advanced phase measurement application manual …€¦ · micro motion® advanced phase measurement...

48
Application Manual MMI-20030076, Rev AB February 2017 Micro Motion ® Advanced Phase Measurement Application Manual

Upload: phungtram

Post on 10-Aug-2018

274 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Application ManualMMI-20030076, Rev AB

February 2017

Micro Motion® Advanced Phase Measurement

Application Manual

Page 2: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Safety messages

Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefullybefore proceeding to the next step.

Emerson Flow customer service

Email:

• Worldwide: [email protected]

• Asia-Pacific: [email protected]

Telephone:

North and South America Europe and Middle East Asia Pacific

United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727

Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804

Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468

Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682

Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000

Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803

Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600

Egypt 0800 000 0015 Singapore +65 6 777 8211

Oman 800 70101 Thailand 001 800 441 6426

Qatar 431 0044 Malaysia 800 814 008

Kuwait 663 299 01

South Africa 800 991 390

Saudi Arabia 800 844 9564

UAE 800 0444 0684

Page 3: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Contents

Chapter 1 Before you begin ............................................................................................................ 11.1 About this application manual ..................................................................................................... 11.2 About the software ...................................................................................................................... 11.3 Software requirements ................................................................................................................ 21.4 Terms and definitions .................................................................................................................. 3

Chapter 2 Measurement options and configuration ....................................................................... 72.1 Liquid with gas .............................................................................................................................72.2 Net oil ........................................................................................................................................112.3 Gas with liquid ........................................................................................................................... 14

Chapter 3 Additional configuration .............................................................................................. 193.1 Configure viewing and reporting for process variables ............................................................... 193.2 Configure the totalizer history log to include Advanced Phase Measurement

contract totals ........................................................................................................................... 223.3 Configure events ....................................................................................................................... 22

Chapter 4 Transmitter operation ..................................................................................................234.1 Read process variables ............................................................................................................... 234.2 Read contract totals ...................................................................................................................23

Appendices and referenceAppendix A Application parameters and data ................................................................................. 25

A.1 Advanced Phase Measurement configuration parameters ......................................................... 25A.2 Modbus interface to configuration parameters ..........................................................................25A.3 Modbus interface to process data .............................................................................................. 27

Appendix B Typical applications and other information .................................................................. 31B.1 Advanced Phase Measurement with a two-phase separator .......................................................31B.2 Advanced Phase Measurement with a three-phase separator .................................................... 32B.3 Advanced Phase Measurement at the wellhead ......................................................................... 33B.4 Determining the Drive Gain Threshold .......................................................................................33

Appendix C Best practices for two-phase measurement performance ............................................. 35C.1 Entrained gas performance ........................................................................................................35C.2 Entrained liquid (mist) performance .......................................................................................... 37C.3 Density determination ...............................................................................................................38

Contents

Application Manual i

Page 4: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Contents

ii Micro Motion® Advanced Phase Measurement

Page 5: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

1 Before you beginTopics covered in this chapter:

• About this application manual

• About the software

• Software requirements

• Terms and definitions

1.1 About this application manualThis application manual explains how to configure and use the Advanced PhaseMeasurement licensed software option on select Model 5700 transmitters. It also containslimited, supplemental installation and configuration information specifically related to theAdvanced Phase Measurement software. Refer to the sensor and transmitter installationmanuals and the transmitter configuration and use manual for complete information.

1.1.1 Additional documentation from Micro Motion

Additional documentation for Advanced Phase Measurement installationsTable 1-1:

Document Use

Sensor installation manual for your sensor Installation and wiring for the sensor

Transmitter installation manual for theModel 5700 transmitter

Installation and wiring for the Model 5700 trans-mitter

Configuration manual for the Model 5700 trans-mitter

Configuration, operation, maintenance, andtroubleshooting for features that are not relatedto Advanced Phase Measurement

ProLink III user manual Installation and use of ProLink III

Modbus Interface Tool Programming the Modbus host

Enhanced Density Measurement Manual Configuring and using the enhanced density ap-plication

1.2 About the softwareMicro Motion Advanced Phase Measurement software improves long-term flow reportingand measurement performance in processes with intermittent periods of two-phase flow,including liquids with entrained gas or gas with entrained liquid. And if it is combined with

Before you begin

Application Manual 1

Page 6: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

the Net Oil or Concentration Measurement software options, the software can also reportliquid concentration, Net Oil, and/or Gas Void Fraction (GVF) during the same two-phaseconditions.

There are three measurement options for the Advanced Phase Measurement software:Liquid with Gas, Net Oil, and Gas with Liquid.

NoteEach option is licensed separately and must ordered with the transmitter.

Advanced Phase Measurement software optionsTable 1-2:

License option (orderingcode) Process fluid Desired measurement

Liquid with Gas (PL) (1) Liquid with entrained gas Liquid flow rate and totals, GVF

Net Oil (PO)(2) Mixture of oil and water Net oil (dry oil at reference con-ditions) and net water flowrates and totals, GVF

Gas with Liquid (PG) Gas with entrained liquid (mist) Gas flow rate and totals

(1) Can be combined with code PO or CM (Concentration Measurement).

(2) Can be combined with Advanced Phase Measurement licence code PL.

1.3 Software requirementsEnsure that your installation meets these requirements.

Transmitter

Advanced Phase Measurement software is available only on the Model 5700 transmitter,either integrally mounted, in a 9-wire remote configuration, or in combination with any800 Series Enhanced Core Processor.

Sensor

Advanced Phase Measurement software is compatible with any sensor that is supported bythe Model 5700 transmitter except for H300, F300, or any T-Series sensor due to limitedperformance with two-phase fluids. The software is not compatible with the 700 SeriesStandard Core processor.

Installation and wiring

• Follow the installation and wiring instructions in the sensor and transmitterinstallation manuals.

• Identify and follow any application-specific installation needs as identified in thismanual for your application type.

Before you begin

2 Micro Motion® Advanced Phase Measurement

Page 7: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

• Refer to the entrained gas/mist/empty-full-empty best installation practices. Referto Appendix C.

• If you plan to use the net oil measurement option with an external water cutmonitor:

- For the Model 5700 configurable transmitter,Channel D on the transmitter mustbe enabled, and must be configured as a mA input, wired to the water cutmonitor, and configured appropriately. HART integration is not currentlysupported.

- For the Model 5700 Ethernet transmitter, an external host system is required toaccept the water cut monitor input and feed it to the transmitter.

NoteWater cut cannot be input to version 1.x of the Model 5700 fieldbus transmitter.

Configuration and operation

The Advanced Phase Measurement software can be configured using the interface optionalready being used to configure the transmitter:

• Transmitter display

• ProLink III v3.2 or later

• Field communicator (Model 5700 configurable and Model 5700 fieldbustransmitters only).

• A fieldbus host (Model 5700 fieldbus transmitter only)

• A web browser (Model 5700 Ethernet transmitter only)

Because much of the Advanced Phase Measurement software process data is notassignable to an output (e.g., contract totals), retrieving the data requires one of thefollowing external host systems, depending upon the transmitter output type:

• The Model 5700 configurable transmitter requires a Modbus/RS-485 connection toChannel E, which must be enabled and wired to a Modbus host to collect andprocess data.

• The Model 5700 Ethernet transmitter requires an Ethernet/IP connection to thetransmitter, and an appropriate host program to collect the Advanced PhaseMeasurement software process data.

• The Model 5700 fieldbus transmitter requires a FOUNDATION fieldbus H1connection to the transmitter and a fieldbus host that reads Advanced PhaseMeasurement software process data.

1.4 Terms and definitionsThe Advanced Phase Measurement software application and this manual use the termsdefined here.

Before you begin

Application Manual 3

Page 8: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Correction The process of calculating the value of a process variable atreference temperature, starting from the value of the processvariable at line temperature (the measured value).

