Compact reasonably priced device intended for gamma and X-radiation radiation ambient dose equivalent and ambient dose equivalent rate measurement. ▪ Low weight and small size ▪ Field operation capability over a wide information about measurement date and and battery life indicator is displayed on quantum is indicated by a sound signal ▪ Bright white backlit LCD-screen ▪ Automatic compensation of intrinsic change of dose rate (measurement ▪ Up to 100 measurement results can be detector background ▪ In search mode each registered gamma ▪ Sound and visual alarm in case threshold level is exceeded for dose and dose rate ▪ Dose rate and dose thresholds can be selected in entire measurement range and saved when the radiation monitor is switched off ▪ Rapid reaction to statistically significant process restart) temperature range stored in non-volatile memory with time ▪ Measurement results, current time, date matrix LCD screen ▪ Radiation protective measures in case of nuclear disasters ▪ Civil protection ▪ Radioecology ▪ Customs service ▪ Fire-fighting service facilities, health care and other institutions ▪ Dosimetric monitoring in manufacturing Device operating principle is based on the process of count rate measurement of impulses, generated in Geiger-Muller counter tube under the influence of gamma radiation. Count rate is converted automatically into measurable physical values throughout the range. Energy compensating filter allows correcting energy dependence of sensitivity efficiently in entire energy range of gamma radiation. Microprocessor-based unit is responsible for controlling the radiation monitor operating modes, calculations, storing and displaying measurement results and for self-checking function. АТ6130C Radiation Monitor Applications Features Pocket Radiation Monitor Operating principle INSTRUMENTS AND TECHNOLOGIES FOR NUCLEAR MEASUREMENTS AND RADIATION MONITORING

Upload: others

Post on 23-Sep-2020

10 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: АТ6130C Radiation Monitor...GOST 27451-87, GOST 28271-89, GOST 17225-85, Safety requirements of IEC 61010-1:2010, EMC requirements of EN 55011:2009, IEC 61000-4-2:2008, IEC 61000-4-3:2008

Compact reasonably priced device intended for gamma and X-radiation

radiation ambient dose equivalent and ambient dose equivalent rate

measurement

Low weight and small size

Field operation capability over a wide

information about measurement date and

and battery life indicator is displayed on

quantum is indicated by a sound signal

Bright white backlit LCD-screen

Automatic compensation of intrinsic

change of dose rate (measurement

Up to 100 measurement results can be

detector background

In search mode each registered gamma

Sound and visual alarm in case threshold

level is exceeded for dose and dose rate

Dose rate and dose thresholds can be

selected in entire measurement range and

saved when the radiation monitor is

switched off

Rapid reaction to statistically significant

process restart)

temperature range

stored in non-volatile memory with

time

Measurement results current time date

matrix LCD screen

Radiation protective measures in case of

nuclear disasters

Civil protection

Radioecology

Customs service

Fire-fighting service

facilities health care and other institutions

Dosimetric monitoring in manufacturing

Device operating principle is based on the process of count rate

measurement of impulses generated in Geiger-Muller counter tube under the

influence of gamma radiation

Count rate is converted automatically into measurable physical values

throughout the range Energy compensating filter allows correcting energy

dependence of sensitivity efficiently in entire energy range of gamma radiation

Microprocessor-based unit is responsible for controlling the radiation monitor

operating modes calculations storing and displaying measurement results and

for self-checking function

АТ6130C Radiation Monitor

Applications

Features

Pocket

Radiation

Monitor

Operating principle

INSTRUMENTS AND TECHNOLOGIES FOR NUCLEAR

MEASUREMENTS AND RADIATION MONITORING

Standard sensitivity response upon gamma radiation incidence angle respect

to the calibration direction

K(E)K(0662)

08

09

10

11

12

13

14

006 0088 0662 12 2 3

E MeV

Normal energy response of monitor 137sensitivity respect to Cs

gamma radiation of 662 keV

Radiation direction

Radiation direction

Specification

Ambient gamma and X radiation dose 001 microSvh ndash 1 mSvh

Ambient gamma and X radiation dose 1 nSv ndash 100 mSv

Limit of intrinsic relative error plusmn20

Response time for dose rate le7s

(in the range from 01 microSvh to 1 mSvh)

Energy range 50 keV ndash 3 MeV

equivalent rate measurement range (or 10 microremh ndash 100 mremh)

in natural background conditions

equivalent indication range (or 01 microrem ndash 10 rem)

equivalent measurement range (or 10 microrem ndash 10 rem)

