Spirometer Calibration Report
Reusable (dark plastic)
Calibration test SUCCESSFULLY COMPLETED. The Device is now using the following correction values:(Expiration: 0,71% ,Inspiration: -1,05%)
prin
ted
by w
insp
iroP
RO
5.3
.0 -
29/0
9/20
14 1
4:10
:44
- Mod
.D10
Date 29/09/2014
Device
SPIROBANK IISerial Number 003885Version 3.6
Calibration Test Results
Inspiration Correction -1,05 %
BTPS 1,063
Expiration Correction 0,71 %
Test Target 3 L
Time 14:05
Expiration (L)3,04
Calibration Test Details
Turbine Reusable
2,70 3,09 3,09 3,08 3,08 3,03
2,92 2,78 3,06 3,06 3,01 2,96Inspiration (L)
* Volume Out Of Range (Excluded from Calibration)
11 /Signature
User WINSPIRO
IMPORTANT: Flow and volume values measured during calibration are at ATP (Ambient Temperature and Pressure) conditions. All spirometry parameters instead are converted to BTPS conditions (Body Temperature and Pressure, Saturated). Eg a 3L syringe (ATP) measured during calibration will produce an FVC of 3.08 L (BTPS).