Uncorrecteddensity

The density of the process fluid at line temperature.

Corrected density The density of the process fluid at reference temperature.

Uncorrectedvolume

The volume of the process fluid at line temperature.

Corrected volume The volume of the process fluid at reference temperature.

Mixture The process fluid before separation, e.g., a combination of oil andwater, or gas, oil, and water.

Water cut The volume fraction of water in the liquid mixture, in %.

External water cut A water cut value measured by an external device and supplied tothe Advanced Phase Measurement software via the mA input,Ethernet, or fieldbus.

Density-basedwater cut

The water cut value calculated by Advanced Phase Measurementsoftware using the measured line density referenced to a density/water cut curve based on user-entered dry oil and water densitiesat reference temperature.

Densitydetermination

The procedure of obtaining the density of dry oil and water atreference temperature—both are required for use with the Net Oilsoftware option.

Net A measurement of a single component of the process fluid, e.g.,oil only, water only.

Entrained,entrainment

The presence of small amounts of gas in a liquid stream, or liquidin a gas stream.

Remediated An adjustment applied to a measured process variable by theAdvanced Phase Measurement algorithm to correct for errorsassociated with two-phase fluid conditions.

Unremediated Measured variables that are not adjusted by the Advanced PhaseMeasurement algorithm. In two-phase conditions, these processvariables represent the whole mixture, or bulk fluid (e.g. water, oil,and gas).

Contract total Up to four user-specified totals that are used to measureproduction.

Contract period The 24-hour period over which contract totals are incremented.Contract totals are reset at the beginning of each contract period.

Contract hour Contract totals are recorded and then reset at the beginning ofthe contract hour.

Gas void fraction The ratio of gas volume to total mixture volume at line conditions.Also called Gas Volume Fraction.

Before you begin

4 Micro Motion® Advanced Phase Measurement

Page 9: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Drive gainthreshold

Maximum drive gain value expected for single-phase fluid undertypical operating conditons. Above the threshold value, themeasurement will be remediated in accordance with theconfigured Advanced Phase Measurement software settings. See Section B.4 for more information.

Pre-mist averagingperiod

Period of stable gas flow conditions prior to a liquid entrainmentevent. Averaging Period is user configurable with units = "n"seconds. The average flow rate during this period will be reporteduntil the entrained liquid event has passed.

Post-mistadjustment delay

Period of stable gas flow conditions after a liquid entrainmentevent. The average flow rate during Post-Mist Adjustment Delay isaveraged with the Pre-Mist Averaging Period and the reportedflow rate is adjusted a maximum of +/-10% of reading until anydifference has been appropriately corrected.

Period averagedoutput (PAO)

Rolling averages, with adjustable averaging period, that helpidentify trends in noisy data. Available PAOs are: mass flow rate,density, volume flow rate, net oil flow rate and water cut at lineand reference conditions, gas void fraction, and temperature. TheAverage Reporting Interval is used to adjust the averaging period.

Shrinkage The change in liquid volume between the measurement point anda stock tank due to lighter hydrocarbons evaporating. This iscaused by the stock tank being at a lower pressure, further belowthe bubble point of the oil. The shrinkage factor is a user-inputfactor, based on a PVT (pressure-volume-temperature) test of theoil.

Shrinkage MeterFactor

Used just like the normal Meter Factor, when proving the meteragainst liquid measurement at stock tank conditions.

Before you begin

Application Manual 5

Page 10: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Before you begin

6 Micro Motion® Advanced Phase Measurement

Page 11: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

2 Measurement options andconfigurationTopics covered in this chapter:

• Liquid with gas

• Net oil

• Gas with liquid

2.1 Liquid with gasThis measurement option improves flow measurement in liquid processes withintermittent entrained gas, or with known-density liquids under continuous entrained gasconditions.

NoteThe Liquid with Gas measurement option can also be combined with the Net oil measurementoption or concentration measurement. See http://www2.emersonprocess.com/siteadmincenter/PM%20Micro%20Motion%20Documents/Enhanced-Density-Manual-20002315.pdf.

Liquid with gas measurement process

The presence of entrained gas (or bubbles) can cause significant errors when measuringthe volume flow of liquid through a Coriolis meter. Because bubbles displace some of theliquid in a flow stream, the measured volume of the mixture may differ from the actualamount of liquid that emerges from the pipe downstream.

So how can you tell when a liquid contains gas? When bubbles are present in a liquidstream, Coriolis meters will report an increase in drive gain coinciding with a decrease inboth fluid density(1) and mass flow rate(2) due to the lower amount of mass contained inthe liquid-gas mixture. Therefore, in order to measure only the liquid portion of thestream, the volume of the bubbles must be ignored or subtracted from the mixturetotal.(3) Advanced Phase Measurement software performs exactly this function, using drivegain as the diagnostic indication that bubbles or entrained gas is present in the liquid flowstream, and then substituting a liquid-only density in place of the live measurement untilthe gaseous event has subsided. When the gassy portion has passed, indicated by anassociated drop in drive gain, the software returns to reporting the live measured volumeflow rate.

(1) High frequency sensors may erroneously report a higher fluid density when entrained gas is present, and therefore are not recommended foruse on liquids with entrained gas. High frequency sensors include F300, H300, and all T-Series sensors.

(2) The accuracy and repeatability of the mass flow and density measurements for liquids with entrained gas is dependent on the sensor-fluiddecoupling ratio, which is a complex function of fluid velocity, fluid viscosity, fluid density, the difference between the liquid and gasdensities, the operating sensor frequency, and the Gas Volume Fraction (GVF) of gas. For best measurement performance, GVF should bekept below 15%.

(3) The unmeasured gases can be (and are often) collected and processed separately downstream if desired (using a separator for example).

Measurement options and configuration

Application Manual 7

Page 12: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

How entrained gas affects drive gain and density measurementFigure 2-1:

A

B

C

DG%t

ρ

D

• DG% = drive gain percentage• t = Time• ρ = DensityA. Drive gain indicationB. Measured densityC. Entrained gas ocurring during these intervalsD. Drive gain threshold

Entrained gas in liquids affects drive gain and density measurement. The green line (A) shows the drivegain indication which is stable under most single-phase conditions. However, if gas is entrained in theliquid, the density reading (B) will drop and the drive gain reading will increase. When the entrained gasbubbles go away, the drive gain will return to its normal indication and the density measurement willreflect the density of the liquid.

The Advanced Phase Measurement software identifies entrained gas in liquid flow bydetecting the sharp increases in drive gain and corresponding decreases in densitymeasurement. The software continuously monitors the most recent drive gain data (up to60 minutes) to determine drive gain threshold. If the measurement exceeds the drive gainthreshold, the fluid is deemed to contain entrained gas, and remediation occurs.

Production type options

Option Description

Continuous(default)

Select this option only when flow rates are expected to be stable undernormal operating conditions. The Advanced Phase Measurementsoftware assumes the liquid properties and flow rate through the pipe isconsistent, and hence is able to remediate the mass flow values inaddition to density and volume values.

Variable Select this option when flow rates are not stable, such as batching, two-stage valve control, beam pumps, or other variable processes. Whenvariable flow is selected, Advanced Phase Measurement will remediateonly density and volume values.

Measurement options and configuration

8 Micro Motion® Advanced Phase Measurement

Page 13: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Remediation options

If the drive gain threshold is exceeded, the volume calculation for the period of high drivegain can be handled in either of two ways.

Option Description

Hold LastValue

Use a held density value from an earlier point in the process to calculatevolume. If this option is chosen, the density from the point just beforethe entrained gas event is effectively held constant throughout theevent.