Ambient gamma and X radiation dose 01 microSvh ndash 1 mSvh

137relative to 662 keV ( Cs)

Ambient gamma and X radiation dose 01 microSv ndash 100 mSv

equivalent rate indication range (or 1 microremh ndash 100 mremh)

Continuous run time ge700 h

137 -1Typical sensitivity to Cs gamma radiation 28 cps(microSvh )

change from 1 to 10 microSvh (accuracy error leplusmn10)

Radiation overloading Radiation monitor can withstand

100-fold rise of dose rate

measurement upper range limit

for 5 minutes with readings not

lower than maximum

Working temperature range -20degC to +55degC

Energy dependence plusmn30

of dose rate measurement

оle35 С without condensation

Drop protection From le15 m to hard surface

Relative humidity with air temperature le95

Protection class IP40

Burn-up life ge100 Sv

or 2 x AA-size rechargeable

cells with nominal voltage 12 V

Overall dimensions weight 111x70x28 mm 02 kg

Power supply 2 x AA-size batteries (LR 6)

Design and specifications are subject to change without notice

Units of measure are chosen during ordering procedure and cannot be altered later

АТ6130C Radiation Monitor

5 Gikalo st Minsk 220005 Republic of BelarusTelFax +375-17-270-81-42E-mail infoatomtexcom

The radiation monitors comply with

GOST 27451-87 GOST 28271-89 GOST 17225-85

Safety requirements of IEC 61010-12010

EMC requirements of EN 550112009

IEC 61000-4-22008 IEC 61000-4-32008

Page 2: АТ6130C Radiation Monitor...GOST 27451-87, GOST 28271-89, GOST 17225-85, Safety requirements of IEC 61010-1:2010, EMC requirements of EN 55011:2009, IEC 61000-4-2:2008, IEC 61000-4-3:2008

Standard sensitivity response upon gamma radiation incidence angle respect

to the calibration direction

K(E)K(0662)

08

09

10

11

12

13

14

006 0088 0662 12 2 3

E MeV

Normal energy response of monitor 137sensitivity respect to Cs

gamma radiation of 662 keV

Radiation direction

Radiation direction

Specification

Ambient gamma and X radiation dose 001 microSvh ndash 1 mSvh

Ambient gamma and X radiation dose 1 nSv ndash 100 mSv

Limit of intrinsic relative error plusmn20

Response time for dose rate le7s

(in the range from 01 microSvh to 1 mSvh)

Energy range 50 keV ndash 3 MeV

equivalent rate measurement range (or 10 microremh ndash 100 mremh)

in natural background conditions

equivalent indication range (or 01 microrem ndash 10 rem)

equivalent measurement range (or 10 microrem ndash 10 rem)

Ambient gamma and X radiation dose 01 microSvh ndash 1 mSvh

137relative to 662 keV ( Cs)

Ambient gamma and X radiation dose 01 microSv ndash 100 mSv

equivalent rate indication range (or 1 microremh ndash 100 mremh)

Continuous run time ge700 h

137 -1Typical sensitivity to Cs gamma radiation 28 cps(microSvh )

change from 1 to 10 microSvh (accuracy error leplusmn10)

Radiation overloading Radiation monitor can withstand

100-fold rise of dose rate

measurement upper range limit

for 5 minutes with readings not

lower than maximum

Working temperature range -20degC to +55degC

Energy dependence plusmn30

of dose rate measurement

оle35 С without condensation

Drop protection From le15 m to hard surface

Relative humidity with air temperature le95

Protection class IP40

Burn-up life ge100 Sv

or 2 x AA-size rechargeable

cells with nominal voltage 12 V

Overall dimensions weight 111x70x28 mm 02 kg

Power supply 2 x AA-size batteries (LR 6)

Design and specifications are subject to change without notice

Units of measure are chosen during ordering procedure and cannot be altered later

АТ6130C Radiation Monitor

5 Gikalo st Minsk 220005 Republic of BelarusTelFax +375-17-270-81-42E-mail infoatomtexcom

The radiation monitors comply with

GOST 27451-87 GOST 28271-89 GOST 17225-85

Safety requirements of IEC 61010-12010

EMC requirements of EN 550112009

IEC 61000-4-22008 IEC 61000-4-32008