![Page 1: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/1.jpg)
Aerogen Nebulizer
Adult Ventilator Use
![Page 2: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/2.jpg)
Parts1. Aerogen Neb2. T-Piece*3. Controller box4. Controller cable5. AC Adapter6. Mounting bracket7. Continuous Neb
tube set
*Adult sized
![Page 3: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/3.jpg)
Set-up
![Page 4: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/4.jpg)
Aerogen Controller
![Page 5: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/5.jpg)
Adding meds
• Uncap thesilicone plugand introducethe medication
• Make sure theAerogen isupright.
![Page 6: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/6.jpg)
Syringe Pump
• Power on
• Press #4
![Page 7: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/7.jpg)
Syringe Pump
• Hit the oval keybelow “MORE” atthe bottom of thescreen
• Choose “1” ML/HR
• On the nextscreen, choose “1”B-D (not shown)
![Page 8: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/8.jpg)
Syringe Pump
• Load prepared syringe(note: this is a syringe andtubing specially designedfor this application,available in the RTdepartment)
• Use the lever on the endto set move arm and locksyringe in place.
• Hit “ENTER” when done
![Page 9: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/9.jpg)
Syringe Pump
![Page 10: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/10.jpg)
Syringe Pump
• Type 12 and hit“ENTER”
• At this screen, hit“Disable”, leaving thebox blank.
• Choose “YES” toconfirm settings (Notshown)
![Page 11: Aerogen Nebulizer - RT Connection · 30 Min. - Press & Release Continuous Mode - 3 Sec. hold from off Battery Status Indicator W DC Input . Title: Microsoft PowerPoint - Aerogen Nebulizer.pptx](https://reader034.vdocuments.mx/reader034/viewer/2022042712/5f9ae358fdb040049f075be6/html5/thumbnails/11.jpg)
Syringe Pump
• Press “BOLUS” key toprime tubing(remove the air fromtubing.
• Press “EXIT” whenpriming is complete
• Press the START key