Acceptance testing, commissioning and routine QA of the Varian TrueBeam
Richard Popple
![Page 2: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/2.jpg)
Disclosure
• UAB has Varian research contracts • Honorarium from Varian
![Page 3: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/3.jpg)
Not your father’s Chevy…
![Page 4: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/4.jpg)
Not your father’s Chevy…
![Page 5: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/5.jpg)
Not your father’s Chevy…
![Page 6: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/6.jpg)
Not your father’s Chevy Linac…
![Page 7: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/7.jpg)
TrueBeam – distinquishing characteristics • Completely digital control system • Waveguide and filter design allows 5 flattened
photon energies up to 20 MV. • Electron scattering foils are of a new design. • Implements 6 MV and 10 MV flattening filter free
(FFF) beams that provide dose rates up to 2400 monitor units per minute.
![Page 8: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/8.jpg)
Possible Benefits of a FFF
• Efficiency • More accurate beam modeling due to decreased
head scatter • Decreased leakage and dose outside field
The study showed that removing the filter increased the dose rate on the central axis by a factor of 2.31 (6 MV) and 5.45 (18 MV) at a given target current. Because the flattening filter is a major source of head scatter photons, its removal from the beam line could reduce the out-of-field dose.
Vassiliev ON, Titt U, Kry SF, Pönisch F, Gillin MT, Mohan R Med Physics 2006 vol. 33 (4) pp. 820-7
![Page 9: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/9.jpg)
Unflattened Beam Has a Lower Risk of Secondary Tumors
Cashmore J, Int J Radiat Oncol Biol Phys. 2011
![Page 10: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/10.jpg)
Beam Profile – Flattening Filter
![Page 11: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/11.jpg)
Depth Dose with Flattened Beam
2004 Phys. Med. Biol. 49 1535
![Page 12: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/12.jpg)
Calibration
• No difference for FFF • Polarity and recombination corrections small
Energy Dose rate Ppol Pion
6X 600 1.000 1.004
15X 600 1.000 1.005
6X FFF 1400 1.000 1.006
10X FFF 2400 1.000 1.013
![Page 13: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/13.jpg)
Calibration
Kry SF, Popple R, Molineu A, Followill DS. Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams. J Appl Clin Med Phys. 2012 Nov 8;13(6):3803. doi: 10.1120/jacmp.v13i6.3803. PubMed PMID: 23149774.
![Page 14: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/14.jpg)
Calibration
Kry SF, Popple R, Molineu A, Followill DS. Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams. J Appl Clin Med Phys. 2012 Nov 8;13(6):3803. doi: 10.1120/jacmp.v13i6.3803. PubMed PMID: 23149774.
![Page 15: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/15.jpg)
Calibration – use Pb foil for all FFF beams
![Page 16: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/16.jpg)
Calibration sanity check
-2
-1
0
1
2
3
4
5
6X FFF 400 M
U/m
in
6X FFF 1400 M
U/m
in
6X 600 M
U/m
in
10X FFF 400 M
U/m
in
10X FFF 2400 M
U/m
in
15X 600 M
U/m
in
Diff
eren
ce fr
om 6
X (%
)
Exradin A12EDR2MOSFET
![Page 17: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/17.jpg)
IROC Calibration check
• RPC OSLD measurement / Institution • 6X FFF = 0.99 • 10X FFF = 0.99
![Page 18: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/18.jpg)
Profile and depth dose correction for recombination
Measured x-ray distributions of 6FFF and 15FFF beams plotted against charged collected per beam pulse. Also plotted are the results from applying corrections for the ion chamber collection efficiency. The distributions were measured at 100 cm SSD with the 0.1 cm3 chamber at Dmax and biased with 300 V.
S Johnsen “Ion Chamber Collection Efficiency Considerations for Un-Flattened X-Ray Beams,” Med. Phys. 35, 2770 (2008)
![Page 19: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/19.jpg)
Profile and depth dose correction for recombination
Measured 15 FFF x-ray depth-dose distribution and distribution resulting from corrections for ion chamber collection efficiency. Each curve is normalized to 100% at Dmax. Data is for a 0.1 cm3 chamber biased to 300 V; 10x10 cm2 field at 100 cm SSD.