Hold Last Value in operationFigure 2-2:

A

B

C

DG%t

ρ

D

• DG% = drive gain percentage• t = Time• ρ = DensityA. Drive gainB. Measured densityC. Held density value during entrained gas intervalsD. Drive gain thresholdThis figure shows how the Hold Last Value feature works in APM. The green line (A)shows the drive gain value and the blue line (B) shows the density reading. If theliquid gets entrained with gas bubbles, the drive gain increases above the drive gainthreshold (D). Then the software “looks back” in time and selects a density value (C)that does not have a high drive gain. It then substutes that value for the measureddensity until the drive gain goes back below the threshold value (D). This substituteddensity is also referred to as the remediated density. The value of the drive gainthreshold (D) and the “look back” time are automatically determined by thesoftware.

Measurement options and configuration

Application Manual 9

Page 14: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Option Description

Use InputDensity of DryOil Convertedto LineConditions

Convert the density of dry oil at reference temperature (a user-configured value) to density at line temperature, and calculate volume.This option assumes that all volume during the entrained gas event isdry oil.(4)

2.1.1 Configure Liquid with gas measurementVerify or make changes to the configuration settings.

NoteThe parameter names shown appear on the local display. Parameter names using other interfacesmay vary. The full list of all parameters and their locations for most typical configuration tools, see Appendix A.

Configuration settings for liquid with gas measurementTable 2-1:

Parameter Setting Additional information

Mass Flow Cutoff A non-zero value In most installations, the default value is satis-factory.

Two-Phase Flow Low Limit 0 g/cm³ These are the default values. This effectively dis-ables standard two-phase flow processing,which ensures that APM software processingwill be used.

Two-Phase Flow High Lim-it

5 g/cm³

Fluid Type Liquid with Gas

Production Type • Continuous Flow if the flow in your system is generally constant, e.g.,natural production, ESP.

• Variable Flow if the flow in your system frequently starts and stops orfluctuates between high and low flow rates, e.g., pump jack, test sep-arator, batching, 2-stage valve control.

Gas @ Line • Default (zero) if the line pressure is less than 250 psi absolute.• If the density of the entrained gas at typical line conditions is greater

than 250 psi absolute, set Gas @ Line to the density of the entrainedgas at typical line conditions.

Reporting Interval The time period, in minutes, over which process variables will be aver-aged. The averages are available to the transmitter for retrieval by thehost system.

Contract totals(1)

Time Zone The time zone that you are using for the contract period.

Special Time Zone The number of hours to add or subtract from UTC.(2)

(4) This option is only available when combined with Net Oil (order option PO). Consult factory to enable a user-input liquid-only density inapplications with continuous entrained gas, that do not include the Net Oil option.

Measurement options and configuration

10 Micro Motion® Advanced Phase Measurement

Page 15: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Configuration settings for liquid with gas measurement (continued)Table 2-1:

Parameter Setting Additional information

Contract Start Hour The time of day at which the contract starts. Enter the time in a 24-hourHH:MM format, where 00:00 = midnight (e.g., 18:30 for 6:30 pm

(1) You cannot configure contract totals using the display. If you want to configure contract totals, you must useone of the other available software configuration tools.

(2) This change affects the transmitter clock used for all purposes, not only for the contract period.

2.2 Net oilThe Net Oil algorithm calculates the water fraction of the liquid stream so that net oil andnet water can be determined. Net Oil is the volume of oil, corrected to a referencetemperature, that is contained within the gross volume of produced fluid.

NoteThe Net Oil measurement option can also be combined with the standalone liquid with gasremediation function, which gives it the added capability to fremediate against intermittententrained gas.

This algorithm requires the following data:

• Flow rate and temperature, which are measured by the meter.

• Density of both dry oil and water from this well at reference conditions. These aredetermined by the operator and entered during configuration. See Section C.3 formore information about density determination.

• Current water cut, either

- measured by a water cut monitor and supplied to the Advanced PhaseMeasurement software via the mA input or host system(5), or

- calculated by the Advanced Phase Measurement software from current densitydata via the density-based net oil calculation. If density-based water cut ischosen, the software uses the following equation to calculate the water cut.

(5) Not available for the Model 5700 fieldbus version 1.x.

Measurement options and configuration

Application Manual 11

Page 16: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Calculation of density-based water cut Equation 2-1:

WaterCut= ρl–ρOρW–ρO• ρl = Density of the liquid (oil/gas mixture) as measured by the sensor• ρO = Density of produced oil (calculated from user-supplied value)• ρW = Density of produced water (calculated from user-supplied value)

2.2.1 Configure net oil measurementPrerequisites

If you plan to use a water cut monitor:

• For the Model 5700 configurable transmitter,Channel D on the transmitter must beenabled, and must be configured as a mA input, wired to the water cut monitor, andconfigured appropriately. (HART integration is not currently supported.)

• For the Model 5700 Ethernet transmitter, an external host system is required toaccept the water cut monitor input and feed it to the transmitter.

NoteIt is not possible to use a water cut monitor with the Model 5700 fielbus version 1.x.

Procedure

Verify or make changes to the confiuration settings.

NoteThe parameter names shown appear on the local display. Parameter names using other interfacesmay vary. The full list of all parameters and their locations for most typical configuration tools, see Appendix A.

Configuration settings for net oil measurementTable 2-2:

Parameter Setting Additional information

Mass Flow Cutoff A non-zero value In most installations, the default value is satis-factory.

Two-Phase Flow Low Limit 0 g/cm³ These are the default values. This effectively dis-ables standard two-phase flow processing,which ensures that APM software processingwill be used.

Two-Phase Flow High Lim-it

5 g/cm³

Fluid Type Net Oil (alone) or Net Oil and Liquid with Gas.

Measurement options and configuration

12 Micro Motion® Advanced Phase Measurement

Page 17: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Configuration settings for net oil measurement (continued)Table 2-2:

Parameter Setting Additional information

Oil Density @ Ref The density of dry oil, corrected to reference temperature. Obtained fromdensity determination.

Water Density @ Ref The density of water, corrected to reference temperature. Obtained fromdensity determination.

Production Type • Continuous Flow if the flow in your system is generally constant, e.g.,natural production, ESP.

• Variable Flow if the flow in your system frequently starts and stops orfluctuates between high and low flow rates, e.g., pump jack, test sep-arator, batching, 2-stage valve control. If you select this option thereis an additional parameter, Density Corrective Action that can be set to ei-ther Hold Last Value (default) or Density Oil @ Line. You cannot configureDensity Corrective Action using the display; you must use one of the oth-er available tools.

Gas @ Line The density of en-trained gas.• Default (zero) if the

line pressure is lessthan 250 psi abso-lute.

• If the density of theentrained gas attypical line condi-tions is greaterthan 250 psi abso-lute, set Gas @ Lineto the density ofthe entrained gasat typical line con-ditions.

This setting is only needed when Net Oil is com-bined with Liquid with Gas.

Reference Temp The temperature to which net oil and net water measurements will becorrected.

View Production Meas The type of net oil data that will be shown on the display.• If Corrected to Standard is selected, the display will show Watercut @ Ref,

Net Oil Flow @ Ref, etc.• If Uncorrected is selected, the display will show Watercut @ Line, Net Oil

Flow @ Line, etc.This parameter is applicable only if a net oil process variable is configuredas a display variable.

Reporting Interval The time period, in minutes, over which process variables will be aver-aged. The averages are available for retrieval by the host system.

Contract totals(1)

Time Zone The time zone that you are using for the contract period.

Special Time Zone The number of hours to add or subtract from UTC.(2)

Measurement options and configuration

Application Manual 13

Page 18: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Configuration settings for net oil measurement (continued)Table 2-2:

Parameter Setting Additional information

Contract Start Hour The time of day at which the contract starts. Enter the time in a 24-hourHH:MM format, where 00:00 = midnight (e.g., 18:30 for 6:30 pm

(1) You cannot configure contract totals using the display. If you want to configure contract totals, you must useone of the other available tools.

