Fluoroscopic Image Display
![Page 2: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/2.jpg)
2
Television Cameras
3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors (APSs) or (CMOS)
Coupling I.I. to TV tube, CCD or APS Fiber optics
![Page 3: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/3.jpg)
3
Video Viewing System
Closed circuit television Video camera coupled to output screen and
monitor Video cameras
Vidicon or Plumbicon tube CCD APS
![Page 4: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/4.jpg)
4
Viewing
The output phosphor of the II is connected by fiber optic cables directly to a TV camera tube when the viewing is done through a television monitor.
The most commonly used camera tube - vidiconInside the glass envelope that surrounds the TV
camera tube is a cathode, an electron gun, grids and a target.
Past the target is a signal plate that sends the signal from the camera tube to the external video device
![Page 5: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/5.jpg)
5
Vidicon (tube) TV Camera
![Page 6: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/6.jpg)
6
Type of TV camera VIDICON TV camera
improvement of contrast improvement of signal to noise ratio high image lag
PLUMBICON TV camera (suitable for cardiology) lower image lag (follow up of organ motions) higher quantum noise level
CCD TV camera (digital fluoroscopy) digital fluoroscopy spot films are limited in resolution,
since they depend on the TV camera (no better than about 2 lp/mm) for a 1000 line TV system
![Page 7: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/7.jpg)
7
![Page 8: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/8.jpg)
8
Camera tube have a diameter of approximately 1 inch and a length of 6 inches.
![Page 9: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/9.jpg)
9
![Page 10: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/10.jpg)
10
Parts of the camera tube
Glass envelope Electron gun (Cathode) Control grid Electrostatic grids Target
![Page 11: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/11.jpg)
11
Camera Tube steps
Light is received by the camera tube. The light from the II is received at the face plate of
the target assembly. Electrons are formed into an electron beam (by the
control grid) at the electron gun. Electrons are burned off by thermionic emission
then focused and accelerated to the target. (made of antimony trisulfide)
![Page 12: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/12.jpg)
12Vidicon Target Assembly
![Page 13: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/13.jpg)
13
The electrons scan the signal plate similar to reading a page.
Starting in the upper left across to the right, then back to the left to right.
This is called an active trace. The movement of the electron beam
produces a RASTER pattern. The same pattern occurs in the TV
monitor.
![Page 14: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/14.jpg)
14
The signal plate sends the electrical video signal to the control unit which amplifies the signal and synchronizes the pulses between the camera tube and the TV monitor.
TV tube and monitor must be synchronized and duplicated to build a coherent image.
![Page 15: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/15.jpg)
15Synchronization (Sync Signals)
![Page 16: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/16.jpg)
Charge-Coupled Devices (CCDs)
1980s CCD were developed and replaced TV tubes and miniaturized imaging devices.
Light photons enter the silicon layer, ionization of the light separates the e-. A layer of microscopic electrodes beneath the silicon acts as a ground for the freed electrons. Movement of charges can be measured by a circuit.
16
Each electrode is connected to a storage capacitor (TFT)
![Page 17: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/17.jpg)
17
Video Camera Charged Coupled Devices (CCD)
Operate at lower voltages than video tubes More durable than video tubes
Semiconducting device Emits electrons in proportion to amount of
light striking photoelectric cathode Fast discharge eliminates lag
![Page 18: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/18.jpg)
18
CCD’s
![Page 19: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/19.jpg)
Advantages of CCDs
High spatial resolution High SNR High DQE No spatial distortion Unlimited life
Linear response to radiation
Lower patient dose Wider dynamic range
and better contrast resolution than conventional fluoroscopy
19
![Page 20: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/20.jpg)
20
TV camera and video signal (V)
On most fluoroscopy units, the resolution of the system is governed by the number of lines of the television system. MTF
Thus, it is possible to improve the high contrast resolution by increasing the number of television lines.
Some systems have 1,000 lines and prototype systems with 2,000 lines are being developed.
Color TV requires 3 vidicon tubes for each of the wavelengths (red, green, and blue)
![Page 21: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/21.jpg)
21
TV Monitors
![Page 22: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/22.jpg)
Monitors Cathode ray tube (CRT)
Liquid crystal display (LCD) Active matrix liquid crystal display
Emissive displays = produce their own light (diodes)
Not found application for medical imaging so far. Plasma screen Light-emitting diode (LED)
![Page 23: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/23.jpg)
23
MONITOR
CRT – Cathode Ray Tube Much larger than camera tube – but similar
function The electrons are synchronized by the control
unit – so they are of the same intensity and location as the electrons generated by the pick up (camera) tube.
