040D0060-02, Rev. A, 2009-09

ADVIA® Chemistry Interface

Operating Instructions

Page 2: Advia Chemistry Analyzer Advance Infor

©2009 Siemens Healthcare Diagnostics Inc. All rights reserved.

No part of this manual or the products it describes may be reproduced by any means or in any form without prior consent in writing from Siemens Healthcare Diagnostics.

ADVIA, ADVIA CentraLink, ADVIA WorkCell, and LabCell are trademarks of Siemens Healthcare Diagnostics.

The information in this manual was correct at the time of printing. However, Siemens continues to improve products and reserves the right to change specifications, equipment, and maintenance procedures at any time without notice.

If the ADVIA LabCell or ADVIA WorkCell CDX system is used in a manner different than specified by Siemens, the protection provided by the equipment may be impaired. Observe all warning and hazard statements.

Page 3: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 3

040D0060-02, Rev. A

Contents

1 Overview ...........................................................................5

Enhanced LAS Software Features .................................................. 5

Hardware Overview.......................................................................... 6

Software Overview ........................................................................... 8 Automation System Software .................................................................. 8 LineMaster Software ............................................................................... 9 Router Software .....................................................................................15 ADVIA Chemistry System Software........................................................15 Operating Sequence ..............................................................................16

2 Operating the System ....................................................18

Turning the Interface On................................................................ 18

Turning the Interface On (Enhanced LAS Software) ................... 20

Running Sample Tubes from the Instrument Sample Tray......... 23

Turning the Interface Off................................................................ 24

3 Maintenance ...................................................................25

Daily Maintenance.......................................................................... 25

4 Troubleshooting.............................................................26

Processing Events ......................................................................... 26

Using the Diagnostics Screens..................................................... 27 Using the Diagnostics Screen.................................................................27 Using the Diagnostics Screen (Enhanced LAS Software).......................30

Resolving Interface Fatal Error Conditions.................................. 33 Resolving an Interface Fatal Error Condition ..........................................33 Resolving an Interface Fatal Error Condition (Enhanced LAS Software) 35

Recovering from Pause States...................................................... 36 Recovering from the Pause State...........................................................37 Recovering from the Pause State (Enhanced LAS Software) .................38

Page 4: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 4

040D0060-02, Rev. A

Unlocking and Flushing the Gate During Processing................. 39

Troubleshooting Skipped Sample Tube Conditions ................... 41 Troubleshooting a Skipped Sample Tube Condition ...............................41

Replenishing Reagents During Processing (Enhanced LAS Software).............................................................. 42

5 Sample Management......................................................43

6 System Configuration ....................................................44 Enhanced Routing (Enhanced LAS Software) ........................................44

Appendix A: Safety Information ........................................45

Appendix B: Warranty and Support Information..............46

Contact Information ....................................................................... 46

Appendix C: Supplies.........................................................47

Appendix D: Specifications ...............................................47

Appendix E: Symbols.........................................................47

Appendix F: Glossary.........................................................48

Page 5: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 5

040D0060-02, Rev. A

1 Overview

The ADVIA Chemistry system is a clinical chemistry analyzer for in vitro diagnostic use that runs tests on human serum, plasma, or urine in random access, batch, and STAT (interrupt) modes. The ADVIA 1650, ADVIA 1800, and ADVIA 2400 Chemistry systems are collectively called the ADVIA Chemistry system. The ADVIA Chemistry Interface Kit allows you to include the ADVIA Chemistry system on your ADVIA ® LabCell® or ADVIA WorkCell™ CDX system.

This document is a supplement to the ADVIA Automation Solutions Operator’s Guide. For operating procedures of the ADVIA Chemistry system, refer to the appropriate ADVIA Chemistry Operator’s Guide.

IMPORTANT: This document provides separate information and instructions for Enhanced LAS software only where features and procedures significantly differ. Sections specifically for Enhanced LAS software are clearly labeled; if no designation appears, information and procedures apply to the ADVIA Chemistry systems with and without Enhanced LAS software.

Enhanced LAS Software Features This document includes information about Enhanced LAS software, an upgrade for the the ADVIA 1800 Chemistry system and the ADVIA 2400 Chemistry system. The ADVIA 1650 Chemistry system cannot run Enhanced LAS software. Enhanced LAS software provides the following features:

• The ADVIA Chemistry system transmits reagent status and event messages to the LineMaster. These messages provide important information about instrument, reagent, and sample status.

• When the instrument detects a sample integrity problem, such as a clot or short sample, the system sends a message to the LineMaster and routes the sample to SIQ for inspection.

• If there is no workorder, a sample is out of sequence, or the barcode is not read, the system does not aspirate the sample tube before it moves from the interface gate to the main track.

• If a reagent is not available, a test is not calibrated, or a test fails QC, the enhanced routing feature allows a sample tube to route to another instrument that can perform the test. If no other instrument with that test is available, the sample tube routes to the Partial Processed tray. This feature improves Turn Around Time.

• If an ancillary reagent is not available, the instrument pauses. Sample tubes can route to another instrument or to the Partial Processed tray. When the ancillary reagent is replenished, the instrument returns to normal operation.

• Tests can be disabled due to missing reagents, Cal errors, ISE cal errors, optional QC out, and ISE urine cal error. This information is communicated to the LineMaster.

Page 6: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 6

040D0060-02, Rev. A

Hardware Overview The ADVIA Chemistry interface contains the following components:

1 Monitor

2 Workstation

3 Automation interface computer

4 ADVIA Chemistry system computer

5 On/Off switch

6 Top cover

Figure 1. ADVIA Chemistry Interface (front)

Page 7: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 7

040D0060-02, Rev. A

1 Main breaker

2 Auxiliary power

3 Interface box (not used with Enhanced LAS software)

4 Waste container with waste pump

5 Water input

6 Power junction box

Figure 2. ADVIA Chemistry Interface (back)

Page 8: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 8

040D0060-02, Rev. A

Software Overview

Automation System Software The automation system uses 2 software applications:

• The LineMaster workspace serves as the primary point of interaction between the operator and the automation system. The LineMaster computer communicates with and controls the active status of all instruments and components. The LineMaster workspace reports analytical instrument events, tube status, and automation system status.

• The Router software controls the divert gates and interface gates. It defines the route of a sample tube on the track and balances the workload among instruments that perform the same tests.