(2) This change affects the transmitter clock used for all purposes, not only for the contract period.

2.3 Gas with liquidThis measurement option improves mass flow measurement in gaseous processes withintermittent entrained liquids (mist).

Gas with liquid measurement process

The presence of entrained liquids (or mist) can cause significant errors when measuringthe mass flow of gas through a Coriolis meter. Because droplets contribute a relativelylarge amount of mass to a gas mixture, but do not always move uniformly through thepipe with the gas, even a small amount of condensate can cause measurementdiscrepancies between what is measured by the meter and what emerges from the pipedownstream.

So how can you tell when a gas contains mist? When liquids are present in a gas stream,Coriolis meters will report an increase in both drive gain and fluid density(6) in combinationwith an increase in mass flow rate – due to the higher mass of the combined gas-liquidmixture. Therefore, in order to measure only the gaseous portion of the stream, the massof the liquids must be ignored or subtracted from the mixture total.(7) The Advanced PhaseMeasurement software performs exactly this function, using drive gain as the diagnosticindication that mist or entrained liquids are present in the gas flow stream, and thensubstituting a gas-only flow rate in place of the live measurement until the liquid event hassubsided. When the mist event is over, indicated by an associated drop in drive gain, thesoftware returns to reporting the live measured mass flow rate. If the flow rate after theevent differs significantly from the rate prior to the event, the software will apply anadjustment to the mass flow output until the totalizers accurately represent the flow ratechange that occurred during the mist event.(8)

The following figure shows how the change in density and drive gain are affected by mist.

(6) Coriolis meters do not measure the density of gases accurately, but the density reading can be combined with drive gain as a usefuldiagnostic to detect changes in fluid properties.

(7) The unmeasured liquids can be (and are often) collected and processed separately downstream if desired.

(8) The outputs are adjusted by a maximum of ±10% of live reading.

Measurement options and configuration

14 Micro Motion® Advanced Phase Measurement

Page 19: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Effect of transient mist on drive gain and flow measurementFigure 2-3:

B

A C C

FLOWDG%

t

D

• FLOW = mass flow rate• DG% = drive gain percentage• t = TimeA. Drive gain indicationB. Measured mass flow rateC. Entrained liquid (mist) occurring during these intervalsD. Drive gain threshold

The green line (A) shows the drive gain indication during flow. If mist gets entrained into the gas, the drivegain and the mass flow rate (B) will both increase. The red line (D) shows a drive gain threshold abovewhich mist is entrained in the gas. APM will automatically choose an appropriate drive gain threshold (D).

The Advanced Phase Measurement software identifies entrained mist in gas flow bydetecting the sharp increases in drive gain and corresponding increases in mass flowmeasurement. The software continuously monitors the most recent drive gain data (up to60 minutes) to determine drive gain threshold If the measurement exceeds the drive gainthreshold, the fluid is deemed to contain entrained liquid, and remediation occurs.

Gas with liquid remediation

The following figure illustrates Advanced Phase Measurement software processing whenmist is detected in the gas stream.

Measurement options and configuration

Application Manual 15

Page 20: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Advanced Phase Measurement software: Gas with liquid remediationFigure 2-4:

B

A

C

FLOWDG%

t

D

E

F FG

EC

G

• FLOW = mass flow rate• DG% = drive gain percentage• t = TimeA. Drive gain indicationB. Measured mass flow rateC. Pre-mist averaging period and source of average flow rate during this periodD. Drive gain thresholdE. Post-mist adjustment delay and source of average flow rate during this periodF. Held flow rate during entrained mist intervalsG. Post-mist adjustment applied to measured flow rate (if applicable)

The Advanced Phase Measurement software uses the drive gain and mass flow readings to correct for thepresence of entrained mist in gas. The green line (A) shows the drive gain during flow. The blue line (B)shows the measured mass flow reading both with, and without entrained gas. When mist is present, thedrive gain will go above the drive gain threshold (D). This threshold value is automatically determined byAPM. When the drive gain exceeds this threshold, APM will look back in time (C) seconds, and determine anaverage, pre-mist, mass flow rate. It will then substitute mass flow rate (F) until the drive gain goes backbelow the threshold. Then during time period (E), a post-mist, mass flow rate average is determined. If thepost- and pre-mist averages are different, a mass flow adjustment (G) is made to the mass rate and totalafter time period (E).

The first mist event had equal mass flow rates before and after the mist event, so (G) is equal to (F). In thesecond mist event, the post-mist flow rate was greater than the pre-mist rate, so a (G) was applied thatwas greater than (F). This held flow rate (F) is also known as the remediated mass flow rate.

2.3.1 Configure gas with liquid measurementVerify or make changes to the confiuration settings.

Measurement options and configuration

16 Micro Motion® Advanced Phase Measurement

Page 21: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

NoteThe parameter names shown appear on the local display. Parameter names using other interfacesmay vary. The full list of all parameters and their locations for most typical configuration tools, see Appendix A.

Configuration settings for gas with liquid measurementTable 2-3:

Parameter Setting Additional information

Mass Flow Cutoff A non-zero value In most installations, the default value is satis-factory.

Two-Phase Flow Low Limit 0 g/cm³ These are the default values. This effectively en-sures that Advanced Phase Measurement soft-ware processing will be used.

Two-Phase Flow High Lim-it

5 g/cm³

Fluid Type Gas with Liquid

Pre-Mist Ave Period The number of seconds over which mass flow data will be averaged, toproduce the substitute mass flow value.

Post-Mist Adj Delay The number of seconds that the system will wait before beginning meas-urment remediation

Reporting Interval The time period, in minutes, over which process variables will be aver-aged. The averages are available for retrieval by the host system.

Contract totals(1)

Time Zone The time zone that you are using for the contract period.

Special Time Zone The number of hours to add or subtract from UTC.(2)

Contract Start Hour The time of day at which the contract starts. Enter the time in a 24-hourHH:MM format, where 00:00 = midnight (e.g., 18:30 for 6:30 pm)

(1) You cannot configure contract totals using the display. If you want to configure contract totals, you must useone of the other available tools.

(2) This change affects the transmitter clock used for all purposes, not only for the contract period.

Measurement options and configuration

Application Manual 17

Page 22: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Measurement options and configuration

18 Micro Motion® Advanced Phase Measurement

Page 23: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

3 Additional configurationTopics covered in this chapter:

• Configure viewing and reporting for process variables

• Configure the totalizer history log to include Advanced Phase Measurementcontract totals

• Configure events

3.1 Configure viewing and reporting for processvariablesWhen Advanced Phase Measurement software is enabled, additional process variables areavailable.

• To configure a process variable as a display variable, or to report a process variableover an output, see the transmitter configuration and use manual.

• To query a process variable using Modbus, follow standard Modbus programmingtechniques.

3.1.1 Advanced Phase Measurement specific process variablesThe process variables listed here are available only when Advanced Phase Measurementsoftware is enabled.

The following table lists the process variables by measurement option, and providesinformation on reporting.

