900-LC

P. van Poppel – Field Support Specialist October 2009

Page 2: 900-LC P. van Poppel – Field Support Specialist October 2009

The 900 series LC

Analytical 920 LC Semi-Prep 940 LC

Page 3: 900-LC P. van Poppel – Field Support Specialist October 2009

What will be covered

•Modules/ Options

•Parts Replacement Strategy

•Installation

•IMS diagnostics/ Troubleshooting

Page 4: 900-LC P. van Poppel – Field Support Specialist October 2009

Modules

•Autosampler

•Fluorescence Detector

•Quaternary Pump

•Column Heater/ Manual Injector

•Fraction Collector (external)

•Scale Up module

•MIB

---------------------------------------------------------------------------------------------------

•UV-vis Detector

•PDA Detector

•RI Detector (internal and external)

•ELSD (external)

•Isocratic Pump

•Sample Loading Pump

Page 5: 900-LC P. van Poppel – Field Support Specialist October 2009

Modules

•Autosampler

•Fluorescence Detector

•Quaternary Pump

•Column Heater/ Manual Injector

•Scale Up module

•MIB

•And the IMS

Page 6: 900-LC P. van Poppel – Field Support Specialist October 2009

Parts Replacement Strategy

•Ease of serviceability and a minimal number of parts will help field service representatives repair the instrument in an efficient and cost effective manner.

•We expect to service all modules to an assembly level, not to a sub assembly level. Component level repair of the circuit boards is not recommended.

•Warranty replacement of DOA modules within the system must be handled in accordance with the Instrument Escalation Procedure.Melbourne Field Support will not authorize the replacement of defective instruments unless there is an escalation.

Page 7: 900-LC P. van Poppel – Field Support Specialist October 2009

Installation

•Hardware

•Network

•Software

Page 8: 900-LC P. van Poppel – Field Support Specialist October 2009

Hardware Installation

•Plug & Play

•Pre-Plumbed

•Pre-Wired

Page 9: 900-LC P. van Poppel – Field Support Specialist October 2009

Hardware Installation

Autosampler Detector

Sample vial tray screw

Syringe movement lock down bracket

Page 10: 900-LC P. van Poppel – Field Support Specialist October 2009

Hardware Installation

Shipping Screw

Shipping Screw

Shipping Screw

Fluorescence Detector(bottom view)

Page 11: 900-LC P. van Poppel – Field Support Specialist October 2009

Hardware Installation

Page 12: 900-LC P. van Poppel – Field Support Specialist October 2009

Hardware Installation

Page 13: 900-LC P. van Poppel – Field Support Specialist October 2009

Network Setup

Page 14: 900-LC P. van Poppel – Field Support Specialist October 2009

Network Setup

Page 15: 900-LC P. van Poppel – Field Support Specialist October 2009

Network Setup

Page 16: 900-LC P. van Poppel – Field Support Specialist October 2009

Network Setup

Page 17: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 18: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 19: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 20: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 21: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 22: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 23: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 24: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 25: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 26: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 27: 900-LC P. van Poppel – Field Support Specialist October 2009

Software Installation

Page 28: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS

Page 29: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Actions

Page 30: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Counters

Page 31: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Service Tests

Page 32: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Additional Tests

Page 33: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Log Book

Page 34: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Log Book

Page 35: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Help

Page 36: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Help

Page 37: 900-LC P. van Poppel – Field Support Specialist October 2009

Modules

•Autosampler

•Fluorescence Detector

•Quaternary Pump

•Column Heater/ Manual Injector

•Scale Up module

•MIB

Page 38: 900-LC P. van Poppel – Field Support Specialist October 2009

Auto Sampler

•Poor Reliability

• Injection Valve System• Injection Port System• LC System

•Contamination

• Injection Valve System• Injection Port System• Sample• LC System

•Electronics

Page 39: 900-LC P. van Poppel – Field Support Specialist October 2009

Autosampler Electronics

•Electronics

• Not Initialising

• Error X, Y, Z

• S-Valve

• I-Valve

• System Busy/ Communications Error

Page 40: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Autosampler Tests

•Autosampler Complete Test

•Autosampler Temperature Accuracy Test

Page 41: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Autosampler Actions

•Injection port seal replaced

•Injection valve rotor replaced

•Syringe replaced

•Purge syringe

•Syringe valve rotor replaced

•Injection valve stator replaced

•Syringe valve stator replaced

•Needle replaced

•Wash needle

•Move to injection port

•Syringe needle adjustment

Page 42: 900-LC P. van Poppel – Field Support Specialist October 2009

Fluorescence Detector

General Issues

• Noise

• Drift

• Poor Signal to Noise

• Leak Errors

• Power Problems

Electronics

• Lamp Errors

• Board Errors

• Data Errors

Page 43: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Fluorescence Detector Actions

