700 ftir data sheet - gas analyzers · the 700 ftir can serve a variety of analytical applications,...

2
700 FTIR Fourier Transform Infrared Analyzer Features Applications Examples of Gases Analyzed •Proven, rugged interferometer with gold mirrors •No liquid nitrogen required •0.8 wave number (cm –1 ) resolution •Heated sample cell (50°or 191°C) •Rack-mount PC or laptop-controlled with OPUS software High sensitivity with 10.2-meter optical path •Low acquisition and operating cost • Options include: - Pressure compensation - Analog output module - Intelligent multi-point sampler - Sample accessories •Process control •Stack gases (CEM/MACT) •VOC abatement/ scrubber efficiency •Vehicle emissions •Ammonia slip •Gas purity Agricultural emissions •Greenhouse gas Biomass/landfill gas •Carbon Dioxide •Carbon Monoxide •Nitric Oxide •Nitrogen Dioxide •Sulfur Dioxide •Nitrous Oxide •Hydrogen Chloride •Ammonia •Methane •Propane •Butane •Ethane •Ethanol •Ethylene •Propylene •Toluene •Acetylene •Chloroform •Dichloroethylene •Ethyl Benzene •Methyl Ethyl Ketone •Formaldehyde •Sulfur Hexafluoride •Phosgene •Vinyl Chloride •R134A •Others upon request The 700 Series FTIR Analyzer delivers fast, accurate analysis of virtually any gas that has an infrared absorption spectrum.

Upload: others

Post on 18-Jun-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 700 FTIR Data Sheet - Gas Analyzers · The 700 FTIR can serve a variety of analytical applications, including diesel emissions, CEM monitoring, ammonia slip, SCR inlet/outlet moni-toring,

700 FTIRFourier Transform Infrared Analyzer

Features

Applications

Examples of Gases Analyzed

•Proven,rugged interferometer withgoldmirrors

•Noliquidnitrogen required

•0.8wavenumber (cm–1)resolution•Heatedsamplecell (50°or191°C)

•Rack-mountPCor laptop-controlledwithOPUSsoftware

•Highsensitivitywith 10.2-meteropticalpath•Lowacquisitionandoperatingcost

•Optionsinclude:-Pressurecompensation-Analogoutputmodule-Intelligentmulti-pointsampler-Sampleaccessories

•Processcontrol

•Stackgases(CEM/MACT)

•VOCabatement/ scrubberefficiency

•Vehicleemissions

•Ammoniaslip

•Gaspurity

•Agriculturalemissions

•Greenhousegas

•Biomass/landfillgas

•CarbonDioxide•CarbonMonoxide•NitricOxide•NitrogenDioxide•SulfurDioxide•NitrousOxide•HydrogenChloride•Ammonia•Methane

•Propane•Butane•Ethane•Ethanol•Ethylene•Propylene•Toluene•Acetylene•Chloroform

•Dichloroethylene•EthylBenzene•MethylEthylKetone•Formaldehyde•SulfurHexafluoride•Phosgene•VinylChloride•R134A•Othersuponrequest

The 700 Series FTIR Analyzer delivers fast, accurate analysis of virtually any gas that has an infrared absorption spectrum.

Page 2: 700 FTIR Data Sheet - Gas Analyzers · The 700 FTIR can serve a variety of analytical applications, including diesel emissions, CEM monitoring, ammonia slip, SCR inlet/outlet moni-toring,

California Analytical Instruments1312 West Grove Ave., Orange, CA 92865

Phone: 714-974-5560 • Fax: 714-921-2531www.gasanalyzers.com

Description

Specifications

Method of Operation

TheCaliforniaAnalyticalInstruments700FTIRAnalyzerprovidesfast,continuousandstableanalysisofvirtuallyanygasthathasaninfraredabsorptionspectrum.Theproprietaryheatedsam-plecellenablestheinstrumenttoaccommodatehotsamplescontaininghighlevelsofmoisture.

The700FTIRcanserveavarietyofanalyticalapplications,includingdieselemissions,CEM

monitoring,ammoniaslip,SCRinlet/outletmoni-toring,processmonitoringandothers.

UnlikeotherFTIRanalyzers,the700FTIRdoesnotrequireliquidnitrogen,eliminatingtheneedtoconstantlyfillLN2dewarsandtheassociatedsafetyissues.Itssmallfootprintandlightweightalloweasyinstallationandtransportabilitywhenrequired.

The700FTIRAnalyzerisbasedonFourierTransformInfraredSpectroscopy.Nonsymmetricalgas-phasemoleculesabsorbIRlight,causingthemolecularbondstostretch,bendorrotate.Thisabsorptionisusedtomeasureandquantifyseveralchemicalcomponentssimultaneously.

AnIRsourceemitsradiationintherangeof7500to375cm–1.TheIRradiationissplitinaninterferometer,wherethelightissplittowardtwomovingcorner-cubemirrors.Thetwobeamsrecombineandpass

througha10.2-metermulti-reflectiongascellwherethesampleabsorbslightatmolecule-specificfrequencies.TheremaininglightismeasuredwithaDTGSdetectorandFouriertransformedtoconvertfromthetimedomaintothefrequencydomain.Thisproducesasingle-beamspectrumthatisratioedwithabaselinespectrum,produc-inganabsorbancespectrum.TheabsorbancespectrumisquantifiedwithPLSchemometricstoproduceaconcentrationvalue.

Analysis Method – Fourier Transform Infrared (FTIR) Components – Multiple gases Interferometer – Rocksolid,™ permanent alignment, high stability with cube corner reflectors and non-wear bearing for long life Detector Type – DTGS Ranges – From ppb to percent Response Time – From approximately 20 seconds to 5 minutes, depending upon sensitivity (typically 1 minute)Spectral Resolution – 0.8 cm–1 to 128 cm–1 Spectral Range – 305-7500 cm–1

Control – PC, Windows XP or higher Sample Flow – Typically 0.2 to 5 lpm Ambient Temperature – 5° to 40°C Ambient Humidity – Less than 90% RH (non-condensing)

Power Requirements – 115 VAC/60Hz or 230 VAC/50Hz Dimensions – 10.5”H x 19”W x 28”D Weight – Approximately 90 lbs.

Gas CellConstruction – 316 Stainless Steel (50°C or 191°C) Volume – 550 cc Effective Pathlength – 10.2 meters Mirrors – Gold-plated SS 316Windows – ZnSe standard, others available O-rings – Parafluor Inlet/Outlet Connections – 3/8” tubing Purge Fittings – ¼-inch Swagelok® compression

Specifications subject to change without notice.

700 FTIR Fourier Transform Infrared Analyzer

OPUS Software Package

The700FTIROPUSsoftwareisfullyautomatedforexceptionaleaseofuse.Itoffersfeaturessuchasspectrumcalculator,absorbance-to-transmission

conversion,automaticbaselinecorrectionandpeakpicking.Itallowsmultiplespectratobemanipulatedatthesametime.