Advanced Phase Measurement process variablesTable 3-1:

Process variable

Advanced Phase Measurementoption Viewing and reporting

Liquidwith gas Net Oil

Gas withliquid Display

ModbusandEthernet

Field-bus(1)

mAO1,mAO2,mAO3 FO1, FO2

Gas Void Fraction ✓ ✓ ✓ ✓ ✓

DensityOil@Line ✓ ✓ ✓ ✓

DensityOil@Ref ✓ ✓ ✓ ✓

NetFlowOil@Line ✓ ✓ ✓ ✓ ✓ ✓

NetFlowOil@Ref ✓ ✓ ✓ ✓ ✓ ✓

NetTotalOil@Line ✓ ✓ ✓ ✓

NetTotalOil@Ref ✓ ✓ ✓ ✓

Additional configuration

Application Manual 19

Page 24: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Advanced Phase Measurement process variables (continued)Table 3-1:

Process variable

Advanced Phase Measurementoption Viewing and reporting

Liquidwith gas Net Oil

Gas withliquid Display

ModbusandEthernet

Field-bus(1)

mAO1,mAO2,mAO3 FO1, FO2

NetFlowWater@Line ✓ ✓ ✓ ✓ ✓ ✓

NetFlowWater@Ref ✓ ✓ ✓ ✓ ✓ ✓

NetTotalWater@Line ✓ ✓ ✓ ✓

NetTotalWater@Ref ✓ ✓ ✓ ✓

Watercut@Line ✓ ✓ ✓ ✓ ✓

Watercut@Ref ✓ ✓ ✓ ✓ ✓

APM contract period

Today's total ✓ ✓ ✓ ✓ ✓

Yesterday's total ✓ ✓ ✓ ✓ ✓

(1) For fieldbus version 1.x transmitters, you can only publish through the AI Blocks four process variables and two totalizers/inventories.

3.1.2 Default display variablesThe following table lists the default display variables for Advanced Phase Measurementsoftware. In all cases, the two-line display option is enabled. All of these process variablescan be configured as display variables.

Default display variablesTable 3-2:

Display variable

Advanced Phase Measurement option

Liquid with gas NOC with gas Gas with liquid

Two-line display, Varia-ble 1

Volume Flow Rate Volume Flow Rate Mass Flow Rate

Two-line display, Varia-ble 2

Volume Total Water Cut@Line Mass Total

Display Variable 1 Density Net Oil Flow Rate@Line Temperature

Display Variable 2 Temperature Net Oil Total@Line Density

Display Variable 3 GVF Net Water Total@Line Empty

Display Variable 4 Empty Temperature Empty

Display Variable 5 Empty GVF Empty

Display Variable 6–15 Empty Empty Empty

Additional configuration

20 Micro Motion® Advanced Phase Measurement

Page 25: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

3.1.3 Period averaged output (PAO) configurationPeriod averaged outputs provide an easy and reliable way to avoid data collection systemsfrom recording outliers in erratic data. This is included with Advanced Phase Measurementsoftware since two-phase conditions cause more volatile measurement outputs. Theperiod over which the PAOs are averaged can be adjusted using the Average Reporting Intervalparameter.

Example density measurement without averaged outputs and withaverage outputs

Figure 3-1:

B

A

t 1

0.6

0.7

0.8

0.9

2 3

ρt 1

0.6

0.7

0.8

0.9

2 3

ρ

• ρ = Density• t = TimeA. Density measurement without averaged outputsB. Density measurement with averaged outputs

Additional configuration

Application Manual 21

Page 26: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

3.2 Configure the totalizer history log to includeAdvanced Phase Measurement contract totalsThe transmitter can store Advanced Phase Measurement contract totals to the totalizerhistory log. This allows you to access totals from earlier contract periods. Otherwise, thetransmitter maintains data for only the current contract period (today) and theimmediately preceding contract period (yesterday).

1. Ensure that you have configured the Advanced Phase Measurement contract totalsas desired.

2. Set Log Total 1, Log Total 2, Log Total 3, and/or Log Total 4 to the desired Advanced PhaseMeasurement contract total.

You can configure the totalizer history log to include both Advanced PhaseMeasurement and standard totals.(1)

NoteDo not change the settings of Start Logging Date, Start Logging Time, or Log Interval. Theseparameters are set automatically by the Advanced Phase Measurement software.

3.3 Configure eventsWhen Advanced Phase Measurement software is enabled, additional process variables areavailable to use in event configuration.

See the transmitter configuration manual for instructions on configuring events.

(1) For fieldbus version 1.x transmitters, any two of the publishable totalizers and inventories can be used, but only two at a time.

Additional configuration

22 Micro Motion® Advanced Phase Measurement

Page 27: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

4 Transmitter operationTopics covered in this chapter:

• Read process variables

• Read contract totals

4.1 Read process variablesAdvanced Phase Measurement process variables can be configured as display variables orassigned to outputs. See the transmitter configuration manual for information on usingthe display to read process variables. You can also use the host system for your Modbus,fieldbus, or Ethernet/IP network to read the variables.

4.2 Read contract totalsAdvanced Phase Measurement software operates on a 24-hour contract period. You canread contract totals for the current contract period and for the previous contract period.Depending on the configuration of the totalizer history log, you may be able to readcontract totals for earlier periods.

The contract totals are derived from existing inventories. However, they are resetautomatically at the beginning of each contract period. Therefore, the values shown willprobably not match the values shown for the inventories.

ImportantYou can reset inventories manually, and you can stop and start inventories manually. However, if youdo this, data for the current contract period will not reflect the entire 24-hour period. Data for earliercontract periods is not affected.

• The contract totals for the current contract period are stored in the Today's Total [1-4]parameters.

• The contract totals for the previous contract period are stored in the Yesterday's Total[1-4] parameters.

• The contract totals from earlier contract periods can be read in the totalizer historylog. See the transmitter configuration manual for more information.

Related information

Configure the totalizer history log to include Advanced Phase Measurement contracttotals

Transmitter operation

Application Manual 23

Page 28: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Transmitter operation

24 Micro Motion® Advanced Phase Measurement

Page 29: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Appendix AApplication parameters and data

Topics covered in this appendix:

• Advanced Phase Measurement configuration parameters

• Modbus interface to configuration parameters

• Modbus interface to process data

A.1 Advanced Phase Measurement configurationparametersThe model 5700 transmitter can be configured using many different methods, dependingupon the output option selected when it was ordered. This section provides a reference forthe transmitter Advanced Phase Measurement related parameters for each of thefollowing configuration tool types: the Display , ProLink III, the Field Communicator, anenhanced FF host (FOUNDATION™ fieldbus), a basic FF host, and a Web browser

Where to find the parameters

The Advanced Phase Measurement parameters are typically found in the followinglocations:

• The Display : Menu > Configuration > Process Measurement > Adv Phase Measurement• ProLink III: Device Tools > Configuration > Process Measurement > Advanced Phase

Measurment.• Field Communicator or an enhanced FF host: Configure > Manual Setup > Advanced Phase

Measurment.• Basic FF host: the APM transducer block.

• Web browser: Configuration > Process Measurement > Advanced Phase Measurement.

NoteFor the location of parameters not exclusive to the Advanced Phase Measurement software, refer tothe transmitter configuration and use manual.

A.2 Modbus interface to configuration parametersThis section provides information on the Modbus registers that can be used in AdvancedPhase Measurement software configuration.

Application parameters and data

Application Manual 25

Page 30: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

TipTypically, Advanced Phase Measurement configuration is performed using either ProLink III or thedisplay. This information is provided for completeness.

RestrictionThis section includes only Modbus registers that are specific to or redefined for the Advanced PhaseMeasurement software configuration. To use Modbus to configure other parameters, see theModbus Interface Tool, which is available on the Model 5700 product page(http://www2.emersonprocess.com/en-US/brands/micromotion/coriolis-transmitters/model-5700/Pages/index.aspx).