The operator uses the LineMaster software to perform the following operations for the ADVIA Chemistry interface:

• select the ADVIA Chemistry interface

• change the state of the ADVIA Chemistry interface

• view events

• view and change the Test List screen

• view the Diagnostics screen

Page 9: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 9

040D0060-02, Rev. A

LineMaster Software The LineMaster workspace provides access to the Details, Events, Test List, and Diagnostics screens. To access ADVIA Chemistry system screens, select the ADVIA Chemistry station button.

Figure 3. LineMaster Workspace

For systems with more than 1 ADVIA Chemistry system, each ADVIA Chemistry system station button provides access to information, events, and data unique to that interface.

NOTE: If you use ADVIA Chemistry systems with Enhanced LAS software and ADVIA Chemistry systems without Enhanced LAS software, ensure that station names allow you to easily identify these different systems.

IMPORTANT: For Enhanced LAS software, all ADVIA Chemistry system LineMaster screens display the Diagnostics button above the Test List button.

Page 10: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 10

040D0060-02, Rev. A

Details Screen The Details screen displays when you select the ADVIA Chemistry system station button. The Details screen shows the current operational state of the interface.

Figure 4. ADVIA Chemistry Details Screen

Page 11: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 11

040D0060-02, Rev. A

Events Screen The Events screen displays event messages associated with the ADVIA Chemistry system interface.When an event occurs, a white, yellow, or red status bar appears above the ADVIA Chemistry system station button. A white bar indicates a minor event; a yellow bar indicates that the interface requires operator attention; a red bar indicates a failure that causes the interface to stop. You can review the event log to identify these conditions.

Figure 5. ADVIA Chemistry Events Screen

The Events screen displays only the event messages associated with the ADVIA Chemistry interface. The screen includes the following buttons:

• The Event Details button displays details about the selected event.

• The Acknowledge button removes the selected event from the unacknowledged list.

• The Acknowledge All button removes all events from the unacknowledged list.

You can view the Acknowledged list by selecting the blank field above Unacknowledged. This displays additional minor events that are auto-acknowledged.

You can use the arrow in the Log Time column to change the sort order. The arrow in the Severity column allows you to choose a specific severity level. In addition, you can sort by the Log Time, Machine, Severity, Message, and Code columns by placing the cursor above the column name and typing a keyword or number.

Page 12: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 12

040D0060-02, Rev. A

Test List Screen The Test List screen displays tests that are active on the ADVIA Chemistry interface. Refer to the ADVIA Automation Solutions Operator’s Guide for general information on the Test List screen.

Figure 6. ADVIA Chemistry Test List Screen

Page 13: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 13

040D0060-02, Rev. A

Figure 7. ADVIA Chemistry Test List Screen (Enhanced LAS Software)

The Test List screen for Enhanced LAS software includes the following additional information:

• A new Instrument column shows availability of tests. A test can be unavailable when calibration is necessary, and when the test is not within control range, or when reagents are empty. If the checkbox is cleared, the test cannot run.

• If a common reagent is empty, the Online button turns light blue, indicating that the instrument is in the Pause state. The instrument completes tests for sample tubes that are currently in process if they do not require common reagents. Refer to Recovering from the Pause State (Enhanced LAS Software) for additional information.

• For routing purposes only, you can select a forced routing test. Forced routing tests send a sample tube that requires a rerun to a different instrument or to the same instrument, ignoring normal routing requirements. For repeat/reflex tests, a forced routing test downloads with a workorder from the ADVIA CentraLink® system to the LineMaster. Forced routing tests do not appear in the test map and have no results.

NOTE: You must first configure a forced routing test on the global test list. When you select the down arrow above the Selected column, you can see the selected checkbox, a cleared checkbox, and a blank area. When you select the blank area, the system displays the complete global test list. You can then select a forced routing test from the list.

Page 14: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 14

040D0060-02, Rev. A

Diagnostics Screen The Diagnostics screen displays status information about the ADVIA Chemistry interface. You can use this screen for troubleshooting purposes and on a daily basis when you bring the ADVIA Chemistry interface offline.

To access diagnostic features, select the ADVIA Chemistry system station button and then select Diagnostics .

Figure 8. ADVIA Chemistry Diagnostics Screen

Figure 9. ADVIA Chemistry Diagnostics Screen (Enhanced LAS Software)

For descriptions of all Diagnostics screen fields, refer to Using the Diagnostics Screens.

Page 15: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 15

040D0060-02, Rev. A

Router Software For information on the Router software, refer to the ADVIA Automation Solutions Operator’s Guide.

ADVIA Chemistry System Software For operating procedures of the ADVIA Chemistry system, refer to the ADVIA Chemistry Operator’s Guide.

Page 16: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 16

040D0060-02, Rev. A

Operating Sequence The ADVIA Chemistry interface operates according to the following sequence:

1. A sample tube routes to the ADVIA Chemistry station.

2. At the divert gate, the system reads the sample tube barcode and diverts the sample tube to the side track.

3. The system sends a workorder request to the ADVIA CentraLink system.

4. At the interface gate, the sample tube moves to the pick position.

5. The system reads the barcode again and verifies that the sequence of tubes in the side track is correct.

1 Gate control button

2 Divert gate

3 Side track

4 Interface gate

5 Pick position

6 Place position

7 Main track

Figure 10. ADVIA Chemistry Station

Page 17: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 17

040D0060-02, Rev. A

6. The system aspirates directly from the sample tube.

NOTE: If there is no workorder, the sample is out of sequence, or the barcode is not read, the system aspirates the sample without producing results and moves the sample tube to the main track. For Enhanced LAS software, when these conditions occur, the system does not aspirate the sample tube before it moves from the interface gate to the main track. If this occurs, refer to Troubleshooting for instructions.

7. When processing completes, the sample tube moves back to the side track and then to the main track.

CAUTION

Do not remove samples directly from the side track or interface gate. If you remove a sample directly from the side track or interface gate, the system aspirates the next sample tube without producing results and routes it to the main track. For Enhanced LAS software, the system does not aspirate the sample tube before it moves from the interface gate to the main track.

8. The ADVIA Chemistry system sends the test results directly to the ADVIA CentraLink system.

Page 18: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 18

040D0060-02, Rev. A

2 Operating the System

Turning the Interface On Perform the following steps to turn the interface on:

1. At the ADVIA Chemistry system, lift the gate control button.

The interface gate homes and the gate control button illuminates.

