications where users want to measurethe noted parameters at pressure (25 to 100 psi range)

Compressed Air ModuleAPPLICATION

Wireless ModuleThis

wireless data communication from LASAIR III to PCAPPLICATION wireless unit allow

OPTION: HPD-III

OPTION: WM

Dew Point: -60ºC to 40ºC,Temperature: -20ºC to 75ºCPressure: 0 to 200 psia

OPTION: ENVIRO-SENSORS