3d diode array commissioning: building confidence in 3d qa … · 2020. 10. 1. · arccheck...
TRANSCRIPT
![Page 1: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/1.jpg)
CONE HEALTHCancer Center
3D Diode Array Commissioning: Building Confidence in 3D QA Technology
Caroline Yount, MSCONE HEALTH CANCER CENTER
![Page 2: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/2.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
3D QA
The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose distributions and the methods for IMRT dose delivery demand that the dosimetry measurement techniques… be reviewed and adapted for the unique challenges posed by IMRT.
TG 120
![Page 3: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/3.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
QA Challenges
• Modern technologies are utilizing rotation delivery to speed up treatment time and decrease tissue toxicity (RapidArc®, VMAT, TomoTherapy®,
IMAT, SmartArc)
• Rotational beam delivery creates a challenge for patient specific QA 360 degrees of delivery
![Page 4: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/4.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
3D QA
![Page 5: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/5.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
3D QA
![Page 6: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/6.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
3D QA
![Page 7: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/7.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
3D QA
![Page 8: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/8.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
ArcCHECK Introduction
• Designed for Helical & Arc Delivery
• 1386 diodes in a helical geometry21cm diameter, 21cm length1cm spacing, 3.28cm equivalent depth0.64mm2 active detector area
• 4th Dimension = Time50ms update frequency
• Optional cavity plug insert with ion chamber holder PMMA construction
![Page 9: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/9.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Detector Geometry
• Entrance and exit dose are measured Effectively doubling the detector density in the measurement field.
• Central 10x10 contains approximately 230 detectors • Detectors are arranged on a HeliGrid™
Increases sampling rate and reduces detector overlap from Beams Eye View (BEV)
• Entrance and exit dose can be correlated to determine gantry angle
![Page 10: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/10.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Question
“How can I interpret and trust 3D QA?”
![Page 11: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/11.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Commissioning Measurements
1. Detector spacing2. Linearity3. Sensitivity4. Individual detector response variations5. Absolute dose accuracy
Then… Trend analysis
![Page 12: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/12.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Detector spacing
kVCT MVCT
![Page 13: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/13.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Linearity – Conventional Dose Rates
6MV500MU/min10x10100SAD
200200200300
400500
600700
800
1000
2000
R² = 1
050
010
0015
0020
0025
0030
00
0 500 1000 1500 2000 2500
Dos
e @
a c
entr
al d
iode
MU1.
001.
051.
101.
151.
201.
25
0 1000 2000 3000R
eadi
ng/M
UMU
Ratio
![Page 14: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/14.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
50200200200
500
1000
2000
4000
R² = 1
010
0020
0030
0040
0050
00
0 1000 2000 3000 4000 5000
Dos
e @
a c
entr
al d
iode
MU
Linearity – High Dose Rates
6MV FFF1400MU/min10x10100SAD
1.00
1.05
1.10
1.15
1.20
1.25
0 2000 4000 6000R
eadi
ng/M
UMU
Ratio
![Page 15: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/15.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Linearity – High Dose Rates
10MV FFF2400MU/min10x10100SAD
50200200200500
1000
2000
4000
R² = 1
010
0020
0030
0040
0050
0060
00
0 1000 2000 3000 4000 5000
Dos
e @
a c
entr
al d
iode
MU1.
001.
101.
201.
30
0 2000 4000 6000R
eadi
ng/M
UMU
Ratio
![Page 16: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/16.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Spatial Sensitivity
Table = 0mm Table shifted laterally = 0.5mm
![Page 17: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/17.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Spatial Sensitivity
0.5mm
1mm
2mm
3mm
86%88%90%92%94%96%98%
0 1 2 3 4Shift (mm)
Pass Rate vs Shift
Lateral Table Shifts
![Page 18: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/18.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Spatial Sensitivity
0.5mm
1mm
2mm
3mm
86%88%90%92%94%96%98%
0 1 2 3
Shift (mm)
Pass Rate vs Shift
Longitudinal Table Shifts
![Page 19: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/19.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Angular Sensitivity
0.5°
1°
2°
3°
Gantry Angle Changes
86%88%90%92%94%96%98%
0 1 2 3Change in Angle (degrees)
Pass Rate vs Angle Change
![Page 20: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/20.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Detector Response Variations
• Sensitivity differences between the ArcCHECK detectors
• These differences can be measured• Individual correction factors applied to subsequent raw measurements
![Page 21: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/21.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Detector Response Variations
![Page 22: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/22.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Detector Response Variations
![Page 23: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/23.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Calibration Validity Check
PAAP
![Page 24: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/24.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Calibration Validity Check
In‐house array calibration:
Factory array calibration:
![Page 25: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/25.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Calibration Validity Check
1.5%
New In‐house array calibration:
1%
![Page 26: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/26.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
ArcCHECK Calibration
• Array calibration on standard linac• Dose calibration on standard linac SAD hand calc at
3.28 cm depth (water equivalent depth)
![Page 27: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/27.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Dose Accuracy
Measured 10x10 Planned 10x10
![Page 28: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/28.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Clinical Example ‐ Chest
![Page 29: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/29.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Phantom Plan
![Page 30: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/30.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
QA Result
Measured Calculated
Comparison
Profiles
Ion Chamber = -2%
![Page 31: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/31.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Clinical Example – Head & Neck
![Page 32: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/32.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Phantom Plan
![Page 33: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/33.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
QA Result
Measured Calculated
Comparison
Profiles
Ion Chamber = 0%
![Page 34: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/34.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Clinical Example ‐ Prostate
![Page 35: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/35.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Phantom Plan
![Page 36: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/36.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
QA Result
Measured Calculated
Comparison
Profiles
Ion Chamber = -1.2%
![Page 37: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/37.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Trend Analysis
“When did 3D QA indicate clinical issues?”
![Page 38: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/38.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Trend Analysis
Patient‐specific Chamber Readings
Setup error, high gradients, modulation too high
‐13.5%‐12.0%‐10.5%‐9.0%‐7.5%‐6.0%‐4.5%‐3.0%‐1.5%0.0%1.5%3.0%4.5%6.0%7.5%9.0%10.5%12.0%13.5%
Machine dose rate low ~900 QAs
![Page 39: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/39.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Trend Analysis
Patient‐specific Gamma Pass Rates
Gamma 3%/3mm <1
0%
10%
20%
30%
40%
50%
60%
70% Last 100 QAs
![Page 40: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/40.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Future work
1. Better characterization with FFF2. Higher resolution characterization for SRS3. Applying results to patient’s CT for DVH
analysis
![Page 41: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/41.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Conclusions
1. Commissioning2. Ongoing Quality Improvement
![Page 42: 3D Diode Array Commissioning: Building Confidence in 3D QA … · 2020. 10. 1. · ArcCHECK Introduction • Designed for Helical & Arc Delivery • 1386diodes in a helical geometry](https://reader035.vdocuments.mx/reader035/viewer/2022071413/610a90c18d0b6717ae3877b2/html5/thumbnails/42.jpg)
CONE HEALTHCancer Center
CONE HEALTHCancer Center
Thank you!