Advanced Phase Measurement Modbus configuration parametersTable A-1:

Parameter

Advanced Phase Measurement op-tion Modbus

Liquidwith gas

Net Oilwith gas

Gas withliquid Address Data type

Integer codes / Unit(Range)

Output Type ✓ ✓ ✓ 3940 U16 • 0 = No remediation• 1 = Liquid with gas,

continuous flow• 2 = NOC with gas,

continuous flow• 3 = Liquid with gas,

variable flow• 4 = NOC with gas,

variable flow• 5 = Gas with liquid• 6 = NOC only

Gas Density(1) ✓ ✓ ✓ 3935 Float Configured unit

Density Corrective Ac-tion

✓ 4450 U16 • 0 = Hold Last Value• 1 = Density Oil @ Line

Reference Temperature ✓ 319 Float Configured unit

Dry Oil Density @ Ref ✓ 1959 Float g/cm³ (0.2–1.5)

Water Density @ Ref ✓ 1831 Float g/cm³ (0.5–1.5)

Pre-Mist Averaging Peri-od

✓ 619 U16 seconds (2–128)

Post-Mist Averaging Pe-riod

✓ 620 U16 seconds (2–128)

Average Reporting Inter-val

✓ ✓ ✓ 3900 U16 minutes (1–1440)

Contract Start Time ✓ ✓ ✓ 3966 U16 hours (0–23)

Contract Total 1 ✓ ✓ ✓ 3967 U16 • 4 = Inventory 1• 7 = Inventory 2• 18 = Inventory 3• 64 = Inventory 4

Contract Total 2 ✓ ✓ ✓ 3968 U16

Contract Total 3 ✓ ✓ ✓ 3969 U16

Application parameters and data

26 Micro Motion® Advanced Phase Measurement

Page 31: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Advanced Phase Measurement Modbus configuration parameters (continued)Table A-1:

Parameter

Advanced Phase Measurement op-tion Modbus

Liquidwith gas

Net Oilwith gas

Gas withliquid Address Data type

Integer codes / Unit(Range)

Contract Total 4 • 25 = Inventory 5• 28 = Inventory 6• 31 = Inventory 7

✓ ✓ ✓ 3970 U16

(1) At line conditions

Default totalizer and inventory values by Advanced Phase Measurementoption

Table A-2:

Totalizer/Inventory Liquid with Gas Gas with Liquid Net Oil

1 Mass Flow (Remedi-ated)

Mass Flow (Remedi-ated)

Mass Flow

2 Volume Flow (Remedi-ated)

Volume Flow (Remedi-ated)

Volume Flow

3 Temperature Correc-ted Volume

Temperature Correc-ted Volume

Net Oil @ Ref

4 Gas Standard Volume Gas Standard Volume Net Oil @ Line

5 Standard Volume Standard Volume Net Water @ Ref

6 Net Mass Net Mass Net Water @ Line

7 Net Volume Net Volume Net Volume

A.3 Modbus interface to process dataThis section provides information on the Modbus registers that contain Advanced PhaseMeasurement software process data.

RestrictionThis section includes only process data that is specific to Advanced Phase Measurement software. Touse Modbus to read other process data, see the Modbus Interface Tool, which is available on theModel 5700 product page(http://www2.emersonprocess.com/en-US/brands/micromotion/coriolis-transmitters/model-5700/Pages/index.aspx).

Application parameters and data

Application Manual 27

Page 32: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Advanced Phase Measurement Modbus process variablesTable A-3:

Process variable(1)

Advanced Phase Measurement op-tion Modbus

Liquidwith gas

Net Oilwith gas

Gas withliquid Address Data type Unit

Gas Void Fraction ✓ 3907 Float %

Density Oil @ Line ✓ 345 Float SGU

Density Oil @ Line ✓ 347 Float °API

Net Oil Flow @ Line ✓ 1553 Float Configured unit

Net Oil Flow @ Ref ✓ 1547 Float Configured unit

Net Oil Total @ Line(2) ✓ 1665 Float Automatically derivedfrom configured unit

Net Oil Total @ Line(3) ✓ 4240 Double Automatically derivedfrom configured unit

Net Oil Total @ Ref(2) ✓ 1661 Float Automatically derivedfrom configured unit

Net Oil Total @ Ref(4) ✓ 4236 Double Automatically derivedfrom configured unit

Net Water Flow @ Line ✓ 1561 Float Configured unit

Net Water Flow @ Ref ✓ 1549 Float Configured unit

Net Water Total @Line (2)

✓ 1667 Float Automatically derivedfrom configured unit

Net Water Total @ Line(5) ✓ 4248 Double Automatically derivedfrom configured unit

Net Water Total @ Ref ✓ 1663 Float Automatically derivedfrom configured unit

Net Water Total @ Ref(6) ✓ 4244 Double Automatically derivedfrom configured unit

Watercut @ Line ✓ 1555 Float %

Watercut @ Ref ✓ 1557 Float %

Total Mist Time(7) ✓ 989 U32 Seconds

APM Liquid with Gas re-mediation status

✓ 433, Bit#12

U16 • 0 = Inactive• 1 = Active

Current period data (Period Averaged Outputs)

PAO Mass Flow ✓ ✓ ✓ 3949 Float Configured unit

PAO Density ✓ ✓ ✓ 3951 Float Configured unit

PAO Volume Flow ✓ ✓ ✓ 3953 Float Configured unit

PAO Net Oil Flow @ Line ✓ 3955 Float Configured unit

PAO Net Oil Flow @ Ref ✓ 3957 Float Configured unit

PAO Watercut @ Line ✓ 3959 Float Configured unit

Application parameters and data

28 Micro Motion® Advanced Phase Measurement

Page 33: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Advanced Phase Measurement Modbus process variables (continued)Table A-3:

Process variable(1)

Advanced Phase Measurement op-tion Modbus

Liquidwith gas

Net Oilwith gas

Gas withliquid Address Data type Unit

PAO Gas Void Fraction ✓ 3961 Float Configured unit

PAO Temperature ✓ ✓ ✓ 3963 Float Configured unit

APM contract period

Today's total 1 ✓ ✓ ✓ 3972 Float Automatically derivedfrom configured unit

Today's total 2 ✓ ✓ ✓ 3974 Float Automatically derivedfrom configured unit

Today's total 3 ✓ ✓ ✓ 3976 Float Automatically derivedfrom configured unit

Today's total 4 ✓ ✓ ✓ 3978 Float Automatically derivedfrom configured unit

Yesterday's total 1 ✓ ✓ ✓ 3980 Float Automatically derivedfrom configured unit

Yesterday's total 2 ✓ ✓ ✓ 3982 Float Automatically derivedfrom configured unit

Yesterday's total 3 ✓ ✓ ✓ 3984 Float Automatically derivedfrom configured unit

Yesterday's total 4 ✓ ✓ ✓ 3986 Float Automatically derivedfrom configured unit

Unremediated MassFlow

✓ ✓ ✓ 3943 Float Automatically derivedfrom configured uni

Unremediated Density ✓ ✓ ✓ 3945 Float Automatically derivedfrom configured uni

Unremediated VolumeFlow

✓ ✓ ✓ 3947 Float Automatically derivedfrom configured uni

(1) For fieldbus version 1.x transmitters, you can only publish through the AI Blocks four process variables and two totalizers/inventories.