2. At the ADVIA Chemistry system workspace, select Initialize .

Wait for initialization to complete.

NOTE: Ensure that the system is in Ready or Start mode before you complete this procedure.

3. At the LineMaster workspace, select the ADVIA Chemistry system station button.

Figure 11. ADVIA Chemistry Details Screen

4. At the Details screen, select Offline .

Wait for the Offline button to turn white.

5. Select Diagnostics .

6. At the Diagnostics screen, ensure that the /HNREADY checkbox is selected.

If the checkbox is not selected, refer to Troubleshooting.

7. Select Online .

Page 19: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 19

040D0060-02, Rev. A

8. At the ADVIA Chemistry system workspace, perform the following steps:

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox.

Figure 12. ADVIA Chemistry System Start Conditions Screen

c. Select Start . d. Select OK.

9. At the Diagnostics screen, perform the following steps:

a. Ensure that the /BMONL checkbox is selected and wait 2 minutes. b. Ensure that the /BMREADY checkbox is selected.

If either checkbox is cleared, refer to Troubleshooting.

10. Wait while the cuvette washes complete and the Online button transitions from light blue to white.

11. Wait for the station button to turn green.

Page 20: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 20

040D0060-02, Rev. A

Turning the Interface On (Enhanced LAS Software) Perform the following steps to turn the interface on:

1. At the ADVIA Chemistry system, lift the gate control button.

The interface gate homes and the gate control button illuminates.

2. At the ADVIA Chemistry system workspace, select Initialize .

Wait for initialization to complete.

NOTE: Ensure that the system is in Ready or Start mode before you complete this procedure.

3. At the LineMaster workspace, select the ADVIA Chemistry system station button.

Figure 13. ADVIA Chemistry Details Screen

Page 21: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 21

040D0060-02, Rev. A

4. At the Details screen, select Diagnostics .

Figure 14. ADVIA Chemistry Diagnostics Screen

5. At the Diagnostics screen, select Offline .

Wait for the Offline button to turn white.

6. At the Test List screen, verify that the Instrument column checkbox is selected for all required tests (refer to Figure 7).

7. If the checkbox is not selected, perform the following steps:

a. Select Shutdown . b. Replenish reagents. c. Repeat steps 4-7. d. Select Online .

Page 22: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 22

040D0060-02, Rev. A

8. At the ADVIA Chemistry system workspace, perform the following steps:

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox.

Figure 15. ADVIA Chemistry System Start Conditions Screen

c. Select Start . d. Select OK.

9. At the LineMaster, bring the system online.

The Online button remains light blue.

10. Wait while the cuvette washes complete.

11. Wait for the Online button to turn white.

12. Wait for the station button to turn green.

Page 23: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 23

040D0060-02, Rev. A

Running Sample Tubes from the Instrument Sample Tray Perform the following steps to run sample tubes from the sample tray at the front of the ADVIA Chemistry system for pediatric, STAT, or short samples:

1. At the ADVIA Chemistry system workspace, select Sample Pause .

Wait until the Start button becomes available.

2. Select Start .

Figure 16. ADVIA Chemistry System Start Conditions Screen

3. At the Start Conditions screen, perform the following steps:

a. Load sample tubes in the Sample Manager. b. In the Ordinary sample area, below the Tray no.field, select the Routine smp.

Analyze checkbox. c. Next to Analyze mode, ensure that the Barcode radio button is selected. d. Select Start . e. Select OK. f. At the Sample confirmation screen, select OK.

After the system aspirates all samples on the ADVIA Chemistry system sample tray, processing resumes from the automation system.

Page 24: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 24

040D0060-02, Rev. A

Turning the Interface Off As part of your end–of–day routine, perform the following procedure to turn the interface off.

1. At the ADVIA Chemistry system workspace, ensure that the system is in the WATCH or READY state.

This indicates that all samples are processed. If necessary, wait until the system processes all samples before proceeding.

2. At the LineMaster, bring the system offline.

3. Select Shutdown .

4. If the Shutdown button does not turn white, at the Diagnostics screen, select Terminate .

This forces the interface to the shutdown state.

5. Ensure that no samples are in the side track.

If samples are present, press the gate control button down and then lift it up to flush the samples from the gate and side track. Manage these sample tubes according to your laboratory standard practices. Refer to Unlocking and Flushing the Gate During Processing for additional information.

Page 25: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 25

040D0060-02, Rev. A

3 Maintenance

For ADVIA Chemistry system maintenance information, refer to the ADVIA Chemistry Operator’s Guide.

CAUTION

Do not allow any liquid to flow into the interface gate during cleaning. Wring or dry the soft cloth or cotton-tipped applicator so that it is damp before you begin cleaning.

Daily Maintenance

Inspecting the ADVIA Chemistry Interface Gate Perform the following steps at the interface gate:

1. Inspect the gate for the presence of dust or serum:

a. Inspect the barcode reader. b. Inspect the gate sensors. c. Inspect the index wheel.

2. If you find dust or serum, use a mild soapy solution and a soft cloth or cotton-tipped applicator to remove any soil.

3. Dry the area with a soft cloth.

4. At the ADVIA Chemistry system, ensure that all reagents and consumables are filled.

Inspecting the Divert Gate Perform the following steps at the divert gate:

1. Inspect the divert gate, singulate, and stop fingers for the presence of dust or serum:

2. If you find dust or serum, use a mild soapy solution and a soft cloth or cotton-tipped applicator to remove any soil.

3. Inspect the barcode reader.

4. If necessary, clean the lens with a lens cloth.

5. At the ADVIA Chemistry system, ensure that all reagents and consumables are filled.

Page 26: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 26

040D0060-02, Rev. A

4 Troubleshooting

Processing Events The ADVIA Chemistry Events screen displays event messages associated with a specific ADVIA Chemistry system station and provides recommended corrective actions.

Event messages are listed chronologically, with the most recent event at the top. You can sort event messages by log time, severity, machine, message, and event code. You can also arrange event messages by acknowledged events or unacknowledged events.

When an ADVIA Chemistry system event occurs, a white, yellow, or red alert bar displays above the ADVIA Chemistry system station button.

