OJIYAS M3.5P0.6 THREAD PLUG GAUGE
Procedure: Order Number:
ID Number: Serial Number :
Asset Number: Calibration Date:
Temperature: Relative Humidity:
Thread Plug Accuracy
Function NominalLower
LimitAs Found As Left
Upper
LimitResult Uncertainty
PD (Go Plug) Top 3.1100 mm 3.1100 mm 3.1110 mm 3.1110 mm 3.1151 mm PASS 0.000601
PD (Go Plug) Middle 3.1100 mm 3.1100 mm 3.1110 mm 3.1110 mm 3.1151 mm PASS 0.000601
PD (Go Plug) Bottom 3.1100 mm 3.1100 mm 3.1110 mm 3.1110 mm 3.1151 mm PASS 0.000601PD (No Go Plug) Top 3.2220 mm 3.2169 mm 3.2202 mm 3.2202 mm 3.2220 mm PASS 0.000601
PD (No Go Plug) Middle 3.2220 mm 3.2169 mm 3.2202 mm 3.2202 mm 3.2220 mm PASS 0.000601
PD (No Go Plug) Bottom 3.2220 mm 3.2169 mm 3.2202 mm 3.2202 mm 3.2220 mm PASS 0.000601
Function Uncertainty
Major Diameter (GO Plug) 0.000601
Major Diameter (NO GO Plug) 0.000601
February 28, 2013
49.0%
33K6-4-203-1
43221004
005857
20.8°C
PH-WO-1082173808
43221004
As Found As Left
3.5050 mm
3.3639 mm
3.5050 mm
3.3639 mm
Page 1 of 12
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 2/12
Outside Diameter (GO)
Nominal Test 1 Test 2 Test 3 Test 4 Test 5 Avg.
#REF! 3.5050 3.5050 3.5050 3.5050 3.5050 3.505
Outside Diameter (NO GO)
Nominal Test 1 Test 2 Test 3 Test 4 Test 5 Avg.
#REF! 3.3639 3.3639 3.3639 3.3639 3.3639 3.3639
Pitch Diameter (GO)
Nominal Test 1 Test 2 Test 3 Test 4 Test 5 Avg.
3.1100 3.1110 3.1110 3.1110 3.1110 3.1110 3.110992
3.1100 3.1110 3.1110 3.1110 3.1110 3.1110 3.11099
3.1100 3.1110 3.1110 3.1110 3.1110 3.1110 3.11099
Pitch Diameter (NO GO)
Nominal Test 1 Test 2 Test 3 Test 4 Test 5 Avg.
3.2220 3.2202 3.2202 3.2202 3.2202 3.2202 3.22021
3.2220 3.2202 3.2202 3.2202 3.2202 3.2202 3.22021
3.2220 3.2202 3.2202 3.2202 3.2202 3.2202 3.22021
NOTE THE WIRE USED:::: 0.02162
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 3/12
mm
Go
Component ValueDistribution
TypeDivisor
Divisor
Value
Adjusted
Component
Squared
Value
TYPE A
Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000
Long Term Stability 0.0 Normal 1.0 1.000 0.0000 0.00
TYPE B
Uncertainty as reported 0.0001016 Rectangular 2.0 2.000 0.000051 0.000
Uncertainty as reported 0.0000737 Rectangular 2.0 2.000 0.000037 0.000
Accuracy of Standard 0.000508 Rectangular 1.732 1.732 0.0003 0.000000
Resolution 0.000025 Rectangular 1.732 1.732 0.0000 0.000000
Sum 0.0000
U = Sqrt(Sum) = 0.000
Expanded U = (for k=2) 0.001
NO GO
Component ValueDistribution
TypeDivisor
Divisor
Value
Adjusted
Component
Squared
Value
TYPE A
Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000
Long Term Stability 0.0 Normal 1.0 1.000 0.0000 0.00
TYPE B
Uncertainty as reported 0.0001016 Rectangular 2.0 2.000 0.000051 0.000
Uncertainty as reported 0.0000737 Rectangular 2.0 2.000 0.000037 0.000
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 4/12
Accuracy of Standard 0.000508 Rectangular 1.732 1.732 0.0003 0.000000
Resolution 0.000025 Rectangular 1.732 1.732 0.0000 0.000000
Sum 0.0000
U = Sqrt(Sum) = 0.000
Expanded U = (for k=2) 0.001
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 5/12
unit test 1 test 2 test 3 test 4 test 5 test times
mm 3.50500 3.50500 3.50500 3.50500 3.50500 5
Notes
Need to generate at least 5 easurement for
one particular range.
