Routine Testing with Test Weights
![Page 2: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/2.jpg)
2
FDA – Questions and Answers on cGMP‘s
http://www.fda.gov/cder/guidance/cGMPs/
Supplier`s Question“Many leading analytical balance manufacturers provide built-in "auto calibration" features in their balances. Are such auto-calibration procedures acceptable instead of external performance checks? If not, then what should the schedule for calibration be?”
FDA AnswerExternal performance checks still have to be carried out, but less frequently. Risk analysis (criticality and tolerance of the process) and frequencyof use determine the frequency of performance checks. The calibration of an “auto-calibrator” should be periodically verified –usual frequency is once per year - using […] traceable standards.
![Page 3: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/3.jpg)
3
Our Experience put into Practice
Higher Risk
⇒More Testing
Impact (for business and for consumer)
Wei
ghin
gac
cura
cy
low medium high
1 %
0.1 %
0.01 %
10 %
![Page 4: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/4.jpg)
4
Which tests should I carry out to minimize my risk ?
Risik Management
Let us look at the parameters that reallyinfluence your weighing results. Werecommend that you test only therelevant parameters.
![Page 5: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/5.jpg)
5
Repeatability (RP)
Sensitivity Offset (SE)
Linearity Deviation „Nonlinearity“ (NL)
Eccentric Load Deviation (EC)
Readability
Parameters influencing Weighing Accuracy
![Page 6: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/6.jpg)
6
Rel
ativ
e E
xpan
ded
Wei
ghin
gU
ncer
tain
ty[%
]Influence of Weighing Parameters on Accuracy
0.00001
0.0001
0.001
0.01
0.1
1
0.01 0.1 1 10 100 1000
U_totU_RPU_ECU_NLU_SE
Model XP204: Individual Uncertainty Contributions (@k=2)
Sample Mass [g]
![Page 7: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/7.jpg)
7
Repeatability: Dominant influencing factor near the lowend of the measurement range
Sensitivity and eccentricity:Dominant influencing factors near the upperend of the measurement range
Nonlinearity:Not dominant throughout the wholemeasurement range
Influence of Weighing Parameters on Accuracy
![Page 8: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/8.jpg)
8
All User Tests can be done with 2 Weights
Weight 1: Largest OIML/ASTM weight which isbelow the balance capacity
Weight 2: Largest OIML/ASTM weight whichequals or is below 5% of the capacity
XP205: XP4002S:2000 g and 200 g200 g and 10 g
![Page 9: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/9.jpg)
9
Sensitivity [SE]
Indication
Load
• Ideal characteristic curve
• Characteristic curve withsensitivity deviation
Example: XP205
The calibration of sensitivity is most efficient if done near the upperend of the measurement range
Weight 1:200g F2
![Page 10: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/10.jpg)
10
Sensitivity [SE]For even smaller test weights, sensitivity calibration is useless
- The absolute sensitivity deviation is very small and very difficult if impossible to detect- Near the lower end of the measurement range, the effect of repeatability is dominant
See e.g. USP <41>: For small loads, the determination of measurementuncertainty is executed by means of a repeatability test
Indication
Load
• Ideal characteristic curve
• Characteristic curve withsensitivity deviation
![Page 11: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/11.jpg)
11
Sensitivity TestsTest sensitivity with a test weight close to nominal capacity (1).
Using a smaller test weight (a) results in a smaller offset, which is partially disturbed by repeatability (pink band).
Using an even smaller test weight (b) results in the sensitivity offset being buried entirely in the deviations of repeatability.
Load
• Ideal characteristic curve
• Characteristic curve withsensitivity deviation
1ab
Best
Feasible
Wrong
![Page 12: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/12.jpg)
12
Repeatability Test at 5% of weighing range
Indication
Load
Standard deviation of the repeatability
Lower end (< 5%) of measurementrange: Repeatability almost constantWeight 2
Repeatability is almost constant at the lower end of the measurement range. That is why the weight 2 can be used to determine the minimum weight, which is of course much smaller than weight 2.