![Page 24: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/24.jpg)
Soft copy viewingdigital cathode ray tube (CRT)
![Page 25: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/25.jpg)
25
TV Monitor
The TV monitor contains the picture tube called cathode ray tube (CRT).
It works like the camera tube. With an electron gun and control grids the
electron beam is fired toward the anode. The TV screen contains small fluorescent
crystals
![Page 26: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/26.jpg)
26
![Page 27: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/27.jpg)
27
Video Field Interlacing
![Page 28: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/28.jpg)
28
Different types of scanningINTERLACED
SCANNING
PROGRESSIVESCANNING
12 2
14
4 16
18 6
1
820
13
15
17
10
11
3
21
19
5
7
9
35
1816141210 8 6 4 2
79
11131517
1
625 lines in 40 msi.e. : 25 frames/s
![Page 29: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/29.jpg)
29
Two fields = a frame (525 lines) It take 1/30 of a second. To prevent flicker, two fields are interlaced
to form on television frame. There are 60 fields and 30 frames per
second. Each frame takes 33 ms to form The eye cannot detect flickering above 20
frames/sec.
![Page 30: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/30.jpg)
30
![Page 31: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/31.jpg)
31
RASTER Pattern
The electron beam moves in the same raster pattern as in the camera tube.
The signal consists of many individual pulses corresponding to the individual location on the camera tube target.
The varying voltage pulses are later reassembled into a visible in by the TV monitor.
![Page 32: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/32.jpg)
Refresh rate
The refresh rate is the measure of how fast the monitor rewrites the screen or the number of times that the image is redrawn on the display each second. The refresh rate helps to control the flicker seen by the user; the higher the refresh rate, the less flicker.
![Page 33: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/33.jpg)
33
TV RESOLUTION-Vertical Conventional TV: 525 TV lines to represent
entire image. Example: 9” intensifier (9” FOV)1) 9” = 229 mm
2) 525 TV lines/229 mm = 2.3 lines/mm
3) Need 2 TV lines per test pattern line-pair
4) (2.3 lines/mm) /2 lines/line-pair = 1.15 lp/mm
Actual resolution less because test pattern bars don’t line up with TV lines. Effective resolution obtained by applying a Kell Factor of 0.7.
Example: 1.15 x 0.7 Kell Factor = 0.8 lp/mm
![Page 34: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/34.jpg)
34
Kell Factor
The ability to resolve objects spaced apart in a vertical direction.
More dots = more scan lines = more/better resolution
Kell factor for 525 line system is 0.7
![Page 35: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/35.jpg)
35
KELL FACTOR
VERTICAL RESOLUTIONABILITY TO RESOLVE OBJECTS SPACED APART IN A VERTICAL DIRECTION
MORE DOTS(GLOBULES) = MORE SCAN LINES = MORE/BETTER RESOLUTION
RATIO OF VERTICAL RESOLUITON # OF SCAN LINES
KELL FACTOR FOR 525 LINE SYSTEM
IS 0.7
![Page 36: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/36.jpg)
Dot pitch
Dot pitch is the measurement of how close the dots are located to one another within a pixel
![Page 37: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/37.jpg)
37
TV RESOLUTION-Horizontal Along a TV line, resolution is limited by how fast
the camera electronic signal and monitor’s electron beam intensity can change from minimum to maximum.
This is bandwidth. For similar horizontal and vertical resolution, need 525 changes (262 full cycles) per line. Example (at 30 frames/second):
262 cycles/line x 525 lines/frame x 30 frames/second
= 4.2 million cycles/second or 4.2 Megahertz (MHz)
![Page 38: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/38.jpg)
38
Bandpass/Horizantal Resolution
Horizontal resolution is determined by the bandpass.
Bandpass is expressed in frequency (Hz) and describes the number of times per second the electron beam can be modulated.