Perform the following steps when an event occurs:

1. At the LineMaster workspace, select the ADVIA Chemistry station button.

2. Select Events .

3. Select an event.

4. Select Event Details .

5. Carefully read the event message and follow the recommended corrective actions.

6. Acknowledge the event by selecting the event and then selecting Acknowledge or Acknowledge All .

The Event Log screen displays system fatal, station fatal, operator required, and warning labels in the Status column. Event messages have no status notation.

Page 27: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 27

040D0060-02, Rev. A

Using the Diagnostics Screens The Diagnostics screen displays status information about the ADVIA Chemistry interface. You can use this screen on a daily basis and for troubleshooting purposes. The Diagnostics screen differs for Enhanced LAS software.

Using the Diagnostics Screen

Figure 17. Diagnostics Screen

Inputs from Instrument Area The Inputs from Instrument area displays signals from the ADVIA Chemistry system.

A checkmark next to … Indicates …

/BMINIT the ADVIA Chemistry system is initializing.

/BMONL the ADVIA Chemistry system initiated a run or is running. This signal is required to process samples.

/BMREADY the system is online and is ready to process sample tubes. In normal operation, BMREADY displays a checkmark approximately 2 minutes after BMONL displays. This signal is required to process samples.

/IDREQ a request for a barcode. This is a momentary signal that the operator cannot see.

Page 28: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 28

040D0060-02, Rev. A

A checkmark next to … Indicates …

/BMSMP the ADVIA Chemistry system is sampling. In normal operation, this checkmark displays when the Lock Status checkbox is selected. When the BMSMP is checked, the interface gate does not move.

CAUTION

Do not attempt to unlock a gate while the BMSMP checkbox is selected. Unlocking the gate while the BMSMP checkbox is selected can damage the probe or spill a sample tube. Before unlocking a gate, at the ADVIA Chemistry system, check the status of the probe. Refer to Unlocking and Flushing the Gate During Processing for additional information.

/BMTRBL a fault occurred at the ADVIA Chemistry system. The

ADVIA Chemistry system will not process sample tubes until you resolve the problem.

Outputs to Instrument Area The Outputs to Instrument area displays signals from the LAS interface computer. The /HNREADY and /SMPREQ are used for maintenance and troubleshooting procedures.

A checkmark next to … Indicates …

/HNREADY the ADVIA Chemistry system is online or offline. When this checkbox is cleared, the ADVIA Chemistry interface cannot process sample tubes and does not start. If you attempt to start the system when HNREADY is not available, an HNREADY not available message displays at the ADVIA Chemistry interface.

/SMPREQ a sample tube arrived at the aspiration position of the ADVIA Chemistry system. This is a momentary signal.

The HNNEXT field is not used.

Page 29: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 29

040D0060-02, Rev. A

Gate Information Area At the Gate Information area, a Lock Status checkbox indicates that the ADVIA Chemistry system is aspirating a sample tube and the gate is locked. If you press and then lift the gate control button when the gate is locked, the system will not flush tubes and pucks from the interface gate. Refer to Unlocking and Flushing the Gate During Processing for additional information.

NOTE: You do not need to monitor this signal during normal operation. During emergency maintenance and troubleshooting procedures, you can use the Unlock Gate button to unlock and then flush the gate in a proper sequence. Currently, the Pick Request, OK to Pick, and Picking checkboxes are not useful because they do not reflect real-time activity.

Display Area The Display area includes the I/O Signals, Serial Messages, and Controller Activity checkboxes. Ensure that these checkboxes are selected.

Actions Area The Actions area contains the following buttons.

Button Description

Clear Log Select this button to clear text in the Log area. Your technical support provider uses this text area.

Terminate Select this button only when the interface is unresponsive and requires emergency maintenance. The Terminate button forces the interface to the shutdown state.

Unlock Gate Select this button to clear the Lock Status checkbox. Use this button only for emergency maintenance and troubleshooting procedures.

Log Area In the Log area, serial messages reflect the communication between the automation system and the ADVIA Chemistry system.

Page 30: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 30

040D0060-02, Rev. A

Using the Diagnostics Screen (Enhanced LAS Software)

Figure 18. ADVIA Chemistry Diagnostics Screen (Enhanced LAS Software)

Interface Actions Area The Interface Actions area contains the following buttons.

Button Description

Display Queue Select this button to view all sample tube barcodes currently in the side track.

Reset Queue Select this button to reset the automation system queue. When you use this button, sample tubes are flushed from the side track, circle the track once and then divert to the instrument.

NOTE: If you select this button when the system is online, the Change State window displays, forcing you to bring the system offline.

Terminate Select this button only when the interface is unresponsive and requires emergency maintenance. The Terminate button forces the interface to the shutdown state.

Page 31: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 31

040D0060-02, Rev. A

Button Description

Unlock Gate Select this button to unlock the interface gate. Use this button only for emergency maintenance and troubleshooting procedures. Refer to Unlocking and Flushing the Gate During Processing for additional information.

System Status Select this button to test communication between the instrument and LineMaster.

If you select this button when the system is in the shutdown state, the Change State window displays, forcing you to bring the system offline.

NOTE: Use this button only when the system is in the online or offline state.

Gate Information Area The Gate Information area includes the following checkboxes.

Within … A checkmark next to … Indicates …

Interface Gate area

Puck at Pick a puck is at the pick position of the interface gate.

Tube at Pick a sample tube is at the pick position of the interface gate

Lock Status the instrument is aspirating a sample tube and the gate is locked. If you press and then lift the gate control button when the gate is locked, the system will not flush tubes and pucks from the interface gate.

You do not need to monitor this signal during normal operation. During emergency maintenance and troubleshooting procedures, you can use the Unlock Gate button to unlock and then flush the gate in a proper sequence. Refer to Unlocking and Flushing the Gate During Processing for additional information.

Router area Pick Request the instrument is ready to receive and aspirate sample tubes.

Ok to Pick the automation system transmitted the barcode to the instrument.

Picking the instrument is aspirating a sample.

Page 32: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 32

040D0060-02, Rev. A

Barcode Area You can use the fields in the Barcode area to monitor the quality of the barcodes and the barcode reader in real time. The Barcode Actions area contains the following fields:

Field Description

Last Diverted This field displays the barcode of the last sample tube diverted to the instrument. *

Gate Barcode This field displays the barcode of the sample tube currently being aspirated.*

Last Departed This field displays the barcode of the last sample tube moved to the Place position of the interface gate.*

* These fields display No Read when the system cannot read the barcode.