S Johnsen “Ion Chamber Collection Efficiency Considerations for Un-Flattened X-Ray Beams,” Med. Phys. 35, 2770 (2008)
![Page 20: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/20.jpg)
FFF head scatter
Cashmore Phys Med Biol 2008
![Page 21: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/21.jpg)
FFF surface dose
Vassiliev et al. Phys Med Biol 2006
![Page 22: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/22.jpg)
Photon commissioning
• AAA Data requirements are the same • No additional data required for FFF
![Page 23: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/23.jpg)
10X FFF profile
![Page 24: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/24.jpg)
IROC VMAT Spine
![Page 25: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/25.jpg)
IROC Gated VMAT Lung
![Page 26: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/26.jpg)
IMRT QA results
-4 -3 -2 -1 0 1 2 3 40
10
20
30
40
50
60
70
Num
ber o
f cas
es
Difference (%)
Mean=0.7%s.d.=1.1%
![Page 27: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/27.jpg)
IMRT QA results
0 5 10 15 200
20
40
60
80
Num
ber o
f cas
es
Fraction with γ > 1 for 3%/3mm (%)
![Page 28: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/28.jpg)
Daily QA device
0 50 100 150 2000
1
2
3
4
5
6
7
8 x 107
MU
Bea
m C
heck
er P
lus
sign
al
← Saturation!
10F 2400 MupMin10F 400 MupMin
![Page 29: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/29.jpg)
Varian engineer fixing UAB TrueBeam
![Page 30: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/30.jpg)
Machine performance check
![Page 31: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/31.jpg)
Machine performance check
![Page 32: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/32.jpg)
AAPM TG-142
![Page 33: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/33.jpg)
![Page 34: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/34.jpg)
Orthogonal kV images
![Page 35: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/35.jpg)
Orthogonal kV images
![Page 36: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/36.jpg)
Cone beam CT
![Page 37: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/37.jpg)
Cone beam CT
![Page 38: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/38.jpg)
MV image
![Page 39: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/39.jpg)
MV image
![Page 40: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/40.jpg)
MV image
![Page 41: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/41.jpg)
![Page 42: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/42.jpg)
Gantry sag
![Page 43: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/43.jpg)
Difference between automatic and manual
![Page 44: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/44.jpg)
Automated QA Consortium
Richard Popple
Hania Al-Hallaq & Karl Farrey
Todd Pawlicki & Grace Kim
Jeremy Booth & Mario Perez
Vijeshwar Sharma & Sung Park
John DeMarco
Toby Eckhause, Don Roberts, Brett Schultz
& Jean Moran
Tim Ritter
TBD via Dan Low Courtesy Jean Moran
![Page 45: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/45.jpg)
Phase 1 QA Test Suite
Standardize tests and QA test suite so to evaluate the performance of TrueBeam linacs and compare our results across multiple institutions
Use both trajectory log files and EPID measurements
(a) (b)
Courtesy Jean Moran
![Page 46: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/46.jpg)
Example Fields of QA Test Suite
+ picket fence for VMAT with variable gantry speed, variable gantry speed and dose rate Both HDMLC and Millennium MLC supported
Courtesy Jean Moran
![Page 47: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/47.jpg)
Automated QA Analysis
Software analysis developed in MatLab primarily developed by Toby Eckhause (former post-doc)
Test suite takes ~15 minutes to deliver; 1 flood field + enter room once to setup phantom
All analysis from the original test suite is automated and results can be exported
Less than a minute to analyze all data
Courtesy Jean Moran
![Page 48: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/48.jpg)
Trajectory Log File & EPID Analysis
Leaf Position Deviation A (µm)
0 20 40
-10
-5
0
5
10 -6
-4
-2
0
2
4
6
-10 -5 0 5 101.4
1.5
1.6
1.7
1.8
1.9
2
Leaf A Position (cm)
Leaf
Dev
iatio
n S
tdev
( µm)
-10 -5 0 5 10-8
-6
-4
-2
0
2
4
Leaf A Position (cm)
Leaf
Pos
ition
Dev
iatio
n( µm
)
meanminmax
Test suite included all MLC monthly tests in Table V of TG 142, VMAT tests, plus imaging tests.
Analysis of Trajectory Log Files
Maximum difference for leaf position accuracy From Phase 1 of our project with 8 linacs measured over 6 months
Courtesy Jean Moran
![Page 49: Acceptance testing, commissioning and routine QA of the ... · Acceptance testing, commissioning and routine QA of the Varian TrueBeam . Richard Popple . ... • More accurate beam](https://reader036.vdocuments.mx/reader036/viewer/2022081600/604e29d613d50448d101f3c0/html5/thumbnails/49.jpg)