The higher the bandpass, the better the resolution
![Page 39: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/39.jpg)
39
Different types of scanningINTERLACED
SCANNING
PROGRESSIVESCANNING
USED IN DIGITAL
12 2
14
4 16
18 6
1
820
13
15
17
10
11
3
21
19
5
7
9
35
1816141210 8 6 4 2
79
11131517
1
625 lines in 40 msi.e. : 25 frames/s
![Page 40: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/40.jpg)
40
Digital Uses Progressive Scan
1024 x 1024 Higher spatial resolution As compared to 525
8 images/sec (compared to 30 in 525 system)
![Page 41: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/41.jpg)
1080p 1080i
41
![Page 42: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/42.jpg)
42
Digital Uses Progressive Scan
SNR for TV camera tubes
200:1
SNR necessary for DF
Is 1000:1
Each image takes 33 ms
To display
![Page 43: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/43.jpg)
43
TV SYSTEMS Images are displayed on the monitor as
individual frames – which tricks the eye into thinking the image is in motion (motion integration)
15 f/sec – eye can still see previous image
Weakest Link - 2 lp/mm resolution
Real Time
![Page 44: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/44.jpg)
44
Final Image
The result of hundreds of thousands of tiny dots of varying degrees of brightness.
These dots are arranged in a specific patterns along horizontal scan lines.
Usually 525 scan lines. The electron gun within the picture tube scans from
top to bottom in 1/60 of a second, (262 1/2 lines) called a field.
![Page 45: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/45.jpg)
45
![Page 46: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/46.jpg)
46
Bandpass/Horizontal Resolution
Horizontal resolution is determined by the bandpass.
Bandpass is expressed in frequency (Hz) and describes the number of times per second the electron beam can be modulated.
The higher the bandpass, the better the resolution
![Page 47: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/47.jpg)
47
TV RESOLUTION-Horizontal Along a TV line, resolution is limited by how fast
the camera electronic signal and monitor’s electron beam intensity can change from minimum to maximum.
This is bandwidth. For similar horiz and vertical resolution, need 525 changes (262 full cycles) per line. Example (at 30 frames/second):
262 cycles/line x 525 lines/frame x 30 frames/second
= 4.2 million cycles/second or 4.2 Megahertz (MHz)
![Page 48: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/48.jpg)
active matrix liquid crystal display (AMLCD)
Must have sharper resolutionand high speed, each pixel hasits own TFT
![Page 49: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/49.jpg)
Crystals can be aligned by an external electric field
![Page 50: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/50.jpg)
Nematic liquid crystals
Light is twisted along with the crystals.
Varying the amount of twisting vs alignment allows
for more or lesslight and greys in between.
![Page 51: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/51.jpg)
Active matrix liquid crystal displays are superior to cathode ray tube displays.
AMLCD design – gives out more light,
reduces ambient light interference
Better contrast resolution
Less noise Less maintenance
![Page 52: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/52.jpg)
Luminance
Rate of light emitted from a source. Measured in lumen (Lm)
Luminous flux – light as perceived by the human eye.
![Page 53: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/53.jpg)
CRT vs. AMLCD
Lightweight Portable Less expensive More sizes Smaller profile Less heat Longer life Do not produce veiling glare (light
leaking)
![Page 54: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/54.jpg)
LCDs VS CRTs Perfect geometry Uniform sharpness
& brightness Low surface
reflectance (glare) No image flicker No veiling glare
Geometric corrections needed
Uneven sharpness & brightness
Image flicker Veiling glare
![Page 55: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/55.jpg)
LCDs 3 substantial disadvantages
Off-angle viewing degrades rapidly, brightness is less.
Have 1/5 less light intensity of a viewbox. Reducing the contrast and apparent spatial resolution.
Can not transmit true black density. Requires 30 minutes to warm up and
sensitive to extreme tempertures.
![Page 56: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/56.jpg)
When a digital display device is viewed from the side, illumination and image contrast are reduced.
![Page 57: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/57.jpg)
Spatial Resolution improves with the use of higher-
megapixel digital display devices
A 1-megapixel display will have a 1000×1000-pixel arrangement. A high-resolution monitor will have a 5-megapixel display, or a 2000×2500-pixel arrangement
![Page 58: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/58.jpg)
Image Display Resolution
Minimum display resolution of 2.5 LP/mm LCDs resolution is consistent and always
uniform. CRTs must be frequently checked for
deterioration.
![Page 59: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/59.jpg)
Contrast Resolution Dynamic range – the number of different gray
levels or brightness levels that can be represented in the displayed digital image. (gray scale)
Display monitor is the weakest link in the imaging chain, dynamic range compression.
Bit Depth?
59
![Page 60: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/60.jpg)
Pixels
![Page 61: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/61.jpg)
Digital Images – Bit Depth
Pixel values can be any bit depth (values from 0 to 1023)
Bit depth = # or gray shades available for image display
Image contrast can be manipulated to stretched or contracted to alter the displayed contrast.