Message Log Area The Message Log area can display the following information.

• The Message Log Filter drop-down menu allows you to change the level of message filtering passed to the interface when interface transitions from shutdown to offline, or from shutdown to offline to online.

You can select 1 of the following levels:

• None so that no messages display.

• Low to display events.

• Medium to display the test map when the interface goes from shutdown to offline, as well as events, reagent information, and easily-readable messages that are not in code. Medium is the default selection.

• The Save XML File button immediately saves the XML log file for troubleshooting purposes. It closes the current log and opens a new log file.

• The Clear log button clears the message log.

Page 33: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 33

040D0060-02, Rev. A

Resolving Interface Fatal Error Conditions The ADVIA Chemistry interface automatically transitions to the shutdown state when a fatal error condition occurs or if the interface is unresponsive. A red bar displays over the station button. If this occurs, perform the appropriate recovery procedure. Recovery procedures differ for Enhanced LAS software.

Resolving an Interface Fatal Error Condition • Perform the following steps to resolve a fatal error condition:

1. At the ADVIA Chemistry system workspace, select to view the error report.

Figure 19. ADVIA Chemistry System Operation Panel

2. At the Error Report screen, double-select the Stop or Warning + Stop errors to view details.

Follow instructions for resolving the error if instructions display.

3. Resolve the problem that caused error.

4. Select Initialize .

The system performs wash and homing functions.

5. At the LineMaster workspace, bring the interface offline:

The station button turns white.

6. If the station button does not turn white, perform the following steps:

a. Select Diagnostics . b. At the Diagnostics screen, select Terminate .

If the ADVIA Chemistry system does not go to the white shutdown state, contact your technical service provider.

7. Select Online .

Page 34: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 34

040D0060-02, Rev. A

8. At the ADVIA Chemistry system workspace, perform the following steps (refer to Figure 12):

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox. c. Select Start . d. Select OK.

The system starts its wash cycles, which complete in approximately 2 minutes.

9. At the LineMaster, verify that the ADVIA Chemistry interface is ready to process sample tubes:

a. At the LineMaster workspace, select the ADVIA Chemistry system station button.

b. Select Diagnostics . c. At the Diagnostics screen, in the Inputs from Instrument area, ensure that

/BMONL is selected. d. Ensure that the /BMREADY checkbox is selected.

10. Acknowledge the events:

a. At the ADVIA Chemistry Details screen, select Events . b. Select Acknowledge All .

NOTE: If signals are missing or if additional events occur, check the cable connections on the interface box (refer to Figure 2) and the corresponding cable connections at the back of ADVIA Chemistry interface computer. If the problem persists, contact your technical support provider.

Page 35: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 35

040D0060-02, Rev. A

Resolving an Interface Fatal Error Condition (Enhanced LAS Software) Perform the following steps to resolve a fatal error condition:

1. At the LineMaster, select the ADVIA Chemistry system station button.

2. At the Details screen, select Events .

3. At the the Events screen, find the event that caused the fatal error.

4. At the ADVIA Chemistry system, select to view the error report.

5. At the Error Report screen, double-select the Stop or Warning +Stop error.

6. At the Error Report Details screen, confirm the error and follow the directions that display.

7. Resolve the problem that caused the error.

8. Select Initialize .

Wait until the ADVIA Chemistry system initializes and enters the Ready state.

9. At the LineMaster, bring the system offline.

The station button turns white.

10. If the station button does not turn white, perform the following steps:

a. Select Diagnostics . b. At the Diagnostics screen, select Terminate .

This forces the ADVIA Chemistry system to the white shutdown state. If this does not occur, contact your technical service provider.

11. Select Online .

12. At the ADVIA Chemistry system workspace, perform the following steps (refer to Figure 15):

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox. c. Select Start . d. Select OK.

The system starts its wash cycles, which take approximately 2 minutes to complete. The Online button remains light blue until the wash cycles complete. When the Online button turns white, the ADVIA Chemistry interface is ready to process sample tubes.

NOTE: If the test map does not display or if additional events occur, check the network cable connections at the ADVIA Chemistry system interface switch. In addition, check cable connections at the back of the ADVIA Chemistry interface computer and those at the ADVIA Chemistry system. After checking all cables, restart the ADVIA Chemistry system. If the problem persists, contact your technical support provider.

Page 36: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 36

040D0060-02, Rev. A

Recovering from Pause States The ADVIA Chemistry system can enter the Pause state when it cannot process sample tubes. In the Pause state, the automation system stops diverting sample tubes to the instrument. When the interface is online and the ADVIA Chemistry system is in the Pause state, the Online station button turns a green-blue color.

The ADVIA Chemistry system can enter the pause state under any of the following conditions:

• Low levels of oil reserves, system diluent, ISE buffer, wash solution DI water, and saline solution exist.

• The ADVIA Chemistry system processes sample tubes from the sample tray. When processing from the sample tray completes, the ADVIA Chemistry system leaves the pause state automatically and resumes processing from the automation system.

• The ADVIA Chemistry system loses the /BMREADY signal or the BMONL signal (the checkbox is cleared on the Diagnostics screen). These fields appear only for ADVIA Chemistry systems without Enhanced LAS software.

Page 37: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 37

040D0060-02, Rev. A

Recovering from the Pause State Perform the following procedure to recover from the Pause state:

1. At the ADVIA Chemistry system, perform the following steps:

a. Verify that the ADVIA Chemistry system is not processing sample tubes from the sample tray.

b. Verify that the Error Report screen does not display Stop or Warning +Stop errors. If Stop or Warning + Stop errors exist, resolve the errors.

2. Check the cable connections on the interface box (refer to Figure 2) and the corresponding cable connections at the back of ADVIA Chemistry system.

If necessary, remove and reconnect the cable.

3. At the LineMaster workspace, select the ADVIA Chemistry system station button.

4. Select Diagnostics .

5. Verify that the /BMREADY signal or the BMONL signal are selected.

6. At the ADVIA Chemistry system workspace, perform the following steps (refer to Figure 12):

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox. c. Select Start . d. Select OK.

The system starts its wash cycles, which take approximately 2 minutes to complete.

7. At the LineMaster, bring the system offline.

The station button turns white.