•150W xenon lamp replaced

•Long life xenon lamp replaced

•Mercury lamp replaced

•Flowcell cleaned/replaced

•Wavelength Accuracy

Page 44: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Fluorescence Detector Tests

•Signal to Noise by Raman Light Scattering

•Detector Noise Test

Page 45: 900-LC P. van Poppel – Field Support Specialist October 2009

Quaternary Pump Problems

General Issues

• Reproducibility

• Carryover

• No Peak Observed

• Mixing Ratio Accuracy

• Pressure Fluctuations

• Split Peaks

• Liquid does not Flow

• No Power

• Pump Motor does not Rotate

Page 46: 900-LC P. van Poppel – Field Support Specialist October 2009

Quaternary Pump Problems

Degasser Issues

• Bubbles Generated at Outlet

• Bubbles Generated when Pump sucks

• Liquid does not flow

• Vacuum Pump does not work

• Vacuum Pump does not stop

Page 47: 900-LC P. van Poppel – Field Support Specialist October 2009

Quaternary Pump Electronics

Electronics

• Motor Error

• Leaks

• Data & Board Errors

• Limit Errors

Page 48: 900-LC P. van Poppel – Field Support Specialist October 2009

Quaternary Pump Electronics

Error generating condition

Initialization

Pump operation

Error generating condition

YES

YES

NO

NO

Power On

ROM ERROR RAM ERROR EPROM ERROR MOTOR INITIALIZATION ERROR ELINE ERROR

PRESSURE MAX LIMIT ERROR PRESSURE MIN LIMIT ERROR EXTERNAL STOP

Page 49: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Quaternary Pump Actions

•Inlet check valve replaced

•Outlet check valve replaced

•Piston seal replaced

•Drain valve seal replaced

•Drain valve filter replaced

•Solenoid valve replaced

•Mixer filter replaced

•Solvent filter replaced

•Gradient four way joint replaced

•(Cam Position Adjustment)

Page 50: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Quaternary Pump Tests

•Flow rate accuracy - Test meter method

•Flow rate accuracy test - Time volume method

•Flow Compositional accuracy test

•Flow Leak test

•Flow precision test

•Flow pulsation test

•Flow compositional Ripple Test

•Flow gradient linearity test

•Flow compositional precision test

Page 51: 900-LC P. van Poppel – Field Support Specialist October 2009

Column Heater/ Manual Injector Problems

General Issues

• Heating Problems

• Leaks

• Pressure

Page 52: 900-LC P. van Poppel – Field Support Specialist October 2009

Scale Up Module Problems

General Issues

• Leaks

• Pressure

Page 53: 900-LC P. van Poppel – Field Support Specialist October 2009

MIB Problems

General Issues

• Communications

• Power

Page 54: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Column Heater/ Manual Injector Actions & Tests

•Injector valve stator replaced

•Injector valve rotor replaced

•Column valve stator replaced

•Column valve rotor replaced

•Column heater temperature accuracy and stability

Page 55: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS Scale Up Module Actions & Tests

•Injector valve stator replaced (Manual)

•Injector valve rotor replaced (Manual)

•Injector valve stator replaced (SUM)

•Injector valve rotor replaced (SUM)

•Column valve stator replaced (SUM)

•Column valve rotor replaced (SUM)

Page 56: 900-LC P. van Poppel – Field Support Specialist October 2009

IMS MIB Actions & Tests

•Backup the system configuration to a local folder

•Restore the system configuration from a previous backup

•Full System Test

Page 57: 900-LC P. van Poppel – Field Support Specialist October 2009

Autosampler Practical

Person 1:

• X, Y, Z movement adjustment

Person 2:

• Vial Detection & Injection Valve Sensor adjustment

Person 3:

• X & Y movement adjustment Mechanism Rails

…Injection Port Position, Syringe Valve and Syringe Sensor adjustments…

Page 58: 900-LC P. van Poppel – Field Support Specialist October 2009

Fluorescence Detector Practical

Person 1:

• Replacement of the Xenon Lamp and

• Positional Adjustment of the Xenon Lamp

Person 2:

• Replacement of the Mercury Lamp

• Perform a linearity test

Person 3:

• Clean the Flow Cell

• Perform a wavelength accuracy test

Page 59: 900-LC P. van Poppel – Field Support Specialist October 2009

Pump Practical

Person 1:

• Replacement of the Pump Seal & Drain Valve Seal and Solvent Filter Wash

Person 2:

• Replacement & Cleaning the Inlet and Outlet check valves and Pressure Test Mode in IMS

Person 3:

• Remove the pump and Pump Board