Typically use “window width” and “window level” to alter displayed contrast and brightness
![Page 62: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/62.jpg)
Digital - GrayscaleDigital - Grayscale
Bit depth.Bit depth. Number of gray shadesNumber of gray shades
available for displayavailable for display 8 bit 2568 bit 256 10 bit 102410 bit 1024 12 bit 409612 bit 4096 4 bit 163844 bit 16384
![Page 63: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/63.jpg)
Display Bit Depth1 bit 6 bit 8 bit
2 shades 64 shades 256 shades
![Page 64: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/64.jpg)
DICOM Gray Scale Standard Display Function
EHR
64
![Page 65: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/65.jpg)
DICOM Function
Image exchange for both senders and receivers, support for connecting to a database and retrieving image information. Enabling another device to which images have been locally for retrieval Other dimensions deal with image management, patient scheduling information, image quality, media storage, security,
65
![Page 66: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/66.jpg)
Image Quality
Problem how to achieve consistency in the image presentation on different monitors, as well as on film, independent of the make or type of characteristics of the media?
Solution: DICOM Grayscale Standard Display Function.
66
![Page 67: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/67.jpg)
DICOM Gray Scale Standard Display Function
It specifies exactly what luminance or density level should be produced for a certain input value, based on the Barten curve, which maps the values into a range that is perceptually linear. This means that input values are mapped into a space that is perceived as linear by a human observer.
67
![Page 68: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/68.jpg)
Inconsistent image display. The lump visible on the left is almost
invisible on the right
68
![Page 69: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/69.jpg)
Viewable area
The viewable area is measured diagonally from one corner of the display to the opposite corner.
![Page 70: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/70.jpg)
Window Level & Width Function
![Page 71: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/71.jpg)
71
RECORDING THE IMAGE
STATIC IMAGES
DYNAMIC IMAGES
![Page 72: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/72.jpg)
72
Basic Componets of “old” Fluoroscopy “Imaging Chain”
Fluoro TUBE
Primary
Radiation PATIENT
EXIT Radiation
Image Intensifier
ABC Image Recording Devices
Fiber Optics OR
105 Photospot
CINE
VIDICON
Camera Tube
CONTROL
UNITTV
LENS
SPLIT
![Page 73: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/73.jpg)
73
Recording the Fluoroscopic Image
![Page 74: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/74.jpg)
Conventional Spot film
Changes the operation of the tube from low fluoroscopic mA to high radiographic mA. 05 – 5 ma (usually ave 1 – 2 ma)
Exposure delayed Masking capabilities
![Page 75: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/75.jpg)
75
Fluoroscopy mA vs spot exposure
Radiographic Exposure for cassette spot filmsmA increased to 100 – 200 mAESE 200 mR = 1 image
![Page 76: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/76.jpg)
76
Photons used: Fluoro vs Photons used: Fluoro vs RadiographyRadiography
Spotfilm Fluoroscopy
kVp: 85 85mA: 200 3Time (sec): 0.3 0.2*mAs: 60 0.6Ratio: 100 1
![Page 77: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/77.jpg)
77
Image recording
Cassette loaded spot film Where is the tube? How should you put the IR into the II slot?You can format the image, 2 on 1, 4 on 1 or 1 on 1Cassette loaded spot film increases patient dose
![Page 78: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/78.jpg)
78
CASSETTE SPOT FILMINGvs PHOTOSPOT FILMING
First type of recording used 9x9 cassettes then later up to 14x 14 9 on 1, 4 on 1, 2 on 1 Delay while exposing (anatomy still moving) Radiographic mA - must boost up to 100 – 200 mA for filming And moving cassettes around inside tower Higher patient dose Replaced by Photospot (f/sec) filming
![Page 79: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/79.jpg)
70 & 105 PHOTOSPOT (CAMERA)
Similar to a movie camera, only one frame exposed when activated.
79
![Page 80: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/80.jpg)
80
70 & 105 PHOTOSPOT (CAMERA)
Photo spot camera will take the image right off the output phosphor
This requires less patient dose 70 & 105 mm roll film General rule, larger film format = better
image quality but at increased patient dose.