8. If the station button does not turn white, perform the following steps:

a. Select Diagnostics . b. At the Diagnostics screen, select Terminate .

If the ADVIA Chemistry system does not go offline, contact your technical service provider.

c. Select Offline .

9. Select Online .

10. When the Online button turns white, verify that the ADVIA Chemistry interface is ready to process sample tubes.

Page 38: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 38

040D0060-02, Rev. A

Recovering from the Pause State (Enhanced LAS Software) Perform the following procedure to recover from the Pause state for Enhanced LAS software:

1. At the LineMaster workspace, select the ADVIA Chemistry system station button.

2. At the Details screen, select Events .

3. Select an event appropriate for the instrument that is paused.

4. At the Events screen, select Event Details .

5. At the ADVIA Chemistry system, perform the following steps:

a. Verify that the ADVIA Chemistry system is not processing sample tubes from the sample tray.

b. Verify that the Error Report screen does not display Stop or Warning +Stop errors. If Stop or Warning + Stop errors exist, resolve the errors.

6. Replenish any common reagent that is depleted.

7. At the LineMaster workspace, select the ADVIA Chemistry system station button and verify that the system resumed processing.

8. If the ADVIA Chemistry system is not processing samples, select Initialize .

9. At the ADVIA Chemistry system workspace, perform the following steps (refer to Figure 15):

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox. c. Select Start . d. Select OK.

The system starts its wash cycles, which take approximately 2 minutes to complete.

10. At the LineMaster, bring the system offline.

11. Select Online .

12. Verify that the station button turns white and that the ADVIA Chemistry system begins processing samples.

If the ADVIA Chemistry system does not resume processing, contact your local service provider.

Page 39: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 39

040D0060-02, Rev. A

Unlocking and Flushing the Gate During Processing

CAUTION

Do not flush the gates while the system processes sample tubes unless emergency maintenance or troubleshooting conditions occur. Flushing the gates can damage the system. Always replenish reagents and consumables prior to starting daily operation.

If necessary, perform the following steps to flush the gate while the system processes sample tubes:

1. At the ADVIA Chemistry system workspace, select Pause .

Wait while the system aspirates 1 or 2 more sample tubes in the Pause/shift state, in transition to the pause state.

2. At the LineMaster workspace, select the ADVIA Chemistry station button.

3. At the Details screen, select Offline .

4. Select Diagnostics .

5. At the Diagnostics screen, select Unlock Gate .

6. If the gate remains locked, perform the following steps:

a. Switch to the Router computer. b. At the Router Main screen select F1 MAIN. c. Select the stack light of the ADVIA Chemistry system station. d. At the Station screen, verify that the Picking flag is gray.

Figure 20. Router Station Screen

Page 40: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 40

040D0060-02, Rev. A

e. If the picking flag is yellow, at LineMaster, select Offline . f. Select Online . g. Wait several seconds and then select Offline . h. Select Shutdown .

This clears the picking flag.

7. At the ADVIA Chemistry system, press the gate control button down and then lift it up.

Sample tubes move from the ADVIA Chemistry system interface gate and return to the main track.

NOTE: Allow the system to finish processing the aspirated sample tubes. When processing completes and the system is in the Ready state, replace reagents or consumables according to normal procedures.

8. Perform the necessary emergency maintenance or troubleshooting procedures.

CAUTION

At the ADVIA Chemistry system, ensure that you select the Initialize button. If you do not select the Initialize button, the system skips tubes at the interface gate. Skipping these tubes can damage the sample probe.

9. At the ADVIA Chemistry system, select the Initialize button to clear the internal sample queue.

10. At the LineMaster workspace, select the ADVIA Chemistry system station button.

11. At the Details screen, select Offline .

12. At the ADVIA Chemistry system workspace, perform the following steps (refer to Figure 12):

a. Select Start . b. At the Start Conditions screen, in the Ordinary sample area, below the Outside

analyze field, select the Routine smp. Analyze checkbox. • At the LineMaster Diagnostics screen, wait until the BMREADY checkbox

displays a check. • For Enhanced LAS software, verify that the Diagnostics screen displays the

test map. c. Select Start . d. Select OK.

The system starts its wash cycles, which take approximately 2 minutes to complete.

13. Bring the interface online.

Page 41: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 41

040D0060-02, Rev. A

Troubleshooting Skipped Sample Tube Conditions

Troubleshooting a Skipped Sample Tube Condition When barcode misread or a No Workorder condition occurs, the ADVIA Chemistry system indexes, or skips, sample tubes. In order to maintain the internal queue, the system aspirates the sample tube with the misread barcode. Subsequent sample tubes are not impacted by barcode misreads.

CAUTION

Do not leave sample tubes on the automation system if they have repeatedly skipped. Skipped sample tubes return to the ADVIA Chemistry system when the system reads barcode at the divert gate but misreads it at the interface gate. For ADVIA Chemistry systems without Enhanced LAS software, this causes depletion of the sample. For Enhanced LAS software, the sample tube continues to circle the track but is not depleted. Remove sample tubes that have repeatedly skipped from the automation system.

If the ADVIA Chemistry system alarm alerts you to a skipped sample tube, perform the following procedure.

1. At the ADVIA Chemistry workspace, locate the SID of the skipped sample tube:

a. At the ADVIA Chemistry system, select to view the error report. b. Double-select the message that reads LAS/RH: skip sample ID received. c. Record the Pre-read ID, which is the SID of the skipped sample tube.

2. At the LineMaster workspace, select System > SIQ or press F5.

3. At the SIQ screen, select the Workorders tab.

4. In the SID field, enter the barcode number of the misread sample tube.

5. Select Enter .

6. When the sample tube information displays, select Route to SIQ .

If the sample tube is still on the track, it circles the track again and then routes to the SIQ gate or SIQ tray.

7. Examine the sample tube barcode label for proper positioning and clarity.

8. If the necessary, replace the label according to your laboratory procedures.

9. Reintroduce the sample tube to the automation system via an SIQ gate or Load tray.

Page 42: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 42

040D0060-02, Rev. A

Replenishing Reagents During Processing (Enhanced LAS Software)

If an individual reagent is low or empty, you can replenish reagents while the system processes samples in several minutes. The ADVIA Chemistry system suspends sampling, adds reagent to the last samples, and then allows you to add reagents.