(2) Legacy NOC register

(3) Inventory 4 (only if you are using the default configuration)

(4) Inventory 3 (only if you are using the default configuration)

(5) Inventory 6 (only if you are using the default configuration)

(6) Inventory 5 (only if you are using the default configuration)

(7) Automatically set to 0 on a power cycle

Application parameters and data

Application Manual 29

Page 34: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Application parameters and data

30 Micro Motion® Advanced Phase Measurement

Page 35: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Appendix BTypical applications and other information

Topics covered in this appendix:

• Advanced Phase Measurement with a two-phase separator

• Advanced Phase Measurement with a three-phase separator

• Advanced Phase Measurement at the wellhead

• Determining the Drive Gain Threshold

B.1 Advanced Phase Measurement with a two-phase separator

Advanced Phase Measurement with a two-phase separatorFigure B-1:

A. From wellheadB. SeparatorC. Gas legD. Oil/water legE. Coriolis sensor and transmitter with Advanced Phase Measurement (NOC with gas)F. Coriolis sensor and transmitter with Advanced Phase Measurement (Gas with liquid)G. Modbus host (flow computer)

Typical applications and other information

Application Manual 31

Page 36: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

B.2 Advanced Phase Measurement with a three-phase separator

Advanced Phase Measurement with a three-phase separatorFigure B-2:

A. From wellheadB. SeparatorC. Water legD. Oil legE. Gas legF. Coriolis sensor and transmitter with Advanced Phase Measurement (Liquid with gas, variable flow)G. Coriolis sensor and transmitter with Advanced Phase Measurement (Liquid with gas, variable flow)H. Coriolis sensor and transmitter with Advanced Phase Measurement (Gas with liquid)I. Modbus host (flow computer)

Typical applications and other information

32 Micro Motion® Advanced Phase Measurement

Page 37: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

B.3 Advanced Phase Measurement at the wellhead

Advanced Phase Measurement at the wellheadFigure B-3:

A. WellheadB. Coriolis sensorC. Transmitter with Advanced Phase Measurement (NOC with gas)D. Modbus host (flow computer)

B.4 Determining the Drive Gain ThresholdAn essential function of the Advanced Phase Measurement software is to monitor drivegain, and then use drive gain data to define and adjust the Drive Gain Threshold, whichultimatelly determines when and how measurement remediation is needed.

Typical applications and other information

Application Manual 33

Page 38: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Drive Gain Threshold determinationFigure B-4:

BA

D

E

C

BA

C

DG%t1 t2

D

• DG% = Drive Gain Percentage• t1 = Drive Gain Threshold interval 1• t2 = Drive Gain Threshold interval 2A. Drive Gain Threshold (measurement will be remediated if drive gain exceeds this threshold)B. Drive gain indicationC. Minimum drive gain detected at the beginning of each drive gain threshold interval.D. Drive Gain Threshold Addition (configurable). This value is added to the minimum drive gain (C) to

establish a small buffer so the software only remediates values that exceed the typical drive gainfluctuations for each unique process.

E. Advanced Phase Measurement software remediates density during these intervals.

This figure illustrates how the software determines the Drive Gain Threshold (A) under live conditions. Thegreen dotted line shows the Coriolis sensor’s live drive gain indication (B) over time (t). The softwarecontinuously analyzes the live drive gain using the most recent data (duration set by the Drive GainThreshold Interval) to determine the lowest drive gain that is typical for the process.(1)The live drive gainindication for most applications fluctuates a bit under normal operating conditions, typically within asmall range that is not attributable to two-phase flow or other process upsets. To avoid remediatingduring this typical process noise, a small Drive Gain Threshold Addition (D) is added to the lowest samplepoint (C). The newly established Drive Gain Threshold (A) represents the sampled minimum drive gain plusthe drive gain threshold addition. During the remainder of the threshold Interval, if the indicated drive gain(B) exceeds the established Drive Gain Threshold (A), the transmitter will appropriately remediate themeasured flow rate and/or the fluid density.(2) After the threshold interval is over, the process starts again.

(1) The factory default for Drive Gain Threshold Interval is 60 minutes, which is suitable for most continuous processes, but it may be shortenedor lengthened as required for each application. Contact Micro Motion for assistance.

(2) The measurement variables remediated by APM are configurable per license type and software and output configuration.

Typical applications and other information

34 Micro Motion® Advanced Phase Measurement

Page 39: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Appendix CBest practices for two-phase measurementperformance

Topics covered in this appendix:

• Entrained gas performance

• Entrained liquid (mist) performance

• Density determination

C.1 Entrained gas performanceMeasurement accuracy for liquids with entrained gas is a complex function of GVF,viscosity, velocity, sensor geometry, drive frequency, and orientation.

Common sources for unintentional gas entrainment

• Long drops from fill point to liquid level in tanks

• Agitators and mixers

• Leaks in seals or pumps

• Pumping out of nearly empty tanks

• pressure loss (flashing) for volatile liquids

• Pumping through nearly empty piping

Ways to minimize flow errors

• Use ELITE® (low frequency) sensors whenever possible. F-Series and H-Series sensorsare also acceptable, but less accurate.

• Do not use T-Series sensors or Models F300 or H300 because they have a highoperating frequency.

• Use the enhanced core processor (Model 800) or Model 5700 transmitters: theyperform best in applications with entrained gas.

• Orient the meter properly:

Best practices for two-phase measurement performance

Application Manual 35

Page 40: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Preferred sensor orientation for liquids with entrained gasTable C-1:

Process Preferred orientation

Delta-shaped sensors (CMF010, CMF025,CMF050, CMF100)

Any F-Series or CMFS sensor, and CMF200 orlarger (flow should go up)

• Ensure sensor is filled as quickly as possible, and stays full during measurement:

- For horizontal pipes, maintain a minimum flow velocity of 1 m/s to purge airfrom an empty pipe and keep it full.

- For vertical pipes, flow upward and maintain minimum velocity of 1 m/s toprevent solids from settling out of the fluid.

• Add back pressure, or increase line pressure, to minimize size of bubbles in flowstream.

• Ensure fluid is well mixed. If needed, you can install a blind "T" and/or static mixerjust upstream of the sensor to evenly distribute bubbles through both sensor tubes.If using a blind "T", install it in the same plane as the sensor tubes.

• If re-zeroing in the field is necessary, zeroing must be done on a pure liquid withoutbubbles in order to avoid error. If this cannot be done, use the factory zero.

• Size the meter appropriately to operate normally as close to the sensor nominal flowrate as is practical. Higher velocity leads to better performance.

• Minimize damping on outputs to minimize processing delay from electronics.

• Do not stop the totalizer immediately after batch; allow the totalizer to stabilize forapproximately 1 second.

• Set Flow Cutoff as high as is practical to avoid totalizing at no flow condition ifbubbles remain in the sensor.

NoteFor configuration and further information regarding the Concentration Measurement softwareoption, please refer to the Micro Motion Enhanced Density Application Manual 20002315: http://www2.emersonprocess.com/siteadmincenter/PM%20Micro%20Motion%20Documents/Enhanced-Density-Manual-20002315.pdf

Best practices for two-phase measurement performance

36 Micro Motion® Advanced Phase Measurement

Page 41: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

C.2 Entrained liquid (mist) performanceMeasurement accuracy for gases with entrained liquids (mist) is mostly related to theamount of mass contained in liquid droplets compared to an equivalent volume of gascontaining the same mass. The sensor geometry, drive frequency, and orientation cancause errors that reduce performance.

Common sources for unintentional liquid entrainment

• Temperature loss (condensation)

• Pressure increase

• Poorly managed level control in separators or GLCCs

• Malfunctioning or over-filled liquid traps

Ways to minimize measurement errors

• Use ELITE® (low frequency) sensors whenever possible. F-Series and H-Series sensorsare also acceptable, but less accurate.

• Do not use T-Series sensors or Models F300 or H300 because they have a highoperating frequency.

• Use the enhanced core processor (Model 800) or Model 5700 transmitters: theyperform best in applications with entrained liquid.

• Orient the meter properly:

Preferred sensor orientation when there could be entrained liquidTable C-2:

Process Preferred orientation

Delta-shaped sensors (CMF010, CMF025,CMF050, CMF100)

Any F-Series or CMFS sensor, and CMF200 orlarger (flow should go down)

• Ensure sensor is dried (blown-out) as quickly as possible, and stays dry duringmeasurement.

Best practices for two-phase measurement performance

Application Manual 37

Page 42: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

• Avoid temperature losses; insulation is highly recommended if condensate is causedby cooling temperatures.

• Avoid pressure increases in the system; Ensure that pressure regulators arefunctioning properly.

• If entrained liquid is unavoidable, try to ensure that the process is well mixed.