![Page 81: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/81.jpg)
81
![Page 82: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/82.jpg)
82
CASSETTE SPOT FILMINGvs PHOTOSPOT FILMING
Photospot (f/sec) filming – Set at control panel from 1 f/sec – 12 f/sec Used for rapid sequence:
Upper Esophogram Voiding Cystourethrograms (Peds)
Lower patient dose
![Page 83: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/83.jpg)
Photospot filming
100 mR Frame rate12/s
![Page 84: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/84.jpg)
84
Recording the Fluoroscopic Image
Dynamic systems Cine film systems Videotape recording Static digital spot filming systems
![Page 85: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/85.jpg)
85
Cine Film Systems
Movie camera intercepts image 16 mm and 35 mm formats Record series of static exposures at high speed 30 – 60 frames per second
Offer increased resolution At the cost of increased patient dose
![Page 86: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/86.jpg)
86
Cinefluorgraphy aka CINE
35 or 16 mm roll film (movie film) 35 mm ↑ patient dose / 16 mm – higher quality images produced 30 f/sec in US – (60 frames / sec)
THIS DYNAMIC IMAGE CAPTURE = HIGHEST PATIENT DOSE (10X greater than fluoro)
(VS SINGLE EX DOSE IS ↓)
![Page 87: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/87.jpg)
87Cine Cinefluorography is used most
often in cardiology and neuroradiology.
The procedure uses a movie camera to record the image from the image intensifier.
These units cause the greatest patient doses of all diagnostic radiographic procedures, although they provide very high image quality.
The high patient dose results from the length of the procedure and relatively high inherent dose rate.
For this reason special care must be taken to ensure that patients are exposed at minimum acceptable levels.
Patient exposure can be minimized in a number of ways. The most obvious means of limiting exposure is to limit the time the beam is on.
CINE - 2mR per frame (60f/sec)
400 mr per “look”
![Page 88: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/88.jpg)
88
Synchronization
Camera shutters and x-ray pulsed fluoro happen at the same time
Only exposes pt when shutter is open to record image
Patient radiation dose ↑ as #/f/sec ↑
(filming a TV show – pattern seen)
![Page 89: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/89.jpg)
89
Framing frequency
Number of frames per second Cine – division of 60 (7.5, 15,30,90,120) Organ if interest determines f/s rate Patient exposure?
![Page 90: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/90.jpg)
90
Video disc
Referred to as electronic radiography. Fluoroscopic radiation continues only long enough
to build up a useful image on the display monitor. The image is stored as a single television frame on
the video disc recorder. There is about a 95% reduction in patient dose.
![Page 91: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/91.jpg)
91
Video tape
Utilizes VHS or high-resolution tapes. Patient’s exposure to radiation is not
increased. Used for barium swallows.
![Page 92: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/92.jpg)
Fluoroscopy & Digital Photospot
Real-time fluoroscopy operates at 30 f/sec the human eye can perceive 3-5 frames of image information.
Last image will appear very noisy.
Digital spot images are acquired through the II display system (TV tube or CCD)
92
![Page 93: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/93.jpg)
Fluoroscopy & Digital Photospot
Real-time fluoroscopy operates at 30 f/sec the human eye can perceive 3-5 frames of image information during that second.
Last image hold will appear very noisy.
Digital spot images are acquired through the II TV tube to ADC or CCD.
93
![Page 94: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/94.jpg)
Digital Spot Imaging
Half the image resolution of photospot images, 2.4 lp/mm
Digital spots are acquired with a much higher mA (radiographic mode). 50 – 100 x one frame of fluoroscopy
Digital images can be windowed/leveled, smoothing algorithms can reduce noise and enhance edges of images.
94
![Page 95: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/95.jpg)
• 1 frame of fluoro digital spot
95
![Page 96: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/96.jpg)
Digital Subtraction Angiography
Computer controlled image matrix size, dynamic range and image acquisition rate.
Matrix size requirement will determine imaging rate capability. 512 x 512 = 30 images per sec. 1024 x 1024 only 8 image per second can be acquired.
Limitation is enormous quantities of data transfer.
96
![Page 97: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/97.jpg)
Postprocessing image subtraction - DSA
Temporal subtraction Dual Energy subtraction Hybrid subtraction
Using both subtraction techniques
97
![Page 98: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/98.jpg)
Temporal Subtraction
Time Exposures made over time will be subtracted
from each other. Ex: scout, during, post Automatic injector and x-ray generator work
in tandem.
98
![Page 99: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/99.jpg)
Dual energy subtraction
2 images acquired at very different kVp. Each pixel value can be compared to
determine how much change occurred. Tissues are identified. Specific ranges of
value can be subtracted from the image.
99
![Page 100: 1 Fluoroscopic Image Display. 2 Television Cameras 3 Methods: Thermionic television camera tube Solid state charge-coupled device (CCD) Active pixel sensors](https://reader036.vdocuments.mx/reader036/viewer/2022081501/56649d9c5503460f94a85568/html5/thumbnails/100.jpg)
100