![Page 13: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/13.jpg)
13
Repeatability is constant at the lower end
![Page 14: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/14.jpg)
14
Nonlinearity [NL]
Nonlinearity normally doesnot play a dominant role foranalytical and low-loadprecision balances and isnegligible for small netloads, so that the usernormally does not have to test nonlinearity.
Nonlinearity is assessed in the framework of theservice by an accreditedservice technician.
Indication
Load
![Page 15: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/15.jpg)
15
Eccentricity [EC]
Eccentricity errors can be minimized by use of specialaccessories as e.g. ErgoClips
![Page 16: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/16.jpg)
16
Frequency of User Tests
Higher Risk
⇒More Testing
Impact (for business and for consumer)
Wei
ghin
gac
cura
cy
low medium high
1 %
0.1 %
0.01 %
10 %
![Page 17: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/17.jpg)
17
Test weights for User Tests
Weight 2 Weight 1Weight 1 Weight 2
( )If required
![Page 18: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/18.jpg)
18
CarePacs for Routine Testing
User BenefitsCost savings – only two weights rather than entire weight set
Time savings – easy and quick tests supported by METTLER TOLEDO SOP’s
Option to add 3rd weight for specific testing
Ergonomically designed weight handling accessories included
![Page 19: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/19.jpg)
19
Frequency of User Tests: Sensitivity
Sensitivity calibration is an integral part of the functional control of a measurement device
Sensitivity tests are very simple and fast
For a balance with activated FACT, sensitivity calibrations withexternal weights can be carried out less frequently
![Page 20: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/20.jpg)
20
Frequency of User Tests: Repeatability
The repeatability is important for the determination of the minimumweight
In many cases, economic and safety reasons play an important rolefor the determination of the minimum weight (weighing of precioussubstances; flavor of food, etc.)
Some regulations as e.g. USP explicitely require the determinationof minimum weight
![Page 21: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/21.jpg)
21
Frequency of User Tests: Eccentricity
Eccentricity may play a role if the tare vessel is removed for filling;The tare may be placed on a different location of the weighing pan
Eccentricity may also be important when larger amounts are weighed in, with the tare vessel being placed non-centric on the pan
The use of accessories such as ergo clips minimizes eccentricity errors, in that case it is enough if eccentricity tests are carried out by the MT service technician.
![Page 22: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/22.jpg)
22
Frequency of User Tests: Nonlinearity
The influence of nonlinearity on the weighing result is often negligible
The determination of nonlinearity is time consuming, since not onlyone or two weights but a whole set of weights have to be used
For almost all of the cases, it is justified to leave the determination of nonlinearity to the service techician
For higher risk, the calibration performed the service technician shallbe carried out more frequently; this calibration includes nonlinearity
![Page 23: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/23.jpg)
23
Example of a GWP® Recommendation
![Page 24: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/24.jpg)
24
Test Tolerances
Control limit exceeded: Balance has to be markedas „out of control limits“ and shall not be usedanymore. MT has be contacted.
Warning limit exceeded: No immediate actionnecessary, but balance has to be kept undersurveillance
GWP SOP‘s for Routine Testingprovide more information for the user
![Page 25: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/25.jpg)
25
SOP‘s for Routine Testing
GWP SOP‘s for Routine Testingprovide more information for the user
![Page 26: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/26.jpg)
26
Example of a GWP® Recommendation
The weight classes are chosen so that theuncertainty of the weights can be neglectedcompared to the process accuracyrequirements.
![Page 27: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/27.jpg)
27
Straightforward routine testing with GWP®
![Page 28: 5b Routine Tests with Test Weights - IFFiff.nu/_files/per/gwp2009riskbasedroutinetesting2.pdf · “Many leading analytical balance manufacturers provide built-in ... METTLER TOLEDO](https://reader034.vdocuments.mx/reader034/viewer/2022051722/5aa6579c7f8b9a7c1a8e9090/html5/thumbnails/28.jpg)
Routine Testing with Test Weights