Perform the following procedure to replenish reagents while the system is processing samples:

1. At the ADVIA Chemistry system workspace, select RGT pause .

CAUTION

Do not select SCAN until you replenish reagents. Selecting SCAN before you replenish reagents can damager reagent probes. Always replenish reagents before you select SCAN.

The system aspirates several additional samples before it suspends processing and the reagent trays stop moving. The RGT SCAN screen displays “After Loading Reagents, click SCAN to perform a barcode scan.”

2. Replenish reagents.

3. Replace reagent covers.

4. At the RGT SCAN screen, select SCAN.

Wait until “BCD SCANNING …Please wait for a while” no longer displays and the system resumes processing.

NOTE: The reagent or method may require an RBL or calibration before the method becomes available.

Page 43: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 43

040D0060-02, Rev. A

5 Sample Management

For sample management information, refer to the ADVIA Automation Solutions Operator’s Guide.

Page 44: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 44

040D0060-02, Rev. A

6 System Configuration

Enhanced Routing (Enhanced LAS Software) The enhanced routing feature is available only with Enhanced LAS software and requires a specific configuration. Discuss enhanced routing options with your local service provider before you install Enhanced LAS software

NOTE: With enhanced routing, the system can create duplicate test aspirations, using additional reagents and producing duplicate results. If you use enhanced routing, you may need to change ADVIA CentraLink settings to avoid duplicate results.

If a reagent is not available, a test is not calibrated, or a test fails QC, enhanced routing allows a sample tube to route to another instrument that can perform the test. If no other instrument with that test is available, the sample tube routes to the Partial Processed tray. Enhanced routing can offer faster Turn Around Time for sample tube processing.

If you turn off enhanced routing and a test is not available at an instrument, the system marks all tests Route Complete, although some tests were not performed. In this case, these sample tubes route to the Unload tray rather than to the Partial Processed tray. This is the current behavior for systems without Enhanced LAS Software.

Page 45: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 45

040D0060-02, Rev. A

Appendix A: Safety Information

For all safety information related to the automation system, refer to the ADVIA Automation Solutions Operator’s Guide.

Page 46: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 46

040D0060-02, Rev. A

Appendix B: Warranty and Support Information

For all warranty and support information related to the automation system, refer to the ADVIA Automation Solutions Operator’s Guide.

Contact Information This section provides the following information:

• the address of the Siemens authorized representative, which is the Siemens contact within the European community

• the Siemens addresses for obtaining service and technical information and for ordering supplies

Siemens Authorized Representative Siemens Healthcare Diagnostics Ltd. Sir William Siemens Sq. Frimley, Camberley, UK GU16 8QD

Addresses For technical assistance, contact your local technical support provider.

www.siemens.com/diagnostics

Siemens Healthcare Diagnostics Pty Ltd 885 Mountain Highway Bayswater Victoria 3153 Australia

Page 47: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 47

040D0060-02, Rev. A

Appendix C: Supplies

For all warranty and support information related to the automation system, refer to the ADVIA Automation Solutions Operator’s Guide.

Appendix D: Specifications

Specification Requirements

Dimensions 1711 mm (W) x 867 mm (D) x 1340 mm (L)

Weight 630 kg (uncrated)

Interface Type Direct Interface

Power Requirements 13.6 amps @ 220 volts 50/60Hz

Air Supply NA

DI Water 40 liters per hour

Appendix E: Symbols

For information on automation system symbols, refer to the ADVIA Automation Solutions Operator’s Guide.

Page 48: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 48

040D0060-02, Rev. A

Appendix F: Glossary

For a complete glossary of terms associated with automation systems, refer to the ADVIA Automation Solutions Operator’s Guide.

A ADVIA Automation Solutions

A system of modular components that can be configured to create custom, open laboratory automation systems for sample tube management.

ADVIA Automation Solutions

A term used by Siemens to describe its product line of automated systems, including the ADVIA LabCell and ADVIA WorkCell CDX systems. Each automation system is comprised of modular components used to create laboratory automation solutions for sample tube management.

Automation system An ADVIA LabCell or ADVIA WorkCell CDX system.

ADVIA CentraLink A data management software that manages the results and QC for each instrument, and provides a connection to the customer’s laboratory information system (LIS). ADVIA CentraLink provides a single point of use for results generated by the instruments connected to the automation system.

B Barcode A code consisting of lines of varying widths that can be read

optically with a scanning device.

Barcode reader An optical character reader that can read data represented by bar codes and translate the codes into digital signals.

C Cold storage A refrigerated area designated for storing completed

samples. Samples are assigned by the operator to predefined cold storage locations within one or more refrigerators on a particular door and shelf.

Page 49: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 49

040D0060-02, Rev. A

D Divert gate A device on the main track of the conveyor that moves

sample tubes from the main track onto a side track. A barcode reader scans the sample tube barcode label, and based on information received from the ADVIA CentraLink system, the sample is diverted onto the side track or continues on to the next gate.

E E-stop An emergency button or cord that immediately stops all

mechanical motion.

Event Log A LineMaster screen that displays and archives event messages and recommended corrective actions.

F

G Gate Divert gates and interface gates read barcodes and direct

sample tubes on the conveyor to instruments and modules, where testing and processing take place.

Gate control button An on/off button located on top of the divert gate that turns the station power, divert gate, and interface gate on and off. Press the gate control button to turn the station off. Lift the gate control button to turn the station on.

Global Test List A list of all laboratory tests that can be performed on or off the automation system. The ADVIA CentraLink system, the instrument test lists, and the sort tray use the global test list to determine the routing of samples.

H

Page 50: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 50

040D0060-02, Rev. A

I ID In the global test list, a unique numeric identifier assigned to

a sample for a specific test. The ID is a shortened form of the test name that the laboratory uses to identify a test.

Instrument An analyzer, excluding the robot and other interface components, that conducts medical tests or procedures on samples.

Instrument Test List A subset of the global test list that defines which tests are performed by a specific analyzer or module connected to the automation system.

Interface gate A device on a side track, located directly in front of an analyzer or module, that reads the sample barcode, then selects and positions the sample so it can be processed by a connected analyzer or module and returned to the main track.

Interface computer A computer that builds an interface between the LineMaster software and an instrument or module. An interface computer controls signals and communication data to and from the instrument.