Need long term stability of CDI system.
If not available, use 1 year
stability specifications.
Uncertainty reported in your Calibration Certificate of
the Thread Wire Gauge
Uncertainty reported in your Calibration Certificate of
the Super Micrometer
Accuracy of STD
Resolution of STD 1/4 read
Use 0.000601mm
unit test 1 test 2 test 3 test 4 test 5 test timesmm 3.36390 3.36390 3.36390 3.36390 3.36390 5
Notes
Need to generate at least 5 easurement for
one particular range.
Need long term stability of CDI system.
If not available, use 1 year
stability specifications.
Uncertainty reported in your Calibration Certificate of
the Thread Wire Gauge
Uncertainty reported in your Calibration Certificate of
the Super Micrometer
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 6/12
Accuracy of STD
Resolution of UUT
Use 0.000601mm
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 7/12
average variance var square-root
3.50500 0.0000000 0.0000000
average variance var square-root 3.36390 0.0000000 0.0000000
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 8/12
GO
Component ValueDistribution
TypeDivisor
Divisor
Value
Adjusted
Component
Squared
Value
TYPE A
Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000
Long Term
Stability0.0 Normal 1.0 1.000 0.0000 0.00
TYPE B
Uncertainty
as reported0.0001016 Rectangular 2.0 2.000 0.000051 0.000
Uncertainty
as reported0.0000737 Rectangular 2.0 2.000 0.000037 0.000
Accuracy of
Standard0.000508 Rectangular 1.732 1.732 0.0003 0.000000
Resolution 0.000025 Rectangular 1.732 1.732 0.0000 0.000000
Sum 0.0000
U = Sqrt(Sum) = 0.000
Expanded U = (for k=2) 0.001
NO GO
Component ValueDistribution
TypeDivisor
Divisor
Value
Adjusted
Component
Squared
Value
TYPE A
Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000
Long Term
Stability0.0 Normal 1.0 1.000 0.0000 0.00
TYPE B
Uncertainty
as reported0.0001016 Rectangular 2.0 2.000 0.000051 0.000
Uncertainty
as reported0.0000737 Rectangular 2.0 2.000 0.000037 0.000
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 9/12
Accuracy of
Standard0.000508 Rectangular 1.732 1.732 0.0003 0.000000
Resolution 0.000025 Rectangular 1.732 1.732 0.0000 0.000000
Sum 0.0000
U = Sqrt(Sum) = 0.000
Expanded U = (for k=2) 0.001
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 10/12
unit test 1 test 2 test 3 test 4 test 5
mm 3.11100 3.11099 3.11099 3.11099 3.11099
Notes
Need to generate at least 5 easurement for
one particular range.
Need long term stability of CDI system.
If not available, use 1 year
stability specifications.
Uncertainty reported in your Calibration
Certificate of the Thread Wire Gauge
Uncertainty reported in your Calibration
Certificate of the Super Micrometer
Accuracy of STD
Resolution of UUT
Use 0.000601mm
unit test 1 test 2 test 3 test 4 test 5
mm 3.22021 3.22021 3.22021 3.22021 3.22021
Notes
Need to generate at least 5 easurement for
one particular range.
Need long term stability of CDI system.
If not available, use 1 year
stability specifications.
Uncertainty reported in your Calibration
Certificate of the Thread Wire Gauge
Uncertainty reported in your Calibration
Certificate of the Super Micrometer
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 11/12
Accuracy of STD
Resolution of UUT
Use 0.000601mm
7/29/2019 43221004
http://slidepdf.com/reader/full/43221004 12/12
test times average variance ar square-root
5 3.11099 0.0000000 0.0000045
test times average variance ar square-root
5 3.22021 0.0000000 0.0000000