• Avoid elbows, valves, or other components that may introduce a flow profileaffecting one tube (e.g., a swirling motion entering the flow tubes)

• If re-zeroing in the field is necessary, zeroing must be done on a pure gas withoutliquid in order to avoid error. If this cannot be done, use the factory zero.

• Size the meter appropriately for gas flow. Avoid large turndowns where sensorsensitivity may be reduced.

• Minimize damping on outputs to minimize processing delay from electronics.

• Do not stop the totalizer immediately after batch; allow the totalizer to stabilize forapproximately 1 second.

• Set Flow Cutoff as high as is practical to avoid totalizing at no flow condition ifdroplets remain in the sensor.

C.3 Density determinationIf you are using the Net Oil application, you must know the density of water from the well,corrected to reference temperature, and the density of dry oil from the well, corrected toreference temperature.

ImportantMicro Motion recommends working with a laboratory to obtain the most accurate values. Theaccuracy of the net oil data depends upon the accuracy of these two density values.

C.3.1 Density determination using a petroleum laboratoryTo configure APM for net oil measurement, you must know the density of dry oil atreference temperature, and the density of produced water at reference temperature. Youcan obtain these values from a petroleum laboratory.

NoteEven after separation, oil typically contains some amount of interstitial water. The water cut may beas high as 1% to 3%. For purposes of this application, this is considered dry oil.

ImportantIf you are using a three-phase separator, you can collect the oil sample and the water sampleseparately, after separation, or you can collect one sample before separation and have the laboratoryperform the separation.

If you are using a two-phase separator, you should collect one sample before separation and havethe laboratory perform the separation.

Best practices for two-phase measurement performance

38 Micro Motion® Advanced Phase Measurement

Page 43: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Prerequisites

Sample collection must meet these requirements:

• You must be able to collect a sample that is representative of your process.

• The sample must be collected by a qualified person, using industry-accepted safetystandards.

• You must know the minimum required sample size. This varies depending on thewater cut and the volume of the sample cylinder. Consult the petroleum laboratoryfor specific values.

• If the sample contains oil, you must be able to collect and maintain the sample atline pressure, so that the oil will not lose pressure and outgas. This will change thelaboratory-measured density.

• If you collect the water sample separately, you must be able to protect it fromcontamination and evaporation.

You must know the reference temperature that you plan to use.

The petroleum laboratory must be able to meet these requirements:

• The laboratory density meter must be able to keep the oil sample pressurized at linepressure during the density measurement.

• The sample cylinder must be a constant-pressure type, and must be properly ratedfor the oil–water composition and for sample pressure.

• The oil and water density measurement units must be entered into the APMsoftware in g/cm³ at reference temperature.

• The laboratory report must include the oil density, water density, and the referencetemperature.

Procedure

1. Communicate the handling and measurement requirements and the referencetemperature to the petroleum laboratory.

2. If you are collecting one sample that contains both oil and water, identify the pointin the line where the sample will be taken.

Recommendations:

• Collect the sample at a point where the fluid is well mixed.

• The line pressure at the sample point should be close to the line pressure at thesensor.

• The line temperature at the sample point should be close to the line temperatureat the sensor.

3. If you are using a three-phase separator and collecting the oil and water samplesseparately:

a. Identify the points where the samples will be taken.

Recommendations:

• The sample point for oil must be on the oil leg, as close to the sensor aspossible.

Best practices for two-phase measurement performance

Application Manual 39

Page 44: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

• The line pressure at the oil sample point should be similar to the line pressureat the sensor.

• The sample point for water must be on the water leg, as close to the sensor aspossible.

• The line temperature at the water sample point should be similar to the linetemperature at the sensor.

b. Wait until separation has occurred.

4. Collect the sample or samples, meeting all requirements for pressure and protectionfrom contamination or evaporation.

5. Mark and tag the sample or samples with the well name or number, time and date,sample type, line pressure, and line temeprature.

6. Transport the samples to the laboratory safely, as soon as is practical.

Postrequisites

If the laboratory measurements were not corrected to your reference temperature, usethe Oil & Water Density Calculator to calculate density at reference temperature. This is aspreadsheet tool developed by Micro Motion. You can obtain a copy by visiting http://www.emersonoildensityref.com or from your Micro Motion representative.

C.3.2 Density determination using a three-phase separatorTo configure net oil measurement, you must know the density of dry oil at referencetemperature, and the density of produced water at reference temperature. If you have athree-phase separator, you can use density data and the Oil & Water Density Calculator toobtain these values.

NoteEven after separation, oil typically contains some amount of interstitial water. The water cut may beas high as 1% to 3%. For purposes of this application, this is considered dry oil.

Prerequisites

You must have a three-phase separator in the process. You can use a mobile three-phasetest separator.

You must have a sensor and transmitter installed on the oil leg, and a sensor andtransmitter installed on the water leg.

You must know the reference temperature that you plan to use.

You must have the Oil & Water Density Calculator. This is a spreadsheet tool developed byMicro Motion. You can obtain a copy by visiting http://www.emersonoildensityref.com orfrom your Micro Motion representative.

ImportantThe accuracy of net oil data depends on the accuracy of the density data. Never use an unstabledensity value, or any density value that has an elevated drive gain.

Best practices for two-phase measurement performance

40 Micro Motion® Advanced Phase Measurement

Page 45: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Procedure

1. Wait until separation has occurred.

2. At the transmitter on the oil leg, read and record the density value and thetemperature value.

3. At the transmitter on the water leg, read and record the density value and thetemperature value.

4. Use the Oil & Water Density Calculator to calculate the density of dry oil at referencetemperature and the density of produced water at reference temperature. You canobtain a copy by visiting http://www.emersonoildensityref.com or from yourMicro Motion representative.

TipUnless the oil is light hot condensate, the oil will almost always contain some interstitialwater. This is generally acceptable for allocation measurements. However, if further accuracyis desired, you can determine the water cut and use it in the calculation. To determine orestimate the water cut, take a shakeout sample from one of the following:

• The current flow/dump cycle, at the time of minimum density

• Similar oils produced from the same reservoir

• The tank or tanks that the separator flows into

Enter this water cut into the Oil & Water Density Calculator to calculate the density of dry oilat reference temperature.

Best practices for two-phase measurement performance

Application Manual 41

Page 46: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Best practices for two-phase measurement performance

42 Micro Motion® Advanced Phase Measurement

Page 47: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

Best practices for two-phase measurement performance

Application Manual 43

Page 48: Advanced Phase Measurement Application Manual …€¦ · Micro Motion® Advanced Phase Measurement Application Manual. Safety messages ... B.2 Advanced Phase Measurement with a three-phase

*MMI-20030076*MMI-20030076

Rev AB

2017

Micro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301T +1 303-527-5200T +1 800-522-6277F +1 303-530-8459www.micromotion.com

Micro Motion EuropeEmerson Automation SolutionsNeonstraat 16718 WX EdeThe NetherlandsT +31 (0) 70 413 6666F +31 (0) 318 495 556www.micromotion.nl

Micro Motion AsiaEmerson Automation Solutions1 Pandan CrescentSingapore 128461Republic of SingaporeT +65 6777-8211F +65 6770-8003

Micro Motion United KingdomEmerson Automation SolutionsEmerson Process Management LimitedHorsfield WayBredbury Industrial EstateStockport SK6 2SU U.K.T +44 0870 240 1978F +44 0800 966 181

Micro Motion JapanEmerson Automation Solutions1-2-5, Higashi ShinagawaShinagawa-kuTokyo 140-0002 JapanT +81 3 5769-6803F +81 3 5769-6844

©2017 Micro Motion, Inc. All rights reserved.

The Emerson logo is a trademark and service mark of EmersonElectric Co. Micro Motion, ELITE, ProLink, MVD and MVD DirectConnect marks are marks of one of the Emerson AutomationSolutions family of companies. All other marks are property of theirrespective owners.