J

K

Page 51: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 51

040D0060-02, Rev. A

L Light tower Vertically stacked colored lights that indicate system status

information. The color of an illuminated light indicates the operating status of a robot or gate.

• Red indicates a serious error occurred on the automation system. The conveyor, robots, and instruments and modules stop. The operator must take corrective action.

• Yellow indicates a station is offline. The conveyor continues to run. The operator must take corrective action.

• White indicates all stations are online. The conveyor continues to run. The operator should take corrective action at the station.

• Green indicates the conveyor is running.

LineMaster A computer-based system that allows the operator to control sample management for the entire automation system. Also called the LineMaster workspace, it is the primary point of interaction between the operator and the automation system. The LineMaster server communicates with and controls the active status of all instruments and components connected to the system. The LineMaster reports events, tube status, and automation system status.

LIS Lab Information System

M Module A connected system, such as the Sample Manager, that

performs sample tube management functions but does not analyze samples.

Page 52: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 52

040D0060-02, Rev. A

N Name The name of a test in the global test list. This name must be

the same name used by the ADVIA CentraLink system.

O Offline A station is not operational.

Non-robotic interfaces reach the offline state when the station no longer diverts sample tubes to the side track.

Robotic interfaces reach the offline state after the instrument or module checks all safety features and the robot homes.

Online A station is operational.

The automation system diverts samples to the station’s interface gate, where they are picked by the robot and processed or directly aspirated at the gate.

Overview screen The main screen of the LineMaster that displays the system configuration.

P Pick A robot action that picks up a sample tube from one point

prior to placing it at another point.

Pick position The position at an interface gate at which instrument probes aspirate samples, or from which robots pick sample tubes for processing.

Place position The position at an interface gate to which robots return sample tubes after processing.

Puck A weighted cylindrical carrier that holds a single sample tube as it moves on the main track and side tracks.

Q Queue A line of sample tubes on the side track waiting for

processing. The queue typically holds up to 20 samples, with 15 extra positions reserved for STAT.

Page 53: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 53

040D0060-02, Rev. A

R RCC Robot Control Cabinet. This component supports

instruments and modules containing robots. It houses all the electronic robotic controls and a computer that communicates with the LineMaster software.

Robot A mechanism that picks sample tubes or racks from one position and places them at another on the automation system. Robots are flexible and move in 3 axes: X, theta, and Z. There are 2 types of robots:

• Interface robots pick sample tubes from the interface gate and move them into the instrument or module.

• Internal robots move samples within an instrument or module.

Router Software that routes samples to the appropriate stations by controlling the divert gates and interface gates. The Router software communicates with the ADVIA CentraLink system and receives workorders with tests and ID information. The Router software also determines, based on system configuration inputs, the sequence of stations to which a sample is delivered, and balances the workload between stations with the same instruments or modules.

S Sample A test tube filled with some media requiring testing or

processing. Samples may include serum, whole blood, plasma, cerebrospinal fluid (CSF), urine, and QC materials. All sample tubes must include a unique barcode label by which the automation system can differentiate among patients and among different sample types from the same patient.

Sample Manager A module that moves sample tubes from the input trays to the conveyor for processing, and when processing is complete, from the conveyor to the output trays.

Sample tube A thin glass or plastic container, closed at one end, containing a sample. The conveyor transports a sample tube in a puck on the surface of the track. Sample tubes include a unique barcode by which the automation system differentiates among patients and among different sample types from the same patient.

Page 54: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 54

040D0060-02, Rev. A

Shutdown A station is not in communication with the LineMaster computer.

In the shutdown state, the robot or instrument stops placing and picking samples. Independent of the automation system, the instruments may continue operation. For instruments that aspirate directly from the interface gate, the station waits for an aspiration to complete before shutting down.

Sample ID The barcode number that identifies the sample to the automation system. The barcode number links the sample tube with the workorder.

SIQ Sample-in-question

SIQ station A station gate or a tray defined as SIQ in the Sample Manager that receives sample tubes requiring manual operator intervention.

The automation system sends samples to SIQ when it cannot direct them to a station. This occurs when a sample barcode is not readable, the sample has no workorder, the sample is a duplicate, or sampling problems occurred at a station. After the operator corrects the problem, the sample is checked into the system.

Page 55: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 55

040D0060-02, Rev. A

T Test Analyzing a sample.

Tray A device that holds up to 100 sample tubes. Each tray has a unique barcode on the top and all 4 sides for easy identification in cold storage. The operator places trays into the Sample Manager and defines each tray as 1 of the following:

• Load Mixed tray, which contains sample tubes ready for testing When the automation system includes a Decapper, a Load Mixed tray accommodates sections of capped and uncapped sample tubes.

• STAT tray, which contains sample tubes requiring immediate processing. Sample tubes in STAT trays have higher priority than sample tubes processed from any other tray. When the automation system includes a Decapper, the STAT tray accommodates sections of capped and uncapped sample tubes.

• Unload tray, which contains sample tubes that were processed and are ready for removal.

• Partial Processed tray, which contains sample tubes with incomplete routes.

• SIQ tray, which contains sample tubes that are samples-in-question (SIQ) or unknown samples and performs the same function as an SIQ station.

• Centrifuge Override tray, which contains sample tubes that do not route to the Centrifuge/Decapper interface module. When the automation system includes a Decapper, the Centrifuge Override tray accommodates all capped or all uncapped sample tubes.

• MultiZone tray, which contains a maximum of 5 different zones, each of which can correspond to the Unload tray, Partial Processed tray, and Sort tray types:

• Sort tray, which contains sample tubes sorted by tests.

• AutoSwitch tray, which the system can use for loading and unloading sample tubes. When the automation system includes a Decapper, the AutoSwitch tray accommodates sections of capped and uncapped sample tubes.

• Unknown tray, which contains sample tubes with unknown status. The Sample Manager does not pick sample tubes from an Unknown tray.

Page 56: Advia Chemistry Analyzer Advance Infor

ADVIA Chemistry Interface Operating Instructions 56

040D0060-02, Rev. A

U Username The name of the operator logged on to the automation

system.

V

W Workspace The user interface of the automation system, also called the

LineMaster workspace, which includes a System Management area and a Control Management area. The System Management area displays system information depending on the system button selected and contains the log on badge. The Station Management area allows the operator to select station buttons and view station-specific data.

X

Y

Z