  • Spirometer Calibration Report

    Reusable (dark plastic)

    Calibration test SUCCESSFULLY COMPLETED. The Device is now using the following correction values:(Expiration: 0,71% ,Inspiration: -1,05%)

    prin

    ted

    by w

    insp

    iroP

    RO

    5.3

    .0 -

    29/0

    9/20

    14 1

    4:10

    :44

    - Mod

    .D10

    Date 29/09/2014

    Device

    SPIROBANK IISerial Number 003885Version 3.6

    Calibration Test Results

    Inspiration Correction -1,05 %

    BTPS 1,063

    Expiration Correction 0,71 %

    Test Target 3 L

    Time 14:05

    Expiration (L)3,04

    Calibration Test Details

    Turbine Reusable

    2,70 3,09 3,09 3,08 3,08 3,03

    2,92 2,78 3,06 3,06 3,01 2,96Inspiration (L)

    * Volume Out Of Range (Excluded from Calibration)

    11 /Signature

    User WINSPIRO

    IMPORTANT: Flow and volume values measured during calibration are at ATP (Ambient Temperature and Pressure) conditions. All spirometry parameters instead are converted to BTPS conditions (Body Temperature and Pressure, Saturated). Eg a 3L syringe (ATP) measured during calibration will produce an FVC of 3.